The Strome-Blitzer Cytology Balloon device is indicated for use in the 4-quadrant collection and retrieval of surface cells from the esophagus in adults (22 years of age or older). The device may be delivered transorally under direct endoscopic visualization.
Device Story
Strome-Blitzer Cytology Balloon is an inflatable biopsy platform for esophageal cell collection. Device consists of a catheter with a distal silicone balloon featuring 6 collection pleats containing cytology collection strips. During insertion, strips are covered by pleats; upon inflation via syringe, strips are exposed to collect surface cells. Includes 4 proximal indicator dots for orientation. Used in clinical settings by physicians during esophagoscopy. Device is sterile, single-use. Output is a collected cytology specimen for diagnostic analysis. Benefits include standardized 4-quadrant sampling of esophageal surface cells.
Clinical Evidence
Animal study (n=6 swine) compared subject device to reference cytology brush. Primary endpoints: safety (tissue inflammation/edema) and effectiveness (cellularity/diagnostic utility of samples). Results: no procedure-related complications, no inflammatory response/edema, and samples were morphologically comparable to reference brush. Bench testing included dimensional, functional, tensile strength, and inflation/deflation cycle testing.
Technological Characteristics
Silicone balloon with 6 pleats and cytology collection strips. Catheter-based delivery. EO sterilization. Single-use. Biocompatibility per ISO 10993-1. Dimensions: 7mm uninflated OD, 20mm inflated min OD, 84mm balloon length, 1250mm min working length. Manual inflation/deflation via syringe.
Indications for Use
Indicated for 4-quadrant collection and retrieval of esophageal surface cells in adults (age 22+). Delivered transorally under direct endoscopic visualization.
Regulatory Classification
Identification
An esophagoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the esophagoscope and is intended to examine or treat esophageal malfunction symptoms, esophageal or mediastinal disease, or to remove foreign bodies from the esophagus. When inserted, the device extends from the area of the hypopharynx to the stomach. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 13, 2019
ADN International LLC % Eric Bannon Regulatory Consultant AlvaMed, Inc. 935 Great Plain Avenue, #166 Needham, MA 02492
Re: K182159
> Trade/Device Name: Strome-Blitzer Cytology Balloon Regulation Number: 21 CFR§ 874.4710 Regulation Name: Esophagoscope (Flexible or Rigid) and Accessories Regulatory Class: II Product Code: EOX, FDX, BTR Dated: May 15, 2019 Received: May 16, 2019
Dear Eric Bannon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182159
Device Name Strome-Blitzer Cytology Balloon
Indications for Use (Describe)
The Strome-Blitzer Cytology Balloon device is indicated for use in the 4-quadrant collection and retrieval of surface cells from the esophagus in adults (22 years of age or older). The device may be delivered transorally under direct endoscopic visualization.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5.0 510(K) SUMMARY FOR STROME-BLITZER CYTOLOGY BALLOON CATHETER DEVICE
| Date of Preparation: | August 6, 2018 |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant (Submitter): | ADN International LLC<br>425 West 59th Street<br>New York, NY 10019<br>Tel: (917) 703-2517 |
| Correspondent Contact Information: | Eric Bannon<br>AlvaMed Inc.<br>1116 Great Plain Avenue, #1<br>Needham, MA 02492<br>Tel: (888) 331-3485<br>Fax: (617) 249-0955<br>e-mail:<br>ebannon@alvamed.com |
| Device Proprietary Name: | Strome-Blitzer Cytology<br>Balloon |
| Device Common Name: | Cytology Balloon |
| Device Classification Regulation &<br>Name: | 21 CFR 874.4710,<br>Esophagoscope (flexible or<br>rigid) and accessories |
| Device Classification & Product Code: | Class II, EOX, FDX, BTR |
| Prior FDA Document Numbers: | None |
| Basis of Submission: | Traditional 510(k) based on<br>K911588 (predicate device),<br>K103437 (reference device) |
| Number of Devices in Submission: | 1 |
{4}------------------------------------------------
## 5.1 Indications for Use
The Strome-Blitzer Cytology Balloon device is indicated for use in the 4-quadrant collection and retrieval of surface cells from the esophagus in adults (22 years of age or older). The device may be delivered transorally under direct endoscopic visualization.
# 5.2 Device Description
The Strome-Blitzer Cytology Balloon is an inflatable biopsy platform. It consists of a catheter with a silicone balloon at its distal end. The balloon has 6 collection pleats on its surface. Inside each pleat are 2 cytology collection strips for specimen collection. While the balloon is uninflated, the strips are covered by the balloon pleats. When inflated, the strips are exposed. The balloon inflates and deflates with the use of a syringe and attached catheter, and is a sterile, single-use device. There are 4 location indicator dots on the proximal balloon surface for orientation during esophagoscopy.
