K182099 · Lima Corporate S.P.A. · LPH · Dec 26, 2018 · Orthopedic
Device Facts
Record ID
K182099
Device Name
Delta TT Pro
Applicant
Lima Corporate S.P.A.
Product Code
LPH · Orthopedic
Decision Date
Dec 26, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Delta Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia; - Rheumatoid arthritis; - Post-traumatic arthritis, - Correction of functional deformity; - Fractures, dislocation of the hip and unsuccessful cup arthroplasty; - Revisions in cases of good remaining bone stock. The Delta Acetabular System is intended for cementless use.
Device Story
Delta TT Pro is a cementless acetabular system for total hip arthroplasty. System comprises Ti6Al4V acetabular cup and Vitamin E-stabilized cross-linked UHMWPE (LimaVit) liners; available in neutral, protruded, and high-wall designs. Liners articulate with Biolox Delta or CoCrMo femoral heads coupled to femoral stems. Bone screws provide supplemental cup stability. Used in hospital/clinical settings by orthopedic surgeons. Device replaces diseased/damaged hip joint components to reduce pain and improve function. Substantial equivalence established via bench testing (push-out, lever-out, torque-out, stiffness, impingement, and wear tests).
Clinical Evidence
Bench testing only. Performed mechanical verification and validation including push-out, lever-out, torque-out, stiffness, impingement, and wear testing to confirm product specifications and safety/effectiveness equivalence.
Technological Characteristics
Materials: Ti6Al4V (acetabular cup), Vitamin E-stabilized cross-linked UHMWPE (liners). Design: Cementless, semi-constrained, porous-coated. Configurations: Neutral, protruded, high-wall. Fixation: Uncemented, optional bone screws. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, hip dysplasia), rheumatoid arthritis, post-traumatic arthritis, functional deformity, hip fractures/dislocation, failed cup arthroplasty, or revisions with adequate bone stock.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
December 26, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Limacorporate S.p.A. Roberto Gabetta Regulatory Manager via Nazionale 52 Villanova di San Daniele, 33038 It
# Re: K182099
Trade/Device Name: Delta TT Pro Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, MBL Dated: November 23, 2018 Received: November 30, 2018
# Dear Roberto Gabetta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
(1-800-638-2041 or 301-796-7100).
Sincerely, Daniel S. Ramsey -S 2018.12.26 15:34:29 -05'00'
FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K182099
Device Name Delta TT Pro
## Indications for Use (Describe)
The Delta Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- · Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
- Rheumatoid arthritis;
- · Post-traumatic arthritis,
- · Correction of functional deformity;
- · Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
- · Revisions in cases of good remaining bone stock.
The Delta Acetabular System is intended for cementless use.
| Type of Use (Select one or both, as applicable) | <input type="checkbox"/> |
|-------------------------------------------------|--------------------------|
|-------------------------------------------------|--------------------------|
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Lima Corporate. The logo consists of a red abstract symbol on the left, followed by the words "Lima" in bold black letters and "Corporate" in light gray letters. Below "Lima Corporate" is the text "Orthopaedic" in light gray, followed by a red diamond shape and the word "motion" in black.
Section 7 - 510(k) Summary
(in accordance with 21 CFR 807.92)
510(k) Number K
### I. Applicant Information
| Applicant: | Limacorporate S.p.A. |
|----------------------------|--------------------------------------------------------------------------------------------------|
| | Via Nazionale, 52 |
| | Villanova di San Daniele |
| | Udine 33038 |
| | Italy |
| Contact Person: | Roberto Gabetta |
| | Regulatory Affairs Manager |
| | Tel: +39 0432945511 |
| | Fax: +39 0432945512 |
| | e-mail: <a href="mailto:roberto.gabetta@limacorporate.com">roberto.gabetta@limacorporate.com</a> |
| Application Correspondent: | Same as Applicant. |
Date Prepared:
August 3, 2018
### II. Device Identification
| Proprietary Name: | Delta TT Pro |
|-----------------------|------------------------------------------------|
| Common/Usual Name: | Total Hip Prosthesis |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained |
| | porous-coated uncemented prosthesis |
| Regulation Number: | 21 CFR 888.3358 |
| Product Codes: | LPH, MBL |
| Classification: | Class II |
| Classification Panel: | Orthopedic Devices |
### III. Predicate Devices
The Delta TT Pro device is substantially equivalent to the following FDA cleared predicate devices with regard to indications for use, performance and technological characteristics:
510(k) Number: K112898, K141395
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for LimaCorporate. The logo consists of a red symbol on the left, followed by the word "Lima" in black and the word "Corporate" in gray. Below the word "Corporate" is the phrase "Orthopaedic motion" in gray, with the word "motion" in red.
