K181954 · Medtronic Perfusion Systems · CAC · Aug 22, 2018 · Anesthesiology
Device Facts
Record ID
K181954
Device Name
autoLog IQ Autotransfusion System
Applicant
Medtronic Perfusion Systems
Product Code
CAC · Anesthesiology
Decision Date
Aug 22, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The autoLog IQ autotransfusion system and Medtronic wash kit are intended for use in the collection, concentration, washing, and reinfusion of autologous blood. Areas of application may include, but are not limited to, the following: · Surgeries including general, cardiovascular, orthopedic, vascular, plastic/reconstructive, obstetric/gynecologic, and neurological · Postoperative treatment areas
Device Story
System collects, concentrates, washes, and reinfuses autologous blood; used in OR or postoperative settings by medical personnel. Input: patient blood via suction line/cannula into collection reservoir with anticoagulant. Operation: integrated vacuum pump aspirates blood; system uses roller pump, valves, and centrifuge to process blood. Processing: centrifuge separates red cells from plasma; saline wash removes debris; packed red cells transferred to holding bag. Output: washed, packed red cells for reinfusion. System features self-start switch triggered by blood volume in reservoir; optical sensor detects centrifuge bowl fill level. Benefits: provides autologous blood for transfusion, reducing need for donor blood. Healthcare providers use output for patient reinfusion; waste bag containing plasma/saline/debris disposed per hospital guidelines.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included software unit/integration/system validation, vacuum performance, roller pump/valve/weight sensor functionality, electrical/power subsystem testing, blood washing performance, usability (human factors), and compliance with IEC 60601-1, IEC 60601-1-6, IEC 60529, and IEC 60601-1-2 standards.
Technological Characteristics
Autotransfusion apparatus; Class II. System includes instrument, disposable wash kit, collection reservoir, and suction line. Sensing: optical sensor for centrifuge bowl fill, weight sensor. Actuation: roller pump, vacuum pump, valves. Connectivity: standalone instrument. Power: electrical. Sterilization: disposables are patient-contacting (same as predicate). Software: automated processing cycles (fill, wash, transfer).
Indications for Use
Indicated for collection, concentration, washing, and reinfusion of autologous blood in patients undergoing general, cardiovascular, orthopedic, vascular, plastic/reconstructive, obstetric/gynecologic, and neurological surgeries, or in postoperative treatment areas. No specific age or gender contraindications provided; surgical team must evaluate risks/contraindications per procedure.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 22, 2018
Medtronic Perfusion Systems Sue Fidler Senior Regulatory Affairs Supervisor 7611 Northland Drive Brooklyn Park. Minnesota 55428
Re: K181954
Trade/Device Name: autoLog IQ Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II Product Code: CAC Dated: July 17, 2018 Received: July 23, 2018
Dear Sue Fidler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Fernando Aguel -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181954
Device Name autoLog IQ™ Autotransfusion System
Indications for Use (Describe)
The autoLog IQ autotransfusion system and Medtronic wash kit are in the collection, concentration, washing, and reinfusion of autologous blood. Areas of application may include, but are not limited to, the following: · Surgeries including general, cardiovascular, orthopedic, vascular, plastic/reconstructive, obstetric/gynecologic, and neurological
· Postoperative treatment areas
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CER 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page of
{3}------------------------------------------------
#### 5.0 510(k) Summary
| Date Prepared: | July 17, 2018 |
|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009 |
| Contact Person: | Chelsea Pioske<br>Senior Regulatory Affairs Supervisor<br>Phone: 763.514.9838<br>Fax: 763.367.8361<br>E-mail: chelsea.pioske@medtronic.com<br>Alternate Contacts:<br>Jake Roeller<br>Senior Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: 763.526.0404<br>Fax: 763.367.8360<br>Email: jake.w.roeller@medtronic.com<br>Sue Fidler<br>Senior Regulatory Affairs Program Manager<br>Medtronic Perfusion Systems<br>Phone: 763.514.9839<br>Fax: 763.367.8360<br>Email: susan.c.fidler@medtronic.com |
| Trade Name:<br>Common Name:<br>Product Code:<br>Regulation Number:<br>Classification:<br>Classification Panel:<br>Special Controls: | autoLog IQ Autotransfusion System<br>Autotransfusion apparatus<br>CAC<br>21 CFR 868.5830<br>Class II<br>Anesthesiology<br>To date, no special controls or performance standards have been<br>established for these devices. |
# Name of Predicate Device:
autoLog Autotransfusion System, 510(k) K093535
K181954 510(k) Summary Page 1
{4}------------------------------------------------
### Device Description
The autoLog IQ Autotransfusion System is used to collect autologous blood peri-operatively and post-operatively into a collection reservoir with an appropriate amount of anticoagulant. The system refers to the instrument (the subject of this submission); disposable devices are also required to operate the system: the Medtronic wash kit, collection reservoir, and suction line. This autologous blood is first filtered through the prefilter in the collection reservoir, and then processed by centrifugation, separating the red cells from the plasma. Contaminating debris is subsequently washed out by the introduction of normal saline in a wash cycle. The resulting packed red cells, suspended in normal saline, are pumped to a transfer bag, and may be reinfused to the patient. The autoLog IQ Autotransfusion System (also referred to as the "instrument" or the "system") is not patient contacting. However, the disposable devices needed to operate the system - the Medtronic wash kit, collection reservoir, suction line - are patient contacting, and are the same disposables used by the predicate autoLog Autotransfusion System device (K093535).
