ONIS-PACS is a software device intended to be used by healthcare personnel and intended for viewing, reviewing, performing measurements/quantifications and reporting of medical images and data acquired from DICOM compliant medical imaging systems. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy images and digitized film images must not be used for primary diagnosis or interpretation.
Device Story
ONIS-PACS is a DICOM 3.0-compliant software system for storage, retrieval, visualization, and analysis of medical images. Components include ONIS Viewer (desktop visualization/processing), Local Server (background DICOM communication), ONIS Remote (client-only viewer), WebONIS (ActiveX browser-based viewer), ONIS Site Server (storage/retrieval), and ONIS Organization Server (multi-site access). Input consists of DICOM objects from imaging modalities; processed via workstation or network-distributed servers. Features include 2D/3D visualization, multi-planar reconstruction (MPR), maximum intensity projection (MIP), volume rendering, annotation, and measurement tools. Used in clinical settings by healthcare personnel for diagnostic support and reporting. Output includes displayed images, quantitative measurements, and reports. Benefits include centralized image management and standardized review workflows.
Clinical Evidence
No clinical data. Bench testing only. Software verification and validation testing performed to demonstrate safety and effectiveness for a moderate level of concern device.
Technological Characteristics
Software-only PACS; DICOM 3.0 compliant. Operates on Windows 7, 8, 10, and Windows Server 2008/2012. Supports 2D, mammography, MPR, MIP, and volume rendering. Connectivity via standard DICOM protocols. Includes annotation and measurement tools. No hardware components.
Indications for Use
Indicated for healthcare personnel to view, review, measure, quantify, and report medical images and data from DICOM-compliant imaging systems. Contraindicated for primary diagnosis or interpretation using lossy or digitized film images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
August 13, 2018
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
DigitalCore Co., Ltd % Esin Yesilalan Senior Regulatory Scientist Voisin Consulting, Inc. 222 Third St. Suite 3121 CAMBRIDGE MA 02142
Re: K181318
Trade/Device Name: Onis-Pacs Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: May 17, 2018 Received: May 18, 2018
Dear Esin Yesilalan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices
{1}------------------------------------------------
or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181318
Device Name ONIS-PACS
#### Indications for Use (Describe)
ONIS-PACS is a software device intended to be used by healthcare personnel and intended for viewing, reviewing, performing measurements/quantifications and reporting of medical images and data acquired from DICOM compliant medical imaging systems. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy images and digitized film images must not be used for primary diagnosis or interpretation.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: Arial, sans-serif;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
### I. Submitter Information
| Company Name: | DigitalCore Co.,ltd |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Company Address: | Setagaya-ku Kinuta 5-8-31<br>Ginga-M2<br>157-0073 Tokyo, Japan |
| Company Phone: | Tel: +81 (3) 5727-1064 |
| Company Fax: | +81 (3) 5727-1065 |
| Contact Person: | Cedric Lemoigne, DigitalCore Director<br>cedric@dgcore.co.jp<br>+81-90-5335-8104 (Japan) +33-7-68-58-52-25 (France) |
| Date the summary was prepared: | August 7, 2018 |
## II. Device Identification
| Trade Name: | ONIS-PACS |
|----------------------|--------------------------------------------|
| Common Name: | Picture Archiving and Communication System |
| Classification Name: | System, Image Processing, Radiological |
| Product Code: | LLZ |
| Regulation Number: | 892.2050 |
| Device Class: | Class II |
#### III.Identification of Predicate Devices
| Device Name | iQ-System PACS |
|------------------|--------------------------------------------|
| Manufacturer | IMAGE Information Systems Ltd |
| 510(k) Number | K062488 |
| Regulatory Class | Class II |
| Common Name | Picture Archiving and Communication System |
| Clearance Date | Sep 19 2006 |
## IV. Device Description
ONIS-PACS is a Picture Archiving Communication System (PACS) fully compliant with the DICOM version 3.0 Standard. It is suitable for storing, distributing, retrieving, visualizing, manipulating, performing measurements/quantifications and reporting various DICOM objects.
{4}------------------------------------------------
ONIS Viewer is a desktop application that makes it possible to visualize, manipulate and process medical images of many different modalities. Basic and advanced tools are provided for manipulating and processing images, including multi-planar reconstruction and 3D volume rendering. The software also includes a Local Server service running in the background that can send and receive DICOM images and can respond to DICOM queries. The ONIS Viewer and the Local Server applications must run on the same computer.
