← Product Code [EZD](/productcode/EZD) · K181445

# AMSure Hydrophilic Intermittent Catheter (K181445)

_Amsino International, Inc. · EZD · Oct 1, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181445

## Device Facts

- **Applicant:** Amsino International, Inc.
- **Product Code:** [EZD](/productcode/EZD.md)
- **Decision Date:** Oct 1, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The AMSure® Hydrophilic Intermittent Catheter is inserted into the patient's bladder and indicated for intermittent use for the purpose of drainage of urine from the bladder. The catheter does not contain a balloon on its tip. An optional Water sachet is included to activate the hydrophilic-coated surface prior to use. The target populations for the subject device are Adults and Pediatrics.

## Device Story

AMSure Hydrophilic Intermittent Catheter is a sterile, single-use urethral catheter for intermittent bladder drainage. Device consists of a flexible medical-grade PVC tube and funnel; surface features a hydrophilic polyvinyl pyrrolidone (PVP) coating to reduce friction upon activation with sterile water (provided in optional sachet). Used by clinicians or patients in clinical or home settings. Insertion into bladder allows urine drainage; absence of balloon prevents indwelling use. Hydrophilic coating enhances lubricity, potentially reducing urethral trauma and patient discomfort during insertion. Device is provided sterile.

## Clinical Evidence

Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) conducted per ISO 10993-5 and ISO 10993-10. Performance testing included leakage, kink resistance, strength, connector security, flow rate, friction value, chemical analysis (reducing matter, pH, metal ions, EO residue), sterility, water sachet integrity, and shelf-life stability. Human factor testing also performed.

## Technological Characteristics

Materials: Medical-grade PVC (tubing/gripper), PVP hydrophilic coating, USP grade sterile water. Dimensions: Various French sizes (10Fr-16Fr). Connectivity: None. Sterilization: Sterile, single-use. Principle: Hydrophilic lubrication via water activation.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- Cure Catheter ™ Hydrophilic Coated ([K132500](/device/K132500.md))
- AMSure® PVC Intermittent catheter ([K091306](/device/K091306.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 1, 2018

Amsino International Inc. Zoe Wu RA Specialist 708 Corporate Center Drive Pomona, CA 91768

Re: K181445

Trade/Device Name: AMSure Hydrophilic Intermittent Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD Dated: August 15, 2018 Received: August 21, 2018

Dear Zoe Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Glenn B. Bell -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K181445

Device Name

AMSure® Hydrophilic Intermittent Catheter

Type of Use (Select one or hoth, as annlicable)

Indications for Use (Describe)

INTENDED USE:

The AMSure® Hydrophilic Intermittent Catheter is inserted into the patient's bladder and indicated for intermittent use for the purpose of drainage of urine from the bladder. The catheter does not contain a balloon on its tip. An optional Water sachet is included to activate the hydrophilic-coated surface prior to use. The target populations for the subject device are Adults and Pediatrics.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the AMSINO logo. The logo is blue and white, with the word "AMSINO" in large, bold letters. Below the word "AMSINO" is the phrase "Advancing Healthcare Worldwide" in smaller letters. The logo is simple and professional, and it conveys the company's mission of improving healthcare around the world.

# Section 5: 510(k) Summary

| Submitter Information | 1                                                |
|-----------------------|--------------------------------------------------|
| Submitter:            | Amsino International Inc.                        |
|                       | 708 Corporate Center Drive, Pomona CA 91768, USA |
| Contact Person:       | Zoe Wu                                           |
|                       | RA Specialist                                    |
| Phone:                | +86(21)-69117118                                 |
| Fax:                  | +86(21)-59148142                                 |
| E-mail:               | Zoe_Wu@amsino.com                                |

Date Prepared: Sep 28, 2018

- വ Device Information
Trade Name: AMSure® Hydrophilic Intermittent Catheter Device Common Name: Urological Catheter and Accessories Regulation Name: Urological Catheter and Accessories Regulation Number: 21 CFR §876.5130 Product Code: EZD

