← Product Code [BTR](/productcode/BTR) · K181880

# VivaSight-SL (TVT) system (K181880)

_Etview , Ltd. · BTR · Nov 29, 2018 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181880

## Device Facts

- **Applicant:** Etview , Ltd.
- **Product Code:** [BTR](/productcode/BTR.md)
- **Decision Date:** Nov 29, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5730
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The ETView VivaSight-SL (TVT™) is intended for intubation procedures. The ETView VivaSight-SL (TVT™) is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral intubations. The VivaSight-SL (TVT™) System is indicated for viewing during non-difficult and difficult intubation procedures, for verifying endotracheal tube and endobronchial blocker placement and repositioning, for viewing during suctioning and for general inspection of the airway.

## Device Story

Single-use, cuffed endotracheal tube (ETT) with embedded CMOS video camera in dedicated lumen; provides real-time airway visualization during intubation and mechanical ventilation. Used in clinical settings (OR/ICU) by physicians/clinicians. System connects to VivaSight monitors or Ambu aView monitor via adapter cable. Enables visual navigation, tube placement verification, and airway inspection without disconnecting ventilator. Benefits include reduced need for direct laryngoscopy and continuous airway monitoring.

## Clinical Evidence

Bench testing only. Performance testing conducted per ISO 5361. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Compatibility testing for Ambu aView monitor included video signal, electronics, image quality, IP rating, and mechanical tests.

## Technological Characteristics

Medical grade PVC ETT; CMOS video sensor; CVBS-NTSC video format. Sizes: 7.0, 7.5, 8.0 mm ID. EtO sterilization. Complies with ISO 5361, ISO 11135, ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18.

## Regulatory Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

## Predicate Devices

- VivaSight-SL (TVT) ([K121028](/device/K121028.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

ETView Ltd. % Yoram Levy Osite General Manager Osite 31 Haavoda St. Binyamina. 30500 Il

Re: K181880

Trade/Device Name: VivaSight-SL (TVT) system Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: Class II Product Code: BTR Dated: October 26, 2018 Received: October 30, 2018

Dear Yoram Levy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Todd D. Courtney -S

Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181880

Device Name VivaSight-SL

### Indications for Use (Describe)

The ETView VivaSight-SL (TVT) is intended for intubation procedures. The ETView VivaSight-SL (TVTM) is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral intubations.

The VivaSight-SL (TVT) System is indicated for viewing during non-difficult intubation procedures, for verifying endotracheal tube and endobronchial blocker placement and repositioning, for viewing and for general inspection of the airway.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for VivaSight. The word "VIVA" is in blue, and the word "SIGHT" is in orange. There is an orange dot above the letter "I" in "VIVA."

# 510(K) SUMMARY

## ETVIEW VIVASIGHT-SL SYSTEM (TVTTM)

## 510(k) Number K181880

Applicant's Name: ETView Ltd. 2 Orange St. Rabin House Misgav Business Park M.P Misgav 2017900 Israel Tel: (972)72-260-7063 Fax: (972)72-260-7263

- Contact Person: Yoram Levy, Qsite 31 Haavoda St. Binyamina, Israel 3054431 Tel (972)4-638-8837 Fax (972)4-638-0510 Yoram@qsitemed.com
ETView VivaSight-SL System (TVTTM) Trade Name:

Device Type: Tracheal tube Preparation Date: November 7, 2018 Classification and Classification Name:

> Name: Tracheal tube Product Code: BTR Regulation No: 21 868.5730 Class: II Panel: Anesthesiology

### Indications for Use Statement:

The ETView VivaSight-SL (TVT™) is intended for intubation procedures. The ETView VivaSight-SL (TVT™) is indicated for use as a temporary

{4}------------------------------------------------

artificial airway in adults requiring mechanical ventilation. It is intended for oral intubations.