The collection and concentration/washing of patient blood are handled independently by the system. The collection process requires the installation of a disposable reservoir and the activation and use of the integrated vacuum pump, which is built into the system, or an external vacuum source. The system has an integrated internal regulator which ensures safe vacuum levels. The blood is aspirated through a disposable collection suction tip/cannula and disposable suction line to the reservoir. Anticoagulant is delivered at the cannula through the anticoagulant line.
The system's concentration/washing process requires the installation of the Medtronic wash kit and the activation and use of a pump, valve, and centrifuge. When the self-start switch registers a sufficient quantity of blood in the reservoir, the system automatically starts operation. The processing includes one cycle of filling the centrifuge bowl and concentrating the red cells (while transferring effluent waste to the waste bag), one cycle of washing the red cells with saline and one cycle of transferring the packed, washed red cells to the sterile holding bag. The pump pulls blood from the reservoir into the centrifuge bowl until the optical sensor detects a full bowl; then the roller pump stops and the wash solution is pumped into the packed red cell mass. After the washing cycle, the pump transfers the washed, packed red cells to the holding bag.
Medical personnel complete the final reinfusion of the collected and washed blood to the patient. Hospital personnel are instructed to dispose the waste bag (containing plasma waste, saline, and contaminating debris) in accordance with hospital guidelines.
### Intended Use
The autoLog IQ autotransfusion system and Medtronic wash kit are intended for use in the collection, concentration, washing, and reinfusion of autologous blood. Areas of application may include, but are not limited to, the following:
- . Surgeries including general, cardiovascular, orthopedic, vascular, plastic/reconstructive, obstetric/gynecologic, and neurological
- Postoperative treatment areas
{5}------------------------------------------------
### Contraindications
The process of blood recovery is associated with few complications. The surgical team must consider the risks and relative contraindications of autotransfusion in any surgical procedure before proceeding with autotransfusion.
### Comparison to Predicate Device
A comparison of the autoLog IQ Autotransfusion System to the predicate device (the autoLog Autotransfusion System) indicates the following similarities:
- Intended Use: The intended use is the same as predicate device. ●
- . Design: The general device design is the same as the predicate device.
- Interaction with Disposables: The device uses the same disposable devices as the predicate device, for the same intended purpose, and with the same basic physical interfaces with the disposables as the predicate.
- . Principles of Operation and Technology: The principles of operation are the same as the predicate device.
- . Performance: The performance is substantially equivalent to the predicate and/or reference device.
#### Summary of Testing
Testing has demonstrated that the autoLog IO Autotransfusion System is substantially equivalent to the predicate.
The following tests were conducted:
| Verification/Validation Test | Result |
|----------------------------------------|--------|
| Software Unit, Integration, and System | Pass |
| Setup and Breakdown | Pass |
| Enclosure and Chassis | Pass |
| Cleanability | Pass |
| Unpackaged Transport | Pass |
| Vacuum | Pass |
| Roller Pump, Valve, and Weight Sensor | Pass |
| EE & Power Subsystem | Pass |
| Operating Conditions | Pass |
| Blood Washing Performance | Pass |
| Usability (Human Factors) | Pass |
| IEC 60601-1, IEC 60601-1-6, IEC 60529 | Pass |
| IEC 60601-1-2 | Pass |
| Packaging and Transport | Pass |
### Conclusion
The summary of the data, included in this submission, is sufficient to show that the autoLog IQ Autotransfusion System is substantially equivalent to the legally marketed predicate device, the autoLog Autotransfusion System.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.