ONIS Remote is a desktop application identical to the ONIS Viewer, except that it runs without the Local Server. It must connect to a Site Server to retrieve the studies to be retrieved and processed.
WebONIS is an ActiveX component loaded into an HTML page that provides the same functionality as the ONIS Remote application. It can only be used with the Microsoft Internet Explorer browser. The browser must connect to a web server to load the ActiveX component, and the latter then connects directly to an ONIS Site Server or an ONIS Organization Server.
ONIS Site Server is a server application that supports the storage and retrieval of a wide range of DICOM Storage objects. It also supports the storage and retrieval of graphical annotations and reports when connected to ONIS client applications (ONIS Viewer, ONIS Remote, and WebONIS).
ONIS Organization Server is a server that provides a single access point to multiple Site Servers.
### V. Indications for Use
ONIS-PACS is a software device intended to be used by healthcare personnel and intended for viewing, reviewing, performing measurements/quantifications and reporting of medical images and data acquired from DICOM compliant medical imaging systems. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy images and digitized film images must not be used for primary diagnosis or interpretation.
#### VI. Comparison to Predicate Devices
| Device Name | New Device | Predicate Device | Comparison |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | DigitalCore | IMAGE Information<br>Systems Ltd | - |
| 510(k) Number | To be assigned | K062488 | - |
| Regulatory Class | Class II | Class II | EQUIVALENT |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | EQUIVALENT |
| Product Code | LLZ | LLZ | EQUIVALENT |
| Device Name | New Device | Predicate Device | Comparison |
| Common Name | Picture Archiving<br>Communications System | Picture Archiving<br>Communications System | EQUIVALENT |
| Clearance Date | To be assigned | Sep 19 2006 | - |
| Indications for Use | ONIS-PACS is a<br>software device intended<br>to be used by healthcare<br>personnel and intended<br>for viewing, reviewing,<br>performing<br>measurements/quantific<br>ations and reporting of<br>medical images and data<br>acquired from DICOM<br>compliant medical<br>imaging systems.<br><br>Images and data can be<br>stored, communicated,<br>processed and displayed<br>within the system or<br>across computer<br>networks at distributed<br>locations.<br><br>Lossy images and<br>digitized film images<br>must not be used for<br>primary diagnosis or<br>interpretation. | iQ-System PACS is a<br>software device intended<br>for viewing of images<br>acquired from CT, MR,<br>CR, DR, US and other<br>DICOM compliant<br>medical imaging systems<br>when installed on<br>suitable commercial<br>standard hardware.<br><br>Images and data can be<br>captured, stored,<br>communicated,<br>processed, and<br>displayed within the<br>system and or across<br>computer networks at<br>distributed locations.<br><br>Lossy compressed<br>mammographic images<br>and digitized film screen<br>images must not be<br>reviewed for primary<br>diagnosis or image<br>interpretation. It is the<br>User's responsibility to<br>ensure monitor quality,<br>ambient light conditions,<br>and image compression<br>ratios are consistent with<br>clinical application. | EQUIVALENT - Both<br>devices are software<br>only devices intended to<br>allow<br>clinicians to view and<br>process DICOM images,<br>and communicate results<br>over a network.<br><br>Both devices provide<br>support for<br>mammographic images,<br>and both devices contra-<br>indicate that lossy<br>images should not be<br>used for primary<br>diagnosis nor<br>interpretation. |
| Prescription device | YES | YES | EQUIVALENT |
| Environment of Use | Prescription device,<br>intended for use in a<br>healthcare setting | Prescription device,<br>intended for use in a<br>healthcare setting | EQUIVALENT |
| Technological characteristics | | | |
| System Architecture | Software only device,<br>comprises workstation<br>and web software, as<br>well as server software. | Software only device,<br>comprises workstation<br>and web software, as<br>well as server software. | EQUIVALENT |
| Operating System | Windows 7, 8, 10<br>Microsoft Windows 2008<br>Server Standard<br>Microsoft Windows 2012<br>Server Standard | Windows 2000 / XP | EQUIVALENT - The<br>subject device works on<br>newer versions of the<br>same Microsoft Windows<br>Operating<br>System |
