EliA B2-Glycoprotein I IgA Immunoassay, EliA Cardiolipin IgA Immunoassay
K181329 · Phadia AB · MSV · Jun 25, 2018 · Immunology
Device Facts
Record ID
K181329
Device Name
EliA B2-Glycoprotein I IgA Immunoassay, EliA Cardiolipin IgA Immunoassay
Applicant
Phadia AB
Product Code
MSV · Immunology
Decision Date
Jun 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
EliA β2-Glycoprotein I IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to β2-Glycoprotein I in human serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of antiphospholipid syndrome (APS) as well as thrombotic disorders related to systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA β2-Glycoprotein I IgA uses the EliA IgA method on the instrument Phadia 2500/5000. EliA Cardiolipin IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to cardiolipin in human serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of antiphospholipid syndrome (APS) as well as thrombotic disorders related to systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Cardiolipin IgA uses the EliA IgA method on the instrument Phadia 2500/5000.
Device Story
EliA β2-Glycoprotein I IgA and EliA Cardiolipin IgA are semi-quantitative immunoassays for detecting IgA autoantibodies in human serum/plasma. The device uses polystyrene wells coated with human β2-Glycoprotein I or bovine cardiolipin/β2-glycoprotein I. Patient samples are incubated in wells; specific IgA antibodies bind to the antigen. After washing, enzyme-labeled anti-IgA conjugate is added, followed by a development solution. Fluorescence is measured to quantify antibody concentration. The system is automated on the Phadia 2500/5000 instrument, which handles pipetting, incubation, washing, and signal detection. Results are compared to calibrators to determine IgA levels. The device is used in clinical laboratories by trained personnel. Output aids clinicians in diagnosing APS and SLE-related thrombotic disorders.
Clinical Evidence
Bench testing only. Precision/reproducibility assessed over 7 days (n=84 replicates/sample). Linearity confirmed across reportable ranges. Method comparison study (>100 samples) between Phadia 250 and Phadia 2500/5000 showed high agreement (PPA/NPA/TPA metrics provided). LoD/LoQ determined per CLSI EP17-A2. No new clinical diagnostic studies performed; clinical performance previously established in K112414 and K131821.
Technological Characteristics
Polystyrene wells coated with human β2-Glycoprotein I or bovine cardiolipin/β2-glycoprotein I. Detection via β-galactosidase labeled anti-IgA conjugate and 4-methylumbelliferyl-β-D-galactoside development solution. Fluorescence-based signal detection. Automated system (Phadia 2500/5000) with random access processing. Reagents include PBS, BSA, detergent, and sodium azide preservatives. Sterilization not specified; modular reagent system.
Indications for Use
Indicated for patients suspected of having antiphospholipid syndrome (APS) or thrombotic disorders related to systemic lupus erythematosus (SLE). Used as an aid in diagnosis in conjunction with other clinical and laboratory findings. Prescription use only.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
EliA ß2-Glycoprotein I IgA on Phadia 250 (K112414)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K181329
B. Purpose for Submission:
Adding previously cleared assays on a new instrument platform (Phadia 2500/5000)
C. Measurand:
IgA antibodies to β-2-glycoprotein I
IgA autoantibodies to Cardiolipin
D. Type of Test:
Automated semi-quantitative solid-phase measurement immunoassays
E. Applicant:
Phadia US, Inc
F. Proprietary and Established Names:
EliA B2-Glycoprotein I IgA Immunoassay,
EliA Cardiolipin IgA Immunoassay
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5660 Multiple autoantibodies immunological test system
2. Classification:
Class II
3. Product code:
MSV System, Test, Antibodies, β2-Glycoprotein I (β2-GPI)
MID System, Test, Anti-Cardiolipin Immunological
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended uses:
EliA β2-Glycoprotein I IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to β2-Glycoprotein I in human serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of antiphospholipid syndrome (APS) as well as thrombotic disorders related to systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA β2-Glycoprotein I IgA uses the EliA IgA method on the instrument Phadia 2500/5000.
