Coated Vicryl Plus Antibacterial (Polyglactin 910) Absorable Suture
K181652 · ETHICON, Inc. · GAM · Sep 28, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181652
Device Name
Coated Vicryl Plus Antibacterial (Polyglactin 910) Absorable Suture
Applicant
ETHICON, Inc.
Product Code
GAM · General, Plastic Surgery
Decision Date
Sep 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues.
Device Story
Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is a sterile, synthetic, absorbable surgical suture; composed of 90% glycolide and 10% L-lactide copolymer; coated with a mixture of Polyglactin 370 and calcium stearate; contains IRGACARE® MP (triclosan) as an antibacterial agent. Used by surgeons in clinical settings for soft tissue approximation and ligation. The device functions as a mechanical support for tissue healing; the antibacterial agent is intended to inhibit bacterial colonization on the suture. The device is absorbed by the body over time, eliciting mild tissue reaction. No changes to materials, construction, or manufacturing compared to the predicate; labeling updated to include recent meta-analysis and guidelines regarding triclosan-coated sutures for surgical site infection reduction.
Clinical Evidence
No clinical data provided; bench testing not required as technological characteristics are identical to the predicate device.
Technological Characteristics
Synthetic absorbable suture composed of 90% glycolide/10% L-lactide copolymer. Coated with Polyglactin 370 and calcium stearate. Contains antibacterial agent IRGACARE® MP (triclosan) at ≤ 472 µg/m. Available in dyed (D&C Violet No. 2) or undyed (natural) forms. Nonantigenic and nonpyrogenic.
Indications for Use
Indicated for general soft tissue approximation and/or ligation. Contraindicated for ophthalmic, cardiovascular, and neurological tissues.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorable Suture (K132580)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 28, 2018
Ethicon, Inc. % Ms. Donna Marshall Regulatory Affairs Manager Route 22 West, P.O. Box 151 Somerville, New Jersey 08876-0151
Re: K181652
Trade/Device Name: Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: August 30, 2018 Received: August 31, 2018
Dear Ms. Marshall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K181652
Device Name
Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorable Suture
Indications for Use (Describe)
Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K181652 p. 1 of 2
Image /page/3/Picture/1 description: The image shows the logo for Ethicon, Inc., a Johnson & Johnson company. The word "ETHICON" is written in large, red, sans-serif letters, with the ",INC." in smaller, red letters to the right. Below the company name is the phrase "a Johnson+Johnson company" in a cursive font.
510(k) Summary
| Submitter: | Ethicon Inc. a Johnson & Johnson company<br>P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876-0151<br>USA |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Donna Marshall<br>Manager, Regulatory Affairs<br>Ethicon, Inc. a Johnson & Johnson company<br>Ph: (908) 541-3990<br>Fax: (908) 218-2595<br>e-mail: dmarsha2@its.jnj.com |
| Date Prepared: | September 26, 2018 |
|----------------|--------------------|
|----------------|--------------------|
| Device Trade Name: | Coated VICRYL™ Plus Antibacterial (Polyglactin 910)<br>Synthetic Absorbable Suture |
|----------------------|------------------------------------------------------------------------------------|
| Device Common Name: | Suture, Surgical, Absorbable |
| Class: | II |
| Classification Name: | Absorbable Poly(glycolide/L-lactide) Surgical Suture<br>(21 CFR 878.4493) |
| Product Code: | GAM |
| Panel: | General and Plastic Surgery Devices |
### Predicate Device:
| Device Trade Name | 510(k) Number |
|-------------------------------------------------------------------------|---------------|
| Coated Vicryl™ Plus Antibacterial<br>(Polyglactin 910) Absorable Suture | K132580 |
## Device Description:
Coated VicryI™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is a sterile, synthetic absorbable surgical suture (dyed and undyed) and is composed of a copolymer made from 90% glycolide and 10% L-lactide. Coated Vicryl™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is coated with a mixture composed of equal parts of a copolymer of glycolide and lactide (Polyglactin 370) and calcium stearate. The suture contains IRGACARE®* MP (triclosan), a broad spectrum antibacterial agent, at no more than 472 µg/m. The copolymers in
{5}------------------------------------------------
the product have been found to be nonantigenic, nonpyrogenic and elicit only a mild tissue reaction during absorption. The suture is available undyed(natural) or dyed (D&C Violet No. 2).
## Indications for Use:
Coated Vicry|™ Plus Antibacterial (Polyglactin 910) Absorbable Suture is indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues.
## Performance Data:
The technological characteristics of the proposed device is identical to the predicate device, therefore, performance data are not necessary to establish substantial equivalence.
### Substantial Equivalence:
The subject device is identical to the predicate device with respect to functionality, technological characteristics and intended use. There is no material, device construction, performance specification, packaging, sterilization or manufacturing process changes to the currently marketed devices. The device differs only in the labeling (Instructions for Use) that have been revised to include the references of the latest publication of meta-analysis and guidelines for consideration of the use of triclosan-coated sutures to lower surgical site infection rates in the Actions Section. The proposed device does not raise new questions of safety or effectiveness as the predicate device and therefore substantially equivalent.
### Conclusion:
Based on the intended use, fundamental scientific technology, technological characteristics and the intended use, the following subject device Coated Vicry]™ Plus Antibacterial (Polyglactin 910) Absorable Suture is considered to be substantially equivalent to the predicated device.
* Trademark IRGACARE®* MP (triclosan) "Registered Trademark of BASF Group"
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.