The ADVIA Centaur® Herpes-1 IgG (HSV1) assay is for in vitro diagnostic use in the qualitative determination of IgG antibodies to herpes simplex virus type 1 (HSV-1) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur systems. The test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. The predictive or negative result depends on the prevalence of HSV-1 infection in the population and the pre-test likelihood of HSV-1 infection. The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for immunocompromised patients, pediatic patients or matrices other than human serum and plasma (EDTA and lithium heparin).
Device Story
The ADVIA Centaur Herpes-1 IgG is a chemiluminescent immunoassay used for the qualitative detection of IgG antibodies to HSV-1 in human serum and plasma. It is designed for use on the ADVIA Centaur systems in clinical laboratory settings. The device processes patient samples to determine serological status, aiding clinicians in the diagnosis of HSV-1 infection. Results are provided as an index value, which clinicians use alongside patient history and prevalence data to inform diagnostic decisions. The assay provides a standardized, automated alternative to manual immunoblot methods, potentially improving laboratory efficiency and diagnostic consistency.
Clinical Evidence
Clinical study evaluated 864 specimens (including 274 pregnant women) across 3 sites. Performance compared to a commercial immunoblot and Western Blot reference. Overall sensitivity was 97.5% (95% CI: 95.8%-98.5%) and specificity was 96.2% (95% CI: 93.6%-97.8%). In the pregnant population, sensitivity was 98.7% (95% CI: 95.5%-99.7%) and specificity was 98.3% (95% CI: 94.0%-99.5%). Precision studies (CLSI EP05-A3) and interference testing (CLSI EP7-A2) were also performed.
Technological Characteristics
Chemiluminescent immunoassay for use on ADVIA Centaur systems. Analyzes human serum, EDTA plasma, and lithium heparin plasma. Reagents stable for 60 days onboard; calibration interval 28 days. Automated processing.
Indications for Use
Indicated for sexually active adults and expectant mothers to aid in the presumptive diagnosis of HSV-1 infection. Not for blood/plasma donor screening. Performance not established for immunocompromised or pediatric patients.
Regulatory Classification
Identification
Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).
Predicate Devices
Focus HerpeSelect 1 and 2 Immunoblot IgG (K000238)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 23, 2018
Biokit, S.A. Joan Guixer QA & RA Director Can Male S/N Llica d'Amunt, 08186 Barcelona, Spain
Re: K181333
Trade/Device Name: ADVIA Centaur Herpes-1 IgG Regulation Number: 21 CFR 866.3305 Regulation Name: Herpes simplex virus serological assays Regulatory Class: Class II Product Code: MXJ Dated: May 17, 2018 Received: May 25, 2018
Dear Joan Guixer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
#### Steven R. Gitterman -S for
Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) k181333
Device Name ADVIA Centaur Herpes-1 IgG
#### Indications for Use (Describe)
The ADVIA Centaur® Herpes-1 IgG (HSV1) assay is for in vitro diagnostic use in the qualitative determination of IgG antibodies to herpes simplex virus type 1 (HSV-1) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur systems. The test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. The predictive or negative result depends on the prevalence of HSV-1 infection in the population and the pre-test likelihood of HSV-1 infection.
The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for immunocompromised patients, pediatic patients or matrices other than human serum and plasma (EDTA and lithium heparin).
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Biokit, a Werfen company. The logo consists of a stylized "W" made up of two overlapping blue triangles, followed by the company name "Biokit" in gray text. Below the company name is the text "A Werfen Company" in a smaller font size, also in gray.
# 510(K) SUMMARY
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
| 1. Submitter's Information | Biokit, S.A.<br>Can Malé S/N<br>Lliçà d'Amunt 08186<br>Barcelona (Spain) |
|----------------------------|--------------------------------------------------------------------------|
|----------------------------|--------------------------------------------------------------------------|
| 2. Contact Person | Joan Guixer, QA & RA Director<br>Phone: +34 93 860 90 00 / +34 657 88 33 47<br>Email: jguixer@biokit.com |
|-------------------|----------------------------------------------------------------------------------------------------------|
|-------------------|----------------------------------------------------------------------------------------------------------|
| 3. Preparation Date | May 17th, 2018 |
|---------------------|----------------|
|---------------------|----------------|
| 5. Regulatory<br>Information | Regulation Number | 21 CFR 866.3305 |
|------------------------------|------------------------|---------------------------------------------|
| | Regulation Description | Herpes simplex virus<br>serological assays. |
| | Classification | Class II Special Controls |
| | Product Code | MXJ |
| | Classification Panel | Microbiology |
| 6. Predicate Device | K000238 (Focus HerpeSelect 1 and 2 Immunoblot IgG) |
|---------------------|----------------------------------------------------|
|---------------------|----------------------------------------------------|
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Biokit, a Werfen company. The logo consists of a stylized "W" made up of two overlapping blue shapes, followed by the word "Biokit" in gray. Below "Biokit" is the text "A Werfen Company" in a smaller, gray font.
