← Product Code [FMI](/productcode/FMI) · K181730

# BD Vacutainer Eclipse UltraFill Blood Collection Needle (K181730)

_Becton, Dickinson and Company · FMI · Mar 1, 2019 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181730

## Device Facts

- **Applicant:** Becton, Dickinson and Company
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Mar 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle is a sterile, single use, medical device specifically intended to be used by trained healthcare professionals in accordance with the instructions for use for the collection of multiple blood samples into evacuated blood collection tubes, for the purpose of in vitro diagnostic testing. The needle is designed with an attached safety shield, which can be activated to cover the needle immediately after venipuncture to provide protection from accidental needle sticks.

## Device Story

Device is a sterile, single-use blood collection needle featuring a double-ended stainless steel cannula, threaded polystyrene hub, flow-interrupting sleeve, and integrated safety shield. Used by trained healthcare professionals in clinical settings for venipuncture. Operation involves inserting the intravenous end into a patient's vein; the non-patient end is then used to puncture evacuated blood collection tubes. The internal sleeve acts as a non-return valve, allowing multiple tube collections. After venipuncture, the user manually pivots the safety shield over the needle to prevent accidental needle sticks. The device is available in non-integrated or pre-attached holder configurations. The primary benefit is efficient, multi-sample blood collection with enhanced protection against sharps injuries. The ultra-thin wall cannula design increases inner diameter to improve blood fill time compared to standard thin-wall needles.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified through mechanical testing (integrity, penetration force, fill time, pull force, activation force, snag resistance, sleeve function, stiffness, breakage resistance, and blood splatter) and simulated use assessment for the safety shield. Hemolysis assessment performed on venous blood samples. Compliance with ISO 23908, ISO 9626, ISO 10993 series, ISO 11137, ISO 11737, ISO 6009, and IEC 62366-1.

## Technological Characteristics

Materials: 304 stainless steel cannula, silicone-based lubricants, polystyrene hub/collar. Design: Double-ended cannula, safety shield, sleeve valve. Dimensions: 1 1/4 inch length, ultra-thin wall (UTW). Sterilization: Gamma irradiation (SAL 10^-6). Standards: ISO 23908 (sharps protection), ISO 9626 (tubing), ISO 10993 (biocompatibility), ISO 11137/11737 (sterilization), ISO 6009 (color coding), IEC 62366-1 (usability).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- BD Vacutainer® Eclipse™ Blood Collection Needle ([K982541](/device/K982541.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Becton, Dickinson and Company % John Smith Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004

Re: K181730

Trade/Device Name: BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: February 1, 2019 Received: February 1, 2019

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sarah B. Mollo -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) k181730

Device Name

BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle

Indications for Use (Describe)

The BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle is a sterile, single use, medical device specifically intended to be used by trained healthcare professionals in accordance with the instructions for use for the collection of multiple blood samples into evacuated blood collection tubes, for the purpose of in vitro diagnostic testing. The needle is designed with an attached safety shield, which can be activated to cover the needle immediately after puncture to provide protection from accidental needle sticks.

Type of Use (Select one or both, as applicable)

| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### K181730 510(k) SUMMARY

## BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle

#### Submitter

Becton Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417-1885

Phone: (201) 847-4570 Facsimile: (201) 847-4858

Contact Person: Eileen Hiller, Senior Staff Regulatory Affairs Specialist

Date Prepared: March 1, 2019

Name of Device: BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle

Common or Usual Name: Blood collection needle

Classification Name: Hypodermic Single-Lumen Needle

Classification Regulation: 21 C.F.R. § 880.5570

Regulatory Class: Il

Product Code: FMI

Classification Panel: General Hospital

Predicate Device: BD Vacutaine® Eclipse™ Blood Collection Needle (K982541)

#### Device Description

The BD Vacutainer® Eclipse™ Blood Collection Needle consists of a double-ended hollow stainless steel cannula, a threaded polystyrene hub, a sleeve that interrupts blood flow between filling multiple tubes, a polystyrene collar, and a safety shield which can be activated to cover the needle immediately after venipuncture to protect against accidental needle sticks. The device is available in two versions: non-integrated holder and Pre-Attached Holder. The Pre-Attached model features a pre-attached needle holder, whereas the non-integrated model requires the user to obtain a compatible holder.