{1}
EliA Cardiolipin IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to cardiolipin in human serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of antiphospholipid syndrome (APS) as well as thrombotic disorders related to systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Cardiolipin IgA uses the EliA IgA method on the instrument Phadia 2500/5000.
2. Indications for use:
Same as intended uses
3. Special conditions for use statements:
For prescription use only
4. Special instrument requirements:
Phadia 2500/5000 instrument.
I. Device Description:
- EliA β2-Glycoprotein I IgA wells are coated with human β2-Glycoprotein I antigen, ready to use
- EliA Cardiolipin IgA wells are coated with bovine cardiolipin antigen and bovine β2-glycoprotein I as co-factor, ready to use
EliA Method-Specific Reagents (Phadia 2500/5000):
- EliA Sample Diluent (yellow colored): PBS containing BSA, detergent and 0.095% (w/v) sodium azide, ready to use
- EliA IgA Conjugate (blue colored): β-Galactosidase labeled anti-IgA (mouse monoclonal antibodies) in PBS containing BSA and 0.06% (w/v) sodium azide, ready to use
- EliA IgA Calibrator Strips: Human IgA (0, 0.3, 1.1, 5, 15, 80 μg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide, ready to use
- EliA IgA Curve Control Strips: Human IgA (20 μg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide, ready to use;
- EliA IgA Calibrator Well: Coated with mouse monoclonal antibodies, ready to use
Phadia 2500/5000 General Reagents:
- Development Solution (0.01% 4-Methylumbelliferyl-β-D-galactoside, <0.0010% preservative), ready for use
- Stop Solution (4% Sodium Carbonate), ready for use
- Dilution Wells, high density polyethylene wells ready for use
- Pipette Tips in Racks, ready for use
- Washing Solution (information in separate Washing Solution package insert).
EliA uses a modular reagent system. The EliA reagents are available as modular packages, each purchased separately. All packages are required to carry out an EliA β2-Glycoprotein I IgA test and the EliA Cardiolipin IgA Test.
2
{2}
J. Substantial Equivalence Information:
1. Predicate device names and 510(k) numbers:
K112414, EliA B2-Glycoprotein I IgA Immunoassay
K131821, EliA Cardiolipin IgA Immunoassay
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device
*Phadia 2500/5000 | Predicate
Phadia 250 |
| Intended Use
EliA β-2-Glycoprotein I IgA | EliA β-2-Glycoprotein I IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to β2-Glycoprotein I in human serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of antiphospholipid syndrome (APS) as well as thrombotic disorders related to systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. | Same |
| Intended Use
EliA Cardiolipin IgA | EliA Cardiolipin IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to cardiolipin in human serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of antiphospholipid syndrome (APS) as well as thrombotic disorders related to systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. | Same |
| Signal | Immuno-fluorescence measurement | Same |
| Assay type | ELISA | Same |
| Type of test | Semi-quantitative | Same |
| Antigen
β-2 Glycoprotein I
Cardiolipin | Human purified β-2 Glycoprotein I
Purified bovine cardiolipin and bovine β-2 Glycoprotein I | Same
Same |
| Common, dedicated
Phadia reagents | Same
Introduction of new article numbers for Development Solution, Stop Solution and Washing Solution is | Same |
3
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device *Phadia 2500/5000 | Predicate Phadia 250 |
| | only due to larger filling volumes which are required for the bigger instruments Phadia 2500/5000 | |
| Time to 1st result | About 2 hours | Same |
| Detection antibody (conjugate) | β-Galactosidase labeled anti-IgA (mouse anti human monoclonal antibodies) | Same |
| Incubation temperature | 37°C | Same |
| Sample matrix | Serum or plasma (heparin, EDTA) | Serum or plasma (heparin, EDTA, citrate) |
