Bencox Mirabo Cup Multi Hole of Bencox Total Hip System is intended for Cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis b. Inflammatory degenerative joint disease, such as rheumatoid arthritis c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed total hip arthroplasty
Device Story
Bencox Mirabo Cup Multi Hole is a cementless acetabular cup component for total hip arthroplasty; used in primary or revision surgery. Device functions as a mechanical interface between the patient's acetabulum and the hip prosthesis liner. It is implanted by orthopedic surgeons in a clinical/OR setting. The cup features a multi-hole design (8, 12, or 16 holes) to facilitate fixation. It is manufactured from titanium alloy with a porous titanium coating to promote osseointegration. The device is modular, compatible with previously cleared Bencox polyethylene liners and instrumentation. By providing a stable, porous-coated surface for bone ingrowth, the device aims to restore hip joint function, reduce pain, and improve mobility in patients with degenerative joint disease or failed prior implants.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence supported by engineering risk analysis and acetabular cup deformation testing per ISO 7206-12, confirming performance is consistent with previously cleared predicate devices.
Technological Characteristics
Acetabular cup made of Titanium alloy (ASTM F136) with pure Titanium powder porous coating (ASTM F1580). Multi-hole design (8, 12, 16 holes). Cementless fixation. Gamma sterilized. Compatible with modular polyethylene liners. Shelf life validated to 5 years per ASTM F1980.
Indications for Use
Indicated for patients requiring partial or total hip arthroplasty (primary or revision) due to non-inflammatory degenerative joint disease (avascular necrosis, osteoarthritis, traumatic arthritis), inflammatory degenerative joint disease (rheumatoid arthritis), non-union, femoral neck/trochanteric fractures, or failed previous surgery.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
April 1, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Corentec Co., Ltd J.S Daniel Associate Director-Global RA/QA/CA 8F Chungho Tower, 483, Gangnam-daero, Seocho Gu Seoul. 06541 Kr
Re: K182221
Trade/Device Name: Bencox Mirabo Cup Multi Hole Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH Dated: March 5, 2019 Received: March 7, 2019
Dear J.S Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel S. Ramsey -S 2019.04.01 15:29:25 -04'00'
FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K182221
Device Name
Bencox Mirabo Cup Multi Hole
Indications for Use (Describe)
Bencox Mirabo Cup Multi Hole of Bencox Total Hip System is intended for Cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "orentec" in gray letters with a red circle to the left of the word. Underneath the word is a thin red line, and under that line are the words "IMPLANT SPECIALIST" in gray letters. The logo is for a company called Orentec, which specializes in dental implants.
## K182221
# 510(K) SUMMARY Corentec Co., Ltd.
Bencox Mirabo Cup Multi Hole 30th Mar. 2019
#### ADMINISTRATIVE INFORMATION
| Manufacturer | Corentec Co., Ltd.<br>12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu<br>Cheonan-si, Chungchongnam-do, Rep. of Korea -31056<br>Telephone: +82-41-585-7114; Fax: +82-41-585-7113 |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | J.S. Daniel<br>Associate Director – Global RA/QA/CA<br>Corentec Co., Ltd<br>33-2, Banpo-daero 20-gil, Seocho Gu,,<br>Seoul, Rep. of Korea - 06649<br>Ph: +82 2 6937 4984 (Direct) ; Fax: +82 2 3445 5467<br>Email: jsdaniel@corentec.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Bencox Mirabo Cup Multi Hole |
|-----------------------------|------------------------------|
| Common Name: | Acetabular Cup System |
| Classification Regulations: | 21 CFR 888.3358 |
| Regulatory Class: | II |
| Product Codes: | LPH |
| Classification Panel: | Orthopedic Products Panel |
# INDICATIONS FOR USE
Bencox Mirabo Cup System of Bencox Hip Replacement System is intended for cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "Corentec Implant Specialist". The logo features the word "Corentec" in gray letters, with a red circle in place of the "O". Below the word is a thin red line, and below that is the phrase "Implant Specialist" in smaller, gray letters. The logo is simple and modern, with a focus on the company's name and specialization.
