← Product Code [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY) · K182371

# da Vinci SP Surgical System, EndoWrist SP Instruments, and Accessories (K182371)

_Intuitive Surgical, Inc. · NAY · Mar 14, 2019 · Gastroenterology, Urology · SESU_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K182371

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY.md)
- **Decision Date:** Mar 14, 2019
- **Decision:** SESU
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K182371 · Mar 14, 2019 | da Vinci SP Surgical System, EndoWrist SP Instruments, and Accessories | Intuitive Surgical, Inc. | Systematic literature review of 13 peer-reviewed clinical publications | The sponsor conducted a systematic literature review of real-world clinical data for the predicate device (da Vinci Si) to compare clinical performance parameters (e.g., operative time, complication rates, conversion rates) against the subject device's prospective clinical study results, supporting a finding of substantial equivalence. | Literature review; Real-world clinical outcomes; Predicate device comparison; TORS |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Systematic literature review of da Vinci Si TORS outcomes; Systematic literature review of peer-reviewed publications; Study Period: 2013-2018 | Patients undergoing transoral robotic surgery (TORS) for oropharyngeal cancer using the da Vinci Si Surgical System; Sample Size: N=9315 (aggregate across 13 studies); Number of Sites: Not applicable | Not applicable for this study | Operative time, estimated blood loss, transfusion rate, length of hospital stay, conversion rate, mortality rate, readmission rate, reoperation rate, postoperative complications |

## Indications for Use

To assist in the accurate control of endoscopic instruments in minimally invasive surgery.

## Device Story

The da Vinci SP Surgical System (Model SP1098) is a software-controlled, electromechanical system for single-port minimally invasive surgery. It comprises a Surgeon Console, Patient Cart, and Vision Cart. The surgeon operates from the console, using master controls and foot pedals to manipulate EndoWrist SP instruments and a camera instrument attached to a single robotic arm. The system provides a 3D endoscopic view of the surgical site. The instruments feature articulating distal tips mimicking human wrist, shoulder, and elbow movement to enable triangulation and X-Y-Z movement within the body. The system is used in hospital operating rooms by trained physicians. It facilitates tasks such as grasping, cutting, dissection, electrocautery, and suturing. By enabling single-port access, the device aims to provide minimally invasive surgical benefits for specific urologic and otolaryngology procedures.

## Clinical Evidence

Prospective, multicenter, single-arm clinical study (N=33) evaluated safety and performance in TORS for T1/T2 malignant oropharyngeal tumors. Primary endpoints: device-related serious adverse events (0%) and conversion rate to open surgery (0%). Mean operative time 49.6 ± 33.1 min; mean blood loss 12.2 ± 18.7 mL. Comparison against literature data for the da Vinci Si system (N=9315) showed all 9 clinical parameters for the SP1098 fell within or below the 95% CIs of the predicate literature. Bench, animal, and cadaver testing supported anatomical access and surgical task performance.

## Technological Characteristics

Electromechanical, software-controlled system. Components: Surgeon Console, Patient Cart, Vision Cart. Instruments: EndoWrist SP (reusable, articulating). Sensing/Actuation: Master/slave servomechanism with servo drive and system-level motor control. Connectivity: Networked system components. Sterilization: Reprocessable instruments/accessories. Software: Embedded control system.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- da Vinci SP Surgical System, Model SP1098 ([K173906](/device/K173906.md))
- Intuitive Surgical da Vinci Si Surgical System, Model IS3000 ([K090993](/device/K090993.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 14, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Intuitive Surgical, Inc. Mike Yramategui Fellow Regulatory Engineer 1266 Kifer Road Sunnyvale, California 94086

Re: K182371

Trade/Device Name: da Vinci SP Surgical System, EndoWrist SP Instruments, and Accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: February 11, 2019 Received: February 12, 2019

Dear Mike Yramategui:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Precautions/Warnings/Contraindications section of the device's labeling:

The safety and effectiveness of this device for use in the performance of general laparoscopic surgery procedures have not been established. This device is only intended to be used for single port

{1}------------------------------------------------

urological procedures and for transoral otolaryngology surgical procedures in the oropharynx for benign tumors and malignant tumors classified as T1 and T2 with the da Vinci EndoWrist SP Instruments and the da Vinci SP Surgical System (SP1098).

