The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients. The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment. The monitor is also intended for use during patient transport inside the hospital environment. The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and which are mainly free from cardiac arrhythmia. The PPV measurement is validated for use with adult patients only. The measurement extension adds physiological measurements (Microstream CO2 and optionally temperature, invasive blood pressure and cardiac output) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals. The measurement extension can only function when it is connected to a dedicated host device.
Device Story
IntelliVue MX100/X3 monitors and 867041 Microstream Extension acquire physiological signals (ECG, respiration, SpO2, NBP, invasive pressure, temperature, CO2, cardiac output) via sensors applied to patients. Devices display waves/trends, generate alarms, and support data management. Used by trained healthcare professionals in hospitals or during transport. Modifications add Pulse Pressure Variation (PPV) calculation (derived from beat-to-beat arterial pressure) and cardiac output measurement (via added hardware board). Output allows clinicians to assess patient status and make treatment decisions. Benefits include continuous monitoring of vital signs and derived parameters to support clinical intervention.
Clinical Evidence
Bench testing only. Verification and validation activities included software life cycle processes (IEC 62304), electrical/mechanical safety (IEC 60601-1, 60601-2-34), EMC (IEC 60601-1-2), and environmental/mechanical stress testing. Performance testing verified PPV calculation accuracy, artifact detection, and cardiac output accuracy/repeatability against predicate device specifications. No clinical data required.
Indicated for monitoring physiological parameters in adult, pediatric, and neonatal patients in hospital environments, including transport. ECG for diagnostic rhythm/morphology; ST segment for adults only; IPI for adults/pediatrics (1-12 years); PPV for sedated, mechanically ventilated adults free from arrhythmia; Masimo rainbow SET for SpO2, pulse rate, SpCO, SpMet, SpHb, and RRac in motion/no-motion and varying perfusion states.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3 (K171801)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 4, 2018
Philips Medizin Systeme Boeblingen GmbH Stefan Breuer Senior Regulatory Affairs Engineer Hewlett-Packard-Str. 2 Boeblingen, 71034 De
## Re: K181314
Trade/Device Name: IntelliVue Patient Monitor MX100, IntelliVue Multi-Measurement Module X3, IntelliVue Microstream Extension 867041 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DSJ, DSK, DXN, DXG, KRB, DRT, DRQ, DPS, MLC, DRJ, DQA, DSB, DSA, MSX, DRG, CCK, BZO, FLL Dated: August 1, 2018 Received: August 6, 2018
Dear Stefan Breuer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Arielle Drummond -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181314
# Device Name
IntelliVue Patient Monitor MX100, Multi-Measurement Module X3, Microstream Extension 867041
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020 See PRA Statement below.
Indications for Use (Describe)
The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment.
The monitor is also intended for use during patient transport inside the hospital environment. The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and which are mainly free from cardiac arrhythmia. The PPV measurement is validated for use with adult patients only.
| Type of Use (Select one or both, as applicable) | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> <span>□ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 2 | BSC Publishing Services (301) 443-6740 | EE |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
| FORM FDA 3881 (7/17) | Page 1 of 2 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
{3}------------------------------------------------
#### Indications for Use (continued)
#### Indications for Use Statement for the Microstream Extension 867041
The measurement extension adds physiological measurements (Microstream CO2 and optionally temperature, invasive blood pressure and cardiac output) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals. The measurement extension can only function when it is connected to a dedicated host device.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
| FORM FDA 3881 (7/17) | Page 2 of 2 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
{4}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
## 5.1 Submitter
The submitter of this Premarket Notification is: Stefan Breuer Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 71034 Boeblingen Germany Telephone: +49 7031 463-2321 Fax: +49 7031 463-2442 Email: stefan.breuer@philips.com This summary was prepared on May 15, 2018.
# 5.2 Name and Classification of the Devices
Trade names: IntelliVue Patient Monitor MX100, Multi-Measurement Module X3, Microstream Extension 867041
Common name: Multiparameter Patient Monitor
Trade name: IntelliVue Microstream Extension 867041
Common name: Multifunction Patient Monitor Module.
