Bander Ureteral Diversion Stent Set, Dretler Ureteroscopy Stent Set
Applicant
Cook Incorporated
Product Code
FAD · Gastroenterology, Urology
Decision Date
Apr 23, 2019
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.4620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bander Ureteral Diversion Stent Sets are used for intraoperative placement to stent the ureter and provide drainage of urine from the kidney to the external stoma during ureteroileal conduit construction and continent urinary diversions. The Dretler Ureteroscopy Stent Set is intended to be used for post-ureteroscopy stenting of the ureter.
Device Story
Bander and Dretler Ureteral Stent Sets are passive medical devices used for ureteral drainage. Bander stents (silicone or polyurethane) are placed intraoperatively during urinary diversion procedures to facilitate urine flow from kidney to external stoma. Dretler stents (polyurethane) are placed post-ureteroscopy to maintain ureteral patency. Both feature pigtail coils at the distal end to prevent migration and sideports for drainage. Accessories include wire guides, adapters, and retainers. Devices are placed by physicians in clinical or surgical settings. Output is the physical maintenance of a drainage pathway, which prevents obstruction and promotes healing. Benefits include reduced risk of ureteral stricture or obstruction post-procedure.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included MRI safety, dimensional analysis, tensile strength, shaft/curl elongation, retention strength, flow rate, radiopacity, kink resistance, component compatibility, artificial urine soak, and accelerated aging.
Technological Characteristics
Materials: Silicone or polyurethane. Design: Pigtail ureteral stents with sideports (18-21 gauge). Dimensions: 4.7-8.4 French; 70-125 cm lengths. Features: Radiopaque ink markings, luer-lock/adapter connections. Sterilization: Ethylene Oxide (EtO). Shelf life: 3 years. Packaging: Tyvek polyethylene peel-open pouch.
Indications for Use
Indicated for patients requiring ureteral stenting and drainage during ureteroileal conduit construction, continent urinary diversions, or following ureteroscopy procedures. Prescription use only.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Cook Medical logo, which is a white "COOK" on a red background, with the word "MEDICAL" underneath. Below the logo, the text "510(k) Summary" is displayed in a larger, bold font. The logo and text are centered.
COOK INCORPORA 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
## K181971 Bander Ureteral Diversion Stent Set Dretler Ureteroscopy Stent Set 21 CFR \$807.92 Date Prepared: April 4, 2019
## Submitted By:
Applicant: Primary Contact: Secondary Contact: Applicant Address:
Primary Contact Number: Secondary Contact Number: (812) 335-3575 x105147 Contact Fax Number:
Cook Incorporated Paul Meyer Andrew Breidenbach Cook Incorporated 750 Daniels Way Bloomington, IN 47404 (812) 335-3575 x105299 (812) 332-0281
Device Information:
| Trade Name: | Bander Ureteral Diversion Stent Set<br>Dretler Ureteroscopy Stent Set |
|----------------------|-----------------------------------------------------------------------|
| Common Name: | Ureteral stent |
| Classification Name: | Ureteral stent, |
| Regulation: | 21 CFR §876.4620 |
| Product Code: | FAD, Class II |
# Predicate Device:
The Percuflex™ Urinary Diversion Stent Set (K830803, Boston Scientific) is designed to be used following percutaneous, endoscopic, or operative procedures. The design features a pigtail curl on the distal end to prevent migration and a luer-lock connection on the proximal end. The stent has graduation markings to aid in placement. The set includes two stents (color-coded left and right sides), a guidewire, catheter adapters, and a drainage bag connector.
## De vice Description:
# The Bander Ureteral Diversion Stent
The Bander Ureteral Diversion Stent Set features a pigtail ureteral stent constructed of either silicone or polyurethane. The stent ranges from 4.7 to 8.4 French and the overall lengths range from 70 to 90 centimeters. The pigtail retention coil can be open or closed, tapered or nontapered, but all configurations contain 6 evenly space sideports continue for the first
{1}------------------------------------------------
K181971 Page 2 of 6
10 centimeters of the stent body from the pigtail. In addition to the 90° rotated sideports, some of the stents feature ink markings to aid in placement.
The Bander Ureteral Diversion Stents Sets are supplied with either one or two stents. When two stents are included, a right and left are color-coded red and blue, respectively.
The Bander Ureteral Diversion Stent Sets are also supplied with a straight wire guide.
The accessories supplied with the Bander Ureteral Diversion Stent Sets include an adapter with an O-ring and a catheter retainer.
