← Product Code [KRD](/productcode/KRD) · K181300

# Embosphere Microspheres (K181300)

_Biosphere Medical, S.A. · KRD · Jul 16, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181300

## Device Facts

- **Applicant:** Biosphere Medical, S.A.
- **Product Code:** [KRD](/productcode/KRD.md)
- **Decision Date:** Jul 16, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K181300 · Jul 16, 2018 | Embosphere Microspheres | Biosphere Medical, S.A. | Published clinical literature (retrospective trials) | A literature review of 40 publications (including 32 retrospective trials) was conducted to evaluate the safety and effectiveness of the device for controlling bleeding/hemorrhaging in the peripheral vasculature. | Literature review; Retrospective trials; Peripheral vasculature; Hemorrhage control |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Literature review of Embosphere Microspheres use; Retrospective trials (32) and prospective trials (8); Study Period: 2000-2018 | 662 patients (at least), including adults and pediatric patients (ages 1-17); Sample Size: 662 | Not applicable for this study | Safety and effectiveness in controlling bleeding/hemorrhaging in the peripheral vasculature |

## Indications for Use

Embosphere Microspheres are indicated for use in the embolization of: - Hypervascular tumors, including symptomatic uterine fibroids - Prostatic arteries for symptomatic benign prostatic hyperplasia (BPH) - Arteriovenous malformations - Blood vessels to occlude blood flow to control bleeding/hemorrhaging in the peripheral vasculature

## Device Story

Embosphere Microspheres are permanent, biocompatible, hydrophilic, compressible spheres composed of acrylic polymer and porcine-derived gelatin. Used by interventional radiologists in clinical settings for targeted arterial embolization. Procedure involves arterial access via guidewire and microcatheter under fluoroscopic guidance. Microspheres are mixed with non-ionic contrast agent and delivered to occlude target vessels, stopping blood flow to hypervascular tumors, BPH-affected prostate tissue, arteriovenous malformations, or bleeding peripheral vessels. Output is physical vessel occlusion; clinical benefit is controlled, targeted hemostasis or tissue devascularization.

## Clinical Evidence

Literature review of 40 publications (2000-2018) including 8 prospective and 32 retrospective trials. Total of 662 patients (adults and pediatric, ages 1-17). Data demonstrates effectiveness in controlling peripheral bleeding/hemorrhaging with low complication rates. Adverse events consistent with known risks of catheterization and embolization (e.g., post-embolization syndrome, non-targeted embolization).

## Technological Characteristics

Materials: Acrylic polymer and porcine-derived gelatin. Form: Compressible, hydrophilic, non-resorbable spheres. Sizes: 40-120 to 900-1200 microns. Sterilization: Steam. Packaging: Glass vial or plastic syringe in 0.9% saline. Biocompatibility: ISO 10993-1:2009 compliant.

## Regulatory Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

## Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

## Predicate Devices

- Embosphere Microspheres ([K021397](/device/K021397.md))

## Reference Devices

- Embosphere Microspheres ([K991549](/device/K991549.md))
- Embosphere Microspheres ([DEN160040](/device/DEN160040.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

July 16, 2018

BioSphere Medical, S.A. Rosène Amossé Senior Regulatory Affairs Specialist Parc des Nations - Paris Nord 2 383 Rue de la Belle Etoile 95700 Roissy en France France

Re: K181300

Trade/Device Name: Embosphere Microspheres Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD. NAJ. NOY. HCG Dated: May 15, 2018 Received: May 17, 2018

Dear Rosène Amossé:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K181300

Device Name Embosphere Microspheres

Indications for Use (Describe)

Embosphere Microspheres are indicated for use in the embolization of:

- · Hypervascular tumors, including symptomatic uterine fibroids
- · Prostatic arteries for symptomatic Benign Prostatic Hyperplasia (BPH)
- · Arteriovenous malformations
- · Blood vessels to occlude blood flow to control bleeding/hemorrhaging in the peripheral vasculature

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Change in Indication for use

