← Product Code [MWI](/productcode/MWI) · K181919

# Patient Monitor (K181919)

_Guangdong Biolight Meditech Co., Ltd. · MWI · Apr 5, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181919

## Device Facts

- **Applicant:** Guangdong Biolight Meditech Co., Ltd.
- **Product Code:** [MWI](/productcode/MWI.md)
- **Decision Date:** Apr 5, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

V9 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters of patients, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp). V9 Patient Monitor is used to monitor vital signals for patients and is intended to be used in controlled, hospital environments. It is applicable for adult, pediatric and neonatal patients. It is not intended for helicopter transport, hospital ambulance or home use.

## Device Story

Portable patient monitor; inputs physiological signals via NIBP cuff, temperature sensor, CO2 sensor, and SpO2 sensor; processes signals using integrated modules (BLT, Suntech, Nellcor, or Masimo) to display, store, and alarm on vital signs; used in hospital environments (out-patient departments, wards, NICU) by clinicians; output provides real-time vital sign data to assist clinical decision-making; benefits patient through continuous monitoring of physiological status.

## Clinical Evidence

Bench testing only. No clinical studies were performed. Substantial equivalence is supported by the use of previously FDA-cleared physiological monitoring modules and verification of design specifications against international standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 80601-2-30, IEC 80601-2-61, ISO 80601-2-55, ISO 80601-2-56).

## Technological Characteristics

Portable patient monitor; AC-powered; modular design supporting various SpO2, NIBP, CO2, and temperature sensors. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms), IEC 80601-2-30 (NIBP), IEC 80601-2-61 (SpO2), ISO 80601-2-55 (CO2), and ISO 80601-2-56 (temperature).

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- V6 Vital Signs Monitor ([K153135](/device/K153135.md))

## Reference Devices

- BLT NIBP Module ([K153135](/device/K153135.md))
- Suntech NIBP Module ([K152552](/device/K152552.md))
- Nellcor SpO2 Module ([K052186](/device/K052186.md))
- MASIMO SpO2 Module ([K101896](/device/K101896.md))
- BLT SpO2 Module ([K153135](/device/K153135.md))
- Temperature Module ([K153135](/device/K153135.md))
- CO2 Module ([K153135](/device/K153135.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 5, 2019

Guangdong Biolight Meditech Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 Cn

Re: K181919

Trade/Device Name: Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: February 19, 2019 Received: February 26, 2019

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181919

Device Name Patient Monitor Model: V9

### Indications for Use (Describe)

V9 Patient Monitor is intended to be used for monitoring, displaying, storing and alarming of multiple physiological parameters of patients, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp).

V9 Patient Monitor is used to monitor vital signals for patients and is intended to be used in controlled, hospital environments. It is applicable for adult, pediatric and neonatal patiented for helicopter transport, hospital ambulance or home use.

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)       </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)       </span>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K181919

- 1. Date of Preparation: 4/5/2019
- 2. Sponsor Identification

## Guangdong Biolight Meditech Co., Ltd.

No.2 Innovation First Road, Technical Innovation Coast Hi-Tech Zone, Zhuhai, 519085, Guangdong, China.

Establishment Registration Number: 3007305624

Contact Person: Dan Hou Position: Manager Tel: +86-0756-3399971 Fax: +86-0756-3399989 Email: hou d@blt.com.cn

### 3. Designated Submission Correspondent

Ms. Diana Hong (Primary Contact Person) Mr. Chengyu Wang (Alternative Contact Person)

# Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850 Fax: 001-3609253199 Email: info@mid-link.net

{4}------------------------------------------------

#### 4. Identification of Proposed Device

Trade Name: Patient Monitor Model: V9 Classification Name: Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) Classification: II Product Code: MWI Regulation Number: 21 CFR 870.2300 Review Panel: Cardiovascular

Indications for Use:

V9 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters of patients, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp).

V9 Patient Monitor is used to monitor vital signals for patients and is intended to be used in controlled, hospital environments. It is applicable for adult, pediatric and neonatal patients. It is not intended for helicopter transport, hospital ambulance or home use.

## Device Description

The V9 Patient Monitor can provide the monitoring of physiological parameters such as the Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp). The V9 Patient Monitor is a portable device, consists of main unit and accessories, including NIBP cuff, Temperature sensor, CO2 sensor and SpO2 sensor.

It consists of main unit and accessories, including NIBP cuff, Temperature sensor, CO2 sensor and SpO2 sensor.

