← Product Code [QJQ](/productcode/QJQ) · K182311

# Cervella (K182311)

_Innovative Neurological Devices, LLC · QJQ · Mar 7, 2019 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182311

## Device Facts

- **Applicant:** Innovative Neurological Devices, LLC
- **Product Code:** [QJQ](/productcode/QJQ.md)
- **Decision Date:** Mar 7, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5800
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Cervella Cranial Electrotherapy Stimulator (CES) is indicated for treatment of insomnia, depression, or anxiety.

## Device Story

Cervella is a Cranial Electrotherapy Stimulator (CES) delivering small pulses of electrical current to the brain via conductive electrodes placed bilaterally on the mastoid process. Electrodes are integrated into stereo over-ear noise-canceling headphones. The device is powered by an internal rechargeable Li-ion battery. Operation is controlled via a smartphone app communicating via Bluetooth LE; the app allows users to adjust stimulation intensity, frequency, and duration. The device includes a separate Bluetooth connection for audio/music playback. Treatment history is automatically stored in the app for healthcare provider review. The device is intended for patient use to manage insomnia, depression, or anxiety, potentially benefiting patients by providing a non-pharmacological treatment option with integrated audio features for comfort during sessions.

## Clinical Evidence

Bench testing only. Conformance to ES60601-1:2005/(R)2012, IEC 60601-1-2 Edition 4.0, IEC 60601-2-10 Edition 2.1, and IEC 62133 Edition 2.0 was confirmed.

## Technological Characteristics

Plastic enclosure; Li-ion battery (3.7V) power source; symmetrical biphasic square wave output; 0-500µA current intensity; 0.5Hz, 1.5Hz, 100Hz frequency selections; Bluetooth LE connectivity for app control; Bluetooth for audio. Electrodes placed on mastoid process. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and IEC 62133.

## Regulatory Identification

A cranial electrotherapy stimulator is a prescription device that applies electrical current that is not intended to induce a seizure to a patient's head to treat psychiatric conditions.

## Special Controls

In combination with the general controls of the FD&C Act, the virtual reality behavioral therapy device for pain relief is subject to the following special controls:

*Classification.* (1) Class II (special controls) when intended to treat insomnia and/or anxiety. The special controls for this device are:(i) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device to treat insomnia and/or anxiety.
(ii) Components of the device that come into human contact must be demonstrated to be biocompatible.
(iii) The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC) in its intended use environment.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The technical parameters of the device, including waveform, output mode, pulse duration, frequency, train delivery, maximum charge, and energy, must be fully characterized and verified.
(vi) The labeling for the device must include the following:
(A) The intended use population and the intended use environment;
(B) A warning that patients should be monitored by their physician for signs of worsening;
(C) A warning that instructs patients on how to mitigate the risk of headaches, and what to do should a headache occur;
(D) A warning that instructs patients on how to mitigate the risk of dizziness, and what to do should dizziness occur;
(E) A detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;
(F) Instructions for use that address where to place the electrodes, what stimulation parameters to use, and duration and frequency of treatment sessions. This information must be based on the results of clinical studies for the device;
(G) A detailed summary of the device technical parameters, including waveform, output mode, pulse duration, frequency, train delivery, and maximum charge and energy; and
(H) Information on validated methods for reprocessing any reusable components between uses.
(vii) Cranial electrotherapy stimulator devices marketed prior to the effective date of this reclassification must have an amendment submitted to the previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
(2) Class III (premarket approval) when intended to treat depression.
(c) Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 19, 2020, for any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 19, 2020, been found to be substantially equivalent to any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- CES Ultra ([K062284](/device/K062284.md))
- Alpha-Stim CS ([K903014](/device/K903014.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 7, 2019

Innovative Neurological Devices LLC Bart Waclawik President 13295 Illinois Street, Suite 312 Carmel, Indiana 46032

Re: K182311

Trade/Device Name: Cervella Regulation Number: 21 CFR 882.5800 Regulation Name: Cranial Electrotherapy Stimulator Regulatory Class: Class III Product Code: JXK Dated: February 1, 2019 Received: February 4, 2019

Dear Bart Waclawik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Pamela D. Scott Digitally signed by Pamela D.
Scott Scott Scott Scott -s Date: 2019.03.07 22:01:53 -5 -05'00 for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183211

Device Name Cervella

Indications for Use (Describe)

Cervella Cranial Electrotherapy Stimulator (CES) is indicated for treatment of insomnia, depression, or anxiety.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR § 807.92(c), 510(k) summary is provided.

