K182013 · Agency For Medical Innovations GmbH · GEI · Mar 11, 2019 · General, Plastic Surgery
Device Facts
Record ID
K182013
Device Name
EasyInstruments
Applicant
Agency For Medical Innovations GmbH
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 11, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Easylnstruments are indicated for cutting, grasping, dissecting and coagulation of tissue in laparoscopic surgical procedures.
Device Story
EasyInstruments is a modular, monopolar electrosurgical instrument system for laparoscopic surgery. System consists of three detachable components: EasyHandle (functional handle), EasyShaft (various diameters/lengths), and EasyTips (application-specific attachments). Used by surgeons in laparoscopic procedures to mobilize, manipulate, grasp, hold, fix, retract, dissect, cut, and coagulate tissue. Modular design facilitates cleaning and reprocessing. Ergonomic fingertip trigger on EasyAct variant facilitates tissue coagulation. Device connects to external electrosurgical generators to deliver high-frequency energy for cutting and coagulation. Clinical benefit derived from modularity, allowing for varied shaft lengths (17 and 26.5 cm) and 3 mm shaft options, providing surgeons with increased procedural flexibility.
Clinical Evidence
Bench testing only. Testing included cleaning (AAMI TIR12/TIR30), sterilization (ISO 17665-1, ISO 11135, ISO 11137), biocompatibility (ISO 10993), electrical safety (IEC 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), and usability (IEC 62366-1). Performance testing verified mechanical integrity (grasping strength, cutting, movement) and electrical continuity. Thermal effects on tissue were assessed via histology on ex vivo animal tissues (liver, kidney, muscle) at various power settings.
Technological Characteristics
Modular monopolar electrosurgical instrument. Components: EasyHandle, EasyShaft, EasyTips. Materials: unspecified medical-grade materials consistent with predicates. Energy: high-frequency (HF) electrical energy. Dimensions: 3 mm shaft diameter; 17 cm and 26.5 cm lengths. Connectivity: connects to external electrosurgical generators. Sterilization: validated per ISO 17665-1, ISO 11135, ISO 11137. Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, ISO 10993, ISO 11607.
Indications for Use
Indicated for cutting, grasping, dissecting, and coagulation of tissue during laparoscopic surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Erbe VIO ESU with Accessories, Model VIO 300 D (K023886)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 11, 2019
Agency for Medical Innovations GmbH % Dr. Pierre Bounaud Acknowledge Regulatory Strategies, LLC 2251 San Diego Ave., Suite B-257 San Diego, California 92110
Re: K182013
Trade/Device Name: EasyInstruments Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories. Regulatory Class: Class II Product Code: GEI Dated: January 25, 2019 Received: January 29, 2019
Dear Dr. Bounaud:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Long H. Chen Long H. Chen -S -s -3 Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182013
Device Name EasyInstruments
Indications for Use (Describe)
The Easylnstruments are indicated for cutting, grasping, dissecting and coagulation of tissue in laparoscopic surgical procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for A.M.I. The letters are in blue and are large and bold. There is a small gray circle below each letter. There is a registered trademark symbol in the upper right corner of the letter I.
# 510(k) Summary
DATE PREPARED
February 26, 2019
### MANUFACTURER AND 510(k) OWNER
Agency for Medical Innovations GmbH Im Letten 1, 6800 Feldkirch, Vorarlberg, Austria Telephone: Fax: +43 5522 90505-4006 Official Contact: Martin Hohlrieder, Ph.D., CTO
### REPRESENTATIVE/CONSULTANT
Pierre Bounaud, Ph.D. Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave., Suite B-257, San Diego, CA 92110, USA Telephone: Email: pbounaud@acknowledge-rs.com akomiyama@acknowledge-rs.com
### PROPRIETARY NAME OF SUBJECT DEVICE
EasyInstruments
### COMMON NAME
Electrosurgical, Cutting & Coagulation & Accessories
### DEVICE CLASSIFICATION
Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400, Product Code GEI, Class II)
#### PREMARKET REVIEW
ODE/DSD/GSDB2 General & Plastic Surgery Panel
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for A.M.I. The logo is in blue and consists of the letters A, M, and I. There is a gray dot below each letter. There is a registered trademark symbol to the right of the letter I.
