Compex HD

K182017 · Addent, Inc. · QGO · Apr 29, 2019 · Dental

Device Facts

Record IDK182017
Device NameCompex HD
ApplicantAddent, Inc.
Product CodeQGO · Dental
Decision DateApr 29, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6100
Device ClassClass 1

Indications for Use

The Compex HD is intended to be used by a dentist or health care provider to warm and dispense dental composite materials contained in a compule or PLT (Pre Loaded Tip) to 155 degrees F (68 degrees C).

Device Story

Compex HD is a handheld dental composite warmer; replaces table-top heating units. Device accepts dental composite compules (PLT); uses internal heating element or electrically conductive plastic compule to heat material to 155°F (68°C). Operated by dentist/healthcare provider; powered by rechargeable Li-Ion battery. User activates heater via push-button; LED indicator signals when target temperature is reached. Heated composite exhibits reduced viscosity; allows easier extrusion/dispensing into dental cavities. Device improves flow, depth of cure, and curing time. Unit does not function while charging. Benefits include improved handling of viscous composites and reduced physical effort during restoration. Device is not intended for anesthetic carpules or whitening bleach.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing comparing temperature output of Compex HD and Calset predicate using thermocouples in composite-filled compules. Both devices successfully reached 155°F (68°C).

Technological Characteristics

Handheld device; molded plastic housing with stainless steel components. Heating via PTC elements or electrically conductive plastic compules. Thermistor-based temperature control. Powered by rechargeable Li-Ion battery. Connectivity: USB for charging only. Standards: ANSI/AAMI/IEC 60601-1 (Safety), ANSI/AAMI/IEC 60601-1-2 (EMC).

Indications for Use

Indicated for use by dentists or healthcare providers to warm and dispense dental composite materials (contained in a compule or PLT) to 155°F (68°C) for dental restorative procedures.

