← Product Code [HQF](/productcode/HQF) · K181682

# VISULAS green (K181682)

_Carl Zeiss Meditec, AG · HQF · Mar 6, 2019 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181682

## Device Facts

- **Applicant:** Carl Zeiss Meditec, AG
- **Product Code:** [HQF](/productcode/HQF.md)
- **Decision Date:** Mar 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4390
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

VISULAS green is intended for use in photocoagulation of tissues in the posterior and anterior segments of the eye including trabeculoplasty and iridotomy.

## Device Story

VISULAS green is a 532 nm frequency-doubled solid-state ophthalmic laser system. It delivers thermal energy to ocular tissue via single-spot or multi-spot modes to perform photocoagulation. The system comprises a laser console, touch control panel, foot switch, instrument table, and various applicators (laser slit lamp, laser indirect ophthalmoscope, endoprobes). Operated by clinicians in clinical settings, the device allows for precise control of laser power (50-1500 mW) and pulse duration (10-2500 ms). The output is used to treat ocular diseases by inducing thermal coagulation. The device provides therapeutic benefits by enabling targeted tissue treatment in the eye.

## Clinical Evidence

No clinical data, animal testing, or bench testing were required to establish substantial equivalence. The device relies on software verification and validation testing and compliance with recognized safety standards (ANSI/AAMI ES60601-1, IEC 60601-1-2, ANSI Z80.36, IEC 60825-1, IEC 60601-2-22, IEC 62133).

## Technological Characteristics

532 nm frequency-doubled solid-state laser; 50-1500 mW power; 10-2500 ms pulse duration; 50-1000 μm spot size. Includes laser console, touch control panel, foot switch, and applicators (LSL, VISULINK 532/U, Endoprobes, LIO). Complies with ANSI/AAMI ES60601-1, IEC 60601-1-2, ANSI Z80.36, IEC 60825-1, IEC 60601-2-22, and IEC 62133.

## Regulatory Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

## Predicate Devices

- VISULAS 532s Laser System with the VITE option ([K100035](/device/K100035.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 6, 2019

Carl Zeiss Meditec AG % Lisa Granev Senior Consultant Biologics Consulting Group, Inc. 1555 King Street. Suite 300 Alexandria, VA 22314

Re: K181682

Trade/Device Name: VISULAS green Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF Dated: January 31, 2019 Received: January 31, 2019

Dear Lisa Graney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely yours, Alexander Beylin -S 2019.03.06 10:50:35 -05'00' for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K181682

Device Name VISULAS green

Indications for Use (Describe)

VISULAS green is intended for use in photocoagulation of tissues in the posterior and anterior segments of the eye including trabeculoplasty and iridotomy.

| Type of Use (Select one or both, as applicable)                                                     |                                                                                                    |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92), the 510(k) Summary for VISULAS green is provided below.

#### 1. SUBMITTER

| Applicant:                | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>D-07745 Jena<br>Germany<br>Phone: +49 3641 2200                                                                                                    |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                  | Dr. Christian Münster<br>Director Regulatory and Clinical Affairs<br>Phone: +49 7364 20 6985<br>christian.muenster@zeiss.com                                                                              |
| Submission Correspondent: | Lisa Graney<br>Senior Consultant<br>Biologics Consulting Group<br>1555 King Street, Suite 300<br>Alexandria, VA 22314<br>lgraney@biologicsconsulting.com)<br>Phone: (571) 777-9518<br>Fax: (703) 548-7457 |
| Date Prepared:            | June 25 2018                                                                                                                                                                                              |

### 2. DEVICE

| Device Trade Name:  | VISULAS green                     |
|---------------------|-----------------------------------|
| Device Common Name: | Laser, Ophthalmic                 |
| Classification Name | 21 CFR 886.4390, Ophthalmic Laser |
| Regulatory Class:   | Class II                          |
| Product Code:       | HQF                               |

### 3. PREDICATE DEVICE

Predicate Device:

VISULAS 532s Laser System with the VITE option (K100035)

{4}------------------------------------------------

### DEVICE DESCRIPTION 4.

VISULAS green is an ophthalmic laser with a 532 nm wavelength used for standard photocoagulation treatments of ocular tissue. VISULAS green is operated in single-spot mode or in multi-spot mode.

VISULAS green consists of a laser console, touch control panel, optional laser light applicators such as laser slit lamp or laser indirect ophthalmoscope, foot switch and instrument table. The device can also be used with various accessories, such as laser endoprobes, or the SL cam 5.0.

### ડ. INTENDED USE/INDICATIONS FOR USE

VISULAS green is intended for use in photocoagulation of tissues in the posterior and anterior segments of the eye, including trabeculoplasty and iridotomy.

### TECHNOLOGICAL COMPARISON 6.

A technological comparison between the subject and predicate devices is shown in Table 1. The cited predicate device is VISULAS 532s Laser System with the VITE option (herein after called VISULAS 532s), cleared under K100035.

