Curiteva Cervical Interbody Fusion System

K181261 · Curiteva, LLC · ODP · Jul 9, 2018 · Orthopedic

Device Facts

Record IDK181261
Device NameCuriteva Cervical Interbody Fusion System
ApplicantCuriteva, LLC
Product CodeODP · Orthopedic
Decision DateJul 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Curiteva Cervical Interbody Fusion is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2 - T1 inclusive). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. Implants are intended to be used with autograft bone (comprised of cancellous and/or corticocancellous bone graft) and supplemental spinal fixation systems that have been cleared for use in the cervical spine. Patients should receive at least six (6) weeks of non-operative treatment with the device.

Device Story

Cervical interbody fusion system; box-shaped implants with central chamber for bone graft; superior/inferior surface teeth for stability. Used in spinal surgery to facilitate fusion in DDD patients. Implanted by surgeons; requires supplemental fixation. Benefits include stabilization of disc space and promotion of fusion.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F2077 (static/dynamic compression and torsion), ASTM F2267 (subsidence), and ASTM F1877 (wear debris). Results demonstrate performance equivalent to predicate devices.

Technological Characteristics

Materials: PEEK (ASTM F2026) with Tantalum markers (ASTM F560) or Titanium alloy (Ti-6Al-4V, ASTM F136). PEEK implants may include CpTi plasma coating (ASTM F1580). Form factor: box-shaped with central chamber and surface teeth. Mechanical fixation via supplemental spinal systems.

Indications for Use

Indicated for skeletally mature patients with cervical degenerative disc disease (DDD) at one level (C2-T1) with radicular symptoms; requires 6 weeks non-operative treatment; used with autograft/allograft and supplemental fixation.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Curiteva, LLC Eric Linder Chief Operating Officer 25127 Will McComb Drive Suite 100 Tanner, Alabama 35671 Re: K181261 Trade/Device Name: Curiteva Cervical Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: May 9, 2018 Received: May 11, 2018 Dear Eric Linder: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); July 9, 2018 {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/6 description: The image shows the name "Brent Showalter -S" in a large, sans-serif font. The text is black and stands out against a white background. The letters are evenly spaced and easy to read. The overall impression is clean and professional. for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K181261 Device Name Curiteva Cervical Interbody Fusion System #### Indications for Use (Describe) The Curiteva Cervical Interbody Fusion is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2 - T1 inclusive). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. Implants are intended to be used with autograft bone (comprised of cancellous and/or corticocancellous bone graft) and supplemental spinal fixation systems that have been cleared for use in the cervical spine. Patients should receive at least six (6) weeks of non-operative treatment with the device. Type of Use (Select one or both, as applicable) | <span style="font-size:12px"><b></b></span> <span style="font-size:12px"><b></b></span> <span style="font-size:12px"><b></b></span> <span style="font-size:12px"><b></b></span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ### A. Submitter Information | Submitter: | Curiteva, LLC<br>25127 Will McComb Drive, Suite 100<br>Tanner, AL 35671<br>Phone: (256) 213-1057<br>Fax: (256) 213-1058 | |-----------------|-------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Eric Linder<br>regulatory@curiteva.com | | Date Prepared: | May 9, 2018 | #### B. Device Information | Trade Name: | Curiteva Cervical Interbody Fusion System | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Intervertebral Body Fusion Device | | Classification Name: | Intervertebral Fusion Device With Bone Graft, Cervical | | Device Classification: | Class II (per 21 CFR 888.3080) | | Product Code: | ODP | | Classification Panel: | Division of Orthopedic Devices | | Predicate Device(s): | Primary: Globus Medical PATRIOT Cervical Spacers, (including<br>COLONIAL and COLONIAL TPS) -- K143578<br>Additional: DeGen Medical Latitude-C Cervical Interbody<br>Spacer System -- K151496<br>Additional: Titan Spine Endoskeleton TC -- K100889 | #### C. Device Description The Curiteva Cervical Interbody Fusion System implants are available in a variety of different footprints, styles and sizes to accommodate the individual pathology and anatomical conditions of the patient. The implants are generally box-shaped with an open central chamber to permit packing with bone graft to facilitate fusion. The superior and inferior surfaces of the construct have a pattern of teeth to provide increased stability and to help prevent movement of the device. The Curiteva Cervical Interbody Fusion System implants are manufactured from PEEK (per ASTM F2026) with Tantalum markers (per ASTM F560), or Titanium alloy (Ti-6Al-4V) that conforms to {4}------------------------------------------------ ASTM F136. The PEEK implants are available with or without a medical grade commercially pure titanium (CpTi) plasma coating (per ASTM F1580) on the superior and inferior surfaces. ### D. Indications for Use The Curiteva Cervical Interbody Fusion System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2 – T1 inclusive). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. Implants are intended to be used with autograft and/or allograft bone (comprised of cancellous and/or corticocancellous bone graft) and supplemental spinal fixation systems that have been cleared for use in the cervical spine. Patients should receive at least six (6) weeks of non-operative treatment prior to treatment with the device. ## E. Technological Characteristics As was established in this submission, the subject Curiteva Cervical Interbody Fusion is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and to have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and range of sizes. #### F. Performance Data The Curiteva Cervical Interbody Fusion System was mechanically tested in the following test modes: static and dynamic compression per ASTM F2077, static and dynamic torsion per ASTM F2077, subsidence per ASTM F2267, expulsion, and wear debris characterization per ASTM F1877. The results of this non-clinical testing show that the strength and performance of the Curiteva Cervical Interbody Fusion System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices. ## G. Conclusion Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject Curiteva Cervical Interbody Fusion System has been shown to be substantially equivalent to legally marketed predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.