| Device Name | New Device | Predicate Device | Comparison |
| Data Access and<br>Communication | Accesses DICOM data<br>on removable media<br>(USB drive, optical disk),<br>local and network drives,<br>and also accesses<br>DICOM data from<br>DICOM server<br>applications using<br>DICOM communication<br>protocols | Accesses DICOM data<br>from portable media as<br>well as through server<br>applications using<br>DICOM communication<br>protocols | EQUIVALENT - Both<br>device access data using<br>Standard DICOM<br>protocols |
| Image file import | YES | YES | EQUIVALENT |
| Patient CD/DVD import | YES | YES | EQUIVALENT |
| Export to a portable<br>memory | YES, Export to memory<br>stick, optical media | YES , Export to memory<br>stick | EQUIVALENT |
| Image export to image<br>file or AVI video file | YES | YES | EQUIVALENT |
| Mammography Display | YES | YES | EQUIVALENT |
| Image Visualization | Supports 2D,<br>Mammography, Multi<br>Planar Reconstruction<br>(Axial, Sagittal, Coronal,<br>Oblique, Double<br>Oblique), Maximum<br>Intensity Projection and<br>Volume Rendered views | Supports mammography<br>display, Orthogonal and<br>Oblique Multi Planar<br>Reconstruction,<br>Maximum Intensity<br>Projection, Surface<br>Shaded Display and<br>Volume Rendered views | EQUIVALENT - Both<br>devices support<br>standard image and<br>volume rendering views |
| Support for Hanging<br>Protocols | YES | YES | EQUIVALENT |
| Annotation Tools | Supports text, arrow,<br>angle, line and other<br>user drawn region of<br>interest shapes | Support for standard<br>annotation tools | EQUIVALENT - Both<br>devices provide standard<br>annotation tools |
| Measurement Tools | Supports distance and<br>angle measurements,<br>and pixel statistics for<br>user drawn regions of<br>interest | Supports advanced<br>measurement tools<br>including Region of<br>Interest computations | EQUIVALENT - Both<br>devices provide standard<br>measurement tools |
| Image Manipulation<br>Tools (Window<br>width/level, Pan, Zoom,<br>etc.) | Provides tools for<br>window width/level, pan,<br>zoom, rotate, color LUT,<br>opacity table, image<br>filters | Provides tools for<br>window width/level, pan,<br>zoom, rotate, image<br>filters, volume cropping<br>and clipping, and other<br>volume rendering color,<br>transparency and light<br>setting options | EQUIVALENT - Both<br>devices provide standard<br>image manipulation<br>functionality |
| Data Output | Supports export of<br>images to various image<br>formats, and printing of<br>images and reports. | Supports export of<br>secondary capture<br>images to the local<br>imagebox, filesystem or<br>PACS | EQUIVALENT - Both<br>devices support export of<br>images to<br>multiple<br>destinations |
| Device Name | New Device | Predicate Device | Comparison |
| Review report | YES (HTML format) | YES (SR formats) | EQUIVALENT - Both<br>devices support report<br>generation functionality,<br>though the specific<br>report format may be<br>different |
| Windows print | YES | YES | EQUIVALENT |
| JPEG lossy/lossless<br>compression | YES | YES | EQUIVALENT |
| Edit patient<br>demographics | YES | YES | EQUIVALENT - Both<br>devices allow for edit of<br>patient demographics.<br>ONIS-PACS implements<br>technical controls to<br>ensure that<br>only authorized users<br>can make these<br>modifications. |
| DICOM Query/ Retrieve,<br>DICOM Print | YES | YES | EQUIVALENT |
The table below summarizes the comparison between the ONIS-PACS and the iQ-System.
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Both ONIS-PACS and the predicate device have similar technological characteristics, and any differences are only related to the specific software implementation.
#### VII. Performance Testing
Like the predicate device. ONIS-PACS is a software only medical device. The ONIS- PACS was determined to present a moderate level of concern per the FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005". Software Validation and Verification testing was performed on the ONIS-PACS device to demonstrate safety and effectiveness.
#### VIII. Conclusion
The subject device has the same intended use as the predicate device, and differences in technological characteristics do not raise different questions of safety and effectiveness. On this basis, ONIS-PACS is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.