| <b>7. Indications for Use /<br/>Intended Use</b> | The ADVIA Centaur Herpes-1 IgG (HSV1) assay is for <i>in vitro</i> diagnostic use in the qualitative determination of IgG antibodies to herpes simplex virus type 1 (HSV-1) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur systems. This assay is intended for use as an aid in the determination of serological status to HSV-1 and in the diagnosis of herpes simplex virus infection. |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| COMPARISON PREDICATE | | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Predicate | New Device |
| Trade Names | Focus HerpeSelect 1 and 2<br>Immunoblot IgG | ADVIA Centaur Herpes-1 IgG |
| 510K nº | K000238 | Pending |
| Manufacturer | Focus Diagnostic<br>Cypress, CA 90630 -USA | Siemens Heathcare<br>Diagnostics Inc.<br>511 Benedict Avenue,<br>Tarrytown, NY 10591- USA |
| Similarities | | |
| Intended use | Focus Diagnostics HerpeSelect 1<br>and 2 Immunoblot IgG test is | The ADVIA Centaur Herpes-1<br>IgG (HSV1) assay is for in vitro |
| | intended for qualitatively detecting<br>the presence or absence of human<br>IgG class antibodies to HSV-1 and<br>HSV-2 in human sera. The test is<br>indicated for testing sexually active<br>adults or expectant mothers for<br>aiding in the presumptive diagnosis<br>of HSV-1 and HSV-2 infection. The<br>predictive value of a positive or<br>negative result depends on the<br>population's prevalence and the<br>pretest likelihood of HSV- 1 and<br>HSV-2 infection. The performance<br>of this assay has not been<br>established for use in a pediatric<br>population, for neonatal screening,<br>for testing of<br>immunocompromised patients, for<br>use by a point of care facility or for<br>use with automated equipment | diagnostic use in the<br>qualitative determination of<br>IgG antibodies to herpes<br>simplex virus type 1 (HSV-1)<br>in human serum and plasma<br>(EDTA and lithium heparin)<br>using the ADVIA Centaur<br>systems. This assay is<br>intended for use as an aid in<br>the determination of<br>serological status to HSV-1<br>and in the diagnosis of<br>herpes simplex virus<br>infection. |
| Measurand | To detect the presence or absence<br>of human IgG class antibodies to<br>HSV-1 and HSV-2 | To detect IgG antibodies to<br>herpes simplex virus type 1<br>(HSV-1). |
| Regulation Section | 21 CFR 866.3305 | Same |
| Product Code | LGC | MXJ |
| Classification | Class II Special Controls | Same |
| Assay Type | Qualitative | Same |
| Differences | | |
| Technology | Nitrocellulose immunoblot | Chemiluminescent<br>immunoassay |
| Sample type | Human serum | Human serum and<br>plasma (EDTA and lithium<br>heparin) |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Biokit, which is a Werfen company. The logo consists of a stylized "W" made up of two overlapping blue shapes, followed by the word "Biokit" in gray text. Below the word "Biokit" is the text "A Werfen Company" in a smaller font size, also in gray.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Biokit, a Werfen Company. The logo consists of a blue geometric shape resembling two overlapping chevrons, followed by the company name "Biokit" in gray text. Below the company name, the text "A Werfen Company" is written in a smaller, gray font.
## 9. Performance Summary
## Precision
A precision study was performed according to CLSI EP05-A3 using the Negative and Positive Controls as well as 6 samples prepared at different levels in the assay range. Each material was tested in duplicate, twice a day for 20 days, for a total of 80 replicates per level.
