← Product Code [GDW](/productcode/GDW) · K181657

# easyLCTM Linear Cutter Staplers and Loading Units fot Single Use; easyCSTM Circular Stapler for Single Use (K181657)

_Ezisurg Medical Co., Ltd. · GDW · Feb 1, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181657

## Device Facts

- **Applicant:** Ezisurg Medical Co., Ltd.
- **Product Code:** [GDW](/productcode/GDW.md)
- **Decision Date:** Feb 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The easyLC™ Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses. The easyCS™ Circular Stapler for Single Use has applications throughout the alimentary for end to end, end to side and side to side anastomoses.

## Device Story

Manual surgical staplers for gastrointestinal/alimentary procedures; linear cutter and circular stapler variants. Linear cutter places two double staggered rows of titanium staples while simultaneously cutting tissue; available in 60mm/80mm lengths with 2.5mm/3.8mm/4.8mm staple sizes; reloadable up to 8 times. Circular stapler places a double staggered circular row of titanium staples and uses a circular knife to resect excess tissue; available in 21mm-33mm diameters with 3.5mm/4.8mm staple sizes. Used by surgeons in clinical settings to facilitate tissue joining and resection. Devices provide mechanical staple formation and tissue division; benefit includes standardized, efficient anastomosis creation. Output is physical tissue closure and resection.

## Clinical Evidence

No clinical data. Bench testing only. Tests included pressure resistance, closed staple dimension, staple formation, and firing force on porcine stomach and intestine. Biocompatibility testing (cytotoxicity, irritation, sensitization, pyrogenicity) performed.

## Technological Characteristics

Manual surgical staplers; titanium staples. Linear cutter: 60-80mm length, 2.5-4.8mm staple sizes. Circular stapler: 21-33mm diameter, 3.5-4.8mm staple sizes. Standards: ISO 10993-1/5/10 (biocompatibility), ISO 11137-2 (radiation sterilization), ASTM F 88/F88M-09 (seal strength), USP <85>/<151> (endotoxins/pyrogens).

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- ENDOPATH™ and PROXIMATE™ Linear Cutters and Staplers ([K020779](/device/K020779.md))
- PROXIMATE Curved and Straight Intraluminal Staplers ([K983536](/device/K983536.md))
- PROXIMATE Curved and Straight Intraluminal Staplers ([K100723](/device/K100723.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 1, 2019

Ezisurg Medical Co., Ltd. % Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, CN 200120

Re: K181657

Trade/Device Name: easyLCTM Linear Cutter Staplers and Loading Units for Single Use; easyCS™ Circular Stapler for Single Use

Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: June 12, 2018 Received: June 22, 2018

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

$$\textbf{David }\textbf{Krausse}\text{-}\textbf{S}_{\begin{subarray}{c}\textbf{S}_{\begin{subarray}{c}\textbf{Diag}\\0\le\textbf{W}\end{subarray}}\text{-}\textbf{S}_{\begin{subarray}{c}\textbf{S}_{\begin{subarray}{c}\textbf{Diag}\\0\le\textbf{W}\end{subarray}}\text{-}\textbf{M}\end{subarray}}$$

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K181657

## Device Name

easyLCTM Linear Cutter Staplers and Loading Units for Single Use, easyCSTM Circular Stapler for Single Use

## Indications for Use (Describe)

The easyLCTM Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses.

The easyCSTM Circular Stapler for Single Use has applications throughout the alimentary for end to side and side to side anastomoses.

| Type of Use (Select one or both, as applicable) |                                                                                                                                                                                                                                                               |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                 | <span> <span style="padding-right: 20px;"> <b> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> </span> <span> <span style="font-size: 14px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

{3}------------------------------------------------

# Exhibit #2 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K181657

- 1. Date of Preparation: 12/29/2018
- 2. Sponsor Identification

## Ezisurg Medical Co., Ltd.

Rm.103, Bldg.2,No.1690 Cailun Road, China (Shanghai) Pilot Free Trade Zone, 201203 Shanghai, China.