| Trade Name: | Delta TT Acetabular System |
|----------------------|------------------------------------------------------------------------------------|
| Manufacturer: | Limacorporate S.p.A. |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis |
| Common/Usual Name: | Total Hip Prosthesis |
| Regulation Number: | 21 CFR 888.3358 |
| Product Code: | LPH, MBL, JDI, LZO |
| Classification: | Class II |
| 510(k) Number: | K140669 |
| Trade Name: | G7 Acetabular System |
| Manufacturer: | Biomet Manufacturing Corp. |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis |
| Common/Usual Name: | Total Hip Prosthesis |
| Regulation Number: | 21 CFR 888.3358 |
| Product Code: | LPH, LZO, OQG, KWZ, JDI, OQH, OQI, PBI |
| Classification: | Class II |
### IV. Device Description
The Delta TT Pro is an acetabular system, intended for total hip replacement in cementless applications. The system consists of a Delta TT Pro acetabular cup (Ti6Al4V) and LimaVit liners (cross-linked UHMWPE with Vitamin E), available in three different design: neutral, protruded and high-wall. The LimaVit liners articulate with the Biolox®Delta and CoCrMo femoral heads, which are coupled to LimaCorporate femoral stems. Bone screws may be used to give further stability to the cup.
### Indications for Use Statement V.
The Delta Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- Non-inflammatory degenerative joint disease such as osteoarthritis, ● avascular necrosis and hip dysplasia;
- Rheumatoid arthritis;
- Post-traumatic arthritis, ●
- Correction of functional deformity;
- Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
- Revisions in cases of good remaining bone stock. ●
The Delta Acetabular System is intended for cementless use.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Lima Corporate Orthopaedic. The logo features a red abstract symbol resembling a person in motion, followed by the words "Lima" in bold black and "Corporate" in gray. Below "Lima Corporate" is the word "Orthopaedic" in gray, followed by a red diamond shape and the word "motion" in black.
### VI. Summary of the Technical Characteristics
The intended use, design, and materials of the Delta TT Pro are substantially equivalent to the intended use, design, and materials of the predicate devices. All required design control activities have been completed and the results indicated that the subject device is substantially equivalent to the predicate devices, in terms of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Lima Corporate Orthopaedic Motion. The words "Lima Corporate" are stacked vertically in a bold, sans-serif font. To the right of "Corporate" are the words "Orthopaedic" and "motion" stacked on top of each other in a smaller font. A red symbol is located below the word "Lima".
# Predicate Device Comparison Table
| Device Name | Subject Device:<br>Delta TT Pro | Predicate Device:<br>(K112898,K141395)<br>Delta TT Acetabular System | Predicate Device:<br>(K140669)<br>G7 Acetabular System<br>(OsseoTi acetabular shell with<br>El acetabular liner) | Significant Differences |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | LPH, MBL | LPH, MBL, JDI, LZO | LPH, LZO, OQG, KWZ, JDI,<br>OQH, OQI, PBI | No significant differences.<br>The subject device is<br>classified as Product Code<br>LPH and MBL under<br>regulation 888.3358, whereas<br>the predicate devices are<br>classified under regulation<br>888.3358 with Product Codes<br>LPH, MBL, JDI, LZO (Delta<br>TT) and LPH, LZO, OQG,<br>KWZ, JDI, OQH, OQI, PBI<br>(G7 Acetabular System).<br>However, the three products<br>present substantially<br>equivalent features as well as<br>substantially equivalent<br>safety and effectiveness<br>characteristics. |
| Regulation # | 888.3358 | 888.3358 | 888.3358 | |
| Class | II | II | II | No difference. |
| Device Name | | | Significant Differences | |
| | Subject Device:<br>Delta TT Pro | Predicate Device:<br>(K112898,K141395)<br>Delta TT Acetabular System | Predicate Device:<br>(K140669)<br>G7 Acetabular System<br>(OsseoTi acetabular shell with<br>E1 acetabular liner) | |
