← Product Code [OWI](/productcode/OWI) · K181799

# ResCUBE™ Ligament Fixation System (K181799)

_Cousin Biotech Sas · OWI · Aug 29, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181799

## Device Facts

- **Applicant:** Cousin Biotech Sas
- **Product Code:** [OWI](/productcode/OWI.md)
- **Decision Date:** Aug 29, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3010
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ResCUBE System is a temporary implant for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications: 1. Spinal trauma surgery, that can be treated with sublaminar, interspinous or facet wiring techniques. 2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as scoliosis, kyphosis and spondylolisthesis. 3. Spinal degenerative surgery, as an adjunct to spinal fusions.

## Device Story

ResCUBE Ligament Fixation System is a temporary spinal implant; provides stabilization as bone anchor during bony fusion or fracture repair. Components include ligament for stabilization, malleable leader for surgical guidance, and low-profile connector to maintain tension. Used in spinal trauma, reconstructive, and degenerative surgeries; operated by surgeons. Device functions as mechanical fixation; surgeon tensions ligament to stabilize spinal levels. Benefits include temporary support during healing process. No software or AI components.

## Clinical Evidence

Bench testing only. Static and dynamic tension testing performed in accordance with ISO 14630:2012 and IEC 62366:2007. Results demonstrate compliance with predetermined specifications and substantial equivalence to predicate devices.

## Technological Characteristics

Materials: TA6V ELI titanium alloy (ASTM F136, ISO 5832-3), 304L stainless steel, PET sheath. Form factor: Ligament with malleable leader and low-profile connector. Energy source: None (mechanical). Sterilization: Shipped sterile. Connectivity: None. Software: None.

## Regulatory Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

## Predicate Devices

- NAJA™ Ligament Correction System ([K172206](/device/K172206.md))
- Songer Spinal Cable ([K922952](/device/K922952.md), [K941213](/device/K941213.md))
- JAZZ LOCK ([K153348](/device/K153348.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 29, 2018

Cousin Biotech S.A.S. Mr. Franck Pelletier Regulatory Affairs Director 8 Rue De L'abbe Bonpain Wervicq-Sud 59117 FRANCE

Re: K181799

Trade/Device Name: ResCUBE™ Ligament Fixation System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: OWI Dated: June 29, 2018 Received: July 5, 2018

Dear Mr. Pelletier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Ronald P. Jean -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K181799

Device Name ResCUBETM Ligament Fixation System

#### Indications for Use (Describe)

The ResCUBE System is a temporary implant for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and in the repair of bone fractures. The indications for use include the following applications:

- 1. Spinal trauma surgery, that can be treated with sublaminar, interspinous or facet wiring techniques.
2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as scoliosis, kyphosis and spondylolisthesis.

3. Spinal degenerative surgery, as an adjunct to spinal fusions.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary provided in accordance with 21 CFR §807.92(C)

| Submission Date:                     | June 29, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                           | Cousin Biotech S.A.S.<br>8 Rue De L'abbe Bonpain<br>59117 Wervicq-Sud, France                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Submitter Contact:                   | Ms. Mathilde Collet<br>Regulatory Affairs Engineer<br>Phone: +011 33 (3) 20144130<br>Email: m.collet@cousin-biotech.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Application<br>Correspondent:        | Mr. Franck Pelletier<br>Regulatory Affairs Director<br>Phone: +011 33 (3) 20144119<br>Email: f.pelletier@cousin-biotech.com                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Manufacturing Site:                  | Cousin Biotech S.A.S.<br>Allée des roses<br>F-59117 Wervicq-Sud, France                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade Name:                          | ResCUBE™ Ligament Fixation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Name:                 | Bone Fixation Cerclage, Sublaminar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification<br>Regulation:        | 21 CFR §888.3010 – Bone Fixation Cerclage, Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Product Code:                        | OWI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Substantially<br>Equivalent Devices: | K172206 - Cousin Biotech S.A.S. - NAJA™ Ligament Correction<br>System (Primary)<br>K922952, K941213 - Pioneer Surgical (now RTI Surgical) - Songer<br>Spinal Cable (Additional Predicate)<br>K153348 - Implanet S.A. - JAZZ LOCK (Additional Predicate)                                                                                                                                                                                                                                                                                                                                   |
| Device Description:                  | The ResCUBE™ Ligament Fixation System (ResCUBE System) is a<br>temporary spinal implant intended to provide temporary stabilization as<br>a bone anchor during the development of solid bony fusion and aid in<br>the repair of bone fractures. The ResCUBE System consists of the<br>following components: a ligament allowing correction and stabilization<br>of the levels, after tensioning; a malleable leader allowing guidance of<br>the ligament during surgery and attached to the ligament by a PET<br>sheath; and a low-profile connector holding the tension on the ligament. |

