K181407 · Siemens Medical Solution USA, Inc. · OWB · Aug 15, 2018 · Radiology
Device Facts
Record ID
K181407
Device Name
Artis zee/zeego & Artis Q/Q.zen
Applicant
Siemens Medical Solution USA, Inc.
Product Code
OWB · Radiology
Decision Date
Aug 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions. Artis can also support the acquisition of position triggered imaging for spatial data synthesis. The Artis systems include also the software option DynaCT with following IFU: DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.
Device Story
Artis zee/zeego and Artis Q/Q.zen are modular angiography systems comprising a C-arm, X-ray tube, flat panel detector, generator, and patient table. Systems acquire X-ray projections for 2D fluoroscopy/radiography and 3D reconstruction (DynaCT). Operated by healthcare professionals in clinical/OR environments. New software (VD11D) supports the Pixium 3040CV detector and introduces 'nexaris Angio-CT' configuration, allowing integration with sliding-gantry CT systems. 'Smart Collision' calculation improves safety during gantry movement. Output includes 2D/3D images for diagnosis, surgical planning, and interventional guidance. Benefits include reduced patient repositioning and enhanced spatial data synthesis.
Clinical Evidence
Bench testing only. Performance verified through electrical safety (ES60601-1), EMC (60601-1-2), and software validation (IEC 62304). Human factors usability validation performed. No clinical trial data required for these modifications.
Indicated for single and biplane diagnostic imaging and interventional procedures, including cardiac, neuro, general, rotational, and multipurpose angiography, and whole-body radiographic/fluoroscopic procedures. Suitable for pediatric and obese patients. Includes image-guided surgery and DynaCT 3D reconstruction for hard/soft tissue imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solution USA, Inc. % Ms. Patricia Jones Sr. Regulatory Affairs Specialist 40 Liberty Boulevard 65-1A MALVERN PA 19355
Re: K181407
Trade/Device Name: Artis zee/zeego & Artis O/O.zen Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, IZI, JAA, JAK Dated: May 26, 2018 Received: May 30, 2018
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
August 15th, 2018
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeff Baffaro
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181407
Device Name Artis zee / zeego & Artis Q / Q.zen
#### Indications for Use (Describe)
Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the Artis family include cardiac angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions.
Artis can also support the acquisition of position triggered imaging for spatial data synthesis.
The Artis systems include also the software option DynaCT with following IFU:
DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## SIEMENS
## 510(k) Summary: Artis zee / zeego & Artis Q / Q.zen
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355
#### August 15, 2018 Date Prepared:
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
- 1. General Information: lmporter / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
#### Manufacturing Site:
Siemens Healthcare GmbH Siemensstr. 1 91301 Forchheim, Germany Establishment Registration Number: 3004977335
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 448-6474 Email: patricia.d.jones@siemens-Healthineers.com
#### 3. Device Name and Classification:
| Trade Name: | Artis zee / zeego |
|----------------------------|---------------------------------------------|
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1650 |
| Device Class: | Class II |
| Product Codes: | OWB, IZI, JAA, JAK |
| Trade Name: | Artis Q and Artis Q.zen |
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1650 |
| Device Class: | Class II |
{5}------------------------------------------------
## SIEMEN
### Product Codes:
OWB, JAA, IZI
- 4. Legally Marketed Primary Predicate Device
- Trade Name: 510(k) Clearance Clearance Date Classification Name: Classification Panel: CFR Section: Device Class: Product Code: Subsequent Product Codes: Total Product Life Cycle:
Artis zee/zeego SW VC21 K141574 September 05, 2014 Image-intensified fluoroscopic x-ray system Radiology 21 CFR §892.1650 Class II OWB IZI, JAA, JAK All product Recall incidents are considered
during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues.
