FreshLook Spherical (phemfilcon A) soft (hydrophilic) contact lenses are indicated for the correction of visual acuity in persons with nondiseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of up to 2.0 diopters that does not interfere with visual acuity. The FreshLook Spherical Colors, ColorBlends, and Dimensions lenses act to enhance or alter the apparent color of the eye. The eye care professional may prescribe the lens for frequent with daily removal for cleaning and disinfection. The lens may be disinfected using a chemical disinfection system.
Device Story
FreshLook soft contact lenses (phemfilcon A) provide optical correction for refractive errors; lenses available in spherical designs with optional color enhancement/alteration. Lenses manufactured via double-sided molding; composed of hydrophilic copolymer (HEMA and EOEMA); 55% water content. Supplied sterile in isotonic borate buffered saline with 0.005% Poloxamer; packaged in polypropylene blister shells with aluminum-based foil lidding. Prescribed by eye care professionals for daily wear with frequent replacement; patient removes lenses daily for cleaning and chemical disinfection. Modification involves alternate foil lidding material grade for primary packaging. Device benefits include vision correction and cosmetic eye color enhancement.
Clinical Evidence
No clinical data. Bench testing only. Stability, biocompatibility, and process validation testing were performed to verify that the modified packaging material meets all acceptance criteria and maintains equivalence to the predicate device.
Technological Characteristics
Material: phemfilcon A (HEMA/EOEMA copolymer); FDA Group 4 (ionic, >=50% water). Water content: 55%. Manufacturing: Double-sided molding. Sterilization: Steam (autoclave). Packaging: Polypropylene blister shell with aluminum-based foil lidding. Refractive index: 1.412. Oxygen transmissibility (Dk/L): 20 x 10^-9 (cm/s)(ml O2/ml*mmHg) at 35°C.
Indications for Use
Indicated for correction of visual acuity in persons with nondiseased eyes who are myopic or hyperopic, with up to 2.0 diopters of refractive astigmatism. Lenses may enhance or alter eye color. Prescribed for frequent replacement with daily removal for cleaning and disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 03, 2018
Alcon Laboratories, Inc. Andreas Friese Regulatory Project Director, GRA Alcon Vision Care Alcon/CIBA Vision GmbH Industriering 1 Grosswallstadt, Bavaria 63868 Germany
Re: K181796
Trade/Device Name: FreshLook Handling Tint, FreshLook Colors, FreshLook ColorBlends, FreshLook Dimensions Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: June 28, 2018 Received: July 5, 2018
Dear Andreas Friese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{2}------------------------------------------------
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Scott E. Steffen -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K181796
Device Name
FreshLook Handling Tint, FreshLook Colors, FreshLook ColorBlends, FreshLook Dimensions
#### Indications for Use (Describe)
FreshLook Spherical (phemfilcon A) soft (hydrophilic) contact lenses are indicated for the correction of visual acuity in persons with nondiseased eves that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of up to 2.0 diopters that does not interfere with visual acuity.
The FreshLook Spherical Colors, ColorBlends, and Dimensions lenses act to enhance or alter the apparent color of the eye.
The eye care professional may prescribe the lens for frequent with daily removal for cleaning and disinfection. The lens may be disinfected using a chemical disinfection system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) summary document has been prepared in accordance with section 21 CFR 807.92.
#### Submitter Information I.
| Company: | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX 76134-2099, USA |
|-----------------|----------------------------------------------------------------------------------|
| Contact Person: | Dr. Andreas Friese,<br>Regulatory Project Director |
| Phone: | +49 6022-240-514 |
| Fax: | +49 6022-240-512 |
| Email: | andreas.friese@alcon.com |
| Date Prepared: | August 03, 2018 |
#### Devices Subject to this 510(k) II.
| Trade Names: | Soft Contact Lenses for Daily Wear (phemfilcon A material).<br>FreshLook Handling Tint, FreshLook Colors, FreshLook<br>ColorBlends, FreshLook Dimensions |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Soft Contact Lenses |
| Classification Name: | Soft (Hydrophilic) Contact Lenses |
| Device Classification: | Class II [21 CFR 886.5925] |
| Product Code: | LPL |
### III. Predicate Device
The 510(k) devices are a modification of the same predicate devices, i.e. FreshLook Handling Tint, FreshLook Colors, FreshLook ColorBlends and FreshLook Dimensions (phemfilcon A) soft contact lenses, which are legally commercialized devices in the US under 510(k) P830037.
# IV. Device Description
FreshLook (phemfilcon A) are a family of soft contact lenses, intended for the optical correction of refractive error. The lenses are available in spherical designs and in the
{5}------------------------------------------------
following lens models: FreshLook Handling Tint, FreshLook Colors, FreshLook ColorBlends and FreshLook Dimensions.
FreshLook contact lenses are made of phemfilcon A lens material, a hydrophilic copolymer of 2-hydroxyethyl methacrylate (HEMA) and 2-ethoxyethyl methacrylate (EOEMA). The lenses have a water content of 55% by weight.
FreshLook Colors and FreshLook ColorBlends lenses are made of clear lens material, FreshLook Handling Tint and FreshLook Dimensions are made of tinted (green handling tint) lens material polymer to facilitate lens visibility for handling.