| Subject:<br>Strome-Blitzer<br>Cytology Balloon | Predicate Device:<br>Brandt Esophageal<br>Cytology Balloon | Reference Device:<br>US Endoscopy<br>Cytology Brush | |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | ADN International LLC | Wilson-Cook Medical,<br>Inc. | United States<br>Endoscopy Group, Inc. |
| 510(k) Number | (to be determined) | K911588 | K103437 |
| Product Code | EOX, FDX, BTR | EOX | FDX |
| Regulation<br>Number | 874.4710 | 874.4710 | 876.1500 |
| Regulation<br>Description | Esophagoscope<br>(flexible or rigid) and<br>accessories | Esophagoscope (flexible<br>or rigid) and accessories | Endoscope and<br>accessories |
| Common<br>Name | Cytology balloon | Cytology balloon | Endoscopic cytology<br>brush |
| | Subject:<br>Strome-Blitzer<br>Cytology Balloon | Predicate Device:<br>Brandt Esophageal<br>Cytology Balloon | Reference Device:<br>US Endoscopy<br>Cytology Brush |
| Indications for<br>Use | The Strome-Blitzer<br>Cytology Balloon<br>device is indicated for<br>use in the 4-quadrant<br>collection and<br>retrieval of surface<br>cells from the<br>esophagus in adults<br>(22 years of age or<br>older). The device<br>may be delivered<br>transorally under<br>direct endoscopic<br>visualization. | The Brandt<br>Esophageal Cytology<br>Balloon is indicated for<br>use in the collection<br>and retrieval of surface<br>cells in the esophagus. | The disposable<br>Cytology Brush is<br>intended to be used<br>to retrieve cytological<br>cell samples in the<br>gastrointestinal tract. |
| Sterility | EO Sterilization | Data Not Available | EO sterilization |
| Single-Use | Yes | Yes | Yes |
| Biocompat-<br>ibility | Complies with ISO<br>10993-1 | Since 510(k) cleared, it<br>is assumed that the<br>device is biocompatible | Since 510(k) cleared, it<br>is assumed that the<br>device is biocompatible |
| Design<br>Characteristics | Balloon with pleats<br>that open upon<br>inflation to expose<br>cytology collection<br>strips Device can be<br>delivered<br>transorally (under<br>direct<br>esophagoscopic<br>visualization) | Balloon with nipple-<br>like projections for<br>cytology collection Passed alongside<br>scope: transoral<br>delivery (under<br>direct<br>esophagoscopic<br>visualization) Cytology results<br>were similar<br>between balloon<br>and cytology brush | Brush with<br>abrasive bristles<br>for cytology<br>collection Passes through<br>scope working<br>channel to desired<br>location and is<br>rotated to collect<br>cells |
| | Subject:<br>Strome-Blitzer<br>Cytology Balloon | Predicate Device:<br>Brandt Esophageal<br>Cytology Balloon | Reference Device:<br>US Endoscopy<br>Cytology Brush |
| Dimensions | Uninflated balloon outer diameter: 7 mm (without application of negative pressure) Inflated balloon minimum outer diameter: 20 mm Balloon length (without catheter): 84 mm Minimum total working length: 1250 mm | Uninflated balloon diameter: 10 mm Inflated balloon outer diameter: 35 mm Balloon length: 25 mm | Brush diameter: 1-3 mm (various adult sizes) Brush length: 1200-2400 mm (various adult sizes) |
## 5.3 Substantial Equivalence Information
{5}------------------------------------------------
{6}------------------------------------------------
## 5.4 Comparison to Predicate Device
Both devices are designed for collection of esophageal tissue for cytology analysis. Both devices share similar designs, modes of operation, single-use disposition and sterility, intended use, and materials.
The primary differences between the predicate and subject devices are in their methods of sample collection and routes of delivery. The Strome-Blitzer Cytology Balloon uses cvtology collection strips in pleats on the balloon surface rather than the nipple-like projections used in the predicate Brandt Balloon. The U.S. Endoscopy Cytology Brush (K103437) device was used in testing of the subject device in animals due to the unavailability of the Brandt Balloon and for its similarity to the cytology collection strips used in the subject device.
Finally, as demonstrated in the results of design verification, bench testing, and animal studies, any differences in design do not adversely affect performance compared to the predicate device, and the devices are substantially equivalent.
#### 5.5 Summary of Supporting Data
ADN International LLC conducted the following performance testing for the Strome-Blitzer Cytology Balloon device.
#### 5.5.1 Dimensional and Functional Performance Testing (Bench)
- . Cytology collection strip inspection
- Balloon bond joint inspection
- Distal atraumatic tip inspection ●
{7}------------------------------------------------
K182159 Page 5 of 5 5.0 510(k) Summary Strome-Blitzer Cytology Balloon Traditional 510(k)
- . Device trackability and balloon to shaft joint integrity test
- . Device total length
- Balloon length
- . Catheter shaft internal diameter
- . Catheter shaft outer diameter
- Maximum balloon profile (deflated)
- . Axial orientation indicator dot inspection
- . Balloon inflation-deflation cycle test
- . Device kink resistance to evaluate balloon's ability to contain strips during device insertion and removal
- Balloon pressure evaluation and burst test
- . Shaft to Luer joint tensile strength
- . Shaft to balloon ioint tensile strenath
#### 5.5.2 Animal Study of Safety and Effectiveness
The subject device was compared to the U.S. Endoscopy Cytology Brush due to commercial unavailability of the Brandt Balloon and similarity of cytology collection strips on the subject device to the action of the Cytology Brush.
Six swine underwent transoral esophagoscopy and cytology collection with the Strome-Blitzer Cytology Balloon and the U.S. Endoscopy Cytology Brush. There were no procedure-related complications in this study. Acceptance criteria for the subject device were met: cytologic examination of esophageal smears prepared with the Strome-Blitzer Cytology Balloon resulted in diagnostically useful cytology samples of high cellularity that were morphologically comparable to those generated with the reference cytology brush. Cytology collection procedures were well tolerated by the animals, with no inflammatory response or edema in the tissue of the distal esophagus in control or test areas.
#### 5.6 Conclusion
A comparison of the Strome-Blitzer Cytology Balloon to the legally marketed predicate demonstrates the same intended use. Bench and animal testing support the conclusion that any differences in product design between subject and predicate do not raise new questions of safety and effectiveness for the proposed indication. Comparative testing demonstrates the subject device meets or exceeds established specifications. The device performs as well as or better than the legally marketed predicate, supporting substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.