### Legally Marketed Secondary Predicate Device
| Trade Name: | Artis Q and Artis Q.zen |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance | K123529 |
| Clearance Date | February 26, 2013 |
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1650 |
| Device Class: | Class II |
| Product Codes: | OWB, |
| Subsequent Product Codes: | IZI, JAA |
| Total Product Life Cycle: | All product Recall incidents are considered<br>during the Design Input phase of<br>development to ensure the latest models<br>will not be affected by any of the applicable<br>issues. |
#### 5. Device Description:
The Artis Modular Angiography systems are specialized angiography systems. In general, they are equipped with C-arm, stand, flat panel detector, x-ray tube, high voltage generator, patient table and image post-processing software.
Siemens currently markets the Artis zee / zeego (K141574) and the Artis Q / Q.zen (K123529). Siemens will provide new software VD11D for both, the Artis zee / zeego and Artis Q / Q.zen systems. The new software VD11D will support the detector Pixium 3040CV (also known as "40HDR") already cleared with Artis Q / Q.zen (K123529). Systems, Artis zee / zeego, and Artis Q / Q.zen use the cleared AEC (Automatic Exposure Control) functionality.
{6}------------------------------------------------
## SIEMENS
Siemens intends to market new claims specific to the AEC, which are provided in the product claims list. Siemens will replace the cooling unit in both systems.
The Mivabi Angio-CT configuration has been improved and will be known as and marketed as the nexaris Angio-CT configuration. Siemens nexaris Angio-CT configuration was designed to contain both Angio and CT System within close proximity or within the same environment. The CT-gantry on rails will slide towards the Angiography patient table to perform a CT scan without repositioning the patient. After the CT scan, the CT-gantry could be slide away from the patient table to use the table for Angiography C-arm acquisitions.
This 510(k) submission describes modifications made to the previously cleared predicate devices: Artis zee / zeego SW VC21 (K141574) and Artis Q / Q.zen SW Version VD10 (K123529). The modifications will be marketed as "Artis zee / zeego VD11D" and "Artis Q / Q.zen VD11."
The following modifications/features do not impact any new usage of Contrast Agents.
## Proposed Device Modifications for the Artis zee / zeego and the Artis Q / Q.zen Systems:
- Revised Indication for Use Statement applicable to both systems 1. (Artis zee / zeego and Artis Q / Q.zen)
- 2. New system software version VD11D supports hardware for both systems (Artis zee / zeego and Artis Q / Q.zen)
- New detector pixium 3040CV (also known as "as40HDR") 3. Software and Hardware modification applicable to the Artis zee / zeego system only. (This detector is already cleared for use in the Predicate Device for the Artis Q /Q.zen System).
- 4. New Tube cooling unit for both systems (Artis zee / zeego and Artis Q / Q.zen)
- Product Claims List modification applicable for both systems (Artis 5. zee / zeego and Artis Q / Q.zen)
- Update 510(k) Information provided for all predicate devices. 6.
This 510(k) submission also includes the following modification made to the "Miyabi Angio-CT" configuration, which consists of Artis zee / zeego or Artis Q /Q.zen and a CT system (SOMATOM Family Sliding Gantry). The Miyabi Angio-CT configuration that will include these Subject Device modifications will be marketed as "nexaris Angio-CT."
## Proposed Device Modifications for the nexaris Angio-CT configuration:
Note: These modifications are only applicable when the Artis zee / zeego or Artis Q / Q.zen and a CT system (SOMATOM Family Sliding Gantry) are
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "SIEMENS" in all capital letters. The text is a light gray color. The font is sans-serif and the letters are evenly spaced. The background is white.
configured in the nexaris Angio-CT configuration. The nexaris Angio-CT is only an environment which may contain an Angio and CT system in close proximity.
- New software version VD11D interface software component with 7. Angio and CT systems.
- Improved Collision calculation between the Artis zee / zeego or Artis Q / Q.zen and CT Systems.
- .
.
## 6. Indications for Use:
## Artis zee / zeego Indications for Use Statement:
Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the Artis family include cardiac angiography, neuro-angiography, general angiography, rotational angjography; multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image-quided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery, and interventions.
