← Product Code [OWB](/productcode/OWB) · K181670

# Alphenix, INFX-8000V/B, V8.0 (K181670)

_Canon Medical Systems Corporation · OWB · Jul 25, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181670

## Device Facts

- **Applicant:** Canon Medical Systems Corporation
- **Product Code:** [OWB](/productcode/OWB.md)
- **Decision Date:** Jul 25, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This device is a digital radiography/fluoroscopy system used in a diagnostic interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

## Device Story

Alphenix INFX-8000V/B V8.0 is a digital radiography/fluoroscopy system for interventional angiography. System comprises C-arm/Ω-arm, X-ray tube, beam limiter, X-ray receptor, controller, workstation, and patient table. Used in clinical interventional settings by physicians to visualize blood vessels in heart, brain, abdomen, and lower extremities. Inputs include X-ray radiation; system processes signals via software to produce real-time dynamic images and static radiographs. Features include tablet console for remote operation, AlphaCT (CBCT) programs, Spot ROI fluoroscopy, multi-phase CBCT, and fluoroscopy roadmapping. Output displayed on monitors for physician guidance during minimally invasive procedures. Benefits include improved workflow, image quality, and procedural navigation. System supports high-speed biplane acquisition and automated power management.

## Clinical Evidence

Bench testing only. Verification and validation testing conducted per IEC standards. Phantom testing performed to verify image metrics related to quality improvements. Clinical images deemed unnecessary via design control and risk management activities.

## Technological Characteristics

Digital radiography/fluoroscopy system; C-arm/Ω-arm configuration. Components: X-ray tube, beam limiter, aSi/CMOS hybrid flat panel detector (TFP-1200C). Energy source: Ionizing radiation. Connectivity: Networked workstation, tablet console. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-43, IEC 62304, IEC 62366. Software: V8.0, moderate level of concern.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Infinix, INFX-8000V, V7.0 ([K172863](/device/K172863.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 25, 2018

Canon Medical Systems Corporation % Ms. Janine Reyes Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive TUSTIN CA 92780

Re: K181670

Trade/Device Name: Alphenix, INFX-8000V/B, V8.0 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB Dated: June 22, 2018 Received: June 25, 2018

Dear Ms. Reves:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hole 2. Nils

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K181670

Device Name

Alphenix, INFX-8000V/B, V8.0

Indications for Use (Describe)

This device is a digital radiography/fluoroscopy system used in a diagnostic interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Type of Use (Select one or both, as applicable)

| <label>Prescription Use (Part 21 CFR 801 Subpart D)</label> | <input type="checkbox"/> |
|-------------------------------------------------------------|--------------------------|
| <label>Over-The-Counter Use (21 CFR 801 Subpart C)</label>  | <input type="checkbox"/> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# ON MEDICAL SYSTEMS USA, INC.

*Made For life*

## 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92

## 1. CLASSIFICATION and DEVICE NAME

| Classification Name    | Image-Intensified Fluoroscopic X-ray System |
|------------------------|---------------------------------------------|
| Product Code           | OWB                                         |
| Regulation Number      | 21 CFR 892.1650                             |
| Regulatory Class       | Class II                                    |
| Trade Proprietary Name | Alphenix, INFX-8000V/B, V8.0                |

### 2. SUBMITTER'S NAME

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550

## 3. OFFICIAL CORRESPONDENT

Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance

## 4. CONTACT PERSON, U.S. AGENT and ADDRESS

### Contact Person

Janine F. Reyes Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7853 Fax: (714) 730-1310 jfreyes@us.medical.canon

## 5. MANUFACTURING SITE

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan

#### 6. ESTABLISHMENT REGISTRATION 9614698

- 7. DATE PREPARED June 22, 2018
## 8. TRADE NAME(S)

Alphenix, INFX-8000V/B, V8.0

#### Official Correspondent/U.S. Agent

Paul Biggins Sr. Director, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7808 Fax: (714) 730-1310 pbiggins@us.medical.canon

{4}------------------------------------------------

# ANON MEDICAL SYSTEMS USA, INC.

Made For life

#### 9. DEVICE NAME

Interventional Fluoroscopic X-ray System

#### 10. CLASSIFICATION PANEL

Radiology

#### 11. DEVICE CLASSIFICATION

Class II (per 21 CFR 892.1650)

#### 12. PRODUCT CODE / DESCRIPTION

Product Code: OWB - Image-Intensified Fluoroscopic X-ray System

#### 13. PERFORMANCE STANDARD

This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].

