K181223 · Impact Applications, Inc. · POM · Oct 20, 2018 · Neurology
Device Facts
Record ID
K181223
Device Name
ImPACT
Applicant
Impact Applications, Inc.
Product Code
POM · Neurology
Decision Date
Oct 20, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1471
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ImPACT is intended for use as a computer-based neurocognitive test battery to aid in the assessment and management of concussion. ImPACT is a neurocognitive test battery that provides healthcare professionals with objective measure of neurocognitive functioning as an assessment aid and in the management of concussion in individuals ages 12-59.
Device Story
ImPACT is a computer-based neurocognitive test battery used by healthcare professionals to assess concussion. The device presents stimuli on a computer screen; the test-taker responds using standard computer peripherals. The software captures performance data, including reaction time, memory, attention, spatial processing speed, and symptoms. The system processes these inputs into raw scores, composite scores, and supporting measures (e.g., Invalidity Indicator, Cognitive Efficiency Index, Reliable Change Index). Results are stored in a remote central database and accessed by healthcare providers to assist in clinical decision-making regarding return-to-activity. The device includes risk controls to mitigate environmental factors, such as an Invalidity Indicator to identify invalid baseline exams. Baseline testing can be performed in supervised or unsupervised settings, while post-injury testing requires a supervised environment. The device provides objective data to support a multidisciplinary clinical approach, benefiting patients by providing standardized, researched neurocognitive tasks for concussion management.
Clinical Evidence
A study of 50 participants evaluated test-retest reliability of baseline assessments over a mean of 80 days. Results showed stability across measures: Visual Motor Speed (ICC=0.91), Reaction Time (ICC=0.78), Visual Memory (ICC=0.62), and Verbal Memory (ICC=0.55). No significant practice effects or improvements in composite/symptom scores were observed. A separate usability study of 162 subjects (college, middle/high school, and adults) demonstrated that invalid self-administered test rates in unsupervised environments were comparable to supervised environments reported in literature.
Technological Characteristics
Standalone software application written in HTML5; operates on general-purpose COTS personal computers via web browser. Utilizes remote central database for storage. Risk management conducted per ISO 14971; software development per IEC 62304. Features include neurocognitive test battery (6 modules), symptom questionnaires, and automated scoring (raw/composite scores, Invalidity Indicator, Cognitive Efficiency Index, Reliable Change Index).
Indications for Use
Indicated for individuals ages 12-59 to aid healthcare professionals in the assessment and management of concussion by providing objective measures of neurocognitive functioning.
Regulatory Classification
Identification
The computerized cognitive assessment aid for concussion is a prescription device that uses an individual's score(s) on a battery of cognitive tasks to provide an indication of the current level of cognitive function in response to concussion. The computerized cognitive assessment aid for concussion is used only as an assessment aid in the management of concussion to determine cognitive function for patients after a potential concussive event where other diagnostic tools are available and does not identify the presence or absence of concussion. It is not intended as a stand-alone diagnostic device.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Software, including any proprietary algorithm(s) used by the device to arrive at its interpretation of the patient's cognitive function, must be described in detail in the software requirements specification (SRS) and software design specification (SDS). Software verification, validation, and hazard analysis must be performed.
(2) Clinical performance data must be provided that demonstrates how the device functions as an interpretation of the current level of cognitive function in an individual that has recently received an injury that causes concern about a possible concussion. The testing must:
(i) Evaluate device output and clinical interpretation.
(ii) Evaluate device test-retest reliability of the device output.
(iii) Evaluate construct validity of the device cognitive assessments.
(iv) Describe the construction of the normative database, which includes the following:
(A) How the clinical workup was completed to establish a “normal” population, including the establishment of inclusion and exclusion criteria.
(B) Statistical methods and model assumptions used.
(3) The labeling must include:
(i) A summary of any clinical testing conducted to demonstrate how the device functions as an interpretation of the current level of cognitive function in a patient that has recently received an injury that causes concern about a possible concussion. The summary of testing must include the following:
(A) Device output and clinical interpretation.
(B) Device test-retest reliability of the device output.
(C) Construct validity of the device cognitive assessments.
(D) A description of the normative database, which includes the following:
(
*1* ) How the clinical workup was completed to establish a “normal” population, including the establishment of inclusion and exclusion criteria.(
*2* ) How normal values will be reported to the user.(
*3* ) Representative screen shots and reports that will be generated to provide the user results and normative data.(
*4* ) Statistical methods and model assumptions used.(
*5* ) Whether or not the normative database was adjusted due to differences in age and gender.(ii) A warning that the device should only be used by health care professionals who are trained in concussion management.
(iii) A warning that the device does not identify the presence or absence of concussion or other clinical diagnoses.
(iv) A warning that the device is not a stand-alone diagnostic.
(v) Any instructions technicians must convey to patients regarding the administration of the test and collection of cognitive test data.