The VivaSight-SL (TVT™) System is indicated for viewing during nondifficult and difficult intubation procedures, for verifying endotracheal tube and endobronchial blocker placement and repositioning, for viewing during suctioning and for general inspection of the airway.

## Device Description:

The ETView VivaSight-SL (TVT™) is a single use, cuffed device that functions as a standard endotracheal tube (ETT) and additionally has an embedded video imaging device embedded in a dedicated lumen. The system provides a video image of the patient's trachea, which is displayed on the monitor, for as long as the VivaSight-SL (TVT™) is inside the patient's trachea.

Predicate Device: Substantially equivalent to the following predicate devices:

| Device Name          | Manufacturer | 510k No | Date of<br>Clearance |
|----------------------|--------------|---------|----------------------|
| VivaSight-SL (TVTTM) | ETView       | K121028 | June 21, 2012        |

{5}------------------------------------------------

| Characteristics                       | Proposed ETView VivaSight-<br>SL(TVT) system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ETView VivaSight-SL(TVT)<br>system (K121028)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use                    | The <i>ETView VivaSight-SL</i><br>( <i>TVT</i> TM) is intended for<br>intubation procedures. The<br><i>ETView VivaSight-SL</i> ( <i>TVT</i> TM)<br>is indicated for use as a temporary<br>artificial airway in adults<br>requiring mechanical ventilation.<br>It is intended for oral intubations.<br><br>The <i>VivaSight-SL</i> ( <i>TVT</i> TM)<br>System is indicated for viewing<br>during non-difficult and difficult<br>intubation procedures, for<br>verifying endotracheal tube and<br>endobronchial blocker placement<br>and repositioning, for viewing<br>during suctioning and for general<br>inspection of the airway. | The <i>ETView VivaSight-SL</i><br>( <i>TVT</i> TM) is intended for<br>intubation procedures. The<br><i>ETView VivaSight-SL</i><br>( <i>TVT</i> TM) is indicated for use as<br>a temporary artificial airway in<br>adults requiring mechanical<br>ventilation. It is intended for<br>oral <b>and nasal</b> intubations.<br><br>The <i>VivaSight-SL</i> ( <i>TVT</i> TM)<br>System is indicated for viewing<br>during non-difficult and<br>difficult intubation procedures,<br>for verifying endotracheal tube<br>and endobronchial blocker<br>placement and repositioning,<br>for viewing during suctioning<br>and for general inspection of<br>the airway. |
| Product Code                          | BTR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation No.                        | 21 CFR 868.5730                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification                        | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Supplied/use                          | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Method of<br>Sterilization            | EtO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Characteristics                       | Proposed ETView VivaSight-<br>SL(TVT) system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ETView VivaSight-SL(TVT)<br>system (K121028)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Basic structure                       | Standard ETT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| ETT Material                          | Medical grade Polyvinyl Chloride<br>(PVC)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Imaging sensor                        | CMOS video camera                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Tube Dimensions:<br>Internal Diameter | Three models:<br>• ID 7.0 mm<br>• ID 7.5 mm<br>• ID 8.0 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Tube Dimensions:<br>Outer Diameter    | • OD 10.0 mm<br>• OD 10.5 mm<br>• OD 11.0 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Cuff                                  | Medical grade PVC, High volume,<br>low pressure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Bevel                                 | Opening to the bottom with the<br>camera on the top                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Video transfer                        | Video transfer: Colored real time<br>image acquisition from the aperture<br>of the tracheal lumen on a continuous<br>or intermittent basis without<br>disconnecting the patient from the<br>mechanical ventilator                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Video Format                          | Composite Video Baseband Signal<br>(CVBS)-NTSC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Characteristics                       | Proposed ETView VivaSight-<br>SL(TVT) system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ETView VivaSight-SL(TVT)<br>system (K121028)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Display system                        | VivaSight display system<br>available in the following<br>configurations:<br>VivaSight- Max Monitor- 7"<br>monitor VivaSight Direct- RCA cable<br>configuration for connection to<br>external medical display. Ambu aView monitor (not part<br>of ETView product) -<br>VivaSight is provided with an<br>adapter cable to integrate the<br>SL tubes to Ambu aView<br>monitor.                                                                                                                                                                                                                                                        | VivaSight display system<br>available in the following<br>configurations:<br>VivaSight- Max Monitor- 7"<br>monitor VivaSight Direct- RCA<br>cable configuration for<br>connection to external<br>medical display.                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Number of Murphy<br>Eyes              | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Shelf life/life cycle                 | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 5 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Packaging                             | EtO pouch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Electrical Safety                     | Electrical Safety and<br>Electromagnetic Compatibility data<br>demonstrates that the system (tube<br>and display system) is compliance<br>with IEC 60601-1 and IEC 60601-1-<br>2                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Performance tests                     | Nonclinical laboratory<br>performance testing was<br>performed according to ISO 5361                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

