← Product Code [OCX](/productcode/OCX) · K182319

# Glo-Tip Spray Catheter (K182319)

_Wilson-Cook Medical, Inc. · OCX · Feb 1, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182319

## Device Facts

- **Applicant:** Wilson-Cook Medical, Inc.
- **Product Code:** [OCX](/productcode/OCX.md)
- **Decision Date:** Feb 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Glo-Tip® Spray Catheter is used in conjunction with an endoscope to spray the gastrointestinal mucosa.

## Device Story

The Glo-Tip® Spray Catheter is a sterile, single-use endoscopic accessory. It consists of a long, thin PTFE tube with a metal stylet and a spray tip. The device is inserted through the accessory channel of an endoscope into the GI tract. A clinician injects fluids (e.g., water, saline, or diagnostic dyes) through the catheter to spray the mucosa. This facilitates cleaning of the tissue or aids in disease detection via chromoendoscopy. The device features a tantalum radiopaque crimping band for fluoroscopic visibility. It is intended for use in clinical settings by trained medical professionals.

## Clinical Evidence

Bench testing only. Testing included sterilization validation (SAL 10^-6), biocompatibility (ISO 10993-5 MEM elution, ISO intracutaneous irritation, ISO guinea pig maximization), shelf-life, and functional bench testing (dimensional accuracy, fluoroscopic visibility, packaging integrity, and simulated shipping).

## Technological Characteristics

Materials: PTFE catheter, polycarbonate proximal fitting, tantalum radiopaque band, 303 stainless steel and brass spray tip. Dimensions: 6.9 Fr (2.3 mm) diameter, 235 cm length. Compatible with 2.8 mm endoscope channels. Sterilization: Ethylene Oxide (EO). Connectivity: None. Software: None.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Hobbs Mistifier Medical Endoscopic Spray Catheter ([K030765](/device/K030765.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

February 1, 2019

Wilson-Cook Medical, Inc. % Daniel Dillon Senior Regulatory Scientist MED Institute 1330 Win Hentschel Boulevard West Lafayette, Indiana 47906

Re: K182319/S001 Trade/Device Name: Glo-Tip® Spray Catheter Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCX Dated: January 24, 2019 Received: January 24, 2019

Dear Daniel Dillon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Daniel G. Walter Jr -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K182319

Device Name Glo-Tip® Spray Catheter

Indications for Use (Describe)

The Glo-Tip® Spray Catheter is used in conjunction with an endoscope to spray the gastrointestinal mucosa.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background. The red background is shaped like a rectangle with a diagonal cut on the bottom left corner.

#### 510(k) Summary 001.

# Glo-Tip® Spray Catheter

Traditional 510(k) Premarket Notification October 1, 2018

#### Applicant Information

| Applicant: | Wilson-Cook Medical, Inc. /Cook Endoscopy<br>4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105 |
|------------|----------------------------------------------------------------------------------------------------------------|
| Contact:   | Paula Joyce, Vice President, Regulatory Affairs\Corporate<br>Compliance Officer                                |
| Phone:     | (336) 744-0157 ext. 396245                                                                                     |
| Fax:       | (336) 201-5994                                                                                                 |

#### Device Information

| Trade Name:          | Glo-Tip® Spray Catheter   |
|----------------------|---------------------------|
| Common Names:        | Spray catheter            |
| Classification Name: | Endoscope and accessories |
| Regulation Number:   | 21 CFR 876.1500           |
| Product Code:        | OCX                       |
| Device Class:        | Class II                  |
| Review Panel:        | Gastroenterology-Urology  |

#### Predicate Device

| Name:          | Hobbs Mistifier Medical Endoscopic Spray Catheter |
|----------------|---------------------------------------------------|
| 510(k) Number: | K030765                                           |
| Date:          | Cleared June 9, 2003                              |

#### Device Description

The Glo-Tip® Spray Catheter (subject device) is a sterile, single use device compatible with the accessory channel of an endoscope. The device consists of a long, thin polytetrafluoroethylene (PTFE) tube (catheter) with a stylet (metal wire) that together are introduced into the ductal area or elsewhere in the GI tract. Fluid (e.g., water or saline) can be injected through the catheter to spray the tissue that needs cleaning. The Glo-Tip®

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in large, white, sans-serif font at the top, and the word "MEDICAL" is in smaller, white, sans-serif font at the bottom.

Spray Catheter can also be used in chromoendoscopy (to apply a dye to the mucosa of the gastrointestinal tract through an endoscope to aid in disease detection).

#### Indications for Use

The Glo-Tip® Spray Catheter is used in conjunction with an endoscope to spray the gastrointestinal mucosa.