| Calibrator strips | Human IgA (0, 0.3, 1.5, 5, 15, 80 μg/L) | Same |
| Control | EliA IgA curve control strips, 5μg/L | Same |
| Measuring range | β-2 Glycoprotein I: 0.3-183 U/mL Cardiolipin: 0.3-181 APL-U/mL | Same |
| Cutt-off | β-2 Glycoprotein I: Negative <7 U/mL Equivocal 7-10 U/mL Positive >10 U/mL Cardiolipin: Negative <14 APL-U/mL Equivocal 14-10 APL-U/mL Positive >20 U/mL APL-U/mL | Same |
| Assay set-up | Random access | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Instrument | Instrument Phadia 2500/5000 | Instrument Phadia 250 |
| Sample dilution, pipets | Disposable Pipette Tips in Racks for pipetting samples in Dilution Well | Steel pipette to dilute the samples in Dilution Plates |
| Loading of EliA Carriers | The Phadia 2500/5000 instruments do not have Loading Tray. The EliA carriers are loaded into racks which are directly transferred to the cooled storage compartment | EliA carriers are loaded manually on the Loading Tray from where they can be processed directly or transferred to the cooled storage compartment. |
| Barcode reader | On the Phadia 2500/5000 | The Phadia 250 instrument has a |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | the reagents are on a moving belt which conveys them past the barcode reader. The lot-specific information will be read automatically by the instrument during loading. | built-in barcode reader at the front of the instrument, but the operator needs to scan the barcodes manually by showing the reagents to the barcode reader. Alternatively, the operator can also enter the characters below the barcode manually. |
| Process time / Time to patient result | Phadia 2500/5000 instruments process two Wells in parallel in 48 seconds. Phadia 2500/5000 provides the results at a 24 seconds interval. | Phadia 250 needs one minute to process one Well. Phadia 250 provides the results at a one minute interval. |
| Daily throughput | About 2500/5000 tests | About 250 tests |
*Phadia 2500/5000 – Phadia 2500 and Phadia 5000 are identical instruments except for sample throughput. The difference is that Phadia 2500 consists of 1 process module (2 process lines) and Phadia 5000 consists of 2 process modules (2x2 process lines). For that reason Phadia uses the Phadia 2500 and Phadia 5000 instrument names interchangeably.
## K. Standard/Guidance Document Referenced:
- EP17-A Protocols for Determination of Limits of Detection and Limits of Quantification
- EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures
- EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
## L. Test Principle:
The EliA wells are molded cups comparable to excised wells from a microtiter plate. They are made of polystyrene and are coated with the respective antigen. The wells are at the same time a holder of the coupled antigen for convenient automation and a reaction chamber with reaction/washing solution handling based on pipetting to add and aspiration to remove liquids.
The EliA wells are coated with human β2-Glycoprotein I antigen or with bovine cardiolipin antigen. If present in the patient's specimen, antibodies to these proteins bind to the specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgA antibodies (EliA IgA Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the more specific IgA is present in the specimen. To evaluate test results, the response for patient samples is compared directly to the response for calibrators.
{5}
M. Performance Characteristics:
1. Analytical performance:
All results presented below met the manufacturer's predetermined acceptance criteria
a. Precision/Reproducibility:
The precision of the assay was assessed in a study with a total of 21 runs, one run/day over a period of seven days. Each sample was tested in four replicates/run giving a total of 84 replicates per sample (3 instruments × 7 runs × 4 replicates = 84). The data was calculated against the calibration curve from Day 1. Four serum native patient serum samples were tested, the same samples were tested on all instruments.
