← Product Code [NUH](/productcode/NUH) · K182267

# PAT (Pain Alleviation Technologies) (K182267)

_Marchan Medical, LLC · NUH · Feb 27, 2019 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182267

## Device Facts

- **Applicant:** Marchan Medical, LLC
- **Product Code:** [NUH](/productcode/NUH.md)
- **Decision Date:** Feb 27, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Temporary relief of pain associated with sore and aching muscles and joints due to strain from exercise or normal household and work activities.

## Device Story

PAT is an OTC transcutaneous electrical nerve stimulator (TENS) for temporary pain relief; used by adults in home, clinic, or hospital settings. Device consists of battery-operated enclosure with LCD screen and user buttons. It transmits monophasic micro-current waveforms to affected body parts via cable connected to skin surface electrodes. Features 8 non-programmable modes corresponding to different body areas. User selects mode; device delivers preset electrical pulses to stimulate nerves; intended to alleviate muscle/joint pain from strain. Provides non-invasive, drug-free pain management.

## Clinical Evidence

Bench testing only. No clinical data provided. Equivalence established via comparative bench testing of electrical output parameters (waveform, frequency, current, voltage), compliance with safety standards (AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-10), and verification of device functionality.

## Technological Characteristics

ABS plastic enclosure; monophasic square waveform (50% duty cycle); 3 Hz frequency; 40 μA amplitude; regulated current output. Powered by 3 replaceable AAA batteries. Includes LCD interface. Connectivity via 40-inch double snap cable to 5x5cm electrodes. Complies with AAMI ES 60601-1, IEC 60601-1-2, and IEC 60601-2-10. Type BF patient leakage current rating.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Painmaster MCT Patch ([K130114](/device/K130114.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 27, 2019

Marchan Medical, LLC Mark Chandler Managing Partner 165 S Ashe St. Southern Pines, North Carolina 28387

Re: K182267

Trade/Device Name: PAT (Pain Alleviation Technologies) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH Dated: January 18, 2019 Received: January 28, 2019

Dear Mark Chandler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Scott -S

Sincerely,
Pamela D. Digitally
D. Scott

Digitally signed by Pamela D. Scott -S Date: 2019.02.27 21:10:33 -05'00'

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K182267

Device Name PAT

Indications for Use (Describe)

Temporary relief of pain associated with sore and aching muscles and joints due to strain from exercise or normal household and work activities.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Marchan Medical LLC. The logo features a blue stylized letter "M" above the company name. Below the company name is the website address, marchanmedical.com.

## 510(K) Summary

PAT is an OTC medical device. This device is used for temporary relief of pain of sore and aching muscles and joints, due to strain from exercise or normal household and work activities. The device consists of a small battery operated enclosure with an LCD screen and user buttons to control the device. A monophasic micro-current waveform is transmitted to the affected body part through a cable connected via snaps to skin surface electrodes. There are 8 non-programmable modes representing 8 different body areas to treat.

| Sponsor:                                     | Marchan Medical, LLC<br>165 S Ashe St<br>Southern Pines, NC 28387 |
|----------------------------------------------|-------------------------------------------------------------------|
| Contact Person:                              | Mark Chandler, MD                                                 |
| Phone:                                       | 910-783-5285                                                      |
| Email:                                       | Markmd@marchanmedical.com                                         |
| Date of Original Submission: August 15, 2018 |                                                                   |
| Proprietary Name:                            | PAT                                                               |
| Common Name:                                 | Transcutaneous electrical nerve stimulator (TENS)                 |
| Classification Name:                         | Stimulator, nerve, Transcutaneous, Over-the-Counter               |
| Regulation Number:                           | 21 CFR 882.5890                                                   |
| Regulatory Class:                            | II                                                                |
| Product Code:                                | NUH                                                               |
| Classification Panel:                        | Neurology                                                         |
| Predicate Device:                            | Painmaster MCT Patch (K130114)                                    |

#### Indications for Use:

Temporary relief of pain associated with sore and aching muscles and joints due to strain from exercise or normal household and work activities.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Marchan Medical LLC. The logo features a blue stylized mountain design above the company name. Below the company name is the website address marchanmedical.com.