# DEVICE DESCRIPTION
This submission consists of inclusion of additional variant to the existing devices Bencox Mirabo Cup System which is a cementless hip acetabular system (Metal on Poly Liner or Ceramic on Poly Liner) for hip arthroplasty. The subject device Bencox Mirabo Cup Multi Hole is similar to Bencox Mirabo Cup System cleared under K172806 and K162127 & K120924, with respect to material – Titanium alloy (ASTM F136), coating with pure Titanium powder (ASTM F1580), design, locking system, manufacturing, packaging and sterilization.
# SUBSTANTIAL EQUIVALENCE
Bencox Mirabo Cup Multi Hole is similar to the 510(k) cleared devices as mentioned below with respect to indications, design, specifications, operating principles and material.
| Device | Manufacturer | Trade Name | 510(k) |
|--------------------------|------------------------|----------------------------|------------------|
| Predicate | Corentec Co.<br>Ltd. | Bencox Mirabo Cup [System] | K172806 |
| | Corentec Co.<br>Ltd. | Bencox Mirabo Cup [System] | K162127 &K120924 |
| Additional<br>Predicates | Biomet Inc. | G7 Acetabular Cup [System] | K140669 |
| | | | K121874 |
| Reference | Stryker<br>Orthopedics | Trident PSL | K983382; K143085 |
The subject devices Bencox Mirabo Cup Multi Hole and the predicate device Bencox Mirabo Cups are acetabular cups made of the same material and plasma-sprayed porous coatings with same pore size, porosity and coating thickness. The design, dimensional specification and locking mechanisms are identical to the previously cleared cups of the Bencox Mirabo Cup and hence subject devices Bencox Mirabo Cup Multi Hole will be compatible with same polyethylene acetabular liners, as well as all modular components and instrumentation cleared in the Bencox Mirabo Cup Acetabular System under K172806 and K162127 & K150007 & K120924.
The subject device Bencox Mirabo Cup Multi Hole is available in multi-hole design (8, 12 & 16 holes) in comparison to primary predicate, Bencox Mirabo Cup which is available in limited holes design (3 holes) cleared under K172806 and K162127 & K120924. The subject device Bencox Mirabo Cup Multi-hole design is similar to additional predicate, G7 Acetabular cup with limited & multi-hole designs with 3, 4, 11 & 16 holes.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the "Corentec" logo. The logo consists of a stylized letter "C" with a red circle inside, followed by the text "orentec" in gray. Below the text is a thin red line, and below that is the text "IMPLANT SPECIALIST" in smaller gray letters.
## PERFORMANCE DATA
Performance testing requirements was analyzed by engineering risk analysis which showed that the inclusion of variant, subject devices - Bencox Mirabo Cup Multi Hole, to the cleared predicate devices Bencox Mirabo Cup System under K172806, did not change the worst case configuration tested for all the performance testing as the dimensional specification of both subject and predicate devices are exactly same, except for acetabular cup deformation testing as per ISO 7206-12. Other performance testing conducted for predicate devices Bencox Mirabo Cup System cleared under K172806 is still valid for inclusion of the subject devices, Bencox Mirabo Cup Multi Hole.
The results of the engineering risk analysis & deformation testing as per ISO 7206-12 showed that the subject devices, Bencox Mirabo Cup Multi Hole are expected to be safe and effective for the proposed indications and are substantially equivalent to the predicate device, and could perform similar to Bencox Mirabo Cup System under K172806.
# STERILIZATION & PACKAGING
Both subject devices and predicate devices are subjected to gamma sterilization. The sterilization and packaging validation for a shelf life of 5 years as per ISO & ASTM standards confirming the stability and effectiveness of packaging of the sterilized product during the shelf-life, by evaluating changes by accelerated aging, as per ASTM F1980 Sterilization validation as per ISO 11137-1 & 2 Sterilization of health care products -Radiation ensures sterility of the components for a SAL of 10 valid for predicate devices. Bencox Mirabo Cup System, is also applicable to subject devices, Bencox Mirabo Cup Multi Hole, as there is no change in the worst case tested in predicate devices submission.
#### CONCLUSION
Overall Bencox Mirabo Cup Multi Hole included in this PMN is similar to the identified primary predicate device and additional predicates. The identified minor differences between the subject devices and predicates do not constitute a new intended use and the minor differences in the technological characteristics do not affect or raise new queries of safety and effectiveness based on successful analysis conforming to recognized performance standards for hip replacement devices and has been adequately addressed in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.