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Angela C. Krueger -S for

William H. Maisel, MD, MPH Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K182371

#### Device Name

da Vinci SP Surgical System, Model SP1098, EndoWrist SP Instruments, and Accessories

#### Indications for Use (Describe)

### da Vinci SP® Surgical System, Model SP1098:

The Intuitive Surgical® Endoscopic Instrument Control System (da Vinci SP® Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP® Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant turnors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

### EndoWrist SP® Instruments:

Intuitive Surgical® Endo Wrist SP® Instruments are controlled by the da Vinci SP® Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)

| <img alt="Checked" src="checkbox_checked.png"/> Prescription Use (Part 21 CFR 801 Subpart D)    |
|-------------------------------------------------------------------------------------------------|
| <img alt="Unchecked" src="checkbox_unchecked.png"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(k) Summary

| 510(k) Owner:                      | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086                                                                                                               |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                           | Mike Yramategui<br>Fellow Regulatory Engineer<br>Phone Number: 408-523-2145<br>Fax Number: 408-523-8907<br>Email: Mike.Yramategui@intusurg.com                                   |
| Date Summary Prepared:             | March 12, 2019                                                                                                                                                                   |
| Trade Name:                        | da Vinci SP® Surgical System, Model SP1098,<br>EndoWrist SP® Instruments, and Accessories                                                                                        |
| Common Name:                       | Endoscopic instrument control system, endoscopic<br>instruments and accessories                                                                                                  |
| Classification:                    | Class II<br>21 CFR 876.1500, Endoscope and Accessories                                                                                                                           |
| Product Codes:                     | NAY (System, Surgical, Computer Controlled Instrument)                                                                                                                           |
| Classification Advisory Committee: |                                                                                                                                                                                  |
| Committee:                         | General and Plastic Surgery                                                                                                                                                      |
| Predicate Device:                  | da Vinci SP Surgical System, Model SP1098,<br>EndoWrist SP Instruments, and Accessories (K173906)<br>Intuitive Surgical® da Vinci Si® Surgical System,<br>Model IS3000 (K090993) |

### Device Description

The da Vinci SP Surgical System, Model SP1098 is a software-controlled, electromechanical system designed for surgeons to perform single port minimally invasive surgery. The Model SP1098 Surgical System consists of a Surgeon Console, a Patient Cart, and a Vision Cart, and is used with a Camera Instrument, EndoWrist SP Instruments, and Accessories.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows three pieces of medical equipment against a white background. The first piece of equipment is a surgical console with a viewing screen and controls. The second piece of equipment is a robotic surgical arm. The third piece of equipment is a cart with a monitor and other electronic equipment.

Surgeon Console, Model SS1098

Patient Cart, Model PS1098

Vision Cart, Model VS1098

The surgeon seated at the Surgeon Console controls all movement of the EndoWrist SP Instruments and Camera Instrument by using two Master Controls and a set of foot pedals. The surgeon views the three-dimensional endoscopic image on a High Resolution Stereo Viewer (3D Viewer), which provides him/her a view of patient anatomy and surgical instrumentation, along with icons and other user interface features.

The Vision Cart includes the supporting electronic and video processing equipment for the system.

The Patient Cart is positioned at the operating room table and has four instrument drives on a single arm that is positioned over the target patient anatomy. A Camera Instrument attaches onto one instrument drive and provides the surgeon a high resolution, threedimensional view of the patient anatomy. A suite of EndoWrist SP Instruments can be attached to and detached from the other three instrument drives, enabling the surgeon to perform various surgical tasks. The Camera Instrument and up to three surgical instruments can be used simultaneously, entering the patient through a single port. Accessories including a cannula, an obturator, a seal, an entry guide, disposable tips for selected instruments, instrument sheaths, and a drape are needed to perform procedures with the system.

The EndoWrist SP Instruments come in various configurations such as graspers, scissors, and needle drivers. The EndoWrist SP instruments have a unique articulating design at the distal tip that mimics the human wrist, shoulder, and elbow to enable triangulation and X-Y-Z movement of the instrument in the body. Each instrument is used to perform specific surgical tasks such as grasping, suturing, tissue manipulation, and electrocautery. The EndoWrist SP Instruments can be used only with the SP1098 Surgical System. The instruments are reusable. They are programmed with a maximum number of surgical procedures based upon life testing.

{6}------------------------------------------------

The EndoWrist SP Camera Instrument is a reusable endoscope that provides a stereo image of the surgical site. Like the instruments, the distal end includes multiple joints that provide the flexibility needed for use with a single-port system.