The ADVIA Centaur HSV1 assay is designed to have the following repeatability and withinlab precision requirements:
| | Design Requirement | |
|----------------------------------|-----------------------|--------------------|
| ADVIA Centaur HSV1 Level (Index) | Repeatability<br>% CV | Within-Lab<br>% CV |
| < 0.5 | NAa | NA |
| 0.51–0.79 | ≤ 10.0% | ≤ 15.0% |
| 0.80–1.20 | ≤ 6.0% | ≤ 8.0% |
| 1.21-3.00 | ≤ 5.0% | ≤ 7.0% |
| 3.01-6.00 | ≤ 5.0% | ≤ 7.0% |
| > 6.00 | ≤ 5.0% | ≤ 10.0% |
೩ NA = not applicable
The results for this study are presented below.
| Specimen Type | N | Mean<br>(Index) | Repeatability | | Within-Lab | |
|------------------------------|----|-----------------|---------------|--------|------------|-------|
| | | | SD | CV (%) | SD | CV(%) |
| Negative Control<br>(Plasma) | 80 | 0.26 | 0.01 | NAa | 0.03 | NAa |
| Positive Control<br>(Plasma) | 80 | 2.98 | 0.10 | 3.2 | 0.18 | 6.1 |
| Serum 1 | 80 | 0.25 | 0.03 | NAa | 0.04 | NAa |
| Serum 2 | 80 | 0.57 | 0.03 | 4.8 | 0.03 | 5.9 |
| Serum 3 | 80 | 0.97 | 0.03 | 3.1 | 0.04 | 4.5 |
| Serum 4 | 80 | 1.61 | 0.05 | 3.1 | 0.09 | 5.3 |
| Serum 5 | 80 | 4.08 | 0.20 | 4.9 | 0.26 | 6.4 |
| Serum 6 | 80 | 13.37 | 0.53 | 4.0 | 1.04 | 7.7 |
### Precision results for the HSV-1 Assay
NA = Not applicable
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Biokit, a Werfen company. The logo consists of a stylized "W" made up of two overlapping blue shapes, followed by the word "Biokit" in gray. Below the word "Biokit" is the text "A Werfen Company" in a smaller font size, also in gray.
## Sample matrix
This study was performed on one ADVIA Centaur XP instrument, using 68 sets of matched samples of different matrixes (serum, serum separator tube (SST), EDTA plasma and lithium heparin plasma) from commercial sources
The samples were analyzed in duplicate in randomized order using one reagent lot.
- . Comparing Serum Separator Tube (y) vs Serum (x): (Deming regression) y = -0.01 + 0.99 x, r=0.996, sample range for both matrixes 0.10-19.15 Index Value and 0.10-19.59 Index Value respectively. Note – correlation coefficient is calculated using linear regression.
- . Comparing EDTA Plasma(y) vs Serum (x): (Deming regression) y= -0.02 + 0.96 x, r=0.998, sample range for both matrixes 0.10-19.14 Index Value and 0.11-19.19 Index Value respectively. Note - correlation coefficient is calculated using linear regression.
- . Comparing Lithium Heparin Plasma(y) vs Serum (x): (Deming regression) y= -0.02 + 0.96 x, r=0.998, sample range for both matrixes 0.10-19.15 Index Value and 0.10-18.73 Index Value respectively. Note - correlation coefficient is calculated using linear regression.
The results supports that Serum Separator Tube is equivalent matrix to Serum, EDTA Plasma is equivalent matrix to Serum and Lithium Heparin Plasma is an equivalent matrix to Serum.
## Panels
The commercial sample panels, Seracare Diagnostics and ToRCH-mixed Zeptometrix, and CDC panel were analysed.
The Seracare Diagnostics panel included 25 characterized HSV samples. All of the samples were evaluated with the ADVIA Centaur Herpes-1 IgG assay on ADVIA Centaur XP instrument and 92% total agreement was observed with reference assay 1.
The ToRCH-mixed Zeptometrix panel included 24 characterized HSV samples. All of the samples were evaluated with the ADVIA Centaur Herpes-1 IgG assay on ADVIA Centaur XP instrument and 96% total agreement was observed with reference assay 1.
The CDC panel included 100 blind characterized HSV samples. All of the samples were evaluated with the ADVIA Centaur Herpes-1 IgG assay on ADVIA Centaur XP instrument and 100% total agreement was observed in concordance with the results provided by the CDC.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Biokit, a Werfen Company. The logo consists of a stylized "W" made up of two overlapping blue shapes, followed by the name "Biokit" in gray text. Below the name is the text "A Werfen Company" in a smaller font size, also in gray.