#### l. DATE PREPARED: March 5, 2019

### II. SUBMITTER

INNOVATIVE NEUROLOGICAL DEVICES LLC 13295 Illinois St, Suite 312 Carmel, IN 46032 (855) 413-3300 Phone: Contact Person: Bart Waclawik

### III. OFFICIAL CORRESPONDENCE/CONTACT PERSON

Bart Waclawik President | CEO INNOVATIVE NEUROLOGICAL DEVICES LLC Phone: (855) 413-3300 e-mail: regulatory@cervella.us

#### IV. DEVICE

| Brand Name of Device: | Cervella™                                                                                    |
|-----------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Cranial Electrotherapy Stimulator (CES)                                                      |
| Classification Name:  | Stimulator, Cranial, Electrotherapy, a preamendment<br>Class III device per 21 CFR §882.5800 |
| Regulatory Class:     | III                                                                                          |
| Product Code:         | JXK                                                                                          |

## V. PREDICATE DEVICES

| Device Trade Name: | CES Ultra™                   |
|--------------------|------------------------------|
| Device Company:    | Neuro-Fitness, LLC           |
| 510(k) Numbers:    | K062284                      |
| Device Trade Name: | Alpha-Stim CS                |
| Device Company:    | Electromedical Products Inc. |
| 510(k) Numbers:    | K903014                      |

{4}------------------------------------------------

## VI. DEVICE DESCRIPTION

Cervella Cranial Electrotherapy Stimulator (CES) is a device that delivers small pulses of electrical current through patient's brain. The stimulator is powered by an internal rechargeable battery which provides a low-level constant current to the cranium via a pair of conductive electrodes placed bilaterally on the mastoid process. The electrodes are incorporated into earpads of stereo over-ear noisecancelling headphones. The patient can use the audio (through a separate dedicated Bluetooth connection) and noise cancelling features of the stereo headphones during treatment (e.g. music listening). The stimulator is controlled via a software application (app) installed on the patient's smart device (e.g. smartphone). The app communicates with the device through a dedicated Bluetooth® LE connection that is independent of the optional Bluetooth connection for audio listening purposes. The patient uses the app to adjust the intensity level, frequency, and duration of the treatment. The app also allows the patient to automatically store the treatment history for review by the patient's healthcare provider.

#### INDICATIONS FOR USE VII.

Cervella is indicated for the treatment of insomnia, depression, or anxiety.

#### SUBSTANTIAL EQUIVALENCE VIII.

The Cervella CES is substantially equivalent with respect to indications for use, stimulation parameters (i.e. current levels, frequencies, pulse width and amplitude), and electrode placement to predicate devices CES Ultra cleared by K062284 and Alpha-Stim CS cleared by K903014.

## IX. PERFORMANCE DATA DEMONSTRATING SUBSTANTIAL EQUIVALENCE

The Cervella device has the same intended use and operating principles, with similar design features, and functional and performance characteristics as the previously-cleared devices. Cervella is designed to comply with relevant federal and international safety and performance standards.

Cervella has undergone standard engineering bench testing to confirm conformance to design specifications as well as independent laboratory testing to insure conformance with applicable mandatory and voluntary medical device safety standards as follows:

{5}------------------------------------------------

- ES60601-1:2005/(R)2012 and A1:2012 ●
- IEC 60601-1-2 Edition 4.0 2014-02
- IEC 60601-2-10 Edition 2.1 2016-04
- IEC 62133 Edition 2.0 2012-1.