The device is single-use and supplied sterile. The stainless steel cannula and silicone cannula lubricants are classified as externally communicating with limited (≤24 hours) duration contact; all other device components have limited (≤24 hours) duration surface contact with intact skin of the health care professional user and the patient. Once the vein has been accessed by puncturing with the intravenous end of the cannula, the healthcare professional will begin the blood collection process by placing evacuated blood collection tubes on the cannula's non-patient end, where the sleeve acts as a non-return valve and allows for more than one tube to collect blood from the vein puncture.

{4}------------------------------------------------

#### Intended Use / Indications for Use

The BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle is a sterile, single use, medical device specifically intended to be used by trained healthcare professionals in accordance with the instructions for use for the collection of multiple blood samples into evacuated blood collection tubes, for the purpose of in vitro diagnostic testing. The needle is designed with an attached safety shield, which can be activated to cover the needle immediately after venipuncture to provide protection from accidental needle sticks.

## Comparison with Predicate Device

The BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle has the same intended use and principles of operation, and similar indications for use and technological characteristics, as the previously cleared BD Vacutainer® Eclipse™ Blood Collection Needle (K982541). The subject device has the same intended use as the predicate, namely multiple-sample blood collection via venipuncture. Its indications for use are also substantively the same: the device is indicated for use in clinical environments where venipuncture is performed. The minor differences in the wording of the indications for use statement-namely, the inclusion of language specifying that the device is intended to be used by trained healthcare professionals in accordance with the instructions for use for the collection of multiple blood samples into evacuated blood collection tubes, for the purpose of in vitro diagnostic testing-are for enhanced clarity; they do not affect the device's diagnostic/therapeutic use or raise new questions of safety or effectiveness as compared to the predicate, because they do not alter how the device is used in any way.

Venous flow through the inner diameter of the cannula into attached blood collection tubes is the technological principle for both the subject and predicate devices. The subject and predicate devices are based on the following same technological elements:

- . Same overall design, including safety shield feature and use of a holder
- Same cannula material, geometry and gauge (Outer Diameter)
- . Same method of use and very limited duration of contact (≤ 3 min.) with patient's blood

The primary technological differences between the subject and predicate devices are an increase in the cannula's inner diameter (ultra-thin wall), and introduction of a Pre-Attached Holder configuration with corresponding packaging as another option for the health care practitioner and added more detail and clarifications to the labeling. Minor differences in certain component materials/suppliers have been appropriately verified not to impact device performance. A table comparing the key features of the subject and predicate devices is provided below.

In addition, the minor differences in technological characteristics do not raise any new types of safety or effectiveness questions. Furthermore, performance data (design verification testing) demonstrates that the modified device is as safe and effective as the cited predicate. Thus, the BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle is substantially equivalent to its predicate device.