#### 14. PREDICATE DEVICE

Infinix, INFX-8000V, V7.0 (K172863)

| Product                   | Marketed by               | 510(k) Number | Clearance Date    |
|---------------------------|---------------------------|---------------|-------------------|
| Infinix, INFX-8000V, V7.0 | Canon Medical Systems USA | K172863       | December 14, 2017 |

#### 15. REASON FOR SUBMISSION

Modification of a cleared device

#### 16. SUBMISSION TYPE

Traditional 510(k)

#### 17. DEVICE DESCRIPTION

The Alphenix, INFX-8000V/B, V8.0, is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm/Ω-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

#### 18. INDICATIONS FOR USE

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

#### 19. SUMMARY OF CHANGE(S)

This submission is to report the following items have been changed:

- V8.0 Software: System software changed from V7.0 to V8.0 for improved usability.
- . Tablet Console: used in combination with the Alphenix, INFX-8000V/B, V8.0, the following functions can be performed in the examination room using the tablet console: o Selecting a radiographic program
	- o Selecting an auto-positioning No.
	- o Selecting a function
	- o Performing auxiliary operations on the Angio Workstation
	- o Performing playback, pause, and frame feeding of dynamic images

{5}------------------------------------------------

## NON MEDICAL SYSTEMS USA, INC.

Made For life

o Switching dynamic image files and map image files

- New Examination Menu: enables frequently used functions to be registered into the . favorites menu for improved workflow.
- . Programming Restructuring: AlphaCT encompasses CBCT (Cone Beam CT) programs 3D-DA, LD-3D and LCI.
- . Spot ROI Fluoroscopy for Cardiac: Spot ROI Fluoroscopy is now available with 8" FPD. Spot ROI (region of interest) Fluoroscopy allows for peripheral region visualization outside of the area of interest. Spot ROI Fluoroscopy for 12″ x 12″ and 12″ x 16″ FPD was previously cleared under K152696.
- . Multi-phase CBCT (Cone Beam CT): Multiphase CBCT allows AlphaCT scans to be performed continuously over multiple phases.
- . Sleep Mode: To reduce the power consumption of the system automatically enters sleep mode. Note that it is possible to adjust the period of time before the system enters sleep mode.
- . UPS (Uninterruptible Power System) Connection Kit
- Specification for System Input Power: system input power is changed from 400V and 200V to 400V only.
- . Hybrid (aSi/CMOS) FPD (Flat Panel Detector), TFP-1200C, Image Quality Improvement: Cleared under K170909, hybrid type FPD, TFP-1200C, was modified to improve image quality and includes:

o Noise reduction for TFP-1200C

o Pixel area and FOV change of the aSi/CMOS region

- Space Improvement: To reduce machine room space requirements, racks were introduced for the chiller, water-cooled heat exchanger and large screen monitor system.
- . HSBP (High Speed BiPlane): For biplane systems in which images must be acquired simultaneously on the frontal and lateral sides, the HSBP (High Speed Biplane) function can be used in combination to achieve high-rate image acquisition. This function can only be used in combination with cardiac biplane systems with 8" FPD.
- . Fluoroscopy Roadmapping: DSA workflow improvement enables automatic display of subtracted images after completion of mask generation.

#### 20. SAFETY

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards; IEC 60601-2-43 and IEC60601-2-28. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report.

#### LIST OF APPLICABLE STANDARDS

- IEC 60601-1:2005 +A1:2012
- IEC 60601-1-2:2014
- IEC 60601-1-3:2008 +A1:2013 ●
- IEC 60601-1-6:2010 +A1:2013
- IEC 60601-2-28:2017 ●
- IEC 60601-2-43:2010 +A1:2017 ●

{6}------------------------------------------------

Made For life

## ION MEDICAL SYSTEMS USA, INC.

- IEC 62304:2006 +A1:2015

IEC 62366:2007 +A1:2014

#### 21. TESTING

Risk analysis and verification/validation testing conducted through bench testing demonstrate that the established specifications for the device have been met. Testing included conformity testing to IEC standards and phantom testing was conducted to verify image metrics related to improvements in image quality. Clinical images were deemed not necessary for the aforementioned improvements via design control and risk management activities.

●

This submission contains test data that demonstrates that the system modifications result in performance that is equal to or better than the predicate system. Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems.

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

Cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices " issued on October 2, 2014, is also included as part of this submission.

Additionally, the design controls used for this device included risk management and all known risks were mitigated to an acceptable level.

#### 22. SUBSTANTIAL EQUIVALENCE

The Alphenix, INFX-8000V/B, V8.0, is substantially equivalent to the INFX-8000V, V7.0, (K172863), marketed by Canon Medical Systems USA. The Alphenix, INFX-8000V/B, V8.0, includes system software change from V7.0 to V8.0, Tablet Console, New Examination Menu, Programing Restructuring, Spot ROI Fluoroscopy for 8" FPD, Multi-phase CBCT, Sleep Mode, UPS Connection Kit, Specification for System Input Power Change, TFP-1200C Image Quality Improvement, Space Improvement, HSBP (High Speed BiPlane) and Fluoroscopy Roadmapping DSA workflow improvement.

The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device. There are no new indications for use or intended use of the device.

#### 23. CONCLUSION

The Alphenix, INFX-8000V/B, V8.0, performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product the labeling. The modifications incorporated into the Alphenix, INFX-8000V/B, V8.0, do not change the indications for use or

{7}------------------------------------------------

# NON MEDICAL SYSTEMS USA, INC.

*Made For life*

the intended use of the device. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device. It is the contention of Canon Medical Systems Corporation that all new safety issues have been addressed in the design of this change and that adequate evidence of this is presented with this submission.

---

**Source:** [https://fda.innolitics.com/device/K181670](https://fda.innolitics.com/device/K181670)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
