PPA of 0.9815 (95% CI: [0.9028; 0.9702]); overall false alarm rate of 0.94 per 24 hours (95% CI: [0.71, 1.21]).
Data labeled using gold-standard video-EEG methodology from epilepsy patients experiencing GTCSs in hospital Epilepsy Monitoring Units.
>1 (epileptologists)
Clinical study of 141 patients (80 pediatrics, 61 adults) admitted to Epilepsy Monitoring Units.
3 (readers)
Indications for Use
The Embrace is a prescription only device that is indicated for use as an adjunct to seizure monitoring of adults and children age 6 and up in home or healthcare facilities during periods of rest. The device is worn on the wrist and senses Electrodermal Activity (EDA) and motion data to detect patterns that may be associated with generalized tonic clonic seizures in patients with epilepsy or at risk of having epilepsy. When a seizure event is detected, Embrace sends a command to a paired wireless device that is programmed to initiate an alert to a designated caregiver. The System records and stores data from Accelerometer, EDA, and Temperature sensors for subsequent review by a trained healthcare professional.
Device Story
Wrist-worn biosensor; captures EDA, accelerometer, and temperature data; processes data via on-board algorithm to detect patterns associated with generalized tonic clonic seizures (GTCS); transmits alerts via Bluetooth to paired smartphone app; app notifies caregivers via call/text; enables manual seizure logging; provides activity/sleep context; used in home or healthcare facilities; operated by patient/caregiver; healthcare professionals review stored data; aids in seizure monitoring; facilitates timely caregiver intervention.
Clinical Evidence
Clinical study in 141 patients (80 pediatric, 61 adult) in Epilepsy Monitoring Units. 54 GTCS events recorded across 31 patients. Positive percent agreement (PPA) was 0.9815 (95% CI: 0.9028-0.9702). Overall false alarm rate (FAR) was 0.94 per 24 hours (95% CI: 0.71-1.21).
Technological Characteristics
Wrist-worn biosensor; sensors: EDA, accelerometer, temperature; connectivity: Bluetooth to smartphone; software: moderate level of concern; biocompatibility: ISO 10993-5/10; electrical safety: IEC 60601-1; electromagnetic compatibility: IEC 60601-1-2.
Indications for Use
Indicated for adjunct seizure monitoring in adults and children age 6+ with epilepsy or at risk of epilepsy; used in home or healthcare facilities during rest; detects patterns associated with generalized tonic clonic seizures (GTCS).
Regulatory Classification
Identification
A non-electroencephalogram (non-EEG) physiological signal based seizure monitoring system is a noninvasive prescription device that collects physiological signals other than EEG to identify physiological signals that may be associated with a seizure.
Special Controls
In combination with the general controls of the FD&C Act, the Non-EEG physiological signal based seizure monitoring system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
(i) Hardware specifications must be provided. Appropriate verification, validation, and hazard analysis must be performed.
(ii) Software, including any proprietary algorithm(s) used by the device to achieve its intended use, must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Appropriate software verification, validation, and hazard analysis must be performed.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) The device must be designed and tested for electrical, thermal, and mechanical safety and electromagnetic compatibility (EMC).
(4) Clinical performance testing must demonstrate the ability of the device to function as an assessment aid for monitoring for seizure-related activity in the intended population and for the intended use setting. Performance measurements must include positive percent agreement and false alarm rate.
(5) Training must be provided for intended users that includes information regarding the proper use of the device and factors that may affect the collection of the physiologic data.
(6) The labeling must include health care professional labeling and patient-caregiver labeling. The health care professional and the patient-caregiver labeling must include the following information:
(i) A detailed summary of the clinical performance testing, including any adverse events and complications.
(ii) Any instructions technicians and clinicians should convey to patients and caregivers regarding the proper use of the device and factors that may affect the collection of the physiologic data.
(iii) Instructions to technicians and clinicians regarding how to set the device threshold to achieve the intended performance of the device.