## Interferences
Potential interference in the ADVIA Centaur HSV1 assay from the compounds listed below is designed to be ≤ 10%. Interfering substances at the levels indicated were tested as described in CLSI Document EP7-A2.
The testing was done in three levels of samples in the assay range (high negative, low positive and positive samples) with at least one reagent lot on ADVIA Centaur XP instrument.
Testing confirmed no interference (Demonstrate ≤ 10% change in results) for the ADVIA Centaur Herpes-1 IgG up to the following concentrations:
| Serum specimens that are... | Demonstrate ≤10% change in results up to... |
|-----------------------------|---------------------------------------------|
| Biotin | 3500 ng/mL |
| Hemoglobin | 500 mg/dL |
| Bilirubin complex | 40 mg/dL |
| Bilirubin free | 40 mg/dL |
| Hypoproteinemia | 3 g/dL |
| Hyperproteinemia | 12 g/dL |
| Lipemia | 1000 mg/dL |
| Cholesterol | 400 mg/dL |
## Cross-reactivity
The ADVIA Centaur HSV1 assay was evaluated for potential cross-reactivity with other viral antibodies, disease-state specimens, and other populations. The HSV-1 lgG status of each specimen was verified using the Comparative Assay. Repeat equivocal specimens on the Comparative Assay were sent to a reference laboratory for Western Blot testing. Total percent agreement for the following clinical categories was 95.8% (413/431). The following results were obtained:
### Cross Reactivity Study Performance Results
| | Number | ADVIA Centaur HSV1 | Comparative Assay | |
|--------------------------------------------------------|---------------|--------------------------------|-------------------------------|--|
| Clinical Category | tested | Reactive | Positive | |
| Antibody to hepatitis B surface antigen (Anti-<br>HBs) | 10 | 8 | 8 | |
| Anti-gliadin | 10 | 6 | 6 | |
| Anti-nuclear antibody (ANA) | 10 | 5 | 5 | |
| Clinical Category | Number tested | ADVIA Centaur HSV1<br>Reactive | Comparative Assay<br>Positive | |
| Candida albicans | 10 | 10 | 9 | |
| Chlamydia trachomatis | 10 | 6 | 7a | |
| Cytomegalovirus (CMV IgG) | 10 | 6 | 7b | |
| Elevated IgG | 10 | 8 | 8 | |
| Elevated IgM | 10 | 9 | 10a/c | |
| Epstein-Barr virus (EBV VCA IgG) | 10 | 9 | 9 | |
| Escherichia coli | 10 | 9 | 9 | |
| Flu vaccine recipients | 10 | 3 | 3 | |
| Gardnerella vaginalis | 10 | 7 | 7 | |
| Hepatitis C virus (HCV) | 25 | 19 | 21 | |
| Hepatitis B surface antigen (HBsAg) | 10 | 9 | 9 | |
| Herpes simplex virus type 2 (HSV-2) | 74 | 35 | 33a | |
| Heterophile antibodies | 10 | 1 | 1 | |
| Human anti-mouse antibody (HAMA) | 31 | 24 | 25a | |
| Human herpes virus 6 (HHV-6) | 10 | 6 | 5 | |
| Human herpes virus 8 (HHV-8) | 10 | 9 | 9 | |
| Human immunodeficiency virus (HIV) | 10 | 10 | 10 | |
| Human Papillomavirus (HPV) | 10 | 6 | 6a | |
| Multiple myeloma | 10 | 10 | 10 | |
| Neisseria gonorrhoeae | 25 | 17 | 16 | |
| Parvovirus B19 | 10 | 4 | 4 | |
| Rheumatoid Factor (RF) | 10 | 6 | 6 | |
| Rubella IgG | 10 | 7 | 7 | |
| Syphilis | 10 | 9 | 9a | |
| Systemic lupus erythematosus (SLE) | 26 | 17 | 17 | |
| Toxoplasma IgG | 10 | 10 | 10 | |
| Varicella zoster virus (VZV) | 10 | 5 | 6a | |
| Total Specimens | 431 | 290 | 292 | |
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Biokit. The logo consists of a blue icon on the left and the word "Biokit" in gray on the right. The icon is made up of two overlapping shapes that resemble stylized checkmarks or arrows. The blue color of the icon appears to have a gradient, with a darker shade on the left and a lighter shade on the right.