The following EndoWrist SP Instruments and accessories are listed for use with the da Vinci SP Surgical System, Model SP1098:

## EndoWrist SP Instruments:

- Fenestrated Bipolar Forceps ●
- Maryland Bipolar Forceps
- Medium-Large Clip Applier (a.k.a. ML Clip Applier)
- Monopolar Cautery Instrument ●
- Monopolar Curved Scissors (a.k.a. MCS)
- Needle Driver ●
- Round Tooth Retractor ●
- Cadiere Forceps ●
- EndoWrist SP Camera, 0° (a.k.a. Camera Instrument)

### Accessories for the SP1098 da Vinci SP Surgical System:

- SP Cannula, Circular, 25 x 100 mm (a.k.a. Cannula) ●
- SP Obturator, Circular, 25 x 100 mm (a.k.a. Obturator) ●
- EntryGuide Kit C.6.6.6, 25 x 100 mm (a.k.a. Entry Guide and Cannula Seal) ●
- Instrument Sheath ●
- Camera Sheath ●
- MCS Tip ●
- Cautery Hook Tip ●
- Cautery Spatula Tip ●
- Bipolar Cautery Cord
- EnergyShield Monopolar Cautery Cord ●
- Instrument Arm Drape (a.k.a. Drape)

### Intended Use

To assist in the accurate control of endoscopic instruments in minimally invasive surgery.

{7}------------------------------------------------

## Indications for Use

## da Vinci SP® Surgical System, Model SP1098

The Intuitive Surgical® Endoscopic Instrument Control System (da Vinci SP® Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP® Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

## EndoWrist SP® Instruments

Intuitive Surgical® EndoWrist SP® Instruments are controlled by the da Vinci SP® Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

## Comparison of Technological Characteristics with the Predicate Device

The da Vinci SP Surgical System, Model SP1098 and EndoWrist SP Instruments and Accessories are unchanged from the predicate device, as cleared under K173906, in terms of intended use, design, performance, and technological characteristics. The only difference is that the labeling has been changed to include use in transoral otolaryngology procedures. Performance testing was conducted to evaluate the performance of the SP1098 for conducting all surgical tasks and representative procedures covered by the proposed Indications for Use. Additionally, the testing compared the safety and ability of the SP1098 and the da Vinci Si Surgical System, Model IS3000 to facilitate performance of surgical tasks, as well as facilitate completion of each representative surgical procedure, and to demonstrate functional equivalence between the SP1098 and the IS3000 (additional predicate device) for use in transoral otolaryngology procedures as cleared under K090993.

{8}------------------------------------------------

| Attribute    | Subject Device<br>SP1098 da Vinci SP Surgical System,<br>Instruments, and Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Primary Predicate Device:<br>SP1098 (K173906)<br>Additional Predicate Device:<br>IS3000 (K090993)                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device types | System Carts:<br>• Surgeon Console<br>• Vision Cart, including the<br><i>EnergyShield</i> Monitor<br>• Patient Cart<br>Instruments:<br>• Fenestrated Bipolar Forceps<br>• Maryland Bipolar Forceps<br>• Medium-Large Clip Applier<br>• Monopolar Cautery Instrument<br>• Monopolar Curved Scissors<br>• Needle Driver<br>• Round Tooth Retractor<br>• Cadiere Forceps<br>Endoscope:<br>• Camera Instrument<br>Accessories:<br>• Cannula<br>• Entry Guide<br>• Cannula Seal<br>• Obturator<br>• Instrument Sheath<br>• Camera Sheath<br>• <i>Energy Shield</i> Monopolar Cautery<br>Cord<br>• Drape<br>• MCS Tip | Primary Predicate Device: SAME<br>Additional Predicate Device:<br>SAME types of system carts as subject<br>device.<br>SIMILAR robotic arms, instruments and<br>accessories.<br>The primary difference is that the IS3000<br><i>da Vinci Si</i> Surgical System utilizes four<br>robotic arms and four ports to<br>accommodate three surgical instruments<br>and a camera. In contrast, the subject<br>device (SP1098 <i>da Vinci SP</i> Surgical<br>System) uses a single robotic arm to<br>deliver three surgical instruments and a<br>camera through a single port. |
| Common Name  | Endoscopic instrument control system,<br>endoscopic instruments and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | SAME as subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

| Table 1: Comparison of Subject and Predicate Devices |  |  |
|------------------------------------------------------|--|--|
|------------------------------------------------------|--|--|

{9}------------------------------------------------

| Attribute                               | Subject Device<br>SP1098 da Vinci SP Surgical System,<br>Instruments, and Accessories                                                                                                                                     | Primary Predicate Device:<br>SP1098 (K173906)<br>Additional Predicate Device:<br>IS3000 (K090993) |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Regulation<br>Number & Name             | 21 CFR 876.1500<br>Endoscope and Accessories                                                                                                                                                                              | SAME as subject device                                                                            |
| Classification<br>Advisory<br>Committee | General and Plastic Surgery                                                                                                                                                                                               | SAME as subject device                                                                            |
| Product Code(s)                         | NAY (System, Surgical, Computer<br>Controlled Instrument)                                                                                                                                                                 | SAME as subject device                                                                            |
| Classification                          | Class II                                                                                                                                                                                                                  | SAME as subject device                                                                            |
| Intended Use                            | To assist in the accurate control of<br>endoscopic instruments in minimally<br>invasive surgery.                                                                                                                          | SAME as subject device                                                                            |
| Principles of<br>Operation              | Facilitates accurate movement of surgical<br>instruments and an endoscope through a<br>single surgical port by using a<br>master/slave servomechanism that<br>incorporates servo drive and system-level<br>motor control. | SAME as subject device                                                                            |
| Prescription Use                        | Physician use only                                                                                                                                                                                                        | SAME as subject device                                                                            |
| Where Used                              | Hospital                                                                                                                                                                                                                  | SAME as subject device                                                                            |