FreshLook Colors, FreshLook ColorBlends and FreshLook Dimensions lenses are printed with an intermittent coating containing color additives listed in the color additive provisions of 21 CFR. FreshLook Colors are available in the colors Blue, Green, Hazel, Misty Gray, Sapphire Blue and Violet. FreshLook ColorBlends are available in the colors Amethyst, Blue, Brown, Gray, Green, Honey, Pure Hazel, True Sapphire, Turquoise, Gemstone Green, Brilliant Blue and Sterling Gray. FreshLook Dimensions are available in the colors Caribbean Aqua, Pacific Blue and Sea Green.
FreshLook soft contact lenses are supplied sterile. The lenses immersed in 2.5 ml isotonic borate buffered saline solution are packaged in individual foil-blister packs and are terminally sterilized in a validated autoclave (moist heat, steam under pressure). The packaging saline contains 0.005% Poloxamer.
The foil-blister pack system consists of a thermo-formed array bottom made from polypropylene resin, sealed with an aluminum-based foil lidding. The blister packs are packaged into carton boxes available in different pack sizes.
#### Indications for Use V.
The 510(k) devices are a modification of the same predicate devices. The Indications for Use remain the same, as follows:
FreshLook Spherical (phemfilcon A) soft (hydrophilic) contact lenses are indicated for the correction of visual acuity in persons with nondiseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of up to 2.0 diopters that does not interfere with visual acuity.
The FreshLook Spherical Colors, ColorBlends, and Dimensions lenses act to enhance or alter the apparent color of the eye.
The eye care professional may prescribe the lens for frequent replacement with daily removal for cleaning and disinfection. The lens may be disinfected using a chemical disinfection system.
{6}------------------------------------------------
# VI. Comparison of Technological Characteristics with the Predicate Device
The proposed device modification involves adding an alternate foil lidding material grade from an existing approved supplier for use in the primary packaging (i.e. foilblister pack container system) of FreshLook family soft contact lenses.
The following matrix (Table 1) summarizes the characteristics of the modified device as compared to the predicate device.
| Element of<br>Comparison | Predicate Device | Modified Device |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Administrative / Regulatory Information | | |
| 510(k) Number | 510(k) (P830037) | K181796 |
| Product Name(s) | FreshLook Handling Tint,<br>FreshLook Colors, FreshLook<br>ColorBlends, FreshLook<br>Dimensions | Same |
| Device<br>Classification<br>Information | Class II, Soft (Hydrophilic)<br>Contact Lenses, 21 CFR<br>886.5925 | Same |
| Indications For Use Information | | |
| Intended Use | Daily wear frequent<br>replacement soft contact<br>lenses for the optical<br>correction of refractive error | Same |
| Technology Information | | |
| Lens Material | phemfilcon A | Same |
| Material<br>Classification | FDA Group 4 (≥ 50% H2O,<br>ionic polymer) | Same |
| Water Content | 55% | Same |
| Visibility Tint | FreshLook Handling Tint and<br>FreshLook Dimensions: Light<br>green tint; FreshLook Colors<br>and FreshLook ColorBlends:<br>No tint | Same |
| Manufacturing<br>Method | Double Sided Molding | Same |
| Lens Designs | Spherical | Same |
| Sterilization | Steam sterilization, validated<br>autoclave | Same |
| | | Table 1: Substantial Equivalence Comparison |
|--|--|---------------------------------------------|
|--|--|---------------------------------------------|
{7}------------------------------------------------
| Element of<br>Comparison | Predicate Device | Modified Device |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Primary<br>Packaging System | Foil blister pack container<br>system: Polypropylene blister<br>shell sealed with an<br>aluminum-based foil lidding | Same |
| Primary<br>Packaging: Blister<br>Shell | Injection-molded<br>polypropylene blister shell | Same |
| Primary<br>Packaging: Foil<br>Lidding | Constantia-Hueck foil lidding<br>- Current grade | Constantia-Hueck foil lidding<br>– Modified grade |
| Package Storage /<br>Saline Solution | Isotonic, borate buffered<br>saline containing 0.005%<br>Poloxamer | Same |
| Performance Specifications including any Testing | | |
| Refractive Index | 1.412 | Same |
| Light<br>Transmittance | 95% minimum at 600 nm | Same |
| Oxygen<br>Transmissibility | Dk/L = 20 x 10-9 (cm/s)[ml<br>O2/(ml・mmHg] at 35 °C | Same |
| Biocompatibility | Biocompatible as confirmed<br>by biocompatibility testing | Same |
| Shelf-life | 60 months as confirmed by<br>shelf-life stability testing | Same |
Table 1: Substantial Equivalence Comparison
In accordance with the criteria for claims of substantial equivalence in the FDA guidance Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, May 1994, the information provided supports the claim of substantial equivalence to a lens with an existing USAN and the same manufacturing process.
### VII. Performance Data
Performance testing was conducted in consideration of the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses. The following performance data are provided in support of the substantial equivalence determination:
{8}------------------------------------------------
#### Non-clinical Testing
Successful stability and biocompatibility testing as well as process validation were completed for the modified device to verify equivalence to the predicate device. This resulted in all acceptance criteria being met.
#### Clinical Testing
The scope of the device modification did not require clinical testing to establish safety and effectiveness of the modified device.
### VIII. Conclusions
The cumulative results of all performance testing demonstrate the safety, efficacy and performance of the modified device and, thus, substantial equivalence to the predicate device.
Freshlook brand (phemfilcon A) soft contact lenses in modified primary packaging are substantially equivalent to the predicate lenses of material properties, biocompatibility, clinical performance, and indications for use.
Any differences which may exist between the modified and the predicate device do not adversely affect the established performance characteristics and safety and effectiveness profile.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.