K181447 · Berpu Medical Technology Co., Ltd. · FMI · Mar 12, 2019 · General Hospital
Device Facts
Record ID
K181447
Device Name
Safety insulin needle for single use
Applicant
Berpu Medical Technology Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Mar 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.
Device Story
Safety insulin needle for single use; attaches to pen injector devices for subcutaneous insulin delivery. Device components include needle tube, hub, safety protective cover, self-destruction seat, spring, hub sheath, and safety seat. Safety mechanism automatically engages to reduce accidental needle sticks. Used in clinical or home settings by patients or healthcare providers. Device is EO sterilized. Output is the delivery of insulin; safety shield provides physical protection post-injection. Benefits include reduced risk of needle stick injuries.
Clinical Evidence
No clinical data. Bench testing only. Compliance demonstrated with ISO 7864:2016, ISO 9626:2016, ISO 11608-2, and ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, pyrogenicity). Seal strength and leak testing performed per ASTM F88/F88M-15 and ASTM F1929-15.
Technological Characteristics
Needle tube (304 stainless steel), hub/safety seat (polypropylene), safety cover (MABS). EO sterilization (SAL 10^-6). Complies with ISO 7864, ISO 9626, ISO 11608-2. Manual operation. Various gauges (29G-34G) and lengths (4mm-8mm).
Indications for Use
Indicated for subcutaneous injection of insulin using pen injector devices.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Berpu Medical Technology Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 Cn
Re: K181447
Trade/Device Name: Safety insulin needle for single use Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: February 7, 2019 Received: February 12, 2019
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sarah B. Mollo -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181447
Device Name Safety insulin needle for single use
Indications for Use (Describe)
The safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K181447
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
- 1. Date of Preparation: 01/23/2019
- 2. Sponsor Identification
BERPU MEDICAL TECHNOLOGY CO., LTD NO.14 Xingji Road, Yongxing Street, Longwan District, 325000, Wenzhou, Zhejiang Province
Establishment Registration Number: 3004496829
Contact Person: Jundong Tan Position: Management Representative Tel: +86-577-86651999 Fax: +86-577-86630389 Email: BERPU@BERPU.COM
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Safety insulin needle for single use
Regulatory Information: Classification Name: Hypodermic single lumen needle Classification: II Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital
Intended Use Statement:
The Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.
Indications for Use:
The Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin.
Device Description
The proposed device, Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin. It consists of needle tube, hub, safety protective cover, self-destruction seat, spring, hub sheath, safety seat and sealed paper.
The Safety insulin needle for single use is offered in various gauge sized and length.
The proposed device is available in EO sterilized to achieve a Sterility Assurance Level (SAL) of 10°.
- 5. Identification of Predicate Device
510(k) Number: K152514 Product Name: Clickfine AutoProtect Pen Needle
#### Non-Clinical Test Conclusion 6.
Non-clinical tests were conducted to verify that the proposed device met all design specifications in order to demonstrate that the subject device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
{5}------------------------------------------------
- > ISO 7864:2016, Sterile hypodermic needles for single use.
- > ISO 9626:2016, Stainless steel needle tubing for the manufacture of medical devices.
- > ISO 11608-2 Second edition 2012-04-01, Needle-based injection systems for medical use -Requirements and test methods - Part 2: Needles
- > ISO 10993-1:2009 Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process
- > ISO 10993-4:2017, Biological Evaluation of Medical Device, Part 4: Selection of test for Interactions with Blood.
- > ISO 10993-5: 2009(R), Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity.
- > ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- > ISO 10993-11:2006 Biological evaluation of medical devices-Part 11: Tests for systemic toxicity
- > ISO 10993-7:2008 Biological Evaluation of Medical Devices-Part 7: Ethylene Oxide Sterilization Residuals.
- ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials. >
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ♪ USP <85> Bacterial Endotoxins Test