{0}------------------------------------------------
December 20, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Empatica Srl % Rakesh Lal Consultant Rakesh Lal 318 Rindge Ave 407 Cambridge, Massachusetts 02140
# Re: K181861
Trade/Device Name: Embrace Regulation Number: 21 CFR 882.1580 Regulation Name: Non-Electroencephalogram (EEG) Physiological Signal Based Seizure Monitoring System Regulatory Class: Class II Product Code: POS Dated: November 20, 2018 Received: November 20, 2018
Dear Rakesh Lal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jay R. Gupta -S
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181861
Device Name Embrace
### Indications for Use (Describe)
The Embrace is a prescription only device that is indicated for use as an adjunct to seizure monitoring of adults and children age 6 and up in home or healthcare facilities during periods of rest. The device is worn on the wrist and senses Electrodermal Activity (EDA) and motion data to detect patterns that may be associated with generalized tonic clonic seizures in patients with epilepsy or at risk of having epilepsy. When a seizure event is detected, Embrace sends a command to a paired wireless device that is programmed to initiate an alert to a designated caregiver. The System records and stores data from Accelerometer, EDA, and Temperature sensors for subsequent review by a trained healthcare professional.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "empatica" in a pink sans-serif font. To the right of the word is a pink symbol that resembles a stylized letter E with an arrow pointing to the right. The symbol is also in a sans-serif font and is the same color as the word.
510K Embrace Documentation
# 510(k) Summary
{4}------------------------------------------------
| Submitter Name: | Empatica S.r.l. |
|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | Via Stendhal 36<br>20144 Milano, Italy |
| 510(k) Submission Contact: | Rakesh Lal<br>817-734-8303 |
| Sponsor Contact Person: | Matteo Lai,<br>+393497888125 |
| Date Prepared: | December 19, 2018 |
| Device Trade Name:<br>Device Common Name:<br>Device Classification: | Embrace<br>Non-EEG physiological signal based seizure monitoring system<br>21 CFR 882.1580, Product Code POS |
| Predicate Device:<br>Predicate Device Classification: | Embrace, K172935<br>21 CFR 882.1580, Product Code POS |
| Device Description: | The Embrace is a wearable biosensor device that can capture, store, and<br>wirelessly transmit sensor data via Bluetooth to a paired remote device.<br>Embrace runs an on-board algorithm to continuously process sensor data and<br>make a decision about whether the data might indicate a generalized tonic<br>clonic seizure (GTCS). The algorithm has been validated on data, labeled using<br>the gold-standard video-Electroencephalogram (EEG) methodology designed by<br>a group of epileptologists at a top level 6 epilepsy centers, from epilepsy<br>patients experiencing GTCSs in hospital Epilepsy Monitoring Units. When a<br>likely GTCS is detected, the Embrace sends a message to the Alert smartphone<br>application, which initiates calls and texts to summon the attention of<br>designated caregivers. The device also enables patients to manually record<br>seizure events, and provides contextual information related to activity and<br>sleep. |
| Indications for Use: | The Embrace is a prescription only device that is indicated for use as an adjunct<br>to seizure monitoring of adults and children age 6 and up in home or healthcare<br>facilities during periods of rest. The device is worn on the wrist and senses<br>Electrodermal Activity (EDA) and motion data to detect patterns that may be<br>associated with generalized tonic clonic seizures in patients with epilepsy or at<br>risk of having epilepsy. When a seizure event is detected, Embrace sends a<br>command to a paired wireless device that is programmed to initiate an alert to<br>a designated caregiver. The System records and stores data from<br>Accelerometer, EDA, and Temperature sensors for subsequent review by a<br>trained healthcare professional. |
{5}------------------------------------------------
Substantial Equivalence:
A comparison of Embrace to the predicate device is provided in the table below.