Basic Unit Characteristics

PAT uses Painmaster MCT Patch as the predicate device.

| Parameter                                                                                 | Painmaster MCT Patch<br>Predicate Device<br>(K130114)                                                                                                                                         | PAT<br>Proposed Device<br>(K182267)                                                                                                                                |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                                              | Newmark, Inc.                                                                                                                                                                                 | Shenzhen Bless Electronic<br>Technology Co., Ltd.                                                                                                                  |
| Specifications Developer                                                                  |                                                                                                                                                                                               | Marchan Medical, LLC                                                                                                                                               |
| Product Code                                                                              | NUH                                                                                                                                                                                           | NUH                                                                                                                                                                |
| Environment of use                                                                        | Clinics, hospital and home<br>environment                                                                                                                                                     | Clinics, hospital and home<br>environment                                                                                                                          |
| Indications for Use                                                                       | Temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>upper and lower<br>extremities due to strain<br>from exercise or normal<br>household and work<br>activities. | Temporary relief of pain<br>associated with sore and<br>aching muscles and<br>joints due to strain from<br>exercise or normal<br>household and work<br>activities. |
| Patient Population                                                                        | Adult                                                                                                                                                                                         | Adult                                                                                                                                                              |
| Operation Mode                                                                            | Waveform: monophasic<br>square wave with 50%<br>duty cycle.<br>Amplitude: 42 μA<br>Frequency: 0.5 Hz                                                                                          | Waveform: monophasic<br>square wave. with 50%<br>duty cycle.<br>Amplitude:40 μA<br>Frequency: 3 Hz                                                                 |
| Power Source                                                                              | Primary battery only.<br>One time use.                                                                                                                                                        | Primary battery only.<br>3 replaceable AAA<br>batteries, not<br>rechargeable.                                                                                      |
| Operational Indicator                                                                     | Yes, LED light                                                                                                                                                                                | Yes, LCD screen                                                                                                                                                    |
| Time Range                                                                                | Not adjustable, up to<br>300 hours                                                                                                                                                            | 40-55 minutes depending<br>on area treated                                                                                                                         |
| Housing Materials and<br>Construction                                                     | Plastic (ABS) enclosure                                                                                                                                                                       | Plastic (ABS) enclosure                                                                                                                                            |
| Body contact material                                                                     | Electrode, CE certified                                                                                                                                                                       | Electrode, FDA cleared                                                                                                                                             |
| Maximum Average Current                                                                   | 42 μΑ                                                                                                                                                                                         | 40 μΑ                                                                                                                                                              |
| OTC Use                                                                                   | Yes                                                                                                                                                                                           | Yes                                                                                                                                                                |
| Dimensions (in.)<br>W x H x D                                                             | 1.6 x 0.3 x 1                                                                                                                                                                                 | 6 x 2.75 x 1                                                                                                                                                       |
| Weight                                                                                    | 0.03 lb.                                                                                                                                                                                      | 0.25 lb.                                                                                                                                                           |
| Operating Temperature and<br>Humidity                                                     | 10-45°C, 20%-90%                                                                                                                                                                              | 10-45°C, 20%-90%                                                                                                                                                   |
| Method of Line Current<br>Isolation                                                       | N/A                                                                                                                                                                                           | N/A                                                                                                                                                                |
| Patient Leakage Current                                                                   | Type BF                                                                                                                                                                                       | Type BF                                                                                                                                                            |
| Normal Condition                                                                          | 0                                                                                                                                                                                             | 0                                                                                                                                                                  |
| Single Fault Condition                                                                    | 50 μΑ                                                                                                                                                                                         | 50 μΑ                                                                                                                                                              |
| Average DC current through<br>electrodes when device is on<br>but no pulses being applied | N/A (Pulse is always<br>applied when on)                                                                                                                                                      | 0                                                                                                                                                                  |
| Number of Output Modes                                                                    | 1                                                                                                                                                                                             | 1 (8 different timings<br>based on body part being<br>treated but only one<br>preset waveform)                                                                     |
| Number of Output Channels                                                                 | 1                                                                                                                                                                                             | 1                                                                                                                                                                  |
| Regulated Current of Voltage                                                              | Regulated Current                                                                                                                                                                             | Regulated Current                                                                                                                                                  |