One lot of EliA β2-Glycoprotein I IgA and EliA Cardiolipin IgA was used in these studies because inter-lot-variation was determined in the predicate devices K112414 and K131821. The results for all three instruments are summarized in the tables below:
EliA β2-Glycoprotein I IgA on Phadia 2500/5000:
| Mean (EliA U/mL) | Within-Run | | Between-Run | | Between-Instrument | | Total Imprecision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1.6 | 0.1 | 8.0 | 0.2 | 9.8 | 0.4 | 23.0 | 0.4 | 26.2 |
| 8.0 | 0.3 | 3.7 | 0.3 | 3.5 | 0.3 | 4.2 | 0.5 | 6.6 |
| 10.4 | 0.4 | 3.7 | 0.5 | 5.0 | 0.4 | 3.4 | 0.7 | 7.1 |
| 42.8 | 1.4 | 3.3 | 2.2 | 5.2 | 1.2 | 2.8 | 2.9 | 6.7 |
| 136.6 | 6.6 | 4.8 | 10.3 | 7.5 | 5.1 | 3.7 | 13.2 | 9.7 |
EliA Cardiolipin IgA on Phadia 2500/5000:
| Mean (APL-U/mL) | Within-Run | | Between-Run | | Between-Instrument | | Total Imprecision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 2.1 | 0.2 | 7.2 | 0.2 | 7.9 | 0.3 | 15.1 | 0.4 | 18.5 |
| 13.5 | 0.8 | 5.6 | 0.3 | 2.3 | 0.2 | 1.4 | 0.8 | 6.2 |
| 19.4 | 0.7 | 3.9 | 1.1 | 5.5 | 0.6 | 3.0 | 1.4 | 7.4 |
| 93.0 | 5.4 | 5.8 | 1.4 | 1.5 | 4.8 | 5.2 | 7.4 | 7.9 |
| 153.0 | 10.7 | 7.0 | 12.0 | 7.9 | 8.5 | 5.5 | 18.2 | 11.9 |
b. Linearity/assay reportable range:
Four patient serum samples were diluted in sample diluent and tested in triplicates with one lot of EliA β2-Glycoprotein I IgA and EliA Cardiolipin IgA reagents and one set of system reagents on Phadia 2500/5000. The results of the dilutions were compared with their expected values. The ratio observed/expected (O/E) was calculated. Mean observed value was used in the calculation. The results of the linear regression analysis are summarized below:
{6}
EliA β2-Glycoprotein I IgA on Phadia 2500/5000
| Dilution range (EliA U/mL) | Slope | Intercept | R² |
| --- | --- | --- | --- |
| 0.4 – 66.6 | 1.00 | 0.24 | 1.00 |
| 5.2 – 205.2 | 1.03 | -0.12 | 1.00 |
| 6.4 – 242.8 | 1.02 | -0.46 | 1.00 |
| 5.5 – 198.8 | 1.02 | 1.80 | 1.00 |
The reportable range (Limit of Detection to upper limit) for EliA β2-Glycoprotein I IgA is from 0.3 to 183 U/mL. The measuring range (Limit of Quantitation, to upper limit) is from 1.1 to 183 U/mL.
EliA Cardiolipin IgA on Phadia 2500/5000
| Dilution range (APL-U/mL) | Slope | Intercept | R² |
| --- | --- | --- | --- |
| 0.5 – 49.5 | 1.01 | 0.40 | 1.00 |
| 1.8 – 91.3 | 0.98 | -0.25 | 1.00 |
| 2.4 – 147.0 | 0.98 | -1.45 | 1.00 |
| 4.4 – 221.6 | 1.00 | 3.76 | 1.00 |
The reportable range (Limit of Detection to upper limit) for EliA Cardiolipin IgA is from 0.3 to 181 APL-U/mL. The measuring range (Limit of Quantitation to upper limit) is from 1.0 to 181 APL-U/mL.
The following statements is included in the package inserts: "Please note that concentration values between LoD and LoQ may show a higher uncertainty... Please note that due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the measuring range."
c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability, stability, Expected values (controls, calibrators, or methods) for the EliA IgA method was previously reviewed in K062787.
Controls:
Controls are sold separately from the assay.