| Attribute | Embrace | Embrace (K172935) | Comparison |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Non-EEG physiological<br>signal based seizure<br>monitoring system. | Non-EEG physiological<br>signal based seizure<br>monitoring system. | EQUIVALENT<br>Both devices have the<br>same intended use. |
| Indications for use | The Embrace is a<br>prescription only device<br>that is indicated for use<br>as an adjunct to seizure<br>monitoring of adults and<br>children age 6 and up in<br>home or healthcare<br>facilities during periods of<br>rest. The device is worn<br>on the wrist and senses<br>Electrodermal Activity<br>(EDA) and motion data to<br>detect patterns that may<br>be associated with<br>generalized tonic clonic<br>seizures in patients with<br>epilepsy or at risk of<br>having epilepsy. When a<br>seizure event is detected,<br>Embrace sends a<br>command to a paired<br>wireless device that is<br>programmed to initiate<br>an alert to a designated<br>caregiver. The System<br>records and stores data<br>from Accelerometer,<br>EDA, and Temperature<br>sensors for subsequent<br>review by a trained<br>healthcare professional. | The Embrace is a<br>prescription only device<br>that is indicated for use<br>as an adjunct to seizure<br>monitoring of adults in<br>the home or healthcare<br>facilities during periods of<br>rest. The device is worn<br>on the wrist, and senses<br>Electrodermal Activity<br>(EDA) and motion data to<br>detect patterns that may<br>be associated with<br>generalized tonic clonic<br>seizures in patients with<br>epilepsy or at risk of<br>having epilepsy. When a<br>seizure event is detected,<br>Embrace sends a<br>command to a paired<br>wireless device that is<br>programmed to initiate<br>an alert to a designated<br>caregiver. The System<br>records and stores data<br>from Accelerometers,<br>EDA, and Temperature<br>for subsequent review by<br>a trained healthcare<br>professional. | Patient Population:<br>EQUIVALENT<br>Both devices are<br>indicated for use on<br>patients with epilepsy or<br>at risk of having epilepsy.<br>Embrace 2.0 is<br>additionally indicated for<br>use in children age 6 and<br>up. This difference does<br>not raise new questions<br>of safety and<br>effectiveness.<br><br>Healthcare Environment:<br>EQUIVALENT<br>Both the subject and<br>predicate devices are<br>indicated for use in the<br>home and healthcare<br>settings.<br><br>Seizure Type:<br>EQUIVALENT<br>Both devices are<br>indicated to provide an<br>alert to caregivers when<br>they each detect GTC<br>while the patient is at<br>rest.<br><br>Effect on Clinical<br>Outcome: EQUIVALENT<br>Both devices are<br>indicated as an adjunct to<br>other monitoring devices<br>in the EMU. Neither of<br>the devices is intended to<br>guide therapy decisions. |
| Sensor Technology | Utilizes an electrodermal<br>sensor to acquire | Utilizes an electrodermal<br>sensor to acquire | EQUIVALENT |
| | | | |
| | and an accelerometer<br>sensor to acquire<br>movement data. | and an accelerometer<br>sensor to acquire<br>movement data. | |
| Software Level of<br>Concern | Moderate | Moderate | EQUIVALENT |
| Data Communication | Communicates wirelessly<br>to a smartphone<br>application, which alerts<br>the healthcare provider<br>or caregiver in one or<br>more ways (phone call,<br>text message, etc.). | Communicates wirelessly<br>to a smartphone<br>application, which alerts<br>the healthcare provider<br>or caregiver in one or<br>more ways (phone call,<br>text message, etc.). | EQUIVALENT |
| Algorithm | Uses algorithms to<br>analyze EDA and<br>accelerometer data to<br>detect patterns in the<br>data that may be<br>associated with GTC<br>seizures. | Uses algorithms to<br>analyze EDA and<br>accelerometer data to<br>detect patterns in the<br>data that may be<br>associated with GTC<br>seizures. | EQUIVALENT |
| Biocompatibility | All patient contacting<br>parts are identical to the<br>predicate device. | All patient contacting<br>parts are tested to<br>applicable tests in ISO<br>10993 (Cytotoxicity - ISO<br>10993-5, Sensitization -<br>ISO 10993-10, and Skin<br>Irritation - ISO 10993-10). | EQUIVALENT |
| Electrical Safety | All electrical and<br>electronic parts are<br>identical to the predicate<br>device. | Electrical Safety testing<br>performed to IEC 60601-1<br><br>Electromagnetic<br>compatibility testing<br>performed to IEC 60601-<br>1-2 4th Edition | EQUIVALENT |
| Thermal Safety | Not applicable. The<br>device does not generate<br>any localized heat | Not applicable. The<br>device does not generate<br>any localized heat | EQUIVALENT |
| Chemical Safety | Not applicable. Patient is<br>not exposed to any<br>chemicals during use of<br>the device | Not applicable. Patient is<br>not exposed to any<br>chemicals during use of<br>the device | EQUIVALENT |
| Radiation Safety | Not applicable. Device<br>does not use any ionizing<br>radiation | Not applicable. Device<br>does not use any ionizing<br>radiation | EQUIVALENT |
{6}------------------------------------------------
{7}------------------------------------------------
Performance Testing: Performance testing demonstrates that Embrace complies with IEC 60601-1:2005+AMD1:2012, IEC 60601-1-2:2014, ISO 10993-5:2009, and ISO 10993-10:2010
> Clinical testing was performed to demonstrate the ability of the Embrace to function as an assessment aid for monitoring for seizure related activity in the intended population and for the intended use setting.