Classification:
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|--------------------------------------------------------------------------|
| Cardiovascular<br>Devices | §870.1025, II | MHX | Monitor, Physiological, Patient (with<br>arrhythmia detection or alarms) |
| Cardiovascular<br>Devices | §870.1025, II | DSI | Detector and alarm, arrhythmia |
| Cardiovascular<br>Devices | §870.1025, II | MLD | Monitor, ST Segment with Alarm |
| Cardiovascular<br>Devices | §870.1100, II | DSJ | Alarm, Blood Pressure |
| Cardiovascular<br>Devices | §870.1110, II | DSK | Computer, Blood Pressure |
| Cardiovascular<br>Devices | §870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive |
| Cardiovascular<br>Devices | §870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function |
| Cardiovascular<br>Devices | §870.1915, II | KRB | Probe, Thermodilution |
| Cardiovascular<br>Devices | §870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) |
| Cardiovascular<br>Devices | §870.2060, II | DRQ | Amplifier and Signal Conditioner,<br>Transducer Signal |
| Cardiovascular<br>Devices | §870.2340, II | DPS | Electrocardiograph |
| Cardiovascular<br>Devices | §870.2340, II | MLC | Monitor, ST Segment |
| Cardiovascular<br>Devices | §870.2600, I | DRJ | System, Signal Isolation |
| Cardiovascular<br>Devices | §870.2700, II | DQA | Oximeter |
| Cardiovascular<br>Devices | §870.2770, II | DSB | Plethysmograph, Impedance |
| Cardiovascular<br>Devices | §870.2900, I | DSA | Cable, Transducer and Electrode,<br>incl. Patient Connector |
{5}------------------------------------------------
| Device Panel | Classification | ProCode | Description |
|-------------------------------------------------|----------------|---------|---------------------------------------------------------------------|
| | - | MSX | System, Network and<br>Communication, Physiological<br>Monitors |
| | §870.2910, II | DRG | Transmitters and Receivers,<br>Physiological Signal, Radiofrequency |
| Anesthesiology<br>Devices | §868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase |
| | §868.2375, II | BZQ | Monitor, Breathing Frequency |
| General Hospital and<br>Personal Use<br>Devices | §880.2910, II | FLL | Thermometer, Electronic, Clinical |
# 5.3 Substantial Equivalence
The modified IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3 with Software Rev. M.03 with the derived measurement PPV are substantially equivalent to the legally marketed Intelli Vue Patient Monitor MX100 and Multi-Measurement Module X3, marketed pursuant to K171801, and to IntelliVue Patient Monitor MP5, marketed pursuant to K161531.
The modified IntelliVue Microstream Extension 867041 with Cardiac Output is substantially equivalent to the legally marketed IntelliVue Microstream Extension 867041 and the legally marketed IntelliVue Capnography Extension 867040, both marketed pursuant to K172904.
# 5.4 Description of the Device
# IntelliVue Patient Monitor MX100 and IntelliVue Multi-Measurement Module X3
The IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management.
The devices offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They can be located in the patient vicinity at the bedside or can be used mobile, during patient transport inside hospitals.
The measurement sensors are applied at diverse body locations, depending on the actual physiological parameters monitored.
The MX100 and X3 provide multiple non-invasive and invasive measurements: ECG (including arrhythmia and ST), respiration, SpO2, NBP, dual invasive pressure, temperature, and CO2.
The modification that is subject of this Premarket Notification adds the derived measurement PPV (pulse pressure variation), which is calculated from beat-to-beat arterial pressure values. This is the same measurement as in the predicate device IntelliVue Patient Monitor MP5.
{6}------------------------------------------------
## IntelliVue Microstream Extension 867041
The Microstream Extension 867041 adds to the dedicated host device (IntelliVue Patient Monitor or Multi-Measurement Module) physiological measurements: Microstream CO2, and optionally dual invasive pressure and temperature.
The measurement extension is not a stand-alone device; it can only function when it is connected to a dedicated host device.
The modified Microstream Extension now provides the additional optional measurement cardiac output to the host devices. The cardiac output measurement is the same as in the predicate device IntelliVue Capnography Extension 867040.
# 5.5 Intended Use
# IntelliVue Patient Monitor MX100 and IntelliVue Multi-Measurement Module X3
This is the same Intended Use as that for the legally marketed predicate devices IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3.
The Indications for Use Statement has been modified to add the PPV Indication from the legally marketed predicate device IntelliVue Patient Monitor MP5.
# Indications for Use statement for the IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3:
The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment.
The monitor is also intended for use during patient transport inside the hospital environment.
The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate,
carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow
{7}------------------------------------------------
SET measurement is indicated for use during both no motion and motion conditions. and for patients who are well or poorly perfused.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and which are mainly free from cardiac arrhythmia.
The PPV measurement is validated for use with adult patients only.
# IntelliVue Microstream Extension 867041
This is the same Intended Use as that for the legally marketed predicate device IntelliVue Microstream Extension 867041.
The Indications for Use Statement has been modified to add the cardiac output measurement from the legally marketed predicate device IntelliVue Capnography Extension 867040.
## Indications for Use statement for the Microstream Extension 867041:
The measurement extension adds physiological measurements (Microstream CO2 and optionally temperature, invasive blood pressure and cardiac output) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals.