{5}------------------------------------------------

|                               | VISULAS green                                                                                                                                                | VISULAS 532s                                                                                                                                                                                                                                                                                                                                                                                          |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number                      | Subject Device                                                                                                                                               | K100035                                                                                                                                                                                                                                                                                                                                                                                               |
| Legal manufacturer            | Carl Zeiss Meditec AG                                                                                                                                        | Carl Zeiss Meditec AG                                                                                                                                                                                                                                                                                                                                                                                 |
| Indications for Use Statement | VISULAS green is intended for use in photocoagulating ocular<br>tissues in the treatment of diseases of the eye, including<br>trabeculoplasty and iridotomy. | The VISULAS 532s Laser System with the VITE option is<br>intended for use in retinal, pan-retinal, focal and grid<br>photocoagulation of ocular tissues in the treatment of diseases of<br>the eye including:<br>-Proliferative and non-proliferative diabetic retinopathy<br>-Macular edema<br>-Branch and central retinal vein occlusion<br>-Lattice degeneration<br>-Retinal tears and detachments |
|                               | General Laser Specifications                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                       |
| Laser type                    | solid state laser, frequency-doubled                                                                                                                         | solid state laser, frequency-doubled                                                                                                                                                                                                                                                                                                                                                                  |
| Wavelength                    | 532 nm                                                                                                                                                       | 532 nm                                                                                                                                                                                                                                                                                                                                                                                                |
| Power (cw laser)              | 50 to 1500 mW                                                                                                                                                | 50 to 1500 mW                                                                                                                                                                                                                                                                                                                                                                                         |
| Length of pulse               | 10 - 2500 ms and cw                                                                                                                                          | 10 - 2500 ms and -cw                                                                                                                                                                                                                                                                                                                                                                                  |
| Repeat Mode                   | Yes                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                   |
| Laser spot size settings      | 50 μm to 1000 μm                                                                                                                                             | 50 μm to 1000 μm                                                                                                                                                                                                                                                                                                                                                                                      |
| Aiming beam                   | 620-650 nm                                                                                                                                                   | 620-650 nm                                                                                                                                                                                                                                                                                                                                                                                            |
| Types of laser applicators    | LSL, VISULINK 532/U, Endoprobes, LIO                                                                                                                         | LSL, VISULINK 532/U, Endoprobes, LIO                                                                                                                                                                                                                                                                                                                                                                  |
| Multi spot treatment mode     | Yes                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                   |
| Conditions of Use             |                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                       |
| Site in the body              | Eye                                                                                                                                                          | Eye                                                                                                                                                                                                                                                                                                                                                                                                   |
| Principle of operation        | Photocoagulation of ocular tissue                                                                                                                            | Photocoagulation of ocular tissue                                                                                                                                                                                                                                                                                                                                                                     |

### Table 1. Technological characteristics comparison table of VISULAS green and VISULAS 532s

{6}------------------------------------------------

VISULAS green and the predicate device are intended as ophthalmic lasers for photocoagulation. The indications for use are within the same intended use as the predicate device and do not raise different questions of safety and effectiveness.

VISULAS green and the predicate laser system have minor parameter differences but share the same fundamental principle of operation:

Photocoagulation: Continuous wave operation in which thermal energy created by the absorption of the laser energy by ocular tissue causes coagulation.

VISULAS green is substantially equivalent to the predicate laser system presented in the 510(k) premarket notification with respect to indications for use and technological characteristics. Differences between the subject device and the predicate device do not raise any new issues of safety or effectiveness.

### 7. PERFORMANCE DATA

# Biocompatibility Testing

The materials and manufacturing of the subject device is identical to the predicate device and currently marketed devices. Therefore, no new biocompatibility testing is needed for this device.

# Laser safety, electrical safety and electromagnetic compatibility (EMC)

VISULAS green was evaluated against the following requirements and was found to comply with:

- ANSI/AAMI ES60601-1:2005/(R) 2012 ●
- IEC 60601-1-2:2014 ●
- ANSI Z80.36:2016 ●
- . IEC 60825-1:2007
- IEC 60601-2-22:2012
- IEC 62133: 2012 ●

# Software Verification and Validation Testing

The software of this device is considered as a "Major" level of concern. Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification and validation of VISULAS green demonstrated that the product works as designed.

# Bench Testing

Not Applicable. Bench testing was not necessary to establish the substantial equivalence of this device.

{7}------------------------------------------------

# Animal Testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

# Clinical Data

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

### CONCLUSION 8.

The VISULAS green is substantially equivalent to the predicate device, VISULAS 532s (K100035).

The VISULAS green is similar in technological characteristics, performance, principles of operation (photocoagulation) and has similar indications for use as the predicate device. Any differences between the proposed device and the predicate device do not raise any new issues of safety or effectiveness. Thus VISULAS green is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K181682](https://fda.innolitics.com/device/K181682)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
