K182296 · Parkell, Inc. · EBF · Dec 14, 2018 · Dental
Device Facts
Record ID
K182296
Device Name
HyperFil-LV
Applicant
Parkell, Inc.
Product Code
EBF · Dental
Decision Date
Dec 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HyperFil®-LV is a dual-cure, tooth shade, resin composite which is indicated for Class I, II, III, IV, and V restorations, including root surface cavities, post cementing, and core build-ups.
Device Story
HyperFil-LV is a dual-cure, radiopaque, flowable, bulk-fill resin composite restorative material. It is supplied in dual-chamber 5 mL syringes with mixing tips. The device is used by dental professionals in a clinical setting for direct and indirect restorations. It functions via light-curing (430-480 nm) or self-curing at mouth temperature. The material provides high compressive, tensile, and flexural strength to restore tooth structure. It is bioactive and intended to replace or repair damaged tooth tissue, aiding in structural integrity and aesthetics.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical properties (compressive, tensile, flexural strength, depth of cure, modulus of elasticity, radiopacity, water sorption, and solubility) and biocompatibility evaluations per ISO 10993 standards.
Technological Characteristics
Dual-cure resin composite; flowable viscosity. Materials: EBPADMA, TEGDMA, UDMA, 1,6-Hexanediol dimethacrylate, silica-barium filler. Standards: ANSI/ADA No. 27, ISO 4049:2009, ISO/TS 14569-1:1999, BS ISO 23317:2014, ISO 10993 series. Energy: Light-cure (430-480 nm, >600 mW/cm²) or self-cure. Packaging: Dual-chamber 5 mL syringe with mixing tip. Non-sterile.
Indications for Use
Indicated for Class I, II, III, IV, and V dental restorations, including root surface cavities, post cementing, and core build-ups in patients requiring restorative dental procedures.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
Composite Resin Core Material (K010475) (now marketed as Absolute Dentin®)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2018
Parkell, Inc. David Mott VP, Legal & Regulatory Affairs 300 Executive Drive Edgewood, New York 11717
Re: K182296
Trade/Device Name: HyperFil-LV Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: September 18, 2018 Received: September 20, 2018
Dear David Mott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S.
Digitally signed by
Mary S. Runner -S3
Runner -S3 Date: 2018.12.14
20:03:41 -05'00'
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K182296
Device Name HyperFil®-LV
Indications for Use (Describe)
HyperFil®-LV is a dual-cure, tooth shade, resin composite which is indicated for Class I, II, III, IV, and V restorations, including root surface cavities, post cementing, and core build-ups.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### 1. Submitter:
Parkell, Inc. 300 Executive Drive Edgewood, NY 11717
#### 2. Submitter
Parkell, Inc. 300 Executive Drive Edgewood, NY 11717 Phone: 631-389-1545 Fax: 631-389-1546 dmott@parkell.com
Contact Person: David Mott, Vice President - Legal & Regulatory Date of Original Submission: August 23, 2018 Date of Latest Revision: December 13, 2018
#### 3. Type of Submission: Traditional 510(k).
#### 4. Device
| Device Proprietary Name: | HyperFil®-LV |
|--------------------------|----------------------------|
| Common or Usual Name: | Tooth shade resin material |
| Classification Name: | Tooth shade resin material |
| Regulation Number: | 21 CFR 872.3690 |
| Product Code: | EBF |
| Device Classification | II |
#### 5. Primary Predicate
Celerity (K062518, Parkell, Inc.)
#### 6. Reference Device
Composite Resin Core Material (K010475, Parkell, Inc.) (now marketed by Parkell under the trade name "Absolute Dentin®")
#### 7. Device Description
HyperFil®-LV is a dual-cure, highly radiopaque, flowable, bulk-fill composite restorative material suitable for direct and indirect restorations. HyperFil®-LV is light-curable to a depth of about 3-4 mm in 40 seconds with a dental curing light emitting at 430 - 480 nm with an intensity of >600 mW/cm². It will also self-cure to any depth in about 4-5 minutes at mouth temperature. HyperFil®-LV contains a combination of filling agents which give restorations constructed with HyperFil®-LV high compressive, tensile, and flexural strength.
Image /page/3/Picture/17 description: The image displays the logo for "DirectaDentalGroup" in a large, white, sans-serif font. Below the main logo, there is a line and the text "STOCK SUMMARY". Underneath this, there are four smaller logos: "DIRECTA directa dental.com", "topdental", "ORSING", and "parkell". The background is a dark blue color.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "parkell" in a bold, sans-serif font. To the right of "parkell" is the word "inc." in a smaller, italicized font. The text is black and the background is white.
HyperFil®-LV will be packaged in single dual-chamber 5 mL syringes. The device comprises catalyst and base components, which are combined just prior to use via a mixing tip which is affixed to the dual chamber syringe.
## 8. Indications for Use
HyperFil®-LV is a dual-cure, tooth shade, resin composite which is indicated for Class I, III, IV, and V restorations, including root surface cavities, post cementing, and core build-ups.
## 9. Comparison of Technological Characteristics
The standards applicable to the subject device and to the comparison of technological characteristics referenced within the submission, include: ADSI/ADA Specification No. 27: Resin-Based Filling Materials, ISO 4049:2009 Dentistry – Polymer Based Restorative Materials, ISO/TS 14569-1:1999 Dental Materials – Guidance on testing of wear resistance – Part 1: Wear by tooth brushing, and BS ISO 23317: 2014 Implants for Surgery – In vitro evaluation for apatite-forming ability of implant materials.
