Clyra Wound Irrigation Solution

K181428 · Clyra Medical Technologies, Inc. · FRO · Sep 20, 2019 · SU

Device Facts

Record IDK181428
Device NameClyra Wound Irrigation Solution
ApplicantClyra Medical Technologies, Inc.
Product CodeFRO · SU
Decision DateSep 20, 2019
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Rx: Clyra Wound Irrigation Solution is intended for use by healthcare professionals for cleansing, moistening and debriding to remove wound debris from acute and chronic dermal lesions that are partial or full thickness wounds such as 1st and 2nd degree burns, stage I - IV pressure ulcers, stasis ulcers, abrasions and minor skin irritations, post surgical wounds, grafted and donor sites, in addition to moistening absorbent wound dressings.

Device Story

Clyra Wound Irrigation Solution is a clear, hypotonic, non-sterile topical solution used for wound management. It is applied by healthcare professionals to acute and chronic dermal lesions to cleanse, irrigate, and debride debris. The mechanism of action is primarily the mechanical force of fluid moving across the wound surface, which aids in removing foreign objects like dirt and debris. The solution contains Potassium Iodide and Copper Sulphate, which react to release iodine (250 ppm) to act as a preservative against microbial contamination. The device is supplied in 4 oz. food-grade PET plastic bottles with spray inserts. It is used in clinical settings to prepare wound beds and maintain a moist environment for healing by lubricating absorbent dressings.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), antimicrobial effectiveness (USP <51>), chemical characterization, toxicological risk assessment, and shelf-life stability testing.

Technological Characteristics

Hypotonic aqueous solution containing Potassium Iodide and Copper Sulphate (250 ppm iodine). Supplied in 4 oz. food-grade PET plastic bottles with spray inserts. Non-sterile, preserved. Complies with ISO 10993-1 (biocompatibility) and USP <51> (antimicrobial effectiveness).