Device Class: Class II

Review Panel: Gastroenterology/Urology

Table 5-1

| Product Code | Description                                                          |
|--------------|----------------------------------------------------------------------|
| HG971610     | Hydrophilic Intermittent Catheter, 10Fr                              |
| HG971612     | Hydrophilic Intermittent Catheter, 12Fr                              |
| HG971614     | Hydrophilic Intermittent Catheter, 14Fr                              |
| HG971616     | Hydrophilic Intermittent Catheter, 16Fr                              |
| HG971610SS   | Hydrophilic Intermittent Catheter with water sachet and sleeve, 10Fr |
| HG971612SS   | Hydrophilic Intermittent Catheter with water sachet and sleeve, 12Fr |
| HG971614SS   | Hydrophilic Intermittent Catheter with water sachet and sleeve, 14Fr |
| HG971616SS   | Hydrophilic Intermittent Catheter with water sachet and sleeve, 16Fr |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for AMSINO, a healthcare company. The logo is blue and white, with the word "AMSINO" in large, bold letters. Below the word "AMSINO" is the tagline "Advancing Healthcare Worldwide". The logo is simple and professional, and it conveys the company's mission of improving healthcare around the world.

- 3 Predicate Device Information
	- Cure Catheter ™ Hydrophilic Coated (K132500) -
	- AMSure® PVC Intermittent catheter (K091306) -

#### Device Description: ব

The AMSure® Hydrophilic Intermittent Catheter is a urethral catheter intended to be passed through the urethra during urinary catheterization and into the bladder to drain urine. It is hydrophilic and lubricous coated flexible tube owns a hydrophilic polymer in the surface which makes the catheter highly lubricious and good adhesion upon contact with water, saline solution, body fluids. The catheter and gripper is manufactured with medical grade PVC, the surface is coated with a hydrophilic lowfriction coating (polyvinyl pyrrolidone, or PVP), the water sachet is with USP grade sterile water. Each catheter is provided in sterile, single-use packages, intended to be used by adults and pediatrics.

- 5 Intended use
The AMSure® Hydrophilic Intermittent Catheter is inserted into the patient's bladder and indicated for intermittent use for the purpose of urine from the bladder. The catheter does not contain a balloon on its tip. An optional Water sachet is included to activate the hydrophilic-coated surface prior to use. The target populations for the subject device are Adults and Pediatrics.

#### 6 Product Comparison Summary

The proposed AMSure® Hydrophilic Intermittent Catheter and the predicate devices are intended for patients who drain the urine from bladder. These products are urethral catheters that have the same technological characteristics and material for tubing and funnel, same intended use, similar indications for use, the same function, and the same general characteristics.

The design and materials of the proposed devices are identical to the predicate Cure CatheterTM Hydrophilic Coated (K132500) and AMSure® PVC Intermittent catheter (K091306).

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for AMSINO, a company that is advancing healthcare worldwide. The logo is blue and white, with the word "AMSINO" in large, bold letters. Below the company name is the slogan "Advancing Healthcare Worldwide" in a smaller font. The logo is simple and professional, and it conveys the company's mission to improve healthcare around the world.

- Nonclinical Testing 7
	- � Biocompatibility Testing

Cytotoxicity, Sensitization and Irritation have been conducted per ISO 10993-5 and ISO 10993-10. The Biocompatibility testing demonstrated the biological safety of the proposed devices which may directly contact the patients.

- � Performance Testing
## Performance Testing of AMSure® Hydrophilic Intermittent Catheter was conducted as below:

Table 5-2

| SN | Test items                            |
|----|---------------------------------------|
| 1  | Packaging inspection                  |
| 2  | Appearance inspection                 |
| 3  | Leakage                               |
| 4  | Kink resistance                       |
| 5  | Strength test                         |
| 6  | Connector security                    |
| 7  | Flow test                             |
| 8  | Friction value                        |
| 9  | Chemical performance: Reducing matter |
| 10 | Chemical performance: pH value        |
| 11 | Chemical performance: Metal ions      |
| 12 | Chemical performance: EO residue      |
| 13 | Biological performance: Sterility     |
| 14 | Physical performance: Water sachet    |
| 15 | Chemical performance: sterile water   |
| 16 | Aging (shelf life/stability testing)  |
| 17 | Human factor testing                  |

#### ထ Conclusions

The information provided within this pre-market notification demonstrates that the AMSure® Hydrophilic Intermittent Catheter is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K181445](https://fda.innolitics.com/device/K181445)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