| Table 1 (continued): Comparison of Subject and Predicate Devices |  |  |  |
|------------------------------------------------------------------|--|--|--|
|                                                                  |  |  |  |

{10}------------------------------------------------

## Performance Data

Performance test data included bench, animal and cadaver testing, and a confirmatory clinical study to demonstrate that the subject device is substantially equivalent to the legally marketed predicate device.

## Bench Performance Testing

The addition of transoral otolaryngology surgical procedures to the SP1098 indications does not change any of the safety or performance requirements that were previously verified and/or validated for the SP1098 (K173906) regarding cleaning, sterilization, packaging, shelf life, biocompatibility, software, electrosurgical performance, electromagnetic compatibility, electrical safety, mechanical and electrical performance, reliability, or human factors. However, cleaning validation with a soil representative of TORS was performed (see table below) because it had not yet been completed for all the instruments and accessories at the time of the SP1098 clearance (K173906).

| Item                 | Testing             |
|----------------------|---------------------|
| Camera Instrument    | Cleaning Validation |
| Surgical Instruments | Cleaning Validation |
| Accessories          | Cleaning Validation |

Table 2: Summary of Bench Performance Testing

FDA guidance documents used in bench performance testing: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling. Guidance for Industry and Food and Drug Administration Staff, issued March 17, 2015.

## Cadaver and Animal Performance Testing

Comparative animal and cadaver testing was conducted to demonstrate the equivalence of the SP1098 System to the IS3000 System for performing transoral otolaryngology surgical (TORS) procedures. Cadavers were used to demonstrate device performance for anatomical access and reach. Live animals were used to assess safety and performance in cases where a live tissue model was appropriate. These models replicate factors experienced during normal clinical use, including working with perfused organs, normal tissue manipulation, and ensuring that appropriate hemostasis is achieved and maintained.

Three (3) independent practicing surgeons participated in a pre-clinical study using the SP1098 system to perform a set of transoral otolaryngology surgical procedures using both the SP1098 and the IS3000. Each surgeon performed four (4) procedures in

{11}------------------------------------------------

cadavers and two (2) procedures in a live porcine model with each system, for a total of eighteen (18) procedures performed with each system. Surgical success criteria for each procedure are listed below. In addition, surgeons completed questionnaires that evaluated their ability to perform surgical tasks with the two systems.

| Procedure                                                      | Subject | Surgical Success Criteria                                                                                                                                                                                                           |
|----------------------------------------------------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tongue base<br>resection                                       | Cadaver | • Resection volume is surgically acceptable<br>• Resection boundaries achieved<br>• Critical anatomy identified (e.g., lingual artery, lingual nerve,<br>hypoglossal nerve)<br>• No trauma to epiglottis                            |
| Lateral<br>oropharyngectomy<br>(aka, radical<br>tonsillectomy) | Cadaver | • Resection volume is surgically acceptable<br>• Resection boundaries achieved<br>• Critical anatomy identified (e.g., medial pterygoid muscle,<br>glossopharyngeal nerve)                                                          |
| Tongue base<br>resection                                       | Porcine | • Resection volume is surgically acceptable<br>• Resection boundaries achieved<br>• Critical anatomy identified (e.g., lingual artery, lingual nerve,<br>hypoglossal nerve)<br>• No trauma to epiglottis<br>• Hemostasis maintained |
| Lateral<br>oropharyngectomy<br>(aka, radical<br>tonsillectomy) | Porcine | • Resection volume is surgically acceptable<br>• Resection boundaries achieved<br>• Critical anatomy identified (e.g., medial pterygoid muscle,<br>glossopharyngeal nerve)<br>• Hemostasis maintained                               |

| Table 3: Surgical Success Criteria in Animal and Cadaver Testing |  |
|------------------------------------------------------------------|--|
|------------------------------------------------------------------|--|

## Clinical Study

A prospective, multicenter, single-arm clinical study (NCT03049280) was conducted under an investigational device exemption (IDE #G160251) to confirm the safety and performance of the SP1098 da Vinci SP Surgical System, Instruments and Accessories in transoral robotic surgery (TORS) for malignant oropharyngeal tumors classified as T1 and T2. A total of 33 subjects were enrolled at three institutions in the United States.

