K182090 · Apos Medical Assets , Ltd. · QDT · Nov 16, 2018 · Physical Medicine
Device Facts
Record ID
K182090
Device Name
AposTherapy System
Applicant
Apos Medical Assets , Ltd.
Product Code
QDT · Physical Medicine
Decision Date
Nov 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3475
Device Class
Class 1
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K182090 · Nov 16, 2018
AposTherapy System
Apos Medical Assets , Ltd.
Published clinical literature; Independent clinical studies
Supplemental evidence used to support the safety and effectiveness of the AposTherapy System for knee pain and function outcomes in patients with knee osteoarthritis.
Literature review; Independent studies; Knee osteoarthritis; Clinical outcomes
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published literature and independent studies; Literature review / Independent clinical studies
Patients with knee osteoarthritis
Not applicable for this study
Knee pain and function outcomes
Indications for Use
The AposTherapy System is intended to be used by trained professionals for adjusting the distribution of weight/force(s) that is being applied to a lower limb. The AposTherapy System is intended for patients with knee osteoarthritis, to help temporarily reduce knee pain and improve lower extremity function during activities of daily living.
Device Story
AposTherapy System is a limb orthosis consisting of shoe-like uppers with two adjustable convex units (Pertupods) on the sole; includes screw fixation, spacers, weight discs, and wedges. Used by trained professionals to calibrate weight/force distribution on lower limbs; patient wears device during scheduled daily sessions (e.g., 30 minutes). Device alters ground contact points to redistribute load and introduce controlled instability during gait. Practitioners may use external, legally marketed gait analysis software to inform calibration; software is not supplied with the device. Periodic assessments of pain, function, and gait guide adjustments to the device and therapy program. Benefits include temporary knee pain reduction and improved lower extremity function for knee osteoarthritis patients.
Clinical Evidence
Prospective, randomized, sham-controlled trial (n=220 knee OA patients). Primary endpoint: WOMAC global scores at week 24. Results: AposTherapy arm (1.37±1.15) vs. sham arm (2.47±1.80); mean difference 1.10, Cohen's d 0.73. 83% of AposTherapy arm vs. 42% of sham arm achieved 50% change in WOMAC 24 (OR 6.67; 95% CI: 3.54-12.58). Supplemental evidence from published literature supports consistent improvement in pain and function.
Technological Characteristics
Limb orthosis (Class I, 21 CFR 890.3475). Components: shoe-like uppers, convex Pertupods, screw fixation, spacers (soft, hard, wedge), and weight discs. Principle: mechanical adjustment of ground contact points and gait instability. No direct patient contact (worn over socks). No internal software; external gait analysis software is optional and not provided.
Indications for Use
Indicated for patients with knee osteoarthritis to temporarily reduce knee pain and improve lower extremity function during activities of daily living.
Regulatory Classification
Identification
A limb orthosis (brace) is a device intended for medical purposes that is worn on the upper or lower extremities to support, to correct, or to prevent deformities or to align body structures for functional improvement. Examples of limb orthoses include the following: A whole limb and joint brace, a hand splint, an elastic stocking, a knee cage, and a corrective shoe.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
November 16, 2018
APOS Medical Assets Ltd. % Janice Hogan Partner Hogan Lovells U.S. LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103
Re: K182090
Trade/Device Name: AposTherapy System Regulation Number: 21 CFR 890.3475 Regulation Name: Limb Orthosis Regulatory Class: Class I Product Code: ODT Dated: October 25, 2018 Received: October 25, 2018
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182090
Device Name AposTherapy System
Indications for Use (Describe)
The AposTherapy System is intended to be used by trained professionals for adjusting the distribution of weight/force(s) that is being applied to a lower limb. The AposTherapy System is intended for patients with knee osteoarthritis, to help temporarily reduce knee pain and improve lower extremity function during activities of daily living.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY - K182090 APOS Medical Assets Ltd.'s AposTherapy System
#### Submitter
APOS Medical Assets Ltd. 7 Hapelech Street Tel-Aviv, Israel 6816727
Phone: 917-993-2911
Contact Person: Cliff Bleustein, MD, MBA, Chief Medical Officer
Date: November 16, 2018
Name of Device: AposTherapy System
Common or Usual Name: Orthosis, Shoe with adjustable sole units
Classification Name: 21 C.F.R. 890.3475, Limb orthosis
Regulatory Class: Class I
Product Code: QDT, Shoe with adjustable sole units
Predicate Device: Substantially equivalent to 21 C.F.R. 890.3475
## Device Description
The AposTherapy System consists of a pair of shoe-like uppers with two convex units (Pertupods) on the sole of each shoe, a screw fixation mechanism for securely attaching the Pertupod to the track and, if required, soft spacers, weight discs, hard spacers, and wedge spacers. Legally marketed gait analysis software may be used by practitioners to help inform their calibration of the biomechanical device, but the software is not supplied with the device.
