← Product Code [OVE](/productcode/OVE) · K182151

# MONET Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate (K182151)

_Ctl Medical Corporation · OVE · Oct 11, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182151

## Device Facts

- **Applicant:** Ctl Medical Corporation
- **Product Code:** [OVE](/productcode/OVE.md)
- **Decision Date:** Oct 11, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one-disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is used to facilitate intervertebral body fusion in the cervical spine at the C2 to C7 disc levels using autograft bone. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage. If the MONET Anterior Cervical Interbody Fusion is used without the Supplementary Fixation Plate assembly, note that supplementary fixation is still required.

## Device Story

System comprises interbody fusion cage and supplementary fixation plate; plate and bone screws attach to cage via C-shaped arms and waist tracks. Implanted in cervical spine (C2-C7) to provide mechanical stability and facilitate intervertebral body fusion. Used by surgeons in clinical settings. Provides structural support to disc space; promotes bony fusion. Modular design allows for anatomical variation. Implantation requires device-specific and universal stainless steel instruments.

## Clinical Evidence

Bench testing only. Performed static compression, static compression shear, static torsion, dynamic axial compression, and dynamic torsion per ASTM F2077. Additional testing included plate pull-off and screw push-out.

## Technological Characteristics

Modular assembly: titanium alloy (Ti-6Al-4V ELI, ASTM F136) cage and fixation plate. Instruments: stainless steel (ASTM F899) and silicone. Mechanical stabilization via C-shaped attachment arms and waist tracks. No software or electronic components.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- MONET™ Anterior Cervical Interbody Fusion Cage System ([K172788](/device/K172788.md))
- MATISSE™ Anterior Cervical Interbody Fusion Cage System ([K121569](/device/K121569.md))
- MATISSE™ Anterior Cervical Interbody Fusion Cage System ([K162682](/device/K162682.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 11, 2018

CTL Medical Corporation % Mr. Barry E. Sands President and Founder ROMIS. Inc. 110 Haverhill Road, Suite 526 Amesbury, Massachusetts 01913

Re: K182151

Trade/Device Name: MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE, ODP Dated: August 1, 2018 Received: August 8, 2018

Dear Mr. Sands:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Melissa Hall -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K182151

#### Device Name

MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate

#### Indications for Use (Describe)

MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one-disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is used to facilitate intervertebral body fusion in the cervical spine at the C2 to C7 disc levels using autograft bone. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage. If the MONET Anterior Cervical Interbody Fusion is used without the Supplementary Fixation Plate assembly, note that supplementary fixation is still required.

Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

## CTL Medical Corporation's MONET™ ACIF Cage with Supplementary Fixation Plate

#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

CTL Medical Corporation Andrea Wilcox 4550 Excel Parkway Site 300 Addison, TX 75001 Phone: 214-545-5820 Fax: 888-831-4892

Contact Person: Mr. Barry E Sands President RQMIS Inc. 110 Haverhill Rd., Suite 526 Amesbury, MA 01913 Phone: 978-358-7307 Facsimile: 978-447-0206 Email: barrysands@rqmis.com

Date Prepared: July 30, 2018

#### Name of Device and Name/Address of Sponsor

Trade Name: MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate CTL Medical Corporation 4550 Excel Parkway Site 300 Addison, TX 75001

#### Common or Usual Name

Intervertebral Body Fusion Device

#### Classification Name

21 CFR 888.3080 Intervertebral Body Fusion Device, Cervical

#### Product Codes: ODP, OVE

{4}------------------------------------------------

## Predicate Devices

## Primary:

K172788, MONET™ Anterior Cervical Interbody Fusion Cage System Additional:

K121569, MATISSE™ Anterior Cervical Interbody Fusion Cage System K162682, MATISSE™ Anterior Cervical Interbody Fusion Cage System

## Intended Use / Indications for Use

MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one-disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is used to facilitate intervertebral body fusion in the cervical spine at the C2 to C7 disc levels using autograft bone. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. If the MONET Anterior Cervical Interbody Fusion Cage System is used without the Supplementary Fixation Plate assembly, note that supplementary fixation is still required.

## Device Description

The subject device consists of a plate and bone anchoring screws which are made of titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136. The anterior fixation plates can only be assembled with the predicate K172788 cage prior to final use. The final assembly constructs are implanted using a combination of device specific and universal instruments manufactured from stainless steel materials that conform to ASTM F899.

MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate is considered an enhanced and modular line-extension to already cleared MONET™ Anterior Cervical Interbody Fusion Cage System (K172788).

The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate have the same risk classification, intended use, and generic designation, as the predicate system the MONET™ Anterior Cervical Interbody Fusion Cage System (K172788). The system is offered in various sizes and configurations to accommodate anatomical variation in the cervical vertebral levels and/or patient anatomy.

{5}------------------------------------------------

## Technological Characteristics

MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly consists of two modular components: The first component is the previously cleared K172788, the MONET™ Anterior Cervical Interbody Fusion Cage and the second component is the complimentary reinforcement hardware.

MONET™ ACIF cage and the supplementary fixation plate are mechanically held together by the C-shaped "attachment arms" that mate with the "waist tracks". The Cshaped mating features provide additional stability to the MONET™ ACIF cage by increasing the footprint of the device within the disc space.

When fully assembled, the MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate provides mechanical support/stability to cervical spine segment to allow for bony fusion to occur.

## Material Composition

The subject supplementary fixation plate hardware (modular disc plates and bone anchoring screws) are made of the same implantable grade titanium alloy (Ti-6Al-4V, ASTM F136).

All subject instruments that are included as part of this submission and comes in contact with human tissue are manufactured from stainless steel materials. Instrument handles that will never touch human tissue are manufactured from stainless steel and silicone.

## Performance Data

The following test were performed per ASTM F2077:

- Static Compression .
- Static Compression Shear
- Static Torsion
- Dynamic Axial Compression ●
- Dynamic Torsion

In addition, the following tests were performed.

- Plate Pull-Off
- . Screw Push out

#### Substantial Equivalence

The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate is as safe and effective as the predicate MONET™ and MATISSE™

{6}------------------------------------------------

Anterior Cervical Interbody Fusion Cage System, Titanium and PEEK (K172788, K162682 and K121569). The MONET™ Anterior Cervical Interbody Fusion Cage System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The reinforcement related technological differences between the MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate and its predicate device do not raise different questions of safety or effectiveness. Thus, the MONET™ Anterior Cervical Interbody Fusion Cage System is substantially equivalent.

---

**Source:** [https://fda.innolitics.com/device/K182151](https://fda.innolitics.com/device/K182151)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
