syngo.via View&GO (Version VA10A)

K182208 · Siemens Healthcare GmbH · LLZ · Sep 7, 2018 · Radiology

Device Facts

Record IDK182208
Device Namesyngo.via View&GO (Version VA10A)
ApplicantSiemens Healthcare GmbH
Product CodeLLZ · Radiology
Decision DateSep 7, 2018
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Indications for Use

syngo.via View&GO is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options. syngo.via View&GO supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.

Device Story

syngo.via View&GO is a standalone software-only medical device installed on standard IT hardware; used for viewing, manipulating, communicating, and storing medical images (CT, MR, NM, XA, US, DX). Operates as a reading workplace for clinicians; supports image processing including multiplanar reconstruction (MPR), intensity projections (MIP/MinIP), volume rendering (VRT), and anatomical registration. Features include automatic spine and rib labeling. Output consists of processed images and measurements for clinical interpretation. Does not perform automated diagnostic interpretation (CAD). Healthcare providers use the interface to prepare, read, and annotate images; annotations and measurements can be exported to third-party reporting tools via clipboard functionality. Benefits include streamlined image review and standardized visualization tools for diagnostic support.

Clinical Evidence

Bench testing only. No clinical data. Verification and validation testing confirmed software functionality, performance, and adherence to specifications. Risk analysis and cybersecurity controls implemented.

Technological Characteristics

Software-only device; runs on Microsoft Windows (7/10). Supports DICOM, TCP/IP. Imaging algorithms: MPR, MIP/MinIP, VRT, SSD, DRR, registration, region growing, spine/rib labeling. Connectivity: DICOM-compliant. Cybersecurity: user access control, audit trails, firewall, malware protection. Standards: NEMA PS 3.1-3.20, ISO 10918-1, ISO 15444-1, ISO 14971, IEC 62304, IEC 82304, IEC 62366-1, IEEE 333.2.1.