# The Dretler Ureteroscopy Stent
The Dretler Ureteroscopy Stent Set includes a 4.7, 6.0, or 8.4 French stent is manufactured using a polyurethane material and has a single pigtail at the distal end with 3 evenly spaced sideports. The stent comes with incremental ink markings and is 125 centimeters in length.
The Dretler Ureteroscopy Stent Sets are supplied with an accessory connector with an O-ring and a wire guide.
# Intended Use :
# The Bander Ureteral Diversion Stent
The Bander Ureteral Diversion Stent Sets are used for intraoperative placement to stent the ureter and provide drainage of urine from the kidney to the external stoma during ureteroileal conduit construction and continent urinary diversions.
# The Dretler Ureteroscopy Stent
The Dretler Ureteroscopy Stent Set is intended to be used for post-ureteroscopy stenting of the ureter.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, large, sans-serif font, and the word "MEDICAL" is in white, smaller, sans-serif font.
## Comparison to Predicates:
### The Bander Ureteral Diversion Stent Set
| | | PREDICATE DEVICE | SUBJECT DEVICE |
|---------------------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Percuflex™ Urinary Diversion<br>Stent Set (K830803) | Bander Ureteral Diversion Stent Set<br>(Subject of this Submission) |
| | Manufacturer | Boston Scientific | Cook |
| | Regulation Number | Unclassified | 21 CFR §876.4620 |
| | Product Code | LJE | FAD |
| | Classification Name | Catheter, Nephrostomy | Stent, Ureteral |
| | Device Class | II | Identical |
| Indications for Use | | The Percuflex™ Urinary Diversion<br>Stent Set is used for drainage<br>following percutaneous, endoscopic,<br>or operative procedures.* | Used for intraoperative placement to<br>stent the ureter and provide drainage<br>of urine from the kidney to the<br>external stoma during ureteroileal<br>conduit construction and continent<br>urinary diversions. |
| Model Number | | M0061602XX | 025XXX[-XX] |
| Kit Components | | Stent Wire guide Catheter Connectors* Tube Connectors* | Stent Wire guide Drainage tube adapter Stent retainer |
| Stent | Outer Diameter (Fr) | Pediatric: 4.0, 5.0, 6.0<br>Adult: 6, 7, and 8 | Adult: 4.7, 6.0, 7.2, and 8.4 |
| | Length (cm) | Pediatric: 30<br>Adult: 70 | Adult: 70, 75, and 90 |
| | Material | C-Flex Thermoplastic Elastomer | silicone or polyurethane |
| | Tip Configuration | Closed and Open Tip | Identical |
| | Sideport Size (gage) | Unknown | Curl: 18, 19, or 21<br>Shaft: 17, 18, or 21 |
| | Ink | Unknown | Black Ink |
| Indwelling Time | | ≤ 90 days* | ≤ 90 days |
| Packaging | | Unknown | Tyvek polyethylene peel-open pouch |
| Sterilization | | Unknown | EtO |
| Shelf Life | | Unknown | 3 years |
* If information was not specified within the predicate 510(k), then information was obtained from currently available marketing information and is discussed further in the Substantial Equivalence discussion as appropriate. See AppendixF for predicate marketing data.
The subject device, the Bander Ureteral Diversion Stent Set is similar to the Percuflex™ Urinary Diversion Stent Set (K830803, Boston Scientific) in terms of intended use and technological characteristics. The minor differences from the predicate devices include:
- Similar but not identical indications for use
- Product Code ●
- Additional stent dimensions ●
- Different stent materials
- Indwell time ●
- Packaging ●
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is written in white, and the word "MEDICAL" is written in white below the word "COOK".
K181971 Page 4 of 6 NWW.COOKMEDIC.