Establishment Registration Number: Not yet registered.

Contact Person: Liang Xia Position: Non-active Device Senior Regulatory Affairs Supervisor Tel: +86-21-50456176 Fax: +86-21-50676156 Email: xialiang@ezisurg.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 3609253199 Email: info@mid-link.net

{4}------------------------------------------------

- Identification of Proposed Device 4.
Trade Name: easyLCTM Linear Cutter Staplers and Loading Units for Single Use easyCSTM Circular Stapler for Single Use Common Name: Stapler and Reload

Regulatory Information

Classification Name: Staple, Implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General & Plastic Surgery

Classification Name: Stapler, Surgical; Classification: I Subsequent Product Code: GAG; Regulation Number: 21CFR 878.4800 Review Panel: General & Plastic Surgery

Intended Use Statement:

The easyLC™ Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses.

The easyCS™ Circular Stapler for Single Use has applications throughout the alimentary for end to end, end to side and side to side anastomoses.

Device Description

The easyLCTM Linear Cutter Staplers and Loading Units for Single Use place two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The device is available in 60mm, 80mm length with reload in 2.5mm, 3.8mm and 4.8mm staple size to accommodate various tissue thicknesses. The device may be reloaded and fired up to 8 times in a single procedure.

The easyCS™ Circular Stapler for Single Use places a double staggered, circular row of titanium staples upon activation, which was achieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resect by the circular knife, and then a circular anastomosis is created. The stapler is available in 21mm, 29mm, 31mm and 33mm 

{5}------------------------------------------------

five specifications. Two staple sizes (3.5mm and 4.8mm) are available to accommodate various tissue thicknesses.

- ર . Identification of Predicate Devices
Predicate Device 1 510(k) Number: K020779 Product Name: ENDOPATH™ and PROXIMATE™ Linear Cutters and Staplers

Predicate Device 2 510(k) Number: K983536 Device Name: PROXIMATE Curved and Straight Intraluminal Staplers

Predicate Device 3 510(k) Number: K100723 Device Name: PROXIMATE Curved and Straight Intraluminal Staplers

- Non-Clinical Test Conclusion 6.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- > ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- > USP 38-NF 33 <85> Bacterial Endotoxins Tests
- > ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
- > ISO 11137-2:2013 Sterilization of health care products -Radiation- Part 2: Establishing the sterilization dose
- > USP 39-NF 34 <151> Pyrogen

Bench test was conducted on porcine stomach and intestine for both proposed device and predicate device to determine substantially equivalence. The bench tests include following tests

- > Pressure Resistance Test
- > Closed Staple Dimension Test
- > Staple Formation Test

{6}------------------------------------------------

> Force Required to Fire Stapler Test

Biocompatibility test was conducted on the proposed device, the test include cytotoxicity, irritation, skin sensitization and pyrogenicity.

- 7. Clinical Test Conclusion
No clinical study is included in this submission.