## PREDICATE DEVICE IDENTIFICATION
EasyInstruments device is substantially equivalent to the following predicates:
| 510(k)<br>Number | Predicate Device Name / Manufacturer | Primary<br>Predicate | Reference<br>Predicate |
|------------------|--------------------------------------------------------------------------------|----------------------|------------------------|
| K962119 | ReNew Laparoscopic Instruments / Microline | ✓ | |
| K160706 | ReNew V Handpiece Laparoscopic Instruments /<br>Microline | | |
| K163615 | Snowden-Pencer Microlap 3mm Laparoscopic<br>Instruments / CareFusion 2200 Inc. | | |
| K933002 | Erbotom ICC 350 / Erbe USA, Inc. | | ✓ |
| K023886 | Erbe VIO ESU with Accessories, Model VIO 300 D /<br>Erbe USA, Inc. | | ✓ |
### DEVICE DESCRIPTION
The A.M.I. EasyInstruments device is a modular instrument system for use in laparoscopic highfrequency (HF) surgery and is designed for monopolar application. The Easylnstruments system comprises three-part detachable instruments, which include 1) different functional handles (EasyHandle), 2) instrument shafts of various diameters and lengths (EasyShaft), and 3) a wide range of different, application-specific attachments (EasyTips). It is intended during laparoscopic procedures to mobilize, manipulate, grasp, hold, fix, retract, dissect, cut, and coagulate tissue.
### INDICATIONS FOR USE
The Easylnstruments are indicated for cutting, grasping, dissecting and coagulation of tissue in laparoscopic surgical procedures.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The subject device has a similar design and dimensions and uses similar materials as the devices cleared in K962119, K160706, and K163615. The subject device has equivalent general intended use and similar technological characteristics as the device cleared in K962119, K160706 and K163615. Unlike the predicate devices, the tips, shafts and handles of EasyInstruments are detachable, which allows for easier cleaning and reprocessing of the instruments. A smaller shaft (3 mm) and the associated smaller length of the shaft/handle combination (17 and 26.5 cm) provide the surgeon with more options during laparoscopic procedures. An ergonomic fingertrip trigger is available on the EasyAct device to facilitate coagulation of tissue.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for A.M.I. The letters are large and blue, with a gray dot underneath each letter. The "I" has a registered trademark symbol next to it. The logo is simple and clean, and the colors are eye-catching.
### SUMMARY OF NON-CLINICAL TESTING
Testing was performed per FDA's guidance document Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery (Issued August 15, 2016). A summary of the following tests that were performed was provided in order to demonstrate safety based on current industry standards:
- Cleaning validation per AAMI TIR12 and AAMI TIR30 ●
- . Sterilization validation per ISO 17665-1, ISO 11135, and ISO 11137
- Shelf life and packaging validation per ISO 11607
- Biocompatibility Risk Assessment per ISO 10993-1 ●
- Cytotoxicity testing per ISO 10993-5
- Irritation and sensitization testing per ISO 10993-10
- Electrical safety per IEC 60601-1 and IEC 60601-2-2
- EMC per IEC 60601-1-2
- . Bench performance testing, including performance of disposable tips, performance of reusable tips after reprocessing, performance of EasyHandle reprocessing, and performance of EasyShaft after reprocessing. Performance tests for the tips included visual inspection for damage, opening angle, ease of movement in opening/closing, cutting, dissecting, grasping strength, continuity check between shaft and tip, and compatibility check of the shaft with tips. Performance tests for EasyShaft and EasyHandle included visual inspection (i.e. cracks, damage, warpage), continuity check between shaft and tip, dielectric strength of the shaft, and compatibility check of the shaft with tips.
- Thermal effects on tissue. Side by side testing was performed on ex vivo animal tissues (liver, kidney, and muscle) and thermal damage was assessed by measuring the depth of damage at minimum, default, and maximum power settings using histology.
- Usability testing per IEC 62366-1, including device assembly, device disassembly, device ● reprocessing, and device application in a simulated environment
The results of these tests indicate that the EasyInstruments are substantially equivalent to the predicate devices.
### CONCLUSION
Based on the testing performed, including biocompatibility, cleaning validation, sterilization validation, shelf life validation, packaging validation, electrical safety, EMC testing, non-clinical performance testing (bench), thermal effects on tissue, and usability testing, it can be concluded that the subject device does not raise new issues of safety or efficacy compared to the predicate devices. The equivalent indications for use, technological characteristics, and performance characteristics for the proposed EasyInstruments are assessed to be substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.