The measurement extension can only function when it is connected to a dedicated host device.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
# 5.6 Technological Characteristics
# IntelliVue Patient Monitor MX100 and IntelliVue Multi-Measurement Module X3
The modifications to the IntelliVue Patient Monitor MX100 and the Multi-Measurement Module X3 are limited to minor software changes that do not affect technological characteristics of the devices. The common software of the devices has slightly been modified in order to provide the derived measurement Pulse Pressure Variation (PPV).
The hardware of the subject devices remains unchanged; all accessories are re-used unchanged from the predicate devices.
# IntelliVue Microstream Extension 867041
The modifications to the IntelliVue Microstream Extension 867041 are limited to the addition of the Cardiac Output measurement board. This board is re-used, without any hardware or software change, from the legally marketed predicate device IntelliVue Capnography Extension 867040. All accessories are re-used from the predicate device, without any change.
{8}------------------------------------------------
## All Subject Devices
Design, materials, energy source, portability, user interface, radio technology, measurement principle, and all physical, environmental, and performance specifications of the devices remain all unchanged. The modified IntelliVue Patient Monitor MX100, Multi-Measurement Module X3 and Microstream Extension 867041 have the same relevant technological characteristics as their predicate devices with regard to operating principles, mechanism of action, energy sources, portability, user interface, robustness, dimensions, weight and environmental specifications.
# 5.7 Summarv of V&V Activities
# IntelliVue Patient Monitor MX100 and IntelliVue Multi-Measurement Module X3
The modified devices have been subject to a series of V&V activities:
- Testing according to the recognized consensus standards: ●
- o AAMI ANSI IEC 62304:2006 (Ed. 1) (Software life cycle processes)
- o IEC 60601-2-34: 2011 (Ed. 3) (Invasive blood pressure monitoring equipment)
All applicable requirements have been met.
- . Testing as required by the Hazard Analysis. All specified pass/fail criteria have been met. The test results have confirmed the effectiveness of the implemented design risk mitigation measures.
- . Unit level tests, functional system level tests, and regression system level tests. The modified devices perform according to the specified criteria:
- o Functional unit level tests for Derived Measurement PPV
- o Functional system level tests for Derived Measurement PPV
- Regression Tests for unchanged host monitors o
All specified test requirements have been met. The test results demonstrate that modified and previously available device functions work correctly according to the specifications and labeling claims.
- Performance testing of derived measurement PPV ●
- o Verification of PPV calculation by comparison to predicate device
- o Verification of PPV artifact detection mechanism
- o Verification of calculation of the average PPV
All applicable requirements have been met. The modified devices perform according to the specified criteria that are the same as those for the predicate devices.
## IntelliVue Microstream Extension 867041
The modified devices have been subject to a series of V&V activities:
- Testing according to the recognized consensus standards:
- o AAMI ANSI IEC 62304:2006 (Ed. 1) (Software life cycle processes)
- o AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1) (Basic safety and essential performance)
{9}------------------------------------------------
## o IEC 60601-1-2: 2007 (Ed. 3) (Electromagnetic Compatibility)
All applicable requirements have been met.
- Testing as required by the Hazard Analysis. ● All specified pass/fail criteria have been met. The test results have confirmed the effectiveness of the implemented design risk mitigation measures.
- . Additional environmental testing (temperature, humidity) and mechanical testing (mechanical classes 7M1, 7M2, 7M3, and disinfectant resistance) All specified test requirements have been met. The tests have confirmed that the modified devices work safely and according to their specifications and indicated claims during tests simulating general hospital conditions, handling and transport in hospital environments, disinfection, and storage.
- Functional system level tests, and regression system level tests. ● The modified devices perform according to the specified criteria:
- o Functional system level tests for Cardiac Output (C.O.) measurement
- Regression Tests for unchanged host monitors and unchanged other O measurements in Microstream Extension 867041
All specified test requirements have been met. The test results demonstrate that modified and previously available device functions work correctly according to the specifications and labeling claims.
- Performance testing of Cardiac Output (C.O.)
- Verification of accuracy and repeatability for right heart measurement o
- Verification of accuracy and repeatability for transpulmonary measurement o
- Verification of the visibility of Thermodilution Curve in Cardiac Output O Application Window
o Verification of temperature drift limits and small signal warnings. All specified test requirements have been met. The modified devices perform according to the specified criteria that are the same as those for the predicate devices:
# 5.8 Conclusion
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified devices with respect to the predicates.
Testing comprised electrical and mechanical safety tests, EMC tests, environmental tests, performance tests, functional and regression tests.
Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence.
The results demonstrate that the modified devices meet all defined reliability requirements and performance claims.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.