{4}------------------------------------------------

## 510(k) Summary provided in accordance with 21 CFR §807.92(C)

| Intended Use:             | The ResCUBE System is a temporary implant for use in orthopedic<br>surgery. The system is intended to provide temporary stabilization as a<br>bone anchor during the development of solid bony fusion and aid in the<br>repair of bone fractures. The indications for use include the following<br>applications: |                                                                                         |                                                                                      |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|                           | 1. Spinal trauma surgery, that can be treated with sublaminar,<br>interspinous or facet wiring techniques.                                                                                                                                                                                                       |                                                                                         |                                                                                      |
|                           | 2. Spinal reconstructive surgery, incorporated into constructs for the<br>purpose of correction of spinal deformities such as scoliosis, kyphosis<br>and spondylolisthesis.                                                                                                                                      |                                                                                         |                                                                                      |
|                           |                                                                                                                                                                                                                                                                                                                  | 3. Spinal degenerative surgery, as an adjunct to spinal fusions.                        |                                                                                      |
| Technology<br>Comparison: | The ResCUBE System employs the same technological characteristics<br>as the primary predicate device.                                                                                                                                                                                                            |                                                                                         |                                                                                      |
|                           | Characteristic                                                                                                                                                                                                                                                                                                   | Cousin Biotech S.A.S. / NAJATM<br>Ligament Correction System<br>(K172206)               | Cousin Biotech S.A.S. / ResCUBETM<br>Ligament Fixation System<br>(Proposed Device)   |
|                           | Implant<br>Components                                                                                                                                                                                                                                                                                            | Connector (with lateral hooks)<br>with screw.                                           | Connector (without lateral hooks) with<br>screw.                                     |
|                           |                                                                                                                                                                                                                                                                                                                  | Ligament with metal guide and<br>PET sheath, but no buckle.                             | Ligament with metal guide and PET<br>sheath, but no buckle.                          |
|                           |                                                                                                                                                                                                                                                                                                                  | (Ligament has a stiff section on<br>one end for guiding the band<br>through the spine.) | (Ligament has a stiff section on one end<br>for guiding the band through the spine.) |

Titanium alloy connector is

available in two (2) sizes of 5.5

and 6.0 mm compatible with the

TA6V ELI titanium alloy (ASTM

same diameter rod constructs.

F136 and ISO 5832-3).

304L stainless steel.

PET sheath.

Yes

Sizes

Metal Clamp and Screw

(Connector)

Shipped Sterile

Guide

Titanium alloy connector is available in

one (1) size because it is not used with

TA6V ELI titanium alloy (ASTM F136

rods. The body of the ResCUBE

and ISO 5832-3).

304L stainless steel.

PET sheath.

Yes

connector is identical to the NAJA connector less the lateral hooks.

{5}------------------------------------------------

# 510(k) Summary provided in accordance with 21 CFR §807.92(C)

### Summary of Performance Testing:

| Performance Testing<br>– Bench | Static and dynamic tension testing were performed on ResCUBE<br>System. The ResCUBE System was tested for performance in<br>accordance with its predetermined specifications and the following<br>standards:                                                                                     |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | • IEC 62366: 2007, Medical devices – Application of usability<br>engineering to medical devices; and                                                                                                                                                                                             |
|                                | • ISO 14630: 2012, Non-active surgical implants – General<br>requirements.                                                                                                                                                                                                                       |
|                                | Test results indicate that the ResCUBE System complies with its<br>predetermined specifications and with the standards.                                                                                                                                                                          |
| Conclusion                     | Verification and validation activities were conducted to establish the<br>performance and safety characteristics of ResCUBE System. The<br>results of these activities demonstrate that ResCUBE System is as safe,<br>as effective, and perform as well as or better than the predicate devices. |
|                                | Therefore, the ResCUBE System is considered substantially equivalent<br>to the predicate devices.                                                                                                                                                                                                |

---

**Source:** [https://fda.innolitics.com/device/K181799](https://fda.innolitics.com/device/K181799)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