| Software/Firmware/<br>Microprocessor Control?                                             | Yes                                                                                                                                                                                           | Yes                                                                                                                                                                |
| Automatic Overload Trip                                                                   | Yes                                                                                                                                                                                           | Yes                                                                                                                                                                |
| Automatic No-Load Trip                                                                    | Yes                                                                                                                                                                                           | Yes                                                                                                                                                                |
| Automatic Shut-Off                                                                        | Yes                                                                                                                                                                                           | Yes                                                                                                                                                                |
| Indicator<br>Display                                                                      | On/Off Status? Yes<br>Low Battery? No<br>Voltage/Current Level No                                                                                                                             | Yes                                                                                                                                                                |
| Compliance with Voluntary<br>Standards?                                                   | EN 60601-1<br>HA60601-1-11<br>IEC 60601-1-2<br>IEC 60601-2-10                                                                                                                                 | AAMI ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10                                                                                                                 |
| Compliance with 21 CFR 898                                                                | N/A                                                                                                                                                                                           | N/A                                                                                                                                                                |
| Electrode cable                                                                           | 4 Inch snap                                                                                                                                                                                   | 40 inch double snap,<br>3.5mm                                                                                                                                      |
| Parameter                                                                                 | Predicate Device                                                                                                                                                                              | Proposed Device                                                                                                                                                    |
| Product Name                                                                              | Painmaster MCT Patch                                                                                                                                                                          | PAT                                                                                                                                                                |
| 510(K) number                                                                             | K130114                                                                                                                                                                                       | K182267                                                                                                                                                            |
| Number of output modes                                                                    | 1                                                                                                                                                                                             | 1 (8 different timings based<br>on body part being treated but<br>only one preset waveform )                                                                       |
| Number of output channels                                                                 | 1                                                                                                                                                                                             | 1                                                                                                                                                                  |
| Waveform                                                                                  | Square monophasic,<br>50% duty cycle                                                                                                                                                          | Square monophasic,<br>50% duty cycle                                                                                                                               |
| Maximum output<br>Voltage (max) 500 Ω<br>2K Ω 10K Ω                                       | 500 Ω : 22mV<br>2K Ω: 84mV<br>10K Ω: 370mV                                                                                                                                                    | 500 Ω :20.8 mV<br>2K Ω: 82.0 mV<br>10K Ω: 416 mV                                                                                                                   |
| Maximum output<br>Current (max) 500 Ω<br>2K Ω 10K Ω                                       | 500 Ω : 42 μA<br>2K Ω: 42 μA<br>10K Ω: 37 μA                                                                                                                                                  | 500 Ω : 41.6 μA<br>2K Ω: 41.0 μA<br>10K Ω: 41.6 μA                                                                                                                 |
| Max. Phase charge @500 Ω<br>(Phase charge=pulse duration X<br>current amplitude)          | 44 microcoulombs                                                                                                                                                                              | 6.9 microcoulombs                                                                                                                                                  |
| Max. Average Current @500Ω                                                                | 42 μA                                                                                                                                                                                         | 41.6 μA                                                                                                                                                            |
| Max. Current density @500 Ω)                                                              | 0.002 mA/ cm²                                                                                                                                                                                 | 0.0016 mA/cm²<br>(using included 5X5cm electrodes)                                                                                                                 |
| Maximum Power Density (500 Ω)                                                             | 0.00005mW/cm²                                                                                                                                                                                 | 0.000035 mW / cm²<br>(using included 5X5cm electrodes)                                                                                                             |
| Frequency (Hz)                                                                            | 0.5Hz                                                                                                                                                                                         | 3 Hz                                                                                                                                                               |
| Pulse Duration                                                                            | 1.0 sec                                                                                                                                                                                       | 0.166 sec                                                                                                                                                          |
| Burst Mode                                                                                | N/A                                                                                                                                                                                           | N/A                                                                                                                                                                |
| Timer range (min)                                                                         | 0-300 hours, depending on<br>battery life                                                                                                                                                     | 40-55 min selectable by<br>MODE.                                                                                                                                   |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Marchan Medical LLC. The logo features a blue geometric design above the company name. Below the company name is the website address, marchanmedical.com.

{6}------------------------------------------------

# Output Specifications

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo for Marchan Medical LLC. The logo features a blue geometric shape resembling two overlapping triangles, positioned above the text "Marchan Medical". The text is in a bold, black font. Below the text, there is a website address, also in black.

The conclusions drawn from the nonclinical tests described above demonstrate that the device is as safe and effective as the predicate device. The waveform used in both devices is similar with a square waveform with a 50% duty cycle as discovered with oscilloscope readings. The indications for use are very similar, both devices met the same test standards, and both use very low frequency pulses.

---

**Source:** [https://fda.innolitics.com/device/K182267](https://fda.innolitics.com/device/K182267)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