Artis can also support the acquisition of position triggered imaging for spatial data synthesis.
The Artis systems include also the software option DynaCT with following IFU:
DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures, and treatment follow-up.
### Artis Q / Q.zen Indications for Use Statement:
Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the Artis family include cardiac angiography, neuro-angiography, general angiography, rotational anqiography;
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "SIEMENS" in all capital letters. The letters are a light gray color. The background is white.
multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image-quided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery, and interventions.
Artis can also support the acquisition of position triggered imaging for spatial data synthesis.
The Artis systems include also the software option DynaCT with following IFU:
DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures, and treatment follow-up.
#### 7. Substantial Equivalence:
The Artis zee / zeego and Artis Q / Q.zen SW VD11D is substantially equivalent to the legally marketed predicates listed in the table below:
| This Tradition Bundled 510(k) Submission Contains Two Angiographic Systems | | | |
|----------------------------------------------------------------------------|------------------|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Artis zee / zeego VD11D comparable properties: | | | |
| Predicate Device Name<br>and Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties |
| Primary Predicate<br>Artis zee/zeego SW VC21 | K141574 | 09/05/2014 | • Indications for use<br>• System for Image Acquisition<br>• System for post-processing |
| Secondary Predicates<br>Artis Q and Q.zen - Modular<br>Angiographic System | K123529 | 02/12/2013 | • Detector Pixium 3040CV<br>(as40HDR)<br>• Product Claims |
| Artis Q / Q.zen VD11D comparable properties: | | | |
| Primary Predicate<br>Artis Q and Q.zen - Modular<br>Angiographic System | K123529 | 02/12/2013 | • Indications for use<br>• System for Image Acquisition<br>• System for post processing<br>• Detector Pixium 3040CV<br>(as40HDR)<br>• Product Claims |
| | | Table 1: Predicate Comparable Properties | |
|--|--|------------------------------------------|--|
|--|--|------------------------------------------|--|
{9}------------------------------------------------
# SIEMENS
#### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
Artis zee / zeego and Artis Q / Q.zen Systems are designed as a set of components (C-arm, X-ray tube and housing, flat panel detector, digital imaging system, collimator, generator etc.) that may be combined into different configurations to provide specialized angiography systems. Components used with Artis zee / zeego and Artis Q / Q.zen are either commercially available with current Siemens systems or include modifications to existing components. Technological differences between the Subject Device and the Predicate Device are provided in the table below for all modifications.
| Modification | Subject Device | Primary Predicate Device | Secondary Predicate Device | Comparison Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>For Use<br>Statement | IFU statement<br>revised to simplify<br>and improve clarity<br>of content. | Cleared IFU<br>Statement for<br>K141574 | Not applicable | See row below<br>for results |
| Comparison<br>Result | The revised IFU statement is within the scope of the Intended Use for Interventional<br>fluoroscopic X-Ray Systems per 21 CFR §892.1650. The minor modification of the cleared<br>IFU statement has no impact on the intended use of the device. | | | |
| Software<br>Version | New Software<br>VD11D supports<br>Artis zee / zeego<br>System with Pixium<br>3040CV 30cm x<br>40cm 16 bit detector<br>also known as<br>(as40HDR) | Software VC21 for<br>Artis zee/zeego | Not applicable | Compatible for<br>both Artis zee /<br>zeego and Artis<br>Q / Q.zen<br>systems<br>meets criteria<br>as tested per<br>Software and<br>SSIX Guidance |