## K181300 510(k) Summary

## July 13, 2018

| General<br>Provisions | Submitter Name:                                                                                                                            | Biosphere Medical, S.A.                                                                                                                                                                                   |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                       | Address:                                                                                                                                   | Parc des Nations - Paris Nord 2<br>383, rue de la Belle Etoile<br>95700 Roissy-en-France, France                                                                                                          |
|                       | Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number:                                                                | +33 (0)1 48 17 25 29<br>+33 (0)1 49 38 02 68<br>Alix Fonlladosa<br>9615728                                                                                                                                |
|                       | Correspondent Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number:                             | Biosphere Medical, S.A.<br>Parc des Nations - Paris Nord 2<br>383, rue de la Belle Etoile<br>95700 Roissy-en-France, France<br>+33 (0)1 48 17 25 29<br>+33 (0)1 49 38 02 68<br>Alix Fonlladosa<br>9615728 |
| Subject<br>Device     | Trade Name:<br>Common/Usual Name:<br>Classification Name:                                                                                  | Embosphere® Microspheres<br>Embolization device<br>21 CFR 870.3300 - Vascular Embolization Device                                                                                                         |
| Predicate<br>Device   | Predicate Device:<br>Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer:                                       | Embosphere® Microspheres<br>21 CFR 870.3300 - Vascular Embolization Device<br>K021397<br>Biosphere Medical, S.A.                                                                                          |
|                       | Reference devices:<br>Trade Name:<br>Classification Name:                                                                                  | Embosphere® Microspheres<br>21 CFR 882.5950 - Neurovascular embolization device<br>21 CFR 876.5550 - Prostatic artery embolization device                                                                 |
|                       | Premarket Notification:<br>Manufacturer:                                                                                                   | K991549, DEN160040<br>Biosphere Medical, S.A.                                                                                                                                                             |
| Classification        | Class II<br>21 CFR § 870.3300<br>FDA Product Code: KRD<br>FDA Secondary Product Codes: HCG, NAJ, NOY<br>Division of Cardiovascular Devices |                                                                                                                                                                                                           |

{4}------------------------------------------------

| Intended Use | Embosphere Microspheres are indicated for use in the embolization of:<br>• Hypervascular tumors, including symptomatic uterine fibroids<br>• Prostatic arteries for symptomatic benign prostatic hyperplasia (BPH)<br>• Arteriovenous malformations<br>• <b>Blood vessels to occlude blood flow to control bleeding/hemorrhaging in</b><br><b>the peripheral vasculature</b> (subject of this Traditional 510(k)) |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Embosphere Microspheres are small, compressible, hydrophilic, biocompatible spheres made of acrylic polymer and porcine-derived gelatin. The microspheres are packaged in 0.9% saline and are provided sterile and non-pyrogenic in a vial or in a syringe.

The product is provided in seven size ranges to allow physicians to choose the calibration necessary for the vessel being embolized. The size ranges available are:

- 50-100 microns ●
- 40-120 microns ●
- 100-300 microns
- 300-500 microns ●
- 500-700 microns
- 700-900 microns ●
- 900-1200 microns .

#### Device Description

The principles of operation for the subject device Embosphere Microspheres are the same as the predicate device Embosphere Microspheres (K021397) and reference devices Embosphere Microspheres (K991549, DEN160040). Embosphere Microspheres are permanent implantable devices and are designed for controlled, targeted embolizations for Embosphere Microsphere Microspheres involve arterial embolization; embolization of uterine fibroids, arteriovenous malformations, hypervascular tumors and benign prostatic hyperplasia involve an embolization of the arteries supplying those areas. The procedure of arterial embolization is similar for all arteries. Appropriately sized microspheres for target vessel occlusion are chosen by the trained interventional radiologist. The delivery procedure involves arterial access through an artery, using a quidewire and microcatheter under fluoroscopic guidance. Once the catheter tip is placed in the artery(ies) supplying the targeted tissue Embosphere Microspheres mixed with a non-ionic contrast agent are delivered in a controlled manner under visualization to occlude the feeding vessel(s) to stop blood flow to the targeted area. The device is intended for single use.

{5}------------------------------------------------

The technological characteristics of the subject device Embosphere Microspheres are identical to the predicate Embosphere Microspheres (K021397) and the reference devices Embosphere Microspheres (K991549, DEN160040). Only the indications for use change between the subject device and the legally marketed predicate device Embosphere Microspheres (K021397) and reference devices Embosphere Microspheres (K991549, DEN160040). The indications for use statement of the subject Embosphere Microspheres includes a specific indication by identifying "control bleeding / hemorrhaging in the peripheral vasculature" as being the result of embolization since embolization stops arterial blood flow, as established by 21 CFR § 870.3300.