Information provided in this 510(k) submission demonstrates that HyperFil®-LV is substantially equivalent to the Primary Predicate, Parkell's Celerity (K062518) (now marketed by Parkell under the trade name, HyperFil® dual-cure restorative composite) in terms of critical properties for restorative composites (e.g., compressive, tensile, and flexural strength), wherein HyperFil® LV differs from the Primary Predicate in that has a lower viscosity, making it more flowable than the Primary Predicate, and is bioactive as is demonstrated within this 510(k) submission.
Any differences between the Device and the Primary Predicate, for purposes of this 510(k) submission, do not raise new questions of effectiveness or safety. Thus, this submission demonstrates the substantial equivalence of HyperFil®-LV to the Primary Predicate. A brief comparison of HyperFil®-LV to the Primary Predicate is provided in Table 5-A below.
In addition, information has been included regarding a Reference Device (namely, "Composite Resin Core Material", K010475, Parkell, Inc., which is now marketed by Parkell under the trade name "Absolute Dentin®") in order to address any differences in water sorption and water solubility that may exist between the Device and Primary Predicate.
| Property<br>(pass/fail criteria, if<br>applicable) | HyperFil®-LV<br>(Parkell, Inc.)<br>(K182296) | Primary Predicate<br>(Celerity)<br>(K062518, Parkell, Inc.) |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended uses | A dual-cure, tooth shade, resin<br>composite which is indicated for<br>Class I, II, III, IV, and V restorations,<br>including root surface cavities, post<br>cementing, and core build-ups. | A dual-cure, tooth shade, resin<br>composite for the restoration of<br>Class I, II, III, IV, and V cavities in<br>permanent and deciduous teeth. |
#### Table 5-A:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "parkell" in bold, black letters. To the right of "parkell" and slightly below it, the word "inc." is written in a smaller font size. The font of the word "inc." is also more cursive than the font of the word "parkell."
#### Table 5-A (continued):
| | HyperFil®-LV | Primary Predicate |
|------------------------------------|----------------------------------------------|--------------------------------------------------|
| | (Parkell, Inc.)<br>(K182296) | (Celerity)<br>(K062518, Parkell, Inc.) |
| Classification Product<br>Code | EBF | EBF |
| Regulation Number | 21 CFR 872.3690 | 21 CFR 872.3690 |
| Principle of operation | dual cured resin material | dual cured resin material |
| Packaging | dual-barrel syringe (5mL) with<br>mixing tip | dual-barrel cartridge (10 mL) with<br>mixing tip |
| Dual-cure | yes | yes |
| Radiopaque | yes | yes |
| Sterile | no | no |
| Compressive Strength<br>(>300 MPa) | >300 MPa | >300 MPa |
| Flexural Strength<br>(>100 MPa) | >100 MPa | >100 MPa |
| Tensile Strength<br>(>40 MPa) | >40 MPa | >40 MPa |
| Depth of Cure<br>(>1 mm) | >1 mm | >1 mm |
| Modulus of Elasticity<br>(>10 GPa) | >10 GPa | >10 GPa |
| Radiopacity (>200%) | >200% | >200% |
| Water Sorption<br>(<40 µg/mm³) | <40 µg/mm³ | <40 µg/mm³ |
| Water Solubility<br>(<7.5 µg/mm³) | <7.5 µg/mm³ | <7.5 µg/mm³ |
| Chemical Components | EBPADMA | EBPADMA |
| | TEGDMA | TEGDMA |
| | UDMA | UDMA |
| | 1,6-Hexanediol dimethacrylate | |
| | | Tri-EDMA |
| | Silica-barium Filler | Silica-barium Filler |
As seen above, the differences between the Device and primary predicate are limited, e.g., to minor differences in chemical components and in packaging. These technological differences do not raise different questions of effectiveness or safety and are addressed by comparative performance data addressed within this submission. Minor differences also exist in the water solubility and water sorption values between the Device and Primary Predicate, though all values meet specifications defined for restoratives in ISO 4049:2009 and information is provided for the Reference Device which is directed to water solubility and sorption. In addition, minor differences exist in the indications-for-use language between the Device and primary predicate. Namely, the indications-for-use language for the Device identifies three common procedures
{6}------------------------------------------------
(namely, root surfaces cavities, post cementing, and core build-ups) which are well known in the art to be encompassed or associated with Class I, II, III, IV, and V restorations (for which the primary predicate is cleared). As such, and for this reason, these differences do not impact the safety and effectiveness of the device.
## 10. Biocompatibility
An evaluation of biocompatibility was conducted to determine the safety of HyperFil®-LV in accordance with ISO 10993-1: 2009, ISO 10993-5: 2009, ISO 10993-5: 2009, ISO 10993-11:2017, ISO 10993-12: 2012, and the Final FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (June 16, 2016). The conclusion of the evaluation is that HyperFil®-LV can be considered biocompatible.
## 11. Clinical Performance Data
There was no clinical testing required to support the Device as the intended uses are equivalent to the Primary Predicate. These types of devices, including the Primary Predicate, have been on the market for many years with no reported adverse events. The non-clinical testing detailed in this submission supports the substantial equivalence of the Device.
## 12. Shelf Life
An evaluation of the shelf life of the Device was conducted following real-time storage of the Device at room temperature, as well as after storage under accelerated aging conditions, via physical properties testing in accordance with ISO 4049:2009 Dentistry – Polymer Based Restorative Materials and ASTM D2583-95: Standard Test Method for Indentation Hardness of Rigid Plastics by Means of a Barcol Impressor.
## 13. Statement of Substantial Equivalence:
The information provided above supports that HyperFil --LV is substantially equivalent to the primary predicate. Although minor differences in design and technology exist between the subject and primary predicate, the testing supports that these differences do not raise any new questions of effectiveness or safety. Therefore, it is concluded that HyperFil" -LV is substantially equivalent to the primary predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.