Indications for Use

Indicated for healthcare professionals to cleanse, irrigate, moisten, and debride acute and chronic dermal lesions (partial/full thickness wounds) including 1st/2nd degree burns, stage I-IV pressure ulcers, diabetic ulcers, stasis ulcers, abrasions, minor skin irritations, post-surgical wounds, and grafted/donor sites; also for moistening/lubricating absorbent wound dressings.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 20, 2019 Clyra Medical Technologies Inc. % Tanya Rhodes President Rhodes & Associates, Inc 211 Poinciana Lane Largo, Florida 33770 Re: K181428 Trade/Device Name: Clyra Wound Irrigation Solution Regulatory Class: Unclassified Product Code: FRO Dated: September 15, 2019 Received: September 17, 2019 Dear Tanya Rhodes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Cynthia Chang Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K181428 Device Name Clyra™ Wound Irrigation Solution #### Indications for Use (Describe) Rx: Clyra Wound Irrigation Solution is intended for use by healthcare professionals for cleansing, moistening and debriding to remove wound debris from acute and chronic dermal lesions that are partial or full thickness wounds such as 1st and 2nd degree burns, stage I - IV pressure ulcers, stasis ulcers, abrasions and minor skin irritations, post surgical wounds, grafted and donor sites, in addition to moistening absorbent wound dressings. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for "CLYRA Medical Technologies Inc." The logo consists of a circular emblem on the left and the company name on the right. The emblem is divided into two halves, with the top half being a dark blue color and containing a white plus sign, and the bottom half being a light green color with a leaf-like design. The company name is written in a modern, sans-serif font, with "CLYRA" in a larger, dark blue font and "Medical Technologies Inc" in a smaller, light green font. # Section 6 # 510(k) Summary # Clyra™ Wound Irrigation Solution ## K181428 ### 1. Submission Sponsor Clyra Medical Technologies Inc. 14921 Chestnut St. Westminster California 92683 United States Contact: Steven V. Harrison Title: President #### 2. Submission Correspondent Rhodes & Associates, Inc. 211 Poinciana Lane Largo FL 33770 Office Phone: 727-458-4291 Contact: Tanya Rhodes Title: President #### 3. Date Prepared 9.20.2019 #### 4. Device Identification | Trade/Proprietary Name: | Clyra™ Wound Irrigation Solution | |-------------------------|----------------------------------| | Common/Usual Name: | Wound Cleanser | | Classification Name: | Dressing, Wound, Drug | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a logo for "CLYRA Medical Technologies Inc". The logo features a circular emblem on the left, divided into a blue upper half with a white plus sign and a green lower half with a leaf-like shape. To the right of the emblem is the word "CLYRA" in a stylized, sans-serif font, with the letters in a dark blue color. Below "CLYRA" is the text "Medical Technologies Inc" in a smaller, lighter green font. | Regulation Number: | Unclassified | |-----------------------|------------------------------| | Product Code: | FRO | | Device Class: | Unclassified (Pre-Amendment) | | Classification Panel: | General & Plastic Surgery | #### 5. Legally Marketed Predicate Device(s) K133452 - Puracyn Plus Skin and Wound Care - Innovacyn, Inc. Rialto, CA 92377 ### 6. Indication for Use Statement Clyra Wound Irrigation Solution is intended for use by healthcare professionals for cleansing, irrigating, moistening and debriding to remove wound debris from acute and chronic dermal lesions that are partial or full thickness wounds such as 1st and 2nd degree burns, stage I - IV pressure ulcers, diabetic ulcers, stasis ulcers, abrasions and minor skin irritations, post surgical wounds, grafted and donor sites, in addition to moistening and lubricating absorbent wound dressings. ### 7. Device Description Clyra Wound Irrigation Solution is a clear hypotonic solution topically applied to skin and wound areas. The subject device is a wound management and cleansing solution that is intended for cleansing, irrigating, and debriding dermal wounds in addition to moistening absorbent wound dressings (e.g. gauze). The mechanical action of fluid moving across the wound provides for the mechanism of action and aids in the removal of foreign objects such as dirt and debris. Clyra Wound Irrigation Solution will be supplied in food grade 4 oz. plastic PET bottles with spray inserts and caps. Clyra Wound Irrigation contains Potassium Iodide and Copper Sulphate, which release iodine when combined, resulting in a concentration within the product