Traceability:
The IgA calibrators are traceable via unbroken ochain off calibrations to the International Reference Preparation (IRP) 67/86 of the Human Serum Immunoglobulins A, G, M from World Health Organization (WHO).
There are no international standards for β2-Glycoprotein I antibodies. Results are given in arbitrary EliA Units/mL. The instrument measures specific IgA concentration in μg/mL. By using a conversion factor given by the lot-specific code of the EliA β2-Glycoprotein I IgA Well, these results are automatically converted to U/mL.
{7}
The standardization of the Cardiolipin IgA assay is adjusted to a set of established standard sera (Harris et al., 1987). Results are expressed in APL-U/ml (1 APL-Unit corresponds to the binding activity of 1 µg/ml of cardiolipin IgA antibody that was purified from the standard serum by affinity chromatography).
d. Detection limit:
The limit of blank (LoB) is based on 66 blank replicates. The limit of detection is based on 66 low level replicates. The limit of quantitation is based on 66 determinations and a target uncertainty goal of 20%. The results are summarized in the table below:
| Phadia 2500/5000 | Units | LoB | LoD | LoQ |
| --- | --- | --- | --- | --- |
| EliA β2-Glycoprotein I IgA | EliA U/mL | 0.1 | 0.3 | 1.1 |
| EliA Cardiolipin IgA | APL-U/mL | 0.1 | 0.3 | 1.0 |
e. Analytical specificity:
Interference: Interference studies were previously reviewed in K112414 for β2-Glycoprotein I IgA and in K131821 for Cardiolipin IgA.
Carry-over: Phadia 2500/5000 instruments use disposable tips for pipetting samples and a separate pipete for the conjugate, therefore carry-over from samples to conjugate was not evaluated.
f. Assay cut-off:
The assays' cut-offs are the same as in the predicate devices, K112414 for β2-Glycoprotein I IgA and in K131821 for Cardiolipin IgA and are summarized in the table below:
| | EliA™ β2-Glycoprotein I IgA | EliA Cardiolipin IgA |
| --- | --- | --- |
| Negative | <7 U/mL | < 14 APL-U/mL |
| Equivical | 7-10 U/mL | 14-20 APL-U/mL |
| Positive | >10 U/mL | > 20 APL-U/mL |
2. Comparison studies:
a. Method comparison with predicate device:
Instrument comparison was performed with the predicate device, see 2c below.
b. Matrix comparison:
Matrix comparison between serum and plasma (Li-Heparin, and EDTA) was reviewed in the predicate K112414 for β2-Glycoprotein I IgA and in K131821 for Cardiolipin IgA.
c. Instrument comparison
The purpose of this study is to evaluate conformance and show comparability of β2-Glycoprotein I IgA and Cardiolipin IgA assays on the instruments Phadia 250
{8}
(PH250) versus Phadia 2500/5000 (PH2500/5000). The samples were run in single replicates on one PH 250 instrument and three PH2500/5000 (A, B, C) instruments. Only samples inside the measuring range were included in the calculations. Results were analyzed by Passing-Bablock regression.