Sixty four percent (64%) of subjects had one or more comorbid conditions which included hypertension, diabetes, chronic obstructive pulmonary disease (COPD). connective tissue, liver, renal or peptic ulcer disease. Thirty one (31) (93.9%) subjects underwent neck dissection, and they were either staged prior to TORS [sixteen (16)

{12}------------------------------------------------

(48.5%) subjects] or performed concomitantly with TORS [fifteen (15) (45.4%) subjects].

Safety was determined by incidence rate of device-related serious adverse events. Performance was determined by conversion rate from da Vinci SP to an open approach required to complete the indicated procedure. Results show that the device-related serious adverse event rate was 0%, and the conversion rate was 0%. Additional clinical study parameters were collected to demonstrate substantial equivalence to the predicate. Tables 4 and 5 below summarize the study data.

| Clinical Study Parameters                                     | Study Data  |
|---------------------------------------------------------------|-------------|
| Number of Patients, n                                         | 33          |
| Operative time (min), mean ± SD                               | 49.6 ± 33.1 |
| Estimated blood loss (mL), mean ± SD                          | 12.2 ± 18.7 |
| Transfusion rate, %                                           | 0           |
| Conversion to open rate, %                                    | 0           |
| Device-related complication rate, %                           | 0           |
| Intraoperative complications, n                               | 0           |
| Length of hospital stay (days), mean ± SD                     | 4.2 ± 1.9   |
| Rates of positive surgical margins, n (%)                     | 1 (3.0)     |
| Readmission rate, n (%)                                       | 3 (9.1)     |
| Reoperation rate, n (%)                                       | 2 (6.1)     |
| Mortality rate, %                                             | 0           |
| Postoperative complications rates (non-device related), n (%) | 8 (24.2)    |

Table 4: TORS Clinical Study Summary (da Vinci SP)

Table 5: Postoperative Complications (8 study subjects)

| Type                                                                                          | Number |
|-----------------------------------------------------------------------------------------------|--------|
| Non-serious CTCAE I-II                                                                        |        |
| • Difficulty swallowing with increased secretions                                             | 1      |
| • Fall                                                                                        | 1      |
| • Neck seroma                                                                                 | 1      |
| • Nasal regurgitation                                                                         | 1      |
| • Difficulty opening mouth or trismus                                                         | 1      |
| Serious CTCAE III – IV                                                                        |        |
| • Hypoxemia leading to sequelae of altered mental status and hospital acquired pneumonia      | 1      |
| • Hospital-acquired pneumonia                                                                 | 1      |
| • Hemorrhage                                                                                  | 1      |
| • Clostridium difficile enterocolitis                                                         | 1      |
| • Left middle cerebral artery stroke leading to sequela of hemorrhage post TPA-administration | 1      |

{13}------------------------------------------------

No intraoperative complications were observed and the safety endpoint was met. The bench, animal, cadaver and confirmatory clinical study demonstrate that SP1098 da Vinci SP Surgical System is as safe and effective as the predicate device.

Specific parameters of the clinical study were also compared to recently published clinical data between 2013 and 2018 on the da Vinci Si Surgical System for transoral otolaryngology surgery (TORS). In order to obtain the latest real world data on the predicate device, a systematic literature search of peer-reviewed publications was conducted using specific search criteria and filters.

Data from thirteen (13) publications reporting on transoral robotic surgery for cancer in the oropharynx with da Vinci Si Surgical System met these criteria and were included in the comparison. A detailed summary of this comparison of the da Vinci SP Surgical System data to the published clinical literature data on the da Vinci Si Surgical System is provided in Table 6 below for nine (9) parameters. Each parameter from the literature is shown based on its sample size, weighted average, weighted standard deviation and 95% confidence interval. The weighting has been performed according to the sample size of the study cohort using the weighted version of proc univariate within SAS version 9.4 for the calculations. The abstracted data is summarized in Table 7 at the end of this 510(k) Summary, and a bibliography of the literature is provided in Table 8.