All modules, boards and accessories used in V9 Patient Monitor has already been cleared by FDA, referring to Table 1. In which, two NIBP and three SpO2 modules are optional for proposed device.

| Module      |                     | 510(k) Number |
|-------------|---------------------|---------------|
| NIBP        | BLT NIBP Module     | K153135       |
|             | Suntech NIBP Module | K152552       |
| SpO2        | Nellcor SpO2 Module | K052186       |
|             | MASIMO SpO2 Module  | K101896       |
|             | BLT SpO2 Module     | K153135       |
| Temperature | Temperature Module  | K153135       |
| CO2         | CO2 Module          | K153135       |

| Table 1. List of 510(k) of Modules |  |  |
|------------------------------------|--|--|
|------------------------------------|--|--|

{5}------------------------------------------------

#### 5. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- > IEC 60601-1-8: 2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- > IEC 80601-2-30:2013 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- > IEC 80601-2-61:2011 Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- ISO 80601-2-55:2011 Medical electrical equipment Part 2-55: Particular requirements for the > basic safety and essential performance of respiratory gas monitors
- A ISO 80601-2-56: 2009 Medical electrical equipment - Part 2-56: Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
- > ANSI C63.27-2017 American National Standard For Evaluation Of Wireless Coexistence

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device predicate.

- 6. Clinical Test Conclusion
No clinical study is included in this submission, because all modules have been previously cleared by FDA. The K Number of each module is listed in Table 1 of this document.

- 7. Identification of Predicate Device
Predicate Device 510(k) Number: K153135 Product Name: V6 Vital Signs Monitor

{6}------------------------------------------------

### 8. Substantially Equivalent (SE) Comparison

| ITEM                                           | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device<br>Vital Signs Monitor V6, K153135                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                                   | MWI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | MWI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Regulation Number                              | 21 CFR 870.2300                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 21 CFR 870.2300                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Class                                          | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for Use                            | V9 Patient Monitor is intended to be<br>used for monitoring, displaying,<br>reviewing, storing and alarming of<br>multiple physiological parameters of<br>patients, including Pulse Oxygen<br>Saturation (SpO2), Pulse Rate (PR),<br>Non-invasive Blood Pressure (NIBP),<br>Carbon dioxide (CO2) and Temperature<br>(Temp).<br>V9 Patient Monitor is used to monitor<br>vital signals for patients and is intended<br>to be used in controlled, hospital<br>environments. It is applicable for adult,<br>pediatric and neonatal patients. It is not<br>intended for helicopter transport,<br>hospital ambulance or home use. | The vital signs monitor is intended to be<br>used for monitoring, displaying, reviewing,<br>storing and alarming of multiple<br>physiological parameters of patients,<br>including Pulse Oxygen Saturation (SpO2),<br>Pulse Rate (PR), Non-invasive Blood<br>Pressure (NIBP), Carbon dioxide (CO2)<br>and Temperature (Temp).<br>This vital signs monitor is used to monitor<br>vital signals for patients and is suitable for<br>use in hospital environments including<br>out-patient department, wards and NICU. It<br>is not intended for helicopter transport,<br>hospital ambulance or home use. And it is<br>applicable for adult, pediatric and neonatal<br>patients. |
| Intended Population                            | The monitors are intended for adult,<br>pediatric and neonatal.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | The monitors are intended for adult,<br>pediatric and neonatal.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended<br>Environment                        | The V9 Patient Monitor is used to be<br>used in hospital environments including<br>out-patient outpatient departments,<br>wards and NICU. It is not intended for<br>helicopter transport, hospital ambulance<br>or home use.                                                                                                                                                                                                                                                                                                                                                                                                 | The vital signs monitor is used to be used<br>in hospital environments including<br>out-patient outpatient departments, wards<br>and NICU. It is not intended for helicopter<br>transport, hospital ambulance or home use.                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Power Supply                                   | The monitors can be supplied by AC<br>power.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | The monitors can be supplied by AC<br>power.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Number of Patient<br>be monitored each<br>time | Single patient each time                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Single patient each time                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterile                                        | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Electrical Safety                              | Comply with IEC 60601-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Comply with IEC 60601-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| EMC                                            | Comply with IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comply with IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Alarm Function                                 | Comply with IEC60601-1-8                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Comply with IEC60601-1-8                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| SpO2 Module                                    | BLT, Nellcor and Masimo<br>Comply with IEC 80601-2-61                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | BLT and Nellcor<br>Comply with IEC 80601-2-61                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                | NIBP Module                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | BLT and Suntech<br>Comply with IEC 80601-2-30                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| CO2 Module                                     | Comply with ISO 80601-2-55                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comply with ISO 80601-2-55                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Temp Module                                    | Comply with ISO 80601-2-56                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comply with ISO 80601-2-56                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Patient Contact<br>Accessories                 | SpO2 Sensor, NIBP cuff, CO2 Sensor<br>and Temp Sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | SpO2 Sensor, NIBP cuff, CO2 Sensor and<br>Temp Sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                | Biocompatibility                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | All the patient contact accessories were<br>previously cleared by FDA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

# Table 2 Substantially Equivalent Comparison

{7}------------------------------------------------

The SpO2 and NIBP module of proposed device and predicate device is different. Masimo SpO2 and Suntech NIBP are not included in predicate device, however, the Masimo SpO2 and Suntech NIBP have been cleared by FDA under K131898 and K152552. Therefore, this difference is considered not to affect the substantially equivalency (SE) between the proposed and predicate device.

### 9. Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K181919](https://fda.innolitics.com/device/K181919)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