Conformance to mandatory and voluntary standards, combined with no changes in the Indications for Use and no change in the fundamental scientific technology demonstrates substantial equivalence to the predicate devices.

| Property                      | Cervella                                                                                    | CES Ultra                                                                                    | Alpha-Stim CS                                                                                    | Differences and<br>Comments                                      |
|-------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Indications<br>for Use        | Cervella is<br>indicated for the<br>treatment of<br>insomnia,<br>depression, or<br>anxiety. | CES Ultra is<br>indicated for the<br>treatment of<br>insomnia,<br>depression, or<br>anxiety. | Alpha-Stim CS is<br>indicated for the<br>treatment of<br>insomnia,<br>depression, or<br>anxiety. | None                                                             |
| Waveform                      | Symmetrical<br>Biphasic<br>Square Wave                                                      | Symmetrical Biphasic<br>Square Wave                                                          | Symmetrical<br>Biphasic Square<br>Wave                                                           | None                                                             |
| Current<br>Intensity<br>Range | 0µA – 500µA<br>adjustable in<br>50µA<br>increments                                          | 0µA — 1500µA<br>continually adjustable                                                       | 0µA — 500µA<br>adjustable in 50μA<br>increments                                                  | Same range as<br>Alpha-Stim,<br>narrower range<br>than CES Ultra |
| Pulse Width<br>Range          | 5ms - 1s                                                                                    | 2ms                                                                                          | 250ms - 1s                                                                                       | Pulse width varies<br>depending on<br>frequency<br>selection     |
| Number of<br>electrodes       | Two                                                                                         | Two                                                                                          | Two                                                                                              | None                                                             |
| Electrode<br>Placement        | Head (Mastoid<br>process)                                                                   | Head (Mastoid<br>process or Earlobes)                                                        | Head (Earlobes)                                                                                  | Same placement<br>as CES Ultra                                   |
| Power<br>Source               | Li-ion Battery<br>(3.7V)                                                                    | Alkaline Battery (9V)                                                                        | Alkaline Battery<br>(3V)                                                                         | Cervella battery is<br>rechargeable                              |

# X. SUMMARY OF SIMILARITIES AND DIFFERENCES

{6}------------------------------------------------

# INNOVATIVE NEUROLOGICAL DEVICES LLC | Premarket Notification 510(k) Submission Under 21 CFR § 807.87 for Cervella

| Frequency<br>selections                                                                                           | 0.5Hz, 1.5Hz,<br>100Hz                                              | 100Hz                           | 0.5Hz, 1.5Hz, 100Hz                 | Same frequency<br>selection as<br>Alpha-Stim CS |
|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------|-------------------------------------|-------------------------------------------------|
| Treatment<br>Range                                                                                                | 10 min – 60 min<br>in 10 min<br>intervals                           | 30 min, 60min, or<br>continuous | 10, 20, 60 minutes<br>or continuous | Same min-max<br>range as Alpha-<br>Stim         |
| Unit<br>Controls                                                                                                  | Via Bluetooth-<br>enabled smart<br>device featuring<br>Cervella app | Built into the device           | Built into the device               | Wireless<br>operation                           |
| Treatment<br>history<br>logging                                                                                   | Automatic                                                           | Not available                   | Not Available                       | New feature                                     |
| Ability for<br>patient to<br>use audio<br>and noise<br>cancelling<br>feature of<br>headset<br>during<br>treatment | Optional via<br>dedicated<br>separate<br>Bluetooth<br>connection    | Not available                   | Not Available                       | New feature                                     |
| Dimensions<br>and weight                                                                                          | 7cmx7cmx2cm                                                         | 13.5cmx6.4cmx3.3cm              | 9.8cmx6.3cmx2cm                     | Cervella is slightly<br>smaller                 |
| Enclosure                                                                                                         | Plastic                                                             | Plastic                         | Plastic                             | None                                            |

# XI. CONCLUSIONS

Cervella CES is substantially equivalent to the listed predicate device without raising any new issues of safety or effectiveness. The new device has the same intended use and operating principles, with similar features, and functional and performance characteristics as the predicates. The device was designed to comply with relevant federal and international safety and performance standards and has been tested to ensure conformity with all applicable medical device safety standards.

---

**Source:** [https://fda.innolitics.com/device/K182311](https://fda.innolitics.com/device/K182311)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