{14}------------------------------------------------

|                                                     | Sources of Data           |                                                    |
|-----------------------------------------------------|---------------------------|----------------------------------------------------|
| Clinical Parameter                                  | G160251/K182371           | da Vinci Si -<br>Literature1-13                    |
| Sample Size                                         | N=33                      | N=9315                                             |
| Operative Time (min), mean ± SD<br>95% CI           | 49.6 ± 33.1<br>[0, 114.5] | N=115 4,7,8<br>148.5 (± 71.0)<br>[9.3, 287.7]      |
| Estimated Blood Loss (mL), mean ± SD<br>95% CI      | 12.2 ± 18.7<br>[0, 48.9]  | N=20 8<br>100 a<br>[31.4, 168.6]                   |
| Transfusion rate, (%)<br>95% CI                     | 0%<br>[0, 10.4]*          | N=1806 3,10<br>2.1% (±5.1)<br>[0, 12.1]            |
| Length of Hospital Stay (days), mean ± SD<br>95% CI | 4.2 ±1.9<br>[0.5, 7.9]    | N=2798 2-4,7,8,10<br>4.6 (±1.3) b<br>[2.1, 7.2]    |
| Conversion Rate from TORS (%)<br>95% CI             | 0%<br>[0, 10.4]*          | N=896 2<br>2.1%<br>[1.3, 3.2]**                    |
| Mortality Rate (perioperative) (%)<br>95% CI        | 0%<br>[ 0, 10.4]*         | N=5307 1,3,7,9-11,13<br>0.52% (±0.62)<br>[0, 1.7]  |
| Readmission Rate (%)<br>95% CI                      | 9.1%<br>[3.1, 23.6]*      | N=5265 2,4,5,6,10,13<br>8.8% (±3.1)<br>[2.7, 14.9] |
| Reoperation Rate (%)<br>95% CI                      | 6.1%<br>[1.7, 19.7]*      | N=1417 5,10,12<br>3.9% (±1.6)<br>[0.8, 7.0]        |
| Postoperative Complications (%)<br>95% CI           | 24.2%<br>[12.8, 41.0]*    | N=520 4,5,7<br>55.7% (±18.6) c<br>[19.2, 92.2]     |

#### Table 6: da Vinci SP versus da Vinci Si Literature Comparison

*Wilson (Score) 95% confidence interval, **Agresti Coull 95% confidence interval

Weighted averages, weighted standard deviations and 95% confidence intervals are provided for 7 of the 9 parameters from the literature; estimated blood loss and conversion rate are as reported in one publication.

Studies by Clayburgh D et al. (ref #4), and Chung TK et al. (ref #3), include multiple independent robotic cohorts used in the weighted average calculations for either discrete or continuous parameters.

4Median estimated blood loss is used in place of mean, and standard deviation was estimated using the range/4 reported by More et al.

bMedian length of hospital stay is used in place of mean, standard deviation was estimated using the range/4 for both cohorts reported in Clayburgh D. et al. In the paper by Parhar HS. et al. (ref #10) mean length of hospital stay for 2 cohorts (readmissions and non- readmissions) is used in the overall calculation. Richmon J. et al. (ref#11) only reports a difference in length of hospital stay between robotic and non-robotic cohorts and is therefore not used in the overall calculation.

6Weighted mean postoperative complication rate includes the paper by Lee SY. et al. (ref #7) which reports only TORS related complications.

{15}------------------------------------------------

Comparison of results from the da Vinci SP Surgical System to the published literature on the da Vinci Si Surgical System demonstrate that the point estimate of all 9 parameters from the da Vinci SP study fall within or below the 95% confidence intervals calculated from the published clinical literature on da Vinci Si.

- Operative time
- Transfusion rate
- . Length of hospital stay
- Readmission rate
- Estimated blood loss ●
- Conversion rate from TORS to open ●
- Mortality rate
- Reoperation rate
- . Post-operative complications rate

This comparison was used to demonstrate that the da Vinci SP Surgical System is as safe and effective as the predicate device and does not raise different questions of safety and effectiveness.

# Conclusion

Based on the intended use, indications for use, technological characteristics and performance data, the Intuitive Surgical da Vinci SP Surgical System, Model SP1098, EndoWrist SP Instruments, and Accessories, is substantially equivalent (SE) to the predicate devices. This SE determination is based on performance testing that included: bench, cadaver, and animal testing with simulated and representative transoral otolaryngology surgical procedures, and a confirmatory clinical study. The bench performance testing verified that the instruments and accessories could be cleaned when exposed to soiling expected in transoral otolaryngology surgical procedures. The cadaver performance testing validated the users' ability to use the system to accurately control the endoscopic instruments, to reach the necessary target anatomy, and to perform surgical tasks. The simulated and representative transoral otolaryngology surgical procedures in live animals validated that the system can be used to successfully complete the representative transoral otolaryngology surgical procedures encompassed by the indications for use statement. The confirmatory clinical study showed that the results experienced in the cadaver and animal study reflect the results experienced in a human clinical setting, and that there are no different questions of safety or effectiveness compared to the predicate device. A complete list of postoperative complications experienced in the clinical study is provided in Table 5, and Table 6 provides a comparison of clinical parameters from the study on da Vinci SP (subject device) and published clinical literature on da Vinci Si (predicate device), and supports substantial equivalence for performing transoral otolaryngology surgical procedures.