- Sterilization ●
- Shelf-life ●
Reference devices used to help support a substantial equivalence determination for the Bander Ureteral Diversion Stent include:
- Cook Black Silicone Filiform Double Pigtail Stent Set (K172017) ●
- Surgitek Urinary Diversion Stent (K800109) ●
- Bard Ureteral Stent (K823487) ●
## The Dretler Ureteroscopy Stent Set
| | PREDICATE DEVICE | SUBJECT DEVICE |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| | Percuflex™ Urinary Diversion<br>Stent Set (K830803) | Dretler Ureteroscopy Stent Set<br>(Subject of this Submission) |
| Manufacturer | Boston Scientific | Cook |
| Regulation Number | Unclassified | 21 CFR §876.4620 |
| Product Code | LJE | FAD |
| Classification Name | Catheter, Nephrostomy | Stent, Ureteral |
| Device Class | II | Identical |
| Indications for Use | The Percuflex™ Urinary Diversion<br>Stent Set is used for drainage<br>following percutaneous,<br>endoscopic, or operative<br>procedures.* | Used for post-ureteroscopy stenting of<br>the ureter. |
| Model Number | M0061602XX | 0255XX[-XX] |
| Kit Components | • Stent<br>• Wire guide<br>• Catheter Connectors*<br>• Tube Connectors* | • Stent<br>• Wire guide<br>• Catheter adapter |
| Stent Outer Diameter<br>(Fr) | Pediatric: 4.0, 5.0, 6.0<br>Adult: 6, 7, and 8 | 4.7, 6, and 8.4 |
| Stent Length (cm) | Pediatric: 30<br>Adult: 70 | 125 |
| Stent Material | C-Flex Thermoplastic Elastomer | Polyurethane |
| Stent Tip Configuration | Closed and Open Tip | Open Tip |
| Sideport Size (gage) | Unknown | 18, 19, or 21 |
| Stent Marker Band<br>Material | Unknown | Black Ink |
| Indwelling Time | ≤ 90 days* | ≤ 90 days |
| Packaging | Unknown | Tyvek polyethylene peel-open pouch |
| Sterilization | Unknown | EtO |
| Shelf Life | Unknown | 3 years |
*If information was not specified within the predicate 510(k), then information was obtained from currently available marketing information and is discussed further in the Substantial Equivalence discussion as appropriate. See AppendixF for predicate marketing data.
{4}------------------------------------------------
750 DANIELS WAY, P.O. B BLOOMINGTON IN 47402-0489 II 9 312 339 2235 TOULEREE: 800 457 WWW.COOKMEDICAL
The subject device, the Dretler Ureteral Diversion Stent Set is similar to the Percuflex™ Urinary Diversion Stent Set (K830803. Boston Scientific). The Dretler Ureteral Diversion Stent Set and the predicate are similar in that these devices have similar designs, intended use, and technological characteristics. The modifications from the predicate devices include:
- . A more specific indications for use
- Product Code ●
- Additional stent dimensions ●
- Characteristically similar material ●
- Unknown predicate materials
- Indwell time
- Packaging ●
- Sterilization ●
- . Shelf-life
## Technological Characteristics:
The following tests have been conducted to ensure reliable design and performance under the specified testing parameters on both subject devices. These tests include:
- MRI Safety Testing
- Dimensional
- . Tensile Strength
- I Shaft and Curl Elongation
- Retention Strength ■
- Flow Rate I
- I Radiopacity
- Kink
- Compatibility with components ■
- Artificial Urine Soak
- . Accelerated Aging
The Black Silicone Filiform Double Pigtail Stent Set manufactured by Cook Incorporated (K172017) was used as a reference device to help support a substantial equivalence determination for the Dretler Ureteroscopy Stent Set.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, bold letters at the top. Below that, the word "MEDICAL" is written in white, smaller, bold letters.
COOK INCORPORATI 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
## Conclusion:
Packaging, sterilization, and shelf-life of the predicate device are not known, however, testing on the subject device has been performed on the subject device in accordance with all applicable standards and device requirements. The results of these tests show that the subject devices, Bander Ureteral Diversion Stent Set and Dretler Ureteroscopy Stent Set, met the design input requirements based on the intended use. Therefore, these data support the conclusion that this device does not raise new questions of safety or effectiveness and that the Bander Ureteral Diversion Stent Set is substantially equivalent to the predicate device the Percuflex™ Urinary Diversion Stent Set (K830803).
Furthermore, the results of these tests show that the subject Dreteroscopy Stent Set met the design input requirements based on the intended use. Therefore, these data support the conclusion that this device does not raise new questions of safety or effectiveness and that the subject Dretler Ureteroscopy Stent Set is substantially equivalent to the predicate device, the PercuflexTM Urinary Diversion Stent Set (K830803).
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181971
Device Name Bander Ureteral Diversion Stent Set
Indications for Use (Describe)
Used for intraoperative placement to stent the ureter and provide drainage from the external stoma during ureteroileal conduit construction and continent urinary diversions.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K181971
Device Name Dretler Ureteroscopy Stent Set
Indications for Use (Describe) This device is intended for post-ureteroscopy stenting of the ureter.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
April 23, 2019
Cook Incorporated Paul Meyer Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K181971
> Bander Ureteral Diversion Stent Set, Dretler Ureteroscopy Stent Set Trade/Device Name: Regulation Number: 21 CFR\$ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: April 4, 2019 Received: April 5, 2019
## Dear Paul Meyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls' provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the [kit/tray] have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.