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
#### Substantially Equivalent (SE) Comparison 8.
#### Table 1 General Comparison
| Item | Proposed Device | Predicate Device<br>K152514 | Remark |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Product Code | FMI | FMI | SE |
| Regulation Number | 21 CFR 880.5570 | 21 CFR 880.5570 | SE |
| Class | II | II | SE |
| Indications for Use | The Safety insulin needle for single use is intended for use with pen injector devices for the subcutaneous injection of insulin. | The Clickfine AutoProtect Pen Needle is intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle | SE<br>Analysis 1 |
{6}------------------------------------------------
| | | sticks from the patient end of<br>the needle. The shield also<br>serves to hide the needle before<br>and after injection. | |
|----------------|-------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------|
| Configuration | Needle Tube | Needle Tube | SE |
| | Hub | Hub | Analysis 2 |
| | Safety protective cover | Needle shield | |
| | Self-destruction seat | Housing | |
| | Spring | Spring | |
| | Hub sheath | Outer protective container | |
| | Safety seat | Safety lock indicator | |
| | Sealed paper | Peel tab | |
| Operation Mode | For manual use only | For manual use only | SE |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | SE |
#### SE Analysis 1 - Indications for Use
The Indications for Use of the proposed device are different to that of predicate device. The difference is that the Indications for Use of predicate device include subcutaneous injection of exenatide while the proposed device does not. However, the indications for use of proposed device are completely included by those of the predicate. Additionally, the attached safety shield has the same role during the clinical application as that of the predicate device, although it is not expressed in the indications for use of the proposed device. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
#### SE Analysis 2 - Configuration
The components name of proposed device are different to that of the predicate, however, the components share the same configuration and function. Therefore, this difference in name will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
| Item | Proposed Device | Predicate Device<br>K152514 | Remark |
|--------------------|----------------------------------|---------------------------------|------------|
| Needle Gauge | 29G, 30G, 31G, 32G,<br>33G, 34G | 29G, 30G, 31G | Analysis 3 |
| Needle Length | 4mm, 5mm, 6mm,<br>8mm | 5mm, 6mm, 8mm | Analysis 4 |
| Wall type | Thin-walled<br>Extra-thin-walled | Regular walled<br>Thin-walled | Analysis 5 |
| Needle Performance | Complied with<br>ISO 7864:2016, | Complied with<br>ISO 7864: 1993 | SE |
| Table 2 Performance Comparison | |
|--------------------------------|--|
| | |
{7}------------------------------------------------
| ISO 9626:2016, | ISO 9626: 1991 | |
|----------------|----------------|--|
| ISO 11608-2 | ISO 11608-2 | |
#### Analysis 3 - Needle Gauge
The subject device includes higher needle gauges than that of predicate devices. This difference in needle size will not raise new questions of safety and effectiveness. In addition, all the needle sizes of proposed device have been tested and met ISO 7864 and ISO 9626 standards requirements. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
## Analysis 4 - Needle length
The proposed device has an additional needle length than the of predicate device. This difference in needle size will not raise new questions of safety and effectiveness. In addition, all the needle lengths of the proposed device have been validated and tested and met the ISO 9626 standards requirements. Therefore, this difference will not affect the Substantially Equivalency (SE) between the proposed and predicate device.
#### Analysis 5 - Wall type
The wall thickness of the proposed device is different than the predicate device. the needles including both wall thickness (i.e. thin-walled and extra-thin-walled) of proposed device have been tested and the test results comply with ISO 7864 and ISO 9626 standards requirements. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
| Item | Proposed Device | Predicate Device<br>K152514 | Remark |
|--------------------------|---------------------------------|-----------------------------|---------------|
| Patient-contact material | | | |
| Hub | Polypropylene | Stainless Steel | SE |
| Needle tube | 304 Stainless Steel | Polypropylene | Analysis 6 |
| Safety seat | Polypropylene | Polypropylene | |
| Safety protective cover | MABS | MBS | |
| Biocompatibility | | | |
| Cytotoxicity | No cytotoxicity | Conforms to ISO | SE |
| Irritation | No intracutaneous<br>reactivity | 10993<br>series standard | Analysis 6 |
| Sensitization | No skin sensitization | | |
| Systemic Toxicity | No systemic toxicity | | |
| Hemolysis | No Hemolysis | | |
| Pyrogen | No Pyrogen | | |
| Sterilization | EO Sterilization | Gamma irradiation | SE Analysis 7 |
Table 3 Safety Comparison
{8}------------------------------------------------
| SAL | 10-6 | 10-6 | SE |
|-----------------|-----------------|-----------------|----|
| Single Use | Yes | Yes | SE |
| Endotoxin Limit | 20EU per device | 20EU per device | SE |
#### Analysis 6 - Patient-contact Material and Biocompatibility
Although the material of proposed and the predicate device is different, the patient-contact material of the proposed device material conforms to the ISO 10993 series of standards. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
## Analysis 7 - Sterilization
The proposed and predicate devices are sterilized by different method. However, the sterilization parameter for proposed device was established per ISO11135 to achieve the SAL of 10-6. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis of differences above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.