of 250 parts per million (ppm). The iodine acts as a preservative to inhibit contamination within the solution. #### 8. Substantial Equivalence Discussion The following table compares the Clyra Wound Irrigation to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image is a logo for "CLYRA Medical Technologies Inc". The logo features a circular emblem on the left, divided into a blue upper half with a white plus sign and a green lower half with a leaf design. To the right of the emblem is the word "CLYRA" in a stylized, sans-serif font, with the words "Medical Technologies Inc" underneath in a smaller font. | Manufacturer | Clyra Medical<br>Technologies Inc. | Innovacyn Inc. | Device Comparison | |------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | Clyra Wound<br>Irrigation<br>Solution | Puracyn Plus Skin and<br>Wound Care | | | 510(k) Number | K181428 | K133542 | Not applicable | | Product Code | FRO | FRO | Same | | Regulation Number | Unclassified | Unclassified | Same | | Regulation Name | Dressing, Wound, Drug | Dressing, Wound, Drug | Same | | Prescription Only<br>Indications for Use | Rx: Clyra Wound<br>Irrigation Solution is<br>intended for use by<br>healthcare professionals<br>for cleansing,<br>irrigating,moistening<br>and debriding to remove<br>wound debris from<br>acute and chronic<br>dermal lesions that are<br>partial or full thickness<br>wounds such as 1st and<br>2nd degree burns, stage<br>I - IV pressure ulcers,<br>diabetic ulcers, stasis<br>ulcers, abrasions and<br>minor skin irritations,<br>post surgical wounds,<br>grafted and donor sites,<br>in addition to<br>moistening and<br>lubricating absorbent<br>wound dressings. | Rx: Puracyn Plus™ Skin<br>and Wound Care is<br>intended for use by<br>healthcare professionals<br>for cleansing,<br>irrigating, moistening,<br>and debriding to remove<br>wound debris from<br>acute and chronic<br>dermal lesions that are<br>partial or full thickness<br>wounds such as 1st and<br>2nd degree burns, stage<br>I - IV pressure ulcers,<br>diabetic ulcers, stasis<br>ulcers, abrasions and<br>minor skin irritations,<br>post-surgical wounds,<br>grafted and donor sites,<br>in addition to<br>moistening and<br>lubricating absorbent<br>wound dressings. | Same<br>Each device is indicated<br>for topical wound care/<br>irrigation.<br>The Intended Use<br>Statement utilizes the<br>same indications as<br>previously cleared<br>under predicate<br>devices. | | Manufacturer | Clyra Medical<br>Technologies Inc. | Inovacyn Inc | Device Comparison | | Trade Name | Clyra Wound<br>Irrigation<br>Solution | Puracyn Plus<br>Skin and Wound<br>Care | | | Device Description | Clyra Wound Irrigation<br>Solution is a clear<br>hypotonic solution<br>topically applied to skin<br>and wound areas. The<br>subject device is a<br>wound management and<br>cleansing solution that is<br>intended for cleansing,<br>irrigating, and debriding<br>dermal wounds in<br>addition to moistening<br>and lubricating<br>absorbent wound<br>dressings (e.g. gauze).<br>The mechanical action of<br>fluid moving across the<br>wound provides for the<br>mechanism of action and<br>aids in the removal of<br>foreign objects such as<br>dirt and debris.<br>Clyra Wound Irrigation<br>Solution will be supplied<br>in food grade 4 oz.<br>plastic PET bottles with<br>spray inserts and caps.<br>Clyra Wound Irrigation<br>Solution contains<br>Potassium lodide and<br>Copper Sulphate, which<br>release iodine when<br>combined, resulting in a<br>concentration within the<br>product of 250 parts per<br>million (ppm). The iodine<br>acts as a preservative to<br>inhibit contamination<br>within the solution. | Puracyn Plus Skin and<br>Wound Care is a clear<br>hypotonic solution<br>topically applied to skin<br>and wound areas. The<br>subject device is a<br>wound management<br>and cleansing solution<br>that is intended for<br>cleansing, irrigating, and<br>debriding dermal<br>wounds in addition to<br>moistening and<br>lubricating absorbent<br>wound dressings (e.g.<br>gauze). The mechanical<br>action of fluid moving<br>across the wound<br>provides for the<br>mechanism of action<br>and aids in the removal<br>of foreign objects such<br>as dirt and debris.<br>Puracyn Plus Skin and<br>Wound Care will be<br>supplied in food grade 4<br>oz. plastic PET bottles<br>with spray inserts and<br>caps. | Same.