| Indications for<br>Use | • Non-inflammatory<br>degenerative joint disease<br>such as osteoarthritis,<br>avascular necrosis and hip<br>dysplasia;<br>• Rheumatoid arthritis;<br>• Post-traumatic arthritis;<br>• Correction of functional<br>deformity;<br>• Fractures, dislocation of the<br>hip and unsuccessful cup<br>arthroplasty;<br>• Revisions in cases of good<br>remaining bone stock.<br><br>Intended for cementless use. | • Non-inflammatory<br>degenerative joint disease<br>such as osteoarthritis,<br>avascular necrosis and hip<br>dysplasia;<br>• Rheumatoid arthritis;<br>• Post-traumatic arthritis;<br>• Correction of functional<br>deformity;<br>• Fractures, dislocation of the<br>hip and unsuccessful cup<br>arthroplasty;<br>• Revisions in cases of good<br>remaining bone stock.<br><br>Intended for cementless use. | • Non-inflammatory<br>degenerative joint disease<br>including osteoarthritis and<br>avascular necrosis.<br>• Rheumatoid arthritis.<br>• Correction of functional<br>deformity.<br>• Treatment of non-union,<br>femoral neck fracture, and<br>trochanteric fractures of the<br>proximal femur with head<br>involvement, unmanageable<br>using other techniques.<br>• Revision of previously failed<br>total hip arthroplasty.<br><br>Porous acetabular shells and<br>femoral stems are indicated for<br>uncemented biological fixation.<br>Non coated or polyethylene | No significant differences.<br>The subject and predicate<br>devices have essentially<br>identical Indications for Use,<br>with the minor exception that<br>the G7 Acetabular System is<br>also indicated in cases of non-<br>union and trochanteric<br>fractures (see parts in<br>yellow). |
| | Subject Device: | Predicate Device:<br>(K112898,K141395) | Predicate Device:<br>(K140669) | Significant Differences |
| Device Name | Delta TT Pro | Delta TT Acetabular System | G7 Acetabular System<br>(OsseoTi acetabular shell with<br>E1 acetabular liner) | |
| | | | mating components that are<br>indicated for either cemented or<br>uncemented use. | |
| Environment of Use | Clinical/Hospital Environment. | Clinical/Hospital Environment. | Clinical/Hospital Environment. | No difference. |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Lima Corporate. The text "Lima Corporate" is stacked vertically in black font. To the right of "Corporate" is the text "Orthopaedic motion" in a smaller, gray font, with a red arrow pointing to the left.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Lima Corporate Orthopaedic Motion. The logo is vertically oriented, with the Lima Corporate text stacked on top of each other. Below the text is a red symbol, and to the right of the text is the word Orthopaedic and the word motion with a red symbol.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Lima Corporate. The logo consists of a red abstract symbol resembling a person with outstretched arms, followed by the words "Lima" in black and "Corporate" in gray. Below "Lima Corporate" is the word "Orthopaedic" in gray, followed by a red symbol and the word "motion" in gray.
### VII. Summary of Performance Data
Limacorporate has conducted extensive verification and validation testing of the Delta TT Pro as a Hip Prosthesis intended for use in total hip arthroplasty. All verification and validation testing confirms that the product specifications are met, in support of the substantial equivalence of the intended use and technological characteristic as the predicate devices.
The Delta TT Pro has been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate devices.
The main groups of tests performed include:
- Push out, lever out and torque out disassembly test
- . Stiffness test
- Impingement test ●
- o Wear test
### Substantial Equivalence Conclusions VIII.
Based on the comparison of intended use, principles of operation and technological characteristics, the Delta TT Pro device is deemed to be substantially equivalent to the Delta TT Acetabular System, manufactured by Limacorporate S.p.A. (K112898, K141395) and the G7 Acetabular System, manufactured by Biomet Manufacturing Corp. (K140669).
The minor differences identified between the Delta TT Pro and its predicate devices raise no new issues of safety or effectiveness. The performance and validation data demonstrate that the Delta TT Pro is as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.