- 8. Substantially Equivalent (SE) Comparison

| Item                 | Proposed Device                                                                                                                                                                   | Predicate Device 1                                                                                                                                                                                                                                                                                                                  |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code         | GDW                                                                                                                                                                               | GDW                                                                                                                                                                                                                                                                                                                                 |
| Regulation Number    | 21 CFR 878.4750                                                                                                                                                                   | 21 CFR 878.4750                                                                                                                                                                                                                                                                                                                     |
| Intended Use         | The easyLCTM Linear Cutter Staplers<br>and Loading Units for Single Use have<br>application in gastrointestinal for<br>transection, resection, and/or creation of<br>anastomoses. | The PROXIMATE Linear and Vascular<br>Linear Cutters with Safety Lockout<br>have application in gastrointestinal,<br>gynecologic, thoracic and pediatric<br>surgery for transection, resection, and/or<br>creation of anastomoses and can be used<br>with staple line or tissue buttressing<br>materials, such as bovine pericardium |
| Cutting Mechanism    | Linear                                                                                                                                                                            | Linear                                                                                                                                                                                                                                                                                                                              |
| Operation Principle  | Manual                                                                                                                                                                            | Manual                                                                                                                                                                                                                                                                                                                              |
| Safety Mechanism     | Safety release for prevent from<br>mis-firing                                                                                                                                     | Safety release for prevent from<br>mis-firing                                                                                                                                                                                                                                                                                       |
| Suture Length        | 60mm, 80mm, 100mm                                                                                                                                                                 | 55, 75mm                                                                                                                                                                                                                                                                                                                            |
| Closed Staple Height | 1.0mm, 1.5mm, 2.0mm                                                                                                                                                               | 1.0mm, 1.5mm, 1.8mm, 2.0mm                                                                                                                                                                                                                                                                                                          |
| Closed staple form   | Image: staple form                                                                                                                                                                | Image: staple form                                                                                                                                                                                                                                                                                                                  |
| Endotoxin Limit      | 20 EU                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                |
| Labeling             | Conforms with 21 CFR 801                                                                                                                                                          | Conforms with 21 CFR 801                                                                                                                                                                                                                                                                                                            |

Table 1 Comparison for easyLC™ Linear Cutter Staplers and Loading Units for Single Use

{7}------------------------------------------------

| Item                 | Proposed Device                                                                                                                                            | Predicate Device 2<br>K983536                                                                                                                                             | Predicate Device 3<br>K100723                                                                                                                      |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code         | GDW                                                                                                                                                        | GDW                                                                                                                                                                       | GDW                                                                                                                                                |
| Regulation Number    | 21 CFR 878.4750                                                                                                                                            | 21 CFR 878.4750                                                                                                                                                           | 21 CFR 878.4750                                                                                                                                    |
| Intended Use         | The easyCSTM Circular Stapler for<br>Single Use has applications throughout<br>the alimentary for end to end, end to<br>side and side to side anastomoses. | The PROXIMATER Curved and<br>Straight Intraluminal Staplers have<br>application throughout the alimentary<br>for end to end, end to side and side to<br>side anastomoses. | The Disposable Circular Stapler has<br>application throughout the alimentary<br>tract for end-to-end, end-to-side anc<br>side-to-side anastomoses. |
| Cutting Mechanism    | Circular Knife                                                                                                                                             | Circular Knife                                                                                                                                                            | Circular Knife                                                                                                                                     |
| Operation Principle  | Manual                                                                                                                                                     | Manual                                                                                                                                                                    | Manual                                                                                                                                             |
| Safety Mechanism     | Safety Release is used for preventing<br>from mis-firing.                                                                                                  | Safety Release is used for preventing<br>from mis-firing.                                                                                                                 | Safety Release is used for preventing<br>from mis-firing.                                                                                          |
| Diameter             | 21mm, 25mm, 29mm, 31mm, 33mm                                                                                                                               | 21mm, 25mm, 29mm, 33mm                                                                                                                                                    | 25.5mm, 29.5mm                                                                                                                                     |
| Closed Staple Height | 1.0~2.5mm                                                                                                                                                  | 1.0~2.5mm                                                                                                                                                                 | 1.5~2.0mm                                                                                                                                          |
| Closed staple form   | Image: staple form                                                                                                                                         | Image: staple form                                                                                                                                                        | Image: staple form                                                                                                                                 |
| Endotoxin Limit      | 20 EU                                                                                                                                                      | 20EU                                                                                                                                                                      | 20EU                                                                                                                                               |
| Labeling             | Conforms with 21 CFR 801                                                                                                                                   | Conforms with 21 CFR 801                                                                                                                                                  | Conforms with 21 CFR 801                                                                                                                           |

|  | Table 2 Comparison for easy CSTM Circular Stapler for Single Use |
|--|------------------------------------------------------------------|
|--|------------------------------------------------------------------|

{8}------------------------------------------------

- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K181657](https://fda.innolitics.com/device/K181657)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