<br>Additional information<br>regarding Clyra Wound<br>Irrigation Solution<br>preservative system | | Manufacturer | Clyra Medical<br>Technologies Inc. | Innovacyn Inc | Device Comparison | | Trade Name | Clyra Wound Irrigation<br>Solution | Puracyn Plus Skin<br>and Wound Care | | | Form | The device is supplied in<br>food grade 4 oz plastic<br>PET bottles with spray<br>inserts and caps. This<br>product is preserved and<br>provided non sterile | The device is supplied in<br>food grade 4 oz plastic<br>PET bottles with spray<br>inserts and caps. This<br>product is preserved and<br>provided non sterile. | Same | | Ingredients | Copper Sulphate;<br>Potassium Iodide;<br>Sodium Chloride;<br>Water | Hypochlorous Acid;<br>Sodium Hypochlorite;<br>Phosphate;<br>Sodium Chloride;<br>Water | No significant difference.<br>Materials used in each<br>product are similar in<br>their properties to<br>provide a preserved<br>wound irrigation<br>solution | | Pre-Clinical Testing | Pre-Clinical Testing<br>included pH, Shelf Life,<br>Chemical Stability,<br>Antimicrobial<br>Preservative<br>Effectiveness, | Pre-Clinical Testing<br>included pH, Shelf Life,<br>Chemical Stability,<br>Antimicrobial<br>Preservative<br>Effectiveness | No significant<br>difference. | | Complies with<br>ISO 10993-1 | Yes | Yes | No significant difference | | Complies with<br>Antimicrobial Test<br>USP<51> | Non-sterile<br>Preserved/Conforming<br>to USP <51> | Non-sterile<br>Preserved/Conforming<br>to USP <51> | Same | | Sterile | Non-sterile | Non-sterile | Same | | Shelf-Life | Two (2) Years | Two (2) years | Same | ### Table 6A – Comparison of Characteristics {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for "CLYRA Medical Technologies Inc". The logo features a circular emblem on the left, divided into a blue upper half and a green lower half, with a white medical cross symbol in the center. To the right of the emblem, the word "CLYRA" is written in a stylized, sans-serif font, with the words "Medical Technologies Inc" written in a smaller font below it. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for "CLYRA Medical Technologies Inc". The logo consists of a circular emblem on the left and the company name on the right. The emblem is divided into two halves, with the top half being a dark blue color and the bottom half being a light green color. A white plus sign is in the center of the emblem. The company name is written in a sans-serif font, with "CLYRA" in a dark blue color and "Medical Technologies Inc" in a light green color. ### 9. Non-Clinical Performance Data {8}------------------------------------------------ Image /page/8/Picture/0 description: The image is a logo for "CLYRA Medical Technologies Inc". The logo consists of a circular emblem on the left and the company name on the right. The emblem is divided into two halves, with the top half being a dark blue color and containing a white plus sign, and the bottom half being a light green color with a leaf-like design. The company name is written in a simple, sans-serif font, with "CLYRA" in a larger font size than "Medical Technologies Inc". As part of demonstrating safety and effectiveness of Clyra Wound Irrigation Solution and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Clyra Medical Technologies Inc. completed ISO 10993 biocompatibility testing to establish the safety of Clyra Wound Irrigation Solution for its intended use. - Cytotoxicity Assay (ISO 10993-5:2009): ● - Porcine Wound Healing Study - . Sensitization Test (ISO 10993-10:2010): - Irritation Test (ISO 10993-10:2010): - Material medicated pyrogenicity - Chemical characterization and toxicological risk assessment for systemic toxicity - Antimicrobial Effectiveness Test (AET) (USP<51>): Clyra Wound Irrigation Solution was evaluated for preservative activity in compliance with USP Antimicrobial Effectiveness Test <51> , designed to test efficacy of preservatives. Exposure to Clyra Wound Irrigation Solution caused an inhibition in Pseudomonas aeruginosa, E. coli, Staphylococcus aureus, Candida albicans, and Aspergillus brasiliensis . - Shelf Life Testing A maximum shelf life of 2 years has been assigned to the Clyra Wound ● Irrigation Solution, when stored unopened at ambient temperature, in accordance with the manufacturer's recommendations. ### 10. Clinical Performance Data N/A {9}------------------------------------------------ Image /page/9/Picture/0 description: The image is a logo for "CLYRA Medical Technologies Inc." The logo features a green and blue circle with a white medical cross in the center. To the right of the circle is the company name, "CLYRA" in large, sans-serif font, with the words "Medical Technologies Inc" in a smaller font below. #### 11. Statement of Substantial Equivalence By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise new questions regarding its safety and effectiveness as compared to the predicate device(s). The Clyra Wound Irrigation Solution, as designed and manufactured, is determined to be substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...