Regulatory Classification

Identification

An anesthetic warmer is an AC-powered device into which tubes containing anesthetic solution are intended to be placed to warm them prior to administration of the anesthetic.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 29, 2019 AdDent, Inc. Joshua Friedman President 43 Miry Brook Road Danbury, Connecticut 06810 Re: K182017 Trade/Device Name: Compex HD Regulation Number: 21 CFR 872.6100 Regulation Name: Anesthetic Warmer Regulatory Class: Class I Product Code: OGO Dated: January 25, 2019 Received: January 30, 2019 Dear Joshua Friedman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) ### K182017 Device Name Compex HD #### Indications for Use (Describe) The Compex HD is intended to be used by a dentist or health care provider to warm and dispense dental composite materials contained in a compule or PLT (Pre Loaded Tip) to 155 degrees F (68 degrees C). Federal law restricts this device for sale to a licensed professional. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## K182017 # 510 (k) SUMMARY Owner/Submitter's Name: Joshua Friedman, D.D.S., AdDent, Inc. Address: 43 Miry Brook Road, Danbury, CT 06810 Telephone: (203) 778-0200 Ext.101; (203) 952-3555 (Cell) Fax: (203) 792-2275 Contact Person: Joshua Friedman Email: ifriedman@addent.com Device Name: Compex HD Common Name: Composite Warmer Classification Name: Anesthetic warmer (21 CFR 872.6100 ) Product Code QGO Marketed Device of Equivalence: Calset composite heater, 510(k) # K061395, Classification product code: EBZ (Activator, Ultraviolet, For Polymerization, 21 CFR 872.670), Subsequent product codes: EEG (Heat Source For Bleaching Teeth, 21 CFR 872.6475), EFC (Anesthetic warmer, 21 CFR 872.6100). Prior Submission for this device: FDA 510(k): #K180062 ## Description of the Device: Background: The traditional method of dispensing dental composite contained in a PLT (Pre-Loaded Tip) or compule is to use a mechanical composite dispensing device. The compule is loaded into the tip of the dispenser and hand applied mechanical force is used to extrude the material to the prepared dental cavity. Dispensing these materials at room temperature can be difficult due to their viscous nature. The Compex HD heats the composite material contained in the compule prior to use. The heated composite exhibits improved flow and is dispensed more easily. Composite materials are used to repair or restore teeth in the general population, but are not an integrated part of this device. In addition, research shows that pre-heating composite prior to light curing also improves depth of cure, shortens time of cure along with many other improved physical properties. General Description: The Compex HD uses a heating device to elevate the temperature of dental composite material contained in a compule (PLT) to allow it to flow and be easily extruded. It utilizes a plastic housing in the shape of a traditional dispensing qun. The housing contains a heater element(s) and/or an electrically conductive plastic compule that itself acts as a heater when a current is applied. A control circuit board assembly and rechargeable lithium lon battery are used to apply power to the heater. A separate wall plug-in USB power supply is used to charge the battery. The unit will not function with the USB power supply connected to the unit. {5}------------------------------------------------ Indications for Use (Intended Use) Compex HD: The Compex HD is intended to be used by a dentist or health care provider to warm and dispense dental composite materials contained in a compule or PLT (Pre Loaded Tip) to 155°F (68°C). Indications for Use (Intended Use) Calset Predicate Device: The Calset unit is used to warm dental composite materials and whitening bleach to 130°F (54°C) or 155°F (68°C). It is also used to warm anesthetic carpules to 98°F (37°C) (body temperature). | Indications for Use comparison – Calset predicate device vs. Compex HD | | | |------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Calset (predicate) | Compex HD | | Indications for Use<br>(exact wording) | "The Calset unit is used to warm<br>dental composite materials and<br>whitening bleach to 130°F (54°C) or<br>155°F (68°C). It is also used to<br>warm anesthetic carpules to 98°F<br>(37°C) (body temperature)." | "The Compex HD unit is intended to<br>be used by a dentist or health care<br>provider to warm and dispense<br>dental composite materials<br>contained in a compule or PLT (Pre<br>Loaded Tip) to 155 °F (68°C)." | | Dental application | Yes - "The Calset unit is used to<br>warm dental composite materials" | Yes - "The Compex HD unit is<br>intended to be used by a dentist or<br>health care provider to warm and<br>dispense dental composite<br>materials" | | Used to directly dispense<br>composite material<br>contained in a compule<br>(PLT) | No – composite material are<br>dispensed indirectly using standard<br>mechanical handheld dispensers -<br>refer to explanation and figures below<br>this chart* | Yes - "to warm and dispense dental<br>composite materials contained in a<br>compule or PLT (Pre Loaded Tip)" | | Number of temperature<br>settings | Three - "dental composite<br>materials and whitening bleach to<br>130°F (54°C) or 155°F (68°C). It is<br>also used to warm anesthetic<br>carpules to 98°F (37°C)" | One - "to warm and dispense dental<br>composite materials contained in a<br>compule or PLT (Pre Loaded Tip) to<br>155 °F (68 °C)." | | Temperatures settings | "98°F (37°C), 130°F (54°C), 155°F<br>(68°C)" | "155°F (68°C)" | | Used to warm dental<br>composite materials | Yes - "The Calset unit is used to<br>warm dental composite materials<br>and whitening bleach to 130°F<br>(54°C) or 155°F (68°C)" | Yes - "The Compex HD unit is<br>intended to be used by a dentist or<br>health care provider to warm and<br>dispense dental composite<br>materials contained in a compule or<br>PLT (Pre Loaded Tip) to 155 °F<br>(68 °C)." The temperature option of<br>130°F (54°C) is not included in the<br>Compex HD as it is no longer<br>commonly used to heat composite<br>materials. | | Used to warm anesthetic<br>carpules | Yes - "It is also used to warm<br>anesthetic carpules to 98°F (37°C)"<br>The Calset uses a specially designed<br>tray that accepts anesthetic carpules. | No - The Compex HD unit is by design<br>not intended to accept anesthetic<br>carpules and therefore this feature is<br>eliminated. It will only accept dental<br>materials contained in compules or<br>PLT (Pre Loaded Tip). | | Used to warm whitening<br>bleach | Yes - "The Calset unit is used to<br>warm dental composite materials<br>and whitening bleach to 130°F<br>(54°C) or 155°F (68°C)" Although the<br>Calset is designed to warm whitening<br>bleach, this is no longer a common<br>practice. | No - The Compex HD unit is by design<br>not intended to heat whitening bleach<br>and therefore this feature is eliminated.