EliA β2-Glycoprotein I IgA: 100 samples were used in the study, 63 positive, 25 negative and 12 equivocal samples. The results are summarised below:
| Instrument | Intercept | 95% CI | Slope | 95% CI |
| --- | --- | --- | --- | --- |
| PH2500/5000 A | 0.85 | 0.58 to 1.26 | 0.98 | 0.95 to 1.01 |
| PH2500/5000 B | 0.44 | -0.20 to 0.76 | 1.02 | 0.99 to 1.04 |
| PH2500/5000 C | 0.22 | -0.06 to 0.53 | 1.01 | 0.98 to 1.05 |
Positive percent agreement (PPA), negative percent agreement (NPA), and total percent agreement (TPA) were evaluated with equivocal results considered positive in the table below:
| | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
| --- | --- | --- | --- |
| PPA | 98.7% | 97.3% | 98.7% |
| 95% CI | 92.8% – 100% | 90.6% – 99.7% | 92.8% – 100% |
| NPA | 80.0% | 88.0% | 84.0% |
| 95% CI | 59.3% – 93.2% | 68.8% – 97.5% | 63.9% – 95.5% |
| TPA | 94.0% | 94.9% | 95.0% |
| 95% CI | 87.4% – 97.8% | 88.6% – 98.3% | 88.7% – 98.4% |
Results when equivocal results are considered negative are presented in the table below:
| | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
| --- | --- | --- | --- |
| PPA | 100.0% | 100.0% | 100.0% |
| 95% CI | 94.3% – 100% | 94.2% – 100% | 94.3% – 100% |
| NPA | 94.6% | 97.3% | 100.0% |
| 95% CI | 81.8% – 99.3% | 85.8% – 99.9% | 90.5% – 100% |
| TPA | 98.0% | 99.0% | 100.0% |
| 95% CI | 93.0% – 99.8% | 94.5% – 100% | 96.4% – 100% |
{9}
EliA Cardiolipin IgA: 101 samples were used in the study, 45 positive, 43 negative and 13 equivocal samples. The results are summarised below:
| Instrument | Intercept | 95% CI | Slope | 95% CI |
| --- | --- | --- | --- | --- |
| PH2500/5000 A | 0.30 | -0.01 to 0.56 | 0.98 | 0.96 to 1.01 |
| PH2500/5000 B | -0.27 | -0.64 to 0.01 | 0.99 | 0.95 to 1.02 |
| PH2500/5000 C | -0.40 | -0.90 to -0.15 | 1.06 | 1.04 to 1.10 |
Positive percent agreement (PPA), negative percent agreement (NPA), and total percent agreement (TPA) were evaluated with equivocal results considered positive in the table below:
| criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
| --- | --- | --- | --- |
| PPA | 100.0% | 100.0% | 100.0% |
| 95% CI | 93.8% – 100% | 93.8% – 100% | 93.8% – 100% |
| NPA | 88.4% | 97.7% | 95.3% |
| 95% CI | 74.9% – 96.1% | 87.7% – 99.9% | 84.2% – 99.4% |
| TPA | 95.0% | 99.0% | 98.0% |
| 95% CI | 88.8% – 98.4% | 94.6% – 100% | 93.0% – 99.8% |
Results when equivocal results are considered negative are presented in the table below:
| criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
| --- | --- | --- | --- |
| PPA | 95.6% | 93.3% | 97.8% |
| 95% CI | 84.9% – 99.5% | 81.7% – 98.6% | 88.2% – 99.9% |
| NPA | 94.6% | 94.6% | 94.6% |
| 95% CI | 85.1% – 98.9% | 85.1% – 98.9% | 85.1% – 98.9% |
| TPA | 95.0% | 94.1% | 96.0% |
| 95% CI | 88.8% – 98.4% | 87.5% – 97.8% | 90.2% – 98.9% |
3. Clinical studies:
Clinical performance was previously reviewed in K112414 for $\beta 2$-Glycoprotein I IgA and in K131821 for Cardiolipin IgA.
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
{10}
4. Clinical cut-off:
See assay cut-off
5. Expected values/Reference range:
A total of 400 apparently healthy Caucasian blood donors matched by age and gender were tested on the Phadia 2500/5000 instrument to evaluate expected values in the normal population. The results are summarized below:
| n=400 | β2-Glycoprotein I IgA | | Cardiolipin IgA | |
| --- | --- | --- | --- | --- |
| | EliA U/mL | | APL-U/mL | |
| | PH250 | PH2500/500 | PH250 | PH2500/500 |
| Mean | 1.8 | 2.1 | 3.6 | 3.6 |
| Median | 1.4 | 1.8 | 2.8 | 2.8 |
| 95th percentile | 4.3 | 4.1 | 7.1 | 6.6 |
| 99th percentile | 8.7 | 9.5 | 17.1 | 16.5 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.