A Werfen Company
ª One sample was confirmed positive by Western Blot
Ь One sample was confirmed positive by Comparative Assay
< Two samples were confirmed positive by Comparative Assay
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Biokit, a Werfen Company. The logo consists of two blue chevron-like shapes stacked on top of each other, followed by the word "Biokit" in gray. Below the word "Biokit" is the text "A Werfen Company" in a smaller, gray font. The logo is simple and modern, with a focus on the company name and affiliation.
## Clinical study
A multisite clinical study to compare the reference with ADVIA Centaur Herpes-1 lgG was performed. Sensitivity and specificity were determined by comparing the performance of the ADVIA Centaur HSV1 assay to a commercially available anti-HSV-1 IgG immunoblot method (Comparative Assay) and a validated Western Blot reference confirmatory test (University of Washington, Seattle).A total of 864 specimens (≥ 18 years of age), including specimens from 274 pregnant women, were collected within the United States and tested at 3 independent external laboratories. Samples were analyzed in singlicate. Of these specimens, 520 were reactive, and 344 were nonreactive with the ADVIA Centaur HSV1 assay. The overall agreement was 97.0% (838/864) with a 95% confidence interval (CI) of 95.6%-97.9%.
Of the 864 specimens tested by the Comparative Assay, 22 were equivocal and further tested by the Western Blot test. After Western Blot testing, 19 of the specimens were resolved to be negative and 3 remained equivocal.
| Intended use population | | Comparative Assay.<br>Equivocal results have been resolved by Western Blot (WB) | | | |
|-------------------------|-------------|---------------------------------------------------------------------------------|----------|-----------|-------|
| | | Positive | Negative | Equivocal | Total |
| ADVIA Centaur HSV1 | Reactive | 507 | 13 | 0 | 520 |
| | Nonreactive | 10 | 331 | 3 | 344 |
| | Total | 517 | 344 | 3 | 864 |
The results obtained for the intended use population are presented below
The sensitivity of the ADVIA Centaur HSV1 assay was 97.5% (507/520), with a 95% confidence interval (CI) of 95.8%-98.5%.
The specificity of the ADVIA Centaur HSV1 assay was 96.2% (331/344), with a 95% confidence interval (CI) of 93.6%-97.8%.
Sensitivity and specificity were determined for the pregnant population by comparing the performance of the ADVIA Centaur HSV1 assay to the Comparative Assay and Western Blot confirmatory method.The results of the pregnant women samples are presented below:
| Pregnant women | Comparative Assay<br>Equivocal results have been resolved by Western Blot (WB) | | | | |
|--------------------|--------------------------------------------------------------------------------|-----|-----|---|-----|
| ADVIA Centaur HSV1 | Reactive | 155 | 2 | 0 | 157 |
| | Nonreactive | 2 | 115 | 0 | 117 |
| | Total | 157 | 117 | 0 | 274 |
### Pregnant women Study Contingency (WB resolution) table
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for Biokit. The logo consists of two overlapping blue shapes resembling stylized "V"s on the left. To the right of the shapes is the word "Biokit" in gray, sans-serif font. The overall design is clean and modern.
## A Werfen Company
The sensitivity of the ADVIA Centaur HSV1 assay was 98.7% (155/157) with a 95% confidence interval of 95.5%-99.7%.
The specificity of the ADVIA Centaur HSV1 assay was 98.3% (115/117) with a 95% confidence interval of 94.0%-99.5%.
#### 10. Stability
The onboard stability of the ADVIA Centaur Herpes-1 IgG reagents is 60 days with a calibration interval of 28 days. The onboard stability of the ADVIA Centaur Herpes-1 IgG Calibrators is 8 hours. The opened vial stability of the ADVIA Centaur Herpes-1 IgG Calibrators is 65 days. Unopened reagents and calibrators are stable until the date printed on the box label when stored at 2-8°C.
#### 11. Conclusion
The analytical and clinical study results demonstrate that the ADVIA Centaur Herpes-1 IgG are substantially equivalent to the predicate device, Focus HerpeSelect 1 and 2 Immunoblot IgG (FDA cleared under K000238), and that the assay is safe and effective for its labeled intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.