# Substantial Equivalence to Predicate device

{6}------------------------------------------------

г

{7}------------------------------------------------

{8}------------------------------------------------

## Performance Standards

ETView VivaSight-SL (TVT™) was tested and complies with the following standards:

- . ISO 5361 Anesthetic and respiratory equipment -- Tracheal tubes and connectors
- ISO 11135 Sterilization of health care products Ethylene . oxide
- ISO 14971-1 Risk management for medical devices
- ISO 10993-1 Biological evaluation of medical devices -- Part ● 1: Evaluation and testing
- IEC 60601-1 Medical Electrical Equipment, General ● Requirements for Safety
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: . General Requirements for Safety, Collateral Standard: Electromagnetic Compatibility
- IEC 60601-2-18 Medical Electrical Equipment Part 2-18: ● Particular requirements for the safety of endoscopic equipment

## Principle of Operation:

The operation principle of VivaSight-DL System is similar to the use of standard endotracheal tubes. The use of the device should be performed according to the currently accepted medical technique.

The ETView VivaSight-DL System enables visual navigation of the airway during intubation, reducing the need for direct laryngoscopy and providing absolute verification of tube position, while enabling uninterrupted ventilation. Once the VivaSight-DL is in position, the airways can be inspected for accumulation of secretions and for efficacy of suctioning.

The aView adapter cable is an integrated single use video/power cable with connector between VivaSight-DL and Ambu's aView monitor.

{9}------------------------------------------------

VivaSight-DL displays images of the airway onto the Ambu's aView monitor, for as long as the device remains in place during intubation.

## Performance Testing

ETView performed a set of performance test according to ISO5361 requirements. Performance testing demonstrated that the ETView VivaSight-SL (TVT™) is substantially equivalent to the cleared predicate device.

The addition of the new adapter for connection to Ambu aView Monitor was tested through electrical safety and EMC testing and demonstrates that this addition does not raise different questions of safety or effectiveness.

In addition, bench tests were conducted in order to demonstrate the compatibility of the proposed adapter and Ambu aView monitor with the ETView VivaSight-SL (TVT™). The bench tests include: Video signal test, time restriction test, electronics test, image quality, IP rating and mechanical test.

## Technological Characteristics

Both the proposed ETView VivaSight-SL (TVT™) device and its predicate device (ETView VivaSight-SL K121028) function as a standard endotracheal tube (ETT) that additionally has an embedded video imaging device in a dedicated lumen. The system provides video image of patient's trachea that is displayed for as long as the ETT is inside the patient's trachea.

## Biocompatibility:

Materials of the ETView VivaSight-SL (TVT™) system that are in contact with the human body, are identical to the predicate

{10}------------------------------------------------

(K121028) and were tested and found to be biocompatible in accordance with ISO 10993-1.

## Conclusion:

ETView believes that, based on the information provided in this submission, the proposed ETView VivaSight-SL (TVT™) System is substantially equivalent to its predicate device without raising different questions of safety and/or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K181880](https://fda.innolitics.com/device/K181880)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