{16}------------------------------------------------

| Author                 | Cohort                                                                      | Study<br>Size | Operation<br>Time (min) | Blood<br>Transfusion<br>(%) | EBL (mL)     | Length of Stay (days)                        | Conversions     | Postop<br>Complication Rate<br>(30-days) (%) | Mortality Rate<br>(in-hospital or<br>30 days)<br>(%) | Readmission<br>Rate (30-days)<br>(%) | Reoperation<br>Rate (30-days)<br>(%) |
|------------------------|-----------------------------------------------------------------------------|---------------|-------------------------|-----------------------------|--------------|----------------------------------------------|-----------------|----------------------------------------------|------------------------------------------------------|--------------------------------------|--------------------------------------|
| 1. Baliga S (2018)     | Robotic                                                                     | 2680          | NR                      | NR                          | NR           | NR                                           | NR              | NR                                           | 0.4%                                                 | NR                                   | NR                                   |
| 2. Chen MM (2014)      | Robotic                                                                     | 877           | NR                      | NR                          | NR           | 4.7 (SE 0.2)                                 | 19/896 (2.1%)   | NR                                           | NR                                                   | 4.40%                                | NR                                   |
| 3. Chung TK (2015)     | Robotic partial<br>pharyngectomy -<br>mild/moderate                         | 523           | NR                      | 0%                          | NR           | 3.7 ± 2.2                                    | NR              | NR                                           | 0%                                                   | NR                                   | NR                                   |
|                        | Robotic partial<br>pharyngectomy -<br>major/extreme                         | 118           |                         | 21.20%                      | NR           | 7.6 ± 7.0                                    |                 |                                              | 4.20%                                                |                                      |                                      |
|                        | Robotic partial<br>glossectomy for<br>BOT Ca<br>(mild/moderate)             | 147           |                         | 0%                          | NR           | 3.54 ± 2.61                                  |                 |                                              | 0%                                                   |                                      |                                      |
|                        | Robotic partial<br>glossectomy for<br>anterior tongue ca<br>(mild/moderate) | 68            |                         | 0%                          | NR           | 4.8 ± 3.1                                    |                 |                                              | 0%                                                   |                                      |                                      |
| 4. Clayburgh D (2017)  | Robotic placebo                                                             | 33            | 202 ± 51                | NR                          | NR           | median 5 (range 2-11)                        | NR              | 7/33 (21.2%)                                 | NR                                                   | 0                                    | NR                                   |
|                        | Robotic steroid                                                             | 35            | 211 ± 61                | NR                          | NR           | median 4 (range 2-6)                         |                 | 10/35 (28.6%)                                |                                                      | 2/35 (5.7%)                          |                                      |
| 5. Frenkel CH (2017)   | Robotic                                                                     | 425           | NR                      | NR                          | NR           | NR                                           | 273/425 (64.2%) | NR                                           | 56/425 (13.2%)                                       | 26/425 (6.1%)                        |                                      |
| 6. Goel AN (2018)      | Robotic                                                                     | 2576          | NR                      | NR                          | NR           | NR                                           | NR              | NR                                           | 236/2576 (9.2%)                                      | NR                                   |                                      |
| 7. Lee SY (2013)       | Robotic                                                                     | 27            | 48.3 ± 7.5              | NR                          | NR           | 14.6 ± 4                                     | NR              | TORS related<br>comps: 0%                    | 0%                                                   | NR                                   | NR                                   |
| 8. More YI (2013)      | Robotic                                                                     | 20            | 86 ± 36                 | NR                          | 100 (60-200) | 2 (1-4)                                      | NR              | NR                                           | NR                                                   | NR                                   | NR                                   |
| 9. Motz K (2017)       | Robotic                                                                     | 304           | NR                      | NR                          | NR           | NR                                           | NR              | NR                                           | 1.00%                                                | NR                                   | NR                                   |
| 10. Parhar HS (2018)   | Robotic                                                                     | 955           | NR                      | 13/950<br>(1.4%)            | NR           | Readmit: 5.7 ± 6.8<br>not readmit: 4.3 ± 4.1 | NR              | NR                                           | 5/955 (0.5%)                                         | 117/950 (12.3%)                      | 27/950 (2.8%)                        |
| 11. Richmon J (2014)   | Robotic                                                                     | 116           | NR                      | NR                          | NR           | -1.5                                         | NR              | NR                                           | 0%                                                   | NR                                   | NR                                   |
| 12. Smith RV (2015)    | Robotic                                                                     | 42            | NR                      | NR                          | NR           | NR                                           | NR              | NR                                           | NR                                                   | NR                                   | 3/42 (7.1%)                          |
| 13. Zevallos JP (2016) | Robotic                                                                     | 369           | NR                      | NR                          | NR           | NR                                           | NR              | NR                                           | 4/369 (1.1%)                                         | 14/369 (3.8%)                        | NR                                   |