Indications for Use

Indicated for viewing, manipulation, communication, and storage of medical images in healthcare institutions (Radiology, Nuclear Medicine, Cardiology). Not for digital mammography diagnosis in the U.S.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance. September 7, 2018 Siemens Healthcare GmbH % Alexander Schapovalov Responsible Third Party Reviewer TÜV SÜD America Inc. 1775 Old Highway 8 NEW BRIGHTON, MN 55112 Re: K182208 Trade/Device Name: syngo.via View&GO (Version VA10A) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 2, 2018 Received: August 15, 2018 Dear Alexander Schapovalov: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jeff Rodgers for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K182208 Device Name syngo.via View&GO (Version VA10A) #### Indications for Use (Describe) syngo.via View&GO is a software solution intended to be used for Viewing, communication, and storage of medical images. It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options. syngo.via View&GO supports interpretation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments. The system is not intended for the displaying of digital mammography images for diagnosis in the US. Type of Use (Select one or both, as applicable) | <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |------------------------------------------------------------------------------------| | <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. Date prepared: July 18, 2018 - 1. Submitter: Siemens Healthcare GmbH Henkestrasse 127 91052 Erlangen Germany Establishment Registration Number: 3004977335 ## 2. Contact Person: Mr. Elango Alampalayam Rangappan Regulatory Affairs Manager Siemens Healthcare GmbH, Siemensstr. 1 91301 Forchheim Germany E-mail: ar.elango@siemens-healthineers.com Telephone: +49 (9191) 18-4429 Fax: +49 (9191) 18-4404 ## 3. Device Name and Classification: | Trade Name: | syngo.via View&GO (Version VA10A) | |-----------------------|---------------------------------------------| | Classification Name: | Picture Archiving and Communications System | | Classification Panel: | Radiology | | CFR Section: | 21 CFR §892.2050 | | Device Class: | Class II | | Product Code: | LLZ | ## 4. Legally Marketed Predicate Device: | Trade Name: | syngo.via | |-----------------------|------------------------------------------------------------------------| | 510(k) Clearance: | K150843 | | Clearance Date: | April 24, 2015 | | Classification Name: | Picture Archiving and Communications System | | Classification Panel: | Radiology | | CFR Section: | 21 CFR §892.2050 | | Device Class: | Class II | | Product Code: | LLZ | | Recall Information: | This predicate device doesn't have any open<br>design related recalls. | {4}------------------------------------------------ ## 5. Device Description: Siemens Healthcare GmbH intends to market the Picture Archiving and Communications System, syngo.via View&GO, software version VA10A. This 510(k) submission describes several modifications to the previously cleared predicate device, syngo.via, software version VB10A. syngo.via View&GO is a software only medical device, which is delivered by download to be installed on common IT hardware. This hardware has to fulfil the defined requirements. Any hardware platform that complies to the specified minimum hardware and software requirements and with successful installation verification and validation activities can be supported. The hardware itself is not seen as part of the medical device syngo.via View&GO and therefore not in the scope of this 510(k) submission. syngo.via View&GO provides tools and features to cover the radiological tasks preparation for reading, reading images and support for reporting 4. syngo.via View&GO supports DICOM formatted images and objects. syngo.via View&GO is a standalone viewing and reading workplace. This is capable of rendering the data from the connected modalities for the post processing activities. syngo.via View&GO provides the user interface for interactive image viewing and processing with a limited short term storage which can be interfaced with any Long term storage (e.g. PACS) via DICOM. syngo.via View&GO is based on Microsoft Windows operating systems. syngo.via View&GO supports various monitor setups and can be adapted to a range of image types by connecting different monitor types. The subject device and the predicate device share fundamental scientific technology. # 6. Indications For Use: syngo.via View&GO is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options. syngo.via View&GO supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S. <sup>14</sup> Support for reporting is provided in the clipboard functionality so that the image with measurements and annotations can be copied to any 3td party reporting tools. {5}------------------------------------------------ ## 7. Summary of Similarities and Differences between the Subject Device and the Predicate Device: The similarities and differences between the subject device in this premarket notification and the predicate device are summarized in the following comparison table: | | Subject device<br>syngo.via View&GO VA10A | Predicate device<br>syngo.via VB10A | Comparison | |-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | syngo.via View&GO is a software solution in-<br>tended to be used for viewing, manipulation,<br>communication, and storage of medical images. | syngo.via is a software solution intended to be<br>used for viewing, manipulation, communica-<br>tion, and storage of medical images. | Same | | | It can be used as a stand-alone device or togeth-<br>er with a variety of cleared and unmodified<br>syngo based software options. | It can be used as a stand-alone device or togeth-<br>er with a variety of cleared and unmodified<br>syngo based software options. | | | | syngo.via View&GO supports interpretation<br>and evaluation of examinations within<br>healthcare institutions, for example, in Radiolo-<br>gy, Nuclear Medicine and Cardiology environ-<br>ments. | syngo.via supports interpretation and evaluation<br>of examinations within healthcare institutions,<br>for example, in Radiology, Nuclear Medicine<br>and Cardiology environments. | | | | The system is not intended for