|                            | Technological Characteristics Comparison Table |                                                                                                                                                |                                                                                                                                                |         |
|----------------------------|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------|
|                            | Attribute                                      | Predicate Device<br>Embosphere<br>Microspheres<br>(K021397) and<br>Reference devices<br>Embosphere<br>Microspheres<br>(K991549,<br>DEN160040)  | Subject Device<br>Embosphere<br>Microspheres                                                                                                   | Comment |
| Comparison to<br>Predicate | Shelf Life<br>(single use)                     | Three years<br>(36 months)                                                                                                                     | Three years<br>(36 months)                                                                                                                     | Same    |
|                            | Material<br>(spheres)                          | Acrylic polymer and<br>porcine-derived gelatin                                                                                                 | Acrylic polymer and<br>porcine-derived gelatin                                                                                                 | Same    |
|                            | Physical<br>Characteristics                    | Biocompatible,<br>hydrophilic,<br>compressible, non-<br>resorbable                                                                             | Biocompatible,<br>hydrophilic,<br>compressible, non-<br>resorbable                                                                             | Same    |
|                            | Microspheres<br>Size                           | • 50-100 microns<br>• 40-120 microns<br>• 100-300 microns<br>• 300-500 microns<br>• 500-700 microns<br>• 700-900 microns<br>• 900-1200 microns | • 50-100 microns<br>• 40-120 microns<br>• 100-300 microns<br>• 300-500 microns<br>• 500-700 microns<br>• 700-900 microns<br>• 900-1200 microns | Same    |
|                            | Sterilization                                  | Steam sterilized                                                                                                                               | Steam sterilized                                                                                                                               | Same    |
|                            | Pyrogenicity                                   | Non-pyrogenic                                                                                                                                  | Non-pyrogenic                                                                                                                                  | Same    |
|                            | Performance                                    | Designed for controlled,<br>targeted embolization at<br>the desired level of<br>vessel occlusion                                               | Designed for controlled,<br>targeted embolization at<br>the desired level of<br>vessel occlusion                                               | Same    |
|                            | Principle of<br>Operation                      | The microspheres are<br>administered with<br>contrast medium into the<br>patient's artery via a<br>catheter                                    | The microspheres are<br>administered with<br>contrast medium into the<br>patient's artery via a<br>catheter                                    | Same    |
|                            | Volume of<br>microspheres per<br>container     | 1 ml or 2 ml of<br>microspheres<br>in 0.9% saline solution                                                                                     | 1 ml or 2 ml of<br>microspheres<br>in 0.9% saline solution                                                                                     | Same    |
|                            | Packaging                                      | 8-mL glass vial, or 20-<br>mL plastic syringe<br>Microbial barrier: Blister<br>tray sealed by a<br>Tyvek® peel-away lid                        | 8-mL glass vial, or 20-<br>mL plastic syringe<br>Microbial barrier: Blister<br>tray sealed by a<br>Tyvek® peel-away lid                        | Same    |

{6}------------------------------------------------

No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, vascular and neurovascular embolization devices are subject to the special controls specified in "Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices", issued on December 29, 2004. In addition, the subject device follows the FDA Draft Guidance on "Medical Devices Containing Materials Derived from Animal Sources (Except for In Vitro Diagnostic Devices)", issued on January 23, 2014.

Non-clinical performance testing conducted on the predicate device includes:

- · Chemical analysis
- · Size range
- Catheter compatibility
- Packaging performance
- · Shelf Life
- · Sterilitv
- · Biocompatibility

Biocompatibility evaluation was performed in accordance with ISO 10993-1:2009 for permanent implants in contact with blood for the predicate device Embosphere Microspheres (K021397) and the reference devices Embosphere Microspheres (K991549, DEN160040), and is being adopted for this subject device based upon the fact that all devices are made from the same materials using the same manufacturing and sterilization processes.

### Safety & Performance Tests

No new testing was performed since predicate device Embosphere Microspheres (K021397) / reference devices Embosphere Microspheres (K991549. DEN160040), and subject device Embosphere Microspheres have identical technological characteristics, manufacturing, processing and sterilization.

Clinical data from the literature were reviewed on the use of Embosphere Microspheres for embolization of blood vessels to occlude blood flow to control bleeding / hemorrhaging in the peripheral vasculature. The retrieved clinical data consisted of 40 publications from 2000 to 2018, including eight prospective trials and thirty-two retrospective trials. The total number of patients treated with Embosphere Microspheres was 662 (at least). The majority of patients were adults of both genders. Some publications listed age ranges below adult (from 1 to 17 years).

The extensive clinical data provide relevant information about the safety and effectiveness of the subject Embosphere Microspheres in the proposed indication for use. Embolization of arteries with Embosphere was found to be an effective method to control bleeding / hemorrhaging in the peripheral vasculature, with a low complication rate.

Adverse events that were found related to the procedure are mostly complications related to catheterization, post-embolization syndrome and non- targeted embolization. The raised events found are addressed in the predicate device Embosphere Microspheres (K021397) and the reference devices Embosphere Microspheres (K991549, DEN160040) IFU as warnings or potential complications.

The proposed indication for use does not raise a safety or effectiveness issue that was not raised by the predicate device, or have the potential to significantly increase a safety or effectiveness concern raised by the predicate device. The

{7}------------------------------------------------

function of both the subject device and predicate device (K021397) is to occlude blood vessel(s) in order to stop blood flow to a targeted area, which is the clinical endpoint for both devices. The safety and effectiveness of the device are related to size, shape, compressibility and biocompatibility for both general and specific indications. These characteristics do not change from the predicate to the subject Embosphere Microspheres.

Based on the same intended use (other than the requested change that is the objective of this submission), design, fundamental scientific technological Summary of characteristics, fundamental operational characteristics and safety and Substantial performance testing, the BioSphere Medical subject Embosphere Microspheres device meets the requirements that are considered essential for its intended use Equivalence and is substantially equivalent to the predicate device Embosphere Microspheres (K021397).

---

**Source:** [https://fda.innolitics.com/device/K181300](https://fda.innolitics.com/device/K181300)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