#### Comparison to Predicate Device

The subject device and predicate device have the same intended use and different technological characteristics, as follows:

| Characteristic           | Hobbs Mistifier Medical<br>Endoscopic Spray Catheter<br>(K030765)                                                            | Glo-Tip® Spray Catheter<br>(Subject of This 510(k))                               |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Intended Use             | Used as a flexible endoscopic<br>accessory to apply legally marketed<br>solutions for washing or staining<br>mucosal tissue. | Used in conjunction with an<br>endoscope to spray the<br>gastrointestinal mucosa. |
| Design                   |                                                                                                                              |                                                                                   |
| Catheter working length  | 165 cm or 260 cm                                                                                                             | 235 cm                                                                            |
| Catheter diameter        | 2.6 mm                                                                                                                       | 6.9 Fr (2.3 mm)                                                                   |
| Compatible scope channel | Unknown                                                                                                                      | 2.8 mm                                                                            |
| Spray tip                | Forward-facing, 360°                                                                                                         | Same                                                                              |
| Materials                |                                                                                                                              |                                                                                   |
| Catheter                 | PTFE                                                                                                                         | PTFE                                                                              |
| Proximal-fitting         | Unknown                                                                                                                      | Polycarbonate                                                                     |
| Radiopaque crimping band | Unknown                                                                                                                      | Tantalum                                                                          |
| Spray tip                | "Metal"                                                                                                                      | 303 Stainless Steel and Brass                                                     |
| Packaging                | Unknown                                                                                                                      | Tyvek/Mylar Pouch                                                                 |
| Other                    |                                                                                                                              |                                                                                   |
| Injection Port?          | Yes                                                                                                                          | Same                                                                              |
| Sterile?                 | Yes                                                                                                                          | Same                                                                              |
| Single Use?              | Yes                                                                                                                          | Same                                                                              |

None of the differences in technological characteristics raise different questions of safety and effectiveness. Furthermore, performance data from acceptable scientific methods provide evidence that the subject device is substantially equivalent to the predicate device.

{5}------------------------------------------------

K182319 page 3 of 4

Image /page/5/Picture/1 description: The image is a logo for Cook Medical. The logo consists of the word "COOK" in white, bold, sans-serif font, with the registered trademark symbol next to it. Below the word "COOK" is the word "MEDICAL" in white, bold, sans-serif font. The background is a solid red color.

COOK ENDOSCOPY 0 BETHANIA STAT ON-SALEM, NC 27 TOLL FREE: 800 WWW.COOKMEDICAL

### Performance Data

Performance testing consisting of sterilization, shelf life, biocompatibility, and nonclinical bench testing demonstrates that the Glo-Tip® Spray Catheter meets the performance requirements to fulfill the intended use of the device.

Sterilization validation demonstrates that the device has a sterility assurance level of 10° with acceptable sterilization residuals. Biocompatibility testing is summarized as follows:

| Test method                                                  | Results                                      |
|--------------------------------------------------------------|----------------------------------------------|
| ISO 1X MEM Elution Method                                    | Test article considered<br>not cytotoxic.    |
| ISO Intracutaneous Irritation Study in Rabbits: two extracts | Test article considered<br>a non-irritant.   |
| ISO Guinea Pig Maximization: two extracts                    | Test article considered<br>not a sensitizer. |

| Requirement                                                                                                           | Acceptance Criteria (summary)                                                                                                                                                            | Results                                                  |
|-----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Packaging shall maintain<br>functional integrity of the device<br>following EO sterilization and<br>shipping.         | Meet user requirements and have<br>acceptable visual appearance<br>following two sterilization cycles<br>and simulated shipping conditions.                                              | Packages met acceptance<br>criteria.                     |
| Product must perform intended<br>use over labeled shelf life.                                                         | Same requirements as above,<br>except following two sterilization<br>cycles and aging.                                                                                                   | Product performed as intended<br>over labeled shelf life |
| Sterile product must be<br>adequately packaged to maintain<br>sterile barrier during expected<br>shipping conditions. | Non-opened sterile packaging<br>shall not show any signs of<br>functional damage, such as holes<br>in the pouch and/or voids in the<br>seal area after simulated shipping<br>conditions. | No signs of functional<br>damage.                        |
| Package maintains sterile barrier<br>over labeled shelf life.                                                         | Same requirements as above<br>previous, except following aging.                                                                                                                          | Package has acceptable sterile<br>barrier.               |
| Device must be a Glo-tip device.                                                                                      | Distal tip of device must be<br>fluoroscopically visible.                                                                                                                                | Tip is fluoroscopically visible.                         |

Shelf-life and non-clinical bench testing is summarized as follows:

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in large, white, sans-serif font, and below it, the word "MEDICAL" is in a smaller, white, sans-serif font.

cook endoscopy 4900 bethania station road WINSTON-SALEM, NC 27105 U.S.A. PHONE: 336.744.0157 TOLL FREE: 800.245.4707 WWW.COOKMEDICAL.COM

| Requirement                                                                                                                                                       | Acceptance Criteria (summary)                                                                                                                                             | Results                                               |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| The product label shall contain<br>accurate dimensional/<br>compatibility information                                                                             | Catheter outer diameter: 6.9 Fr<br>with acceptable tolerance.                                                                                                             | Device dimensions are in<br>accordance with labeling. |
|                                                                                                                                                                   | Working length: 235 with<br>acceptable tolerance.                                                                                                                         |                                                       |
| In accordance with ISO<br>8600-1:2013, Clause 4.3: The<br>maximum insertion portion<br>width shall not be larger than<br>that stated in the instruction<br>manual | Maximum insertion portion width<br><2.6 mm when measured in<br>accordance with ISO 8600-4:<br>2014, using a measuring<br>instrument with a minimum<br>accuracy of 0.05 mm |                                                       |

---

**Source:** [https://fda.innolitics.com/device/K182319](https://fda.innolitics.com/device/K182319)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).