| Flat Panel<br>Detector | Pixium 3040CV<br>(as40HDR)<br>30cm x 40cm<br>16 bit | Not applicable | Pixium 3040CV<br>30cm x 40cm<br>16 bit also known as<br>(as40HDR) | Same detector<br>as cleared with<br>Artis Q / Q.zen<br>will be used<br>with Artis zee /<br>zeego meets<br>criteria as<br>tested per<br>Software and<br>SSIX Guidance |
| Tube Cooling<br>Unit | New Tube Cooling<br>Unit: SMC Tube one<br>for all | Tube Cooling Unit:<br>Kluever cooling unit<br>or Lytron cooling<br>system. | Not applicable | Same<br>functionality |
| Modification | Subject Device<br>Artis Q / Q.zen Systems with<br>SW VD11D | Primary Predicate Device<br>Artis Q and Artis Q.zen Modular<br>Angiography System VD10<br>(K123529) | Comparison<br>Results | |
| Indications<br>For Use<br>Statement | IFU statement revised to<br>simplify and improve clarity of<br>content. | | Cleared IFU Statement for K123529 | See row below<br>for results |
| Comparison<br>Result | The revised IFU statement (applicable to both Artis zee / zeego and Artis Q / Q.zen<br>systems) fall within the scope of the Intended Use for Interventional fluoroscopic X-Ray<br>Systems per 21 CFR §892.1650. The minor modification of the cleared IFU statement has<br>no impact on the intended use of the device. | | | |
| Software<br>Version | New Software VD11D supports<br>Artis zee / zeego System with<br>Pixium 3040CV 30cm x 40cm<br>16 bit detector also known as<br>(as40HDR) | | Software Version VD10 cleared with<br>Artis Q / Q.zen | Compatible for<br>both Artis zee /<br>zeego and Artis<br>Q / Q.zen<br>systems meets<br>criteria as<br>tested per<br>Software and<br>SSIX Guidance |
| Tube Cooling<br>Unit | New Tube Cooling Unit: SMC<br>Tube one for all | | Lytron cooling unit | Same<br>functionality |
| Miyabi Angio CT Configuration new marketing name is the nexaris Angio-CT Configuration which is an<br>environment configured with Angio - CT Systems in close proximity previously marketed as Miyabi Angio<br>CT. | | | | |
| Modifications | Subject Device<br>Configuration<br>nexaris Angio-CT<br>Configuration<br>with Artis zee /<br>zeego or Artis Q /<br>Q.zen | Predicate Device<br>Configuration<br>Miyabi Angio CT<br>Configuration with<br>Artis zee / zeego | Predicate Device<br>Configuration<br>Miyabi Angio CT<br>Configuration with<br>Artis Q / Q.zen. | Comparison<br>Results |
| Miyabi Angio-<br>CT System<br>Configuration | New nexaris Angio<br>CT Configuration<br>with Artis zee /<br>zeego or Artis Q /<br>Q.zen Systems | Miyabi Angio CT<br>Configuration with<br>Artis zee / zeego | Miyabi Angio CT<br>Configuration with<br>Artis Q / Q.zen | New marketing<br>name for the<br>previous Miyabi<br>Configuration will<br>be known as the<br>nexaris Angio-CT<br>Configuration with<br>the Artis zee /<br>zeego or Artis Q /<br>Q.zen system and<br>CT Systems with<br>sliding gantry. |
| Collision<br>Calculation | Collision<br>calculation (Smart<br>Collision) of the<br>CT gantry now is<br>based on a CT<br>model | Collision calculation of<br>the CT gantry is<br>based on a simulated<br>moving wall | Collision calculation<br>of the CT gantry is<br>based on a<br>simulated moving<br>wall | Collision<br>calculation has<br>been improved<br>and meets<br>criteria as<br>tested per<br>Software |
### Table 2: Comparison of Technological Characteristics
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are large and evenly spaced, filling most of the frame. The text is a light gray color, contrasting with the white background.
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "SIEMENS" in all capital letters. The text is in a sans-serif font and is a light gray color. The background is white, which makes the text stand out.
#### 9. Nonclinical Performance Testing:
Non-clinical tests were conducted for Artis zee / zeego and Artis Q / Q.zen during product development.