NR: Not reported

{17}------------------------------------------------

| #  | Publication                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|----|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1  | Baliga S, Kabarriti R, Jiang J, et al. Utilization of Transoral Robotic Surgery (TORS) in patients<br>with Oropharyngeal Squamous Cell Carcinoma and its impact on survival and use of<br>chemotherapy. Oral Oncology. 2018;86:75-80.                                                                                                                                                                                                                                                                                                                           |
| 2  | Chen MM, Roman SA, Kraus DH, Sosa JA, Judson BL. Transoral Robotic Surgery: A<br>Population-Level Analysis. Otolaryngology--head and neck surgery : official journal of<br>American Academy of Otolaryngology-Head and Neck Surgery. 2014.                                                                                                                                                                                                                                                                                                                      |
| 3  | Chung TK, Rosenthal EL, Magnuson JS, Carroll WR. Transoral robotic surgery for<br>oropharyngeal and tongue cancer in the United States. The Laryngoscope. 2014.                                                                                                                                                                                                                                                                                                                                                                                                 |
| 4  | Clayburgh D, Stott W, Bolognone R, et al. A randomized controlled trial of corticosteroids for<br>pain after transoral robotic surgery. The Laryngoscope. 2017.                                                                                                                                                                                                                                                                                                                                                                                                 |
| 5  | Frenkel CH, Yang J, Zhang M, Altieri MS, Telem DA, Samara GJ. Compared Outcomes of<br>Concurrent versus Staged Transoral Robotic Surgery with Neck Dissection. Otolaryngology--<br>head and neck surgery : official journal of American Academy of Otolaryngology-Head and<br>Neck Surgery. 2017:194599817706499.                                                                                                                                                                                                                                               |
| 6  | Goel AN, Badran KW, Mendelsohn AH, et al. Readmission after surgery for oropharyngeal<br>cancer: An analysis of rates, causes, and risk factors. The Laryngoscope. 2018.                                                                                                                                                                                                                                                                                                                                                                                        |
| 7  | Lee SY, Park YM, Byeon HK, Choi EC, Kim SH. Comparison of oncologic and functional<br>outcomes after transoral robotic lateral oropharyngectomy versus conventional surgery for T1 to<br>T3 tonsillar cancer. Head & neck. 2013;36(8):1138-1145.                                                                                                                                                                                                                                                                                                                |
| 8  | More YI, Tsue TT, Girod DA, et al. Functional swallowing outcomes following transoral robotic<br>surgery vs primary chemoradiotherapy in patients with advanced-stage oropharynx and<br>supraglottis cancers: Comment. Dysphagia. 2013;28(4):593-594.                                                                                                                                                                                                                                                                                                           |
| 9  | Motz K, Chang HY, Quon H, Richmon J, Eisele DW, Gourin CG. Association of Transoral<br>Robotic Surgery With Short-term and Long-term Outcomes and Costs of Care in Oropharyngeal<br>Cancer Surgery. JAMA otolaryngology-- head & neck surgery. 2017.<br>Motz K, Chang H-Y, Quon H, Richmon J, Eisele DW, Gourin CG. Association of transoral<br>robotic surgery with short-term and long-term outcomes and costs of care in oropharyngeal<br>cancer surgery. Supplementary Online Content. JAMA Otolaryngol Head Neck Surg. Published<br>online March 30, 2017. |
| 10 | Parhar HS, Gausden E, Patel J, et al. Analysis of readmissions after transoral robotic surgery for<br>oropharyngeal squamous cell carcinoma. Head Neck. 2018.                                                                                                                                                                                                                                                                                                                                                                                                   |
| 11 | Richmon J, Quon H, Gourin CG. The effect of transoral robotic surgery on short-term outcomes<br>and cost of care after oropharyngeal cancer surgery. The Laryngoscope. 2013.                                                                                                                                                                                                                                                                                                                                                                                    |
| 12 | Smith RV, Schiff BA, Garg M, Haigentz M. The impact of transoral robotic surgery on the<br>overall treatment of oropharyngeal cancer patients. The Laryngoscope. 2015;125 Suppl 10:S1-<br>s15.                                                                                                                                                                                                                                                                                                                                                                  |
| 13 | Zevallos JP, Mitra N, Swisher-McClure S. Patterns of care and perioperative outcomes in<br>transoral endoscopic surgery for oropharyngeal squamous cell carcinoma. Head Neck. 2014.                                                                                                                                                                                                                                                                                                                                                                             |

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K182371](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY/K182371)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