### Patients:
Clinical Testing:
141 patients diagnosed with epilepsy were admitted to the Epilepsy Monitoring Unit (EMU) for standard care were enrolled in the studies: 80 pediatrics, ages 6-21 years, median: 13 years; 61 adults, ages: 22-63 years, median: 39 years.
### Observed GTCS:
31 EMU patients experienced a total of 54 generalized tonic clonic seizures (GTCSs) while 110 EMU patients did not experience any seizure. Every recorded seizure was classified as epileptic.
### Recorded data:
141 patients provided overall 409 days (9,806 hours), with a median of 49.2 hours of data per patient of ACM and EDA measurements.
### Performance:
On all the 141 patients, the Positive percent agreement (PPA) was found to be 0.9815 (53 out of 54 GTCSs detected), with a 95% confidence interval (CI) of [0.9028; 0.9702], relative to a panel of three readers, and the overall false alarm rate (FAR) was found to be 0.94 false alarms per 24 hours with a 95% confidence interval of [0.71, 1.21], corresponding to a mean FAR (average of FARs across patients) of 1.25. Table 1 and Table 2 report the PPA and the FAR separately on the adult patients (age 22 and up) and on pediatrics subgroups.
| | Table 1 | | | | | | |
|-------|-----------|-------|--------------------|-------|-------------------|--------|-------|
| age | #patients | #GTCS | # detected<br>GTCS | PPA | corrected<br>PPA1 | CI PPA | |
| 6-12 | 6 | 12 | 11 | 0.917 | 0.799 | 0.601 | 0.895 |
| 13-21 | 11 | 20 | 20 | 1 | 0.915 | 0.889 | 0.934 |
| 6-21 | 17 | 32 | 31 | 0.969 | 0.915 | 0.834 | 0.953 |
| >21 | 14 | 22 | 22 | 1 | 0.924 | 0.910 | 0.931 |
| age | #<br>patients | #FA | #days | Overall<br>FAR | CI overall FAR | Mean<br>FAR | CI mean FAR |
|-------|---------------|-----|-------|----------------|----------------|-------------|-------------|
| 6-12 | 39 | 89 | 67.02 | 1.33 | 0.82 1.98 | 1.79 | 0.99 2.73 |
| 13-21 | 41 | 129 | 93.98 | 1.37 | 0.75 2.17 | 1.47 | 0.72 2.41 |
| 6-21 | 80 | 218 | 161 | 1.35 | 0.92 1.87 | 1.63 | 1.06 2.28 |
| >21 | 61 | 165 | 247.6 | 0.67 | 0.43 0.95 | 0.76 | 0.43 1.17 |
Tahla 2
1 A conservative correction was applied to PPA to account for extreme probability values (PPA ~= 1) and the presence of multiple seizures for some patients (Chew, The American Statistician, 25(5), 47-50, 1971; Saha et al, Int J Biostat., 12(2), 2016).
{8}------------------------------------------------
Conclusion:
The subject device has the same intended use as the predicate device, and differences in technological characteristics do not raise different questions of safety and effectiveness. On this basis, Embrace is substantially equivalent to the legally marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.