The device is worn over socks, so it has no direct contact with the patient. Based on a preliminary evaluation of pain, function, and quality of life, a personalized therapy program is created to meet the patient's treatment needs using the personally calibrated device. Practitioners may use basic, legally marketed gait analysis software (not provided with the device) to collect measurements from various movements (e.g., velocity, step length, single limb support) as part of the treatment program. The patient's progress is assessed periodically; the device is then further adjusted and the therapy program updated as needed to achieve the pre-defined goals. The device is intended to be worn only during the time scheduled by the practitioner (e.g., 30 minutes a day).
{4}------------------------------------------------
#### Intended Use / Indications for Use
The AposTherapy System is intended to be used by trained professionals for adjusting the distribution of weight/force(s) that is being applied to a lower limb. The AposTherapy System is intended for patients with knee osteoarthritis, to help temporarily reduce knee pain and improve lower extremity function during activities of daily living.
#### Summary of Technological Characteristics
The intended use of the device is consistent with the classification regulation for a limb orthosis, which defines a class I, 510(k)-exempt device. The adjunctive indication for knee pain due to osteoarthritis does not represent a separate intended therapeutic effect or raise different questions of safety or effectiveness as compared to legally marketed devices of this type, because pain reduction is an outcome of the intended use to re-align limbs and improve their function.
The fundamental technological characteristics of the AposTherapy System are consistent with the classification regulation for a limb orthosis. As defined in 21 C.F.R. § 890.3475, the subject device is worn on the lower extremities, and is used to align body structures for functional improvement. Corrective shoes are specifically listed in the regulation as examples of this type of device.
Moreover, the AposTherapy System's underlying technological principle/mechanism of action is shared with other legally marketed devices. Specifically, Apos Therapy achieves the intended clinical purpose through two mechanisms: 1) adjusting the foot's points of contact with the ground to affect the distribution of weight/force(s) applied to the lower limb, and 2) instability during gait exercise. As noted above, the subject device is in the form of an orthosis shoe, consistent with devices cleared under 21 C.F.R. § 890.3475. The AposTherapy System, like other corrective shoes, is calibrated to the individual patient's needs and adjusted as the patient progresses, and with clinician's assessment of load bearing, gait, and related indicators of function and/or pain at baseline and throughout the course of treatment.
## Performance Data
The AposTherapy System underwent non-clinical testing conducted to ensure that the device can perform as intended. In all instances, the AposTherapy System functioned as intended and each test's success criteria were met.
In addition, clinical testing was performed to demonstrate the device's safety and effectiveness for the proposed indications. A prospective, randomized, sham-controlled clinical trial that enrolled 220 knee OA patients was completed. The data from this study is the primary evidence to support device clearance. The results of the study demonstrate that the AposTherapy System temporarily improves lower extremity function during activities of daily living and reduces knee pain due to OA in the intended treatment population. Specifically, WOMAC global scores at week 24 were 1.37±1.15 and 2.47±1.80 in the AposTherapy and sham arms, respectively (1.10 mean difference, Cohen's d of 0.73). In addition, 83% of the AposTherapy arm (95% Cl: 74% - 89%) and 42% of the sham arm (95% Cl: 33% - 52%) achieved 50% change in WOMAC 24 (adjusted odds
{5}------------------------------------------------
ratio: 6.67 (95% Cl: 3.54-12.58) and adjusted Mantel-Haenszel risk difference of 0.42 (95% Cl: 0.28 – 0.52)).
Additional clinical evidence from the published literature and other independent studies investigating the device's performance for knee pain and function outcomes have been completed and are supplemental evidence in support of device clearance. These supplemental sources also show that the AposTherapy System has been thoroughly investigated and has achieved consistent improvement on knee pain and function outcomes. Consequently, the totality of the data strongly supports clearance of the device for use in patients with knee OA to temporarily improve knee pain and lower extremity function during activities of daily living.
## Conclusions
The AposTherapy System fits within the definition set forth in the classification regulation for a limb orthosis device. Moreover, the AposTherapy System has the same intended use and similar indications for use, technological characteristics, and principles of operation as other legally marketed limb orthoses. Finally, bench and clinical data demonstrate that the AposTherapy System is as safe and effective as other leqally marketed lower limb orthoses. Thus, the AposTherapy System is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.