The Artis zee / zeego and Artis Q / Q.zen were certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance and Electromagnetic Compatibility:
- ES60601-1:2005/(R)2012
- 60601-1-2:2007 ●
- 60601-1-3:2008 ●
- 60601-1-6:2010/A1:2013 ●
- 60825-1:2007 ●
- TR 60878:2015 ●
- 62304:2006
- 80001-1:2010
- 60601-2-28:2010 ●
- 60601-2-43:2010 ●
- 60601-2-54:2009/A1:2015 ●
- 10993-1:2009
- 14971:2007 ●
### Table 3: FDA Guidance Documents
| FDA Guidance Document and Effective Date | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 1. | Guidance for Industry and FDA Staff - User Fees and Refunds for Premarket<br>Notification Submissions 510(k)<br>Document issued on October 2, 2017 | |
| 2. | Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept<br>Policy for 510(k)s<br>Document issued on January 30, 2018 | |
| 3. | Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated<br>510(k)s - Guidance for Industry and FDA Staff<br>Document issued on August 12, 2005 | |
| 4. | Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a<br>change to an existing device.<br>Document issued on October 25, 2017 | |
| 5. | Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program:<br>Evaluating Substantial Equivalence in Premarket Notifications [510(k)]<br>Document Issued on July 28, 2014 | |
| 6. | Guidance for Industry and FDA Staff: Guidance for the Submission Of 510(k)'s for | |
{12}------------------------------------------------
## SIER
| | Solid State X-ray Imaging Devices<br>Document issued on September 1, 2016 |
|-----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 7. | Guidance for Industry and FDA Staff: Guidance for the Content of Premarket<br>Submission for Software in Medical Devices<br>Document issued on May 11, 2005 |
| 8. | Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use in<br>Medical Devices<br>Document issued on September 9, 1999 |
| 9. | Guidance for Industry and FDA Staff: Applying Human Factors and Usability<br>Engineering to Medical Devices.<br>Document issued February 3, 2016 |
| 10. | Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging<br>Device Premarket Notifications.<br>Document issued on November 28, 2017 |
| 11. | Guidance for Industry and FDA Staff: Content of Premarket Submissions for<br>Management of Cybersecurity in Medical devices.<br>Document issued on October 2, 2014 |
| 12. | Guidance for Industry and FDA Staff: Bundling Multiple Devices or Multiple<br>Indications in a Single Submission.<br>Document issued on June 27, 2007 |
## Verification and Validation:
Software Documentation for a Major Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005, and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The modifications of the Artis zee / zeego and Artis Q / Q.zen, software version VD11D are supported with non-clinical information that demonstrates the software safety through verification and validation testing. Verification and validation testing demonstrate that Artis zee / zeego and Artis Q / Q.zen perform as intended. The non-clinical test data demonstrate that Artis zee / zeego and Artis Q / Q.zen device performance is comparable to the predicate device. Software supports all old and modified hardware components for both subject devices with respect to the intended use. The testing supports that all software specifications have met the acceptance criteria. The verification and validation testing support the claims of substantial equivalence.
The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable and do not raise any new issues of safety or effectiveness.
Artis zee / zeego and Artis Q / Q.zen software VD11 was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. The Human Factor Usability Validation showed that Human factors are addressed in the system
{13}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "SIEMENS" in all capital letters. The text is a light gray color. The font is sans-serif and appears to be bolded.
tests according to the operator's manual and in clinical use tests with customer report and feedback form.
Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 80001-1-2010 is the hospital. Provided in the Software Section, is the required cybersecurity information.
### Summary:
Performance tests were conducted to test the functionality of Artis zee / zeego and Artis Q / Q.zen software VD11D. These tests have been performed to assess the functionality of the subject device. Results of all conducted testing and clinical assessment were found acceptable and do not raise any new issues of safety or effectiveness.
#### 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
#### 11. Conclusion as to Substantial Equivalence:
The predicate devices were cleared based on non-clinical supportive information and clinical images and data. Similar non-clinical test results demonstrate that the Artis zee / zeego and Artis Q / Q.zen SW VD11D acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data and software validation data demonstrates that the Subject Devices are as safe and effective when compared to the Predicate Devices that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.