<br>It will only accept dental materials<br>contained in compules or PLT (Pre<br>Loaded Tip). | * With the table top Calset predicate device, the compule containing the composite material (PLT) is loaded in a standard handheld composite dispenser which is then inserted into a pre-heated tray (holster) that sits on top of the Calset heater base. Once the compule reaches the desired application temperature of 155°F (68°C) in approximately three minutes, the dispenser with loaded compule is removed from the heated tray (holster) and the heated composite material is dispensed to the patient. {6}------------------------------------------------ With the Compex HD unit, the heater is integrated into the front of the dispenser. This is in the form of a heating element, or alternatively the compule itself is made from electrically conductive plastic and acts a heating element. The compule (PLT) containing the composite material is loaded in the front of the unit. The heater is activated by a push button switch on the control panel. Once the unit reaches operating temperature of approximately 155°F (68°C), a LED light changes color to indicate that the material is ready to be dispensed directly to the patient. See figure 1. Image /page/5/Figure/1 description: The image is a diagram of a Compex HD device, labeled as Figure 1. The diagram highlights several key components, including the "Compule (PLT)", "Heater location", "Trigger", "battery location", "on-off switch & LED indicators", and "circuit board location". Each of these components is labeled with an arrow pointing to its location on the device. {7}------------------------------------------------ | Technical Characteristics – Calset predicate device vs. Compex HD | | | |-----------------------------------------------------------------------|-----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | | Calset (predicate) | Compex HD | | Device Configuration | Table Top | Hand Held | | Heater element | yes | yes | | Uses removable trays configured for<br>different composite dispensers | yes | no | | Uses hand mechanical force to<br>dispense composite | yes | yes | | Power supply | low voltage detachable<br>wall plug-in power<br>supply | low voltage Li-Ion Battery/<br>low voltage detachable<br>wall plug-in power supply<br>used to charge battery | | Unit operates while connected to<br>mains power supply | Yes - required | No - will not operate | | Control circuit board | yes | yes | | Microprocessor control | yes | yes | | Temperature control | thermistor | thermistor | | Number of temperature settings | 3 | 1 | | Temperatures settings | 98°F (37°C), 130°F<br>(54°C), 155°F (68°C) | 155°F (68°C) | | Push button control | yes- on/off and<br>temperature selection | yes– on/off | | LED indicators | indicate temperature<br>selection and when<br>operating temperature<br>is reached | indicate when<br>operating temperature<br>is reached and battery<br>charge status | | Used to warm anesthetic carpules | yes | no | | Used to warm whitening bleach | yes | no | | Primary housing material | molded plastic | molded plastic | | Other materials (excluding battery) | aluminum/rubber | stainless steel | | Factory temperature calibration | yes | yes | | Disinfection procedure | yes - wiped with cloth | yes - wiped with cloth | | Patient applied part | no | yes | | Biocompatible material(s) | not required | not required | ## Technical Characteristics of the Compex HD compared to the Predicate Device: Discussion: The Compex HD unit performs the same function as the predicate Calset device - to warm composite materials to 155°F (68°C) prior to dispensing the material used to repair the patient's teeth. In both applications the heated composite becomes less viscous and flows more easily, and therefore is dispensed more easily with less physical effort and more control. The Calset must be used with a separate mechanical dispenser such as the CoMax (product code KXR Applicator, Resin, Exempt), also made by AdDent, Inc. The Compex HD dispenses material directly. In one version of the Compex HD the compule containing composite is warmed by being in close proximity to one or more PTC (Positive Temperature Coefficient) electrical heating elements. In another version of the Compex HD the compule itself is molded from electrically conductive plastic that itself becomes a heater when a current is applied to the compule. Unlike the Calset, the Compex HD is not designed to warm or dispense anesthetic solution or whitening bleach, and as a result does not require the other two temperature settings of 98°F (37°C), and 130°F (54°C). {8}------------------------------------------------ Non-Clinical performance data: To demonstrate that the Compex HD and Calset predicate device are substantially equivalent, testing was performed on both devices using thermocouples inserted into composite filled compules to determine the temperature of the composite during heating. The test results demonstrate that both devices heat the composite to a temperature of approximately 155°F (68°C). Refer to Appendix B Product Tests, Test Report # 131 and Test Report # 165 for test results. In addition, both the Calset predicate unit and the Compex HD have been tested and successfully passed the following international standards (refer to Appendix C Product Safety and EMC Test Reports: - . American National Standard/International Standard ANSI/AAMI/IEC 60601-1, "Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (Third Edition, Amendment 1 2012)." - . American National Standard/International Standard ANSI/AAMI/IEC 60601-1-2. "Medical Electrical Equipment, Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances - Requirements and Tests, 4th Ed. (2014)." ### Clinical performance data: Not Applicable. Performance data summary: The non-clinical test results demonstrate that the Compex HD and predicate Calset essentially perform the same function – to heat dental composite material contained in a compule (PLT) to approximately 155°F (68°C). In addition, the process of dispensing the composite material is similar in that the Calset is used with a separate handheld dispenser to hold the compule while it is being heated and then dispensed while the Compex HD has this feature incorporated into a single device. Since there is no difference in the temperature of the applied composite material and the process of dispensing the composite is similar, requiring fewer steps and less time with the Compex HD, the unit is as safe and effective as the predicate Calset device, Elimination of the additional temperature settings of 98°F (37°C), and 130°F (54°C) as featured on the predicate Calset device presents no additional safety hazard. The Compex HD unit is not intended to heat anesthetic carpules or whitening bleach and the physical design of the Compex HD is incompatible with these materials. Use of the 130°F (54°C) temperature setting with composite materials is no longer common. In addition the safety of the Compex HD is demonstrated by successfully passing the required safety and EMC standards referenced above in Non-clinical performance data.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...