the displaying of<br>digital mammography images for diagnosis in<br>the U.S. | The system is not intended for the displaying of<br>digital mammography images for diagnosis in<br>the U.S. | | | | | | | | Software archi-<br>tecture | Standalone workplace system that is logically<br>broken down to syngo.via View&GO subsys-<br>tems syngo modules. | Client-server architecture that is logically bro-<br>ken down to syngo.via subsystems syngo mod-<br>ules. | syngo.via<br>View&GO is a<br>standalone work-<br>place rather syn-<br>go.via is based on<br>Client-Server ar-<br>chitecture | | Image commu-<br>nication | Standard network protocols like TCP/IP and<br>standard communication protocol DICOM. | Standard network protocols like TCP/IP and<br>standard communication protocol DICOM. Ad-<br>ditional fast image transfer protocol for use in-<br>side syngo.via between the client and server. | There is no inter-<br>nal communica-<br>tion as the subject<br>device is a<br>standalone soft-<br>ware system. | | Imaging algo-<br>rithms | - Multiplanar reconstruction (MPR)<br>- Maximum and Minimum Intensity Projec-<br>tion (MIP/MinIP)<br>- Volume Rendering Technique (VRT) with<br>additional edge and surface enhancements<br>and control over rendering parameters<br>- Shaded Surface Display (SSD)<br>- Digitally Reconstructed Radiograph<br>- Editor functionality (e.g. ClipBox)<br>- Registration<br>- Anatomical registration<br>- Region growing<br>- Automatic Spine Labeling, also for ribs in CT<br>thorax scans 15 ("Rib labeling") | - Multiplanar reconstruction (MPR)<br>- Maximum and Minimum Intensity Projec-<br>tion (MIP/MinIP)<br>- Volume Rendering Technique (VRT) with<br>additional edge and surface enhancements<br>and control over rendering parameters<br>- Shaded Surface Display (SSD)<br>- Digitally Reconstructed Radiograph<br>- Editor functionality (e.g. ClipBox)<br>- Registration<br>- Anatomical registration<br>- Region growing<br>- Automatic Spine Labeling, also for ribs in CT<br>thorax scans 16 ("Rib labeling") | Same | | Quantitative<br>algorithms | Distance and angle measurements | Distance and angle measurements | Same | | Supported Im-<br>age Generating<br>Modalities | CT Image (Computed Tomography)<br>- MR Image (Magnetic Resonance)<br>- NM Image (Nuclear Medicine)<br>- XA Image (X-Ray Angiography)<br>- US Image (Ultrasound)<br>- DX Image (Digital Radiography)<br>- DICOM secondary capture objects | CT Image (Computed Tomography)<br>- MR Image (Magnetic Resonance)<br>- NM Image (Nuclear Medicine)<br>- XA Image (X-Ray Angiography)<br>- US Image (Ultrasound)<br>- DX Image (Digital Radiography)<br>- DICOM secondary capture objects | Same | | Image data<br>Compression | Receive & Store:<br>Images are received and stored as received<br>without any change in the compression format.<br><br>Display:<br>Images are displayed as received without any<br>change in the compression.<br><br>Lossy compression images are displayed with<br>an indication to the user with the compression<br>ratio.<br><br>Export:<br>To DICOM Node: Images are sent as per the<br>DICOM negotiation. Uncompressed is pre-<br>ferred and lossy compression is not supported.<br>To Exchangeable media: Images exported as<br>stored in the local storage.<br><br>Supported Compressions for export: lossless<br>compression algorithms, JPEG, JPEG 2000<br>and RLE. | Receive & Store:<br>Images are received and stored as received<br>without any change in the compression format.<br><br>Display:<br>The images are displayed as received without<br>any change in the compression.<br><br>Lossy compression images are displayed with<br>an indication to the user with the compression<br>ratio.<br><br>Export:<br>To DICOM Node and Exchangeable media:<br>Images are sent as per the DICOM negotiation<br>and depending on Compression settings. Lossy<br>compression is not supported.<br><br>Supported Compressions for export: lossless<br>compression algorithms, JPEG, JPEG 2000 and<br>RLE. | syngo.via has the<br>option to config-<br>ure the Image<br>compression type<br>for export but<br>syngo.via<br>View&GO is not<br>having the option<br>to configure. | | Operating systems | Workplace:<br>Microsoft Windows 7 – 64 bit SP1<br>Microsoft Windows 10 – 64 bit | Client:<br>Microsoft Windows XP, Microsoft Windows<br>Vista, or Microsoft Windows 7<br>Server:<br>Microsoft Windows Server 2008 R2 | There is no sepa-<br>rate server in syn-<br>go.via View&GO. | | Software functionality | Graphical user interface with reduced color<br>palette, clearer structure and text labels on<br>icons.<br>Online help system provided in terms of the<br>electronic user documents delivered as part of<br>the installation of the medical device.<br>Reporting functionality is not included in the<br>syngo.via View&GO however the support for<br>creation of reports with any 3rd party tool by<br>enabling clipboard functionality.<br>Patient browser with simplified search func-<br>tionality, clearer structure of search results, un-<br>limited search results, periodic updates of<br>search results.<br>Range pre-sets are not supported.<br>Suggested spine labels to be confirmed by user. | Graphical user interface with reduced color<br>palette, clearer structure and text labels on<br>icons.<br>Online help system with improved search, in-<br>dexing, filtering, library function, document<br>collections, and user-generated content.<br>Reporting with additional functionality to cre-<br>ate one page reports, insert snapshot images,<br>customize reports, and data export in various<br>formats.<br>Patient browser with simplified search func-<br>tionality, clearer structure of search results, un-<br>limited search results, periodic updates of<br>search results, image preview and flexible<br>floating patient browser window.<br>Supports anatomical range pre-sets, support for<br>spine ranges, and annotations as graphical<br>overlays.<br>Suggested spine labels to be confirmed by user,<br>and additional smart placement of labels, also<br>in inter-vertebra regions, support of 2D images, | syngo.via<br>View&GO is a<br>simplified version<br>of syngo.via<br>without the fol-<br>lowing key func-<br>tionalities which<br>are present in<br>syngo.via:<br>● Worklist han-<br>dling<br>● Task flow<br>● Reporting<br>● Findings Nav-<br>igator | | | and additional smart placement of labels, also<br>in inter-vertebra regions, support of 2D images,<br>support of multi-series studies, and added sup-<br>port for rib labels.