{5}------------------------------------------------

|                                    | Subject Device: BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device: BD Vacutainer® Eclipse™ Blood Collection Needle (K982541)                                                                                                                                                                                                                                      | Discussion / Comment                                                                                            |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Regulation Number                  | 21 C.F.R. § 880.5570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 21 C.F.R. § 880.5570                                                                                                                                                                                                                                                                                             | Same                                                                                                            |
| Classification Name                | Needle, Hypodermic, Single-Lumen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Needle, Hypodermic, Single-Lumen                                                                                                                                                                                                                                                                                 | Same                                                                                                            |
| Regulatory Class                   | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Class II                                                                                                                                                                                                                                                                                                         | Same                                                                                                            |
| Product Code                       | FMI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | FMI                                                                                                                                                                                                                                                                                                              | Same                                                                                                            |
| Intended Use / Indications for Use | The BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle is a sterile, single use, medical device specifically intended to be used by trained healthcare professionals in accordance with the instructions for use for the collection of multiple blood samples into evacuated blood collection tubes, for the purpose of in vitro diagnostic testing. The needle is designed with an attached safety shield, which can be activated to cover the needle immediately after venipuncture to provide protection from accidental needle sticks. | The BD Vacutainer® Eclipse™ Blood Collection Needle is a sterile, multiple sample, single-use device for blood collection. The needle is designed with an attached safety shield, which can be activated to cover the needle immediately after venipuncture to provide protection from accidental needle sticks. | The indications for use have been updated for enhanced clarity                                                  |
| Sterility                          | Gamma Irradiation to SAL of 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                  | Same                                                                                                            |
| Single Use?                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                  | Same                                                                                                            |
| Cannula Material                   | 304 Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                  | Same                                                                                                            |
| Cannula Lubricants                 | Silicone-based oils                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                  | Same                                                                                                            |
| Needle Wall Thickness              | Ultra-Thin Wall (UTW)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Thin Wall (TW)                                                                                                                                                                                                                                                                                                   | The needle wall thickness has been decreased to increase the inner-diameter of the needle for faster fill time. |
| Needle Length                      | 1 ¼ inch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                  | Same                                                                                                            |
| Safety Shield Activation Method    | User activates with the thumb after venipuncture, shield "pivots" around the hinge, locking into place at the needle base and at the cannula                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                  | Same                                                                                                            |
| Shelf Life                         | 5 years (Eclipse™ UltraFill™);<br>3 years (Eclipse™ UltraFill™<br>Pre-Attached)                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 5 years                                                                                                                                                                                                                                                                                                          | Same                                                                                                            |

{6}------------------------------------------------

## Performance Data

Design verification and validation were performed to ensure that the Eclipse™ UltraFill™ Needle meets its performance specifications and demonstrates equivalence to the specified predicate device.

- . Samples of venous blood obtained with the BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle were assessed to verify performance as measured by analytical indicators of hemolysis.
- . Simulated use assessment was conducted to validate the performance of the safety shield feature of the BD Vacutainer® Eclipse™ UltraFill™ Needle.
- . Mechanical verification testing was performed to assess product characteristics including mechanical integrity, penetration force, tube fill time, cannula pull force, safety shield activation force, safety shield override force (snag resistance), sleeve function, 3 point stiffness, resistance to breakage, and external blood splatter. The results confirmed that the 21- and 22-gauge BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needles meet preestablished functional design requirements and that the changes incorporated in the subject device as compared to the predicate do not impact the product's ability to be used safely and perform as intended.

The device complies with the following recognized consensus standards:

- . ISO 23908, Sharps Injury Protection - Requirements and Test Methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
- . ISO 9626, Stainless steel needle tubing for the manufacture of medical devices (including amendment I: 2001)
- . AAMI/ANSI/ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . ISO 11137-1, Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11737-1, Sterilization of health care products - Microbiological methods, Part 1: Estimation of the population of microorganisms
- . ISO 11737-2, Sterilization of health care products - Microbiological methods, Part 2: Tests for sterility performed in the validation of sterilization process
- ISO 10993-2, Biological evaluation of medical devices - Part 2: Animal welfare
- . ISO 10993-3, Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- . ISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood
- ISO 10993-5, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity ●
- ISO 10993-6, Biological evaluation of medical devices – Part 6: Tests for local effect affer implantation

{7}------------------------------------------------

- . ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- . ISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- . ISO 10993-12, Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
- . ISO 10993-13. Biological evaluation of medical devices - Part 13: Identification and quantification of degradation products from polymeric medical devices
- ISO 10993-15, Biological evaluation of medical devices Part 15: Identification and . quantification of degradation products from metals and alloys
- . ISO 10993-17、Biological evaluation of medical devices – Part 17: Establishment of allowable limits for leachable substances
- . ISO 10993-18, Biological evaluation of medical devices - Part 18: Chemical characterization of materials
- . ISO 6009, Hypodermic needle for single use - Color coding for identification
- . IEC 62366-1, Medical devices - Part 1: Application of usability engineering to medical devices

## Conclusion

The testing performed demonstrates the BD Vacutainer® Eclipse™ UltraFill™ Blood Collection Needle is substantially equivalent to its predicate device.

---

**Source:** [https://fda.innolitics.com/device/K181730](https://fda.innolitics.com/device/K181730)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