<br>Basic printing with paper is supported.<br><br>Configurable settings for image text.<br>Added textual and graphical annotations, im-<br>proved placement of measurement text, and<br>changed numbering of markers. | support of multi-series studies, and added sup-<br>port for rib labels.<br>Printing with additional options for rearranging<br>images, adding annotations, full/customized<br>image text, print size and orientation.<br><br>Configurable settings for image text.<br>Added textual and graphical annotations, im-<br>proved placement of measurement text, and<br>changed numbering of markers. | | | Impact on Image Generating Devices | None.<br><i>syngo.via</i> View&GO is a pure post processing<br>software and there is no influence on the image<br>generating devices. | None<br><i>syngo.via</i> is a pure post processing software<br>and no influence on the image generating de-<br>vices. | NA as both the<br>devices doesn't<br>impact the Image<br>generating devices. | | CAD Functionalities | None.<br>No automated diagnostic interpretation capa-<br>bilities like CAD are included. All image data<br>are to be interpreted by trained personnel. | None.<br>No automated diagnostic interpretation capa-<br>bilities like CAD are included. All image data<br>are to be interpreted by trained personnel. | NA as both the<br>devices doesn't<br>support any CAD<br>functionalities. | | Software self-test / checks | Intimate the user in case the data transfer is in-<br>terrupted to the connected DICOM node.<br><br>Hardware / Operating system compatibility<br>check during Installation. | Indicates the user in case that data coming<br>from an interface cannot be assigned to a task<br>flow to assign it manually.<br><br>In case of a data mismatch due to incorrect or | Additional checks<br>are there in <i>syngo.via</i> as <i>syn-<br/>go.via</i> has the ad-<br>ditional function- | | | Medical device validation supports the end us-<br>er to qualify their system for the clinical use. | inconsistent use of DICOM rules the adminis-<br>trator can change the data transfer protocol and<br>patient identification rules. | alities like work-<br>flow handling,<br>task flow, report-<br>ing. | | Cyber Securi-<br>ty | - User access control<br>- Audit trails<br>- Documentation of system security infor-<br>mation, Network traffic & Firewall control<br>- Support of virus / malware protection. | - User access control<br>- Audit trails<br>- Documentation of system security infor-<br>mation, Network traffic & Firewall control<br>- Support of virus / malware protection.<br>- System Hardening | syngo.via<br>View&GO<br>doesn't provide<br>the extensive sys-<br>tem hardening<br>which is provided<br>by syngo.via. | | Hardware | Hardware is not understood as part of the med-<br>ical device, but needs to comply with the min-<br>imum requirements as specified by syngo.via<br>View&GO. | Hardware is not understood as part of the med-<br>ical device, but needs to comply with the min-<br>imum requirements as specified by syngo.via. | Same | {6}------------------------------------------------ <sup>15</sup> Rib Labeling as a functionality was already covered by a 510(k) clearance with device SYNGO, CT BONE READING, K123584. <sup>16</sup> Rib Labeling as a functionality was already covered by a 510(k) clearance with device SYNGO, CT BONE READING, K123584. {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ {11}------------------------------------------------ ## 8. Non-clinical Performance Testing: Non-clinical tests were conducted for the device syngo.via View&GO during product development. The modifications described in this Premarket Notification were supported with verification and validation testing. Siemens Healthcare GmbH claims conformance to the following standards: - NEMA PS 3.1 3.20 (2016), Digital Imaging and Communication in Medicine (DI--COM) - ISO IEC 10918-1:1994 + Technical Corrigendum 1:2005 (JPEG) - - ISO/IEC 15444-1:2016 (JPEG 2000) - - ISO 14971:2007 - - IEC 62304:2006-05+AMD1:2015-06 - - IEC 82304:2016-10 - - -IEC 62366-1:2015-02 - IEEE 333.2.1-2015 - ## Software Verification and Validation: Software documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the device syngo.via View&GO during product development. The Risk Analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence. Siemens Healthcare GmbH conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Contained in Section B of this submission are our cybersecurity considerations as they relate to the device syngo.via View&GO. #### Summary: Performance tests were conducted to test the functionality of the device syngo.via View&GO. These tests have been performed to assess the functionality of the subject device. Results of all conducted testing were found acceptable in supporting the claim of substantial equivalence. {12}------------------------------------------------ ## 9. Safety and Effectiveness Information: Software specifications, design descriptions, hazard analysis, and labeling information are submitted in support of this premarket notification. The device labeling contains instructions for use with cautions to provide for safe and effective use of the device. The results of the hazard analysis combined with the appropriate preventive measures taken indicate the device is of moderate level of concern, as per the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005). ## 10. Conclusion as to Substantial Equivalence: The predicate device was cleared based on non-clinical supportive information. The comparison of technological characteristics, device hazards, non-clinical performance data, and software validation data demonstrates that the subject device performs comparably to and is as safe and effective as the predicate device that is currently marketed for the same intended use. In summary, we are of the opinion that the subject device syngo.via View&GO, software version VA10A, does not introduce any new significant potential safety risks and is substantially equivalent to and performs as well as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...