In combination with the general controls of the FD&C Act, the Computerized Cognitive Assessment Aid for Concussion is subject to the following special controls:
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 20, 2018
ImPACT Applications, Inc. Michael Zagorski Director of Regulatory Affairs 9665 Granite Ridge Drive, Suite 550 San Diego, California 92123
Re: K181223
Trade/Device Name: ImPACT Regulation Number: 21 CFR 882.1471 Regulation Name: Computerized Cognitive Assessment Aid For Concussion Regulatory Class: Class II Product Code: POM Dated: June 27, 2018 Received: June 29, 2018
Dear Michael Zagorski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
(1-800-638-2041 or 301-796-7100).
Sincerely,
# Jay R. Gupta -S
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181223
Device Name ImPACT
#### Indications for Use (Describe)
ImPACT is intended for use as a computer-based neurocognitive test battery to aid in the assessment and management of concussion.
ImPACT is a neurocognitive test battery that provides healthcare professionals with objective measure of neurocognitive functioning as an assessment aid and in the management of concussion in individuals ages 12-59.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="2" x2="13" y1="2" y2="13"></line> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="13" x2="2" y1="2" y2="13"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for ImPACT Applications, Inc. The logo features a stylized outline of a human head with a brain inside on the left. To the right of the head is the company name, "ImPACT," in a two-tone blue color scheme, with "Im" in a darker blue and "PACT" in a lighter blue. Below the company name is the text "APPLICATIONS, INC." in smaller, black font.
#### Submission Date: September 19, 2018
### Submitter Information:
| Company: | ImPACT Applications, Inc.<br>9665 Granite Ridge Drive, Suite 550<br>San Diego, CA 92123 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Contract Person: | Michael Zagorski<br>Director of Regulatory Affairs<br>ImPACT Applications, Inc.<br>Tel: 412-567-8400 ext. 939<br>Email: mzagorski@impacttest.com |
| Device Information: | |
| Trade Name: | ImPACT® |
| Common Name: | Computerized Cognitive Test |
| Classification Name: | Computerized cognitive assessment aid for concussion |
| Device Classification: | Class II |
| Product Code: | POM, 21 CFR 882.1471 |
| Panel: | Neurology |
| Predicate Device: | ImPACT, K170209 |
| Reason for submission | Device Modifications |
#### Indications for Use:
ImPACT is intended for use as a computer-based neurocognitive test battery to aid in the assessment and management of concussion.
ImPACT is a neurocognitive test battery that provides healthcare professionals with objective measure of neurocognitive functioning as an assessment aid and in the management of concussion in individuals ages 12-59.
# Device Description:
ImPACT® (Immediate Post-Concussion Assessment and Cognitive Testing) is a computer-based neurocognitive test battery.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for ImPACT Applications, Inc. On the left side of the logo is a blue outline of a head with a brain inside. To the right of the head is the word "ImPACT" in blue letters, with the "Im" in a darker blue than the "PACT". Below the word "ImPACT" is the phrase "APPLICATIONS, INC." in smaller, dark blue letters.
lmPACT is a software-based tool that allows healthcare professionals to conduct a series of neurocognitive tests on individuals to gather basic data related to the neurocognitive functioning of the test subject. This computerized cognitive test battery evaluates and provides a healthcare professional with measures of various neurocognitive functions, including the reaction time, memory, attention, spatial processing speed and symptoms of an individual.
lmPACT provides healthcare professionals with a set of well-developed and researched neurocognitive tasks that have been medically accepted as state-of-the-art best practices and is intended to be used as part of a multidisciplinary approach to making return to activity decisions.
lmPACT for use in supervised and unsupervised settings incorporates several risk controls to mitigate the risk of external factors affecting the validity of the test results. These include:
- Test structure that assesses several different cognitive domains. .
- Subscale scores, composite scores based on factor analysis, and supporting measures (e.g., Invalidity Indicator, Cognitive Efficiency Index, and Reliable Change Index) were all developed and validated to reduce error caused by variation in test environment or effort, and to assist the healthcare provider in interpreting of results.
- An Invalidity Indicator was included to aid in identifying invalid baseline examinations when performance falls below the expected range.
- . Baseline test results are not provided to test-takers (the results are only available for viewing by healthcare providers; a unique id, called an ImPACT Passport ID, is assigned to each test allowing the test-takers to share the results with trained healthcare providers).
- Instructions detailing the proper baseline testing environment available to the test . purchase and at the beginning of the test. These instructions must be viewed before proceeding to the baseline test.
- . A baseline test is not required for clinical assessment of concussion, but provides incremental validity to the management of a suspected concussion when used in conjunction with the normative database available for post-injury testing.
# Device Modifications:
The modified version of ImPACT is substantially equivalent to predicate device cleared under K170209. Both devices have the same intended use for use as computerized neurocognitive tests to aid in the assessment and management of concussion. They are also identical in terms of technological characteristics as both are standalone software applications using general purpose computing platform to electronically record objective performance measurements (speed and accuracy) as the test taker responds to stimuli presented on the screen via input devices. Further, functionality of both devices is identical.
The differences between the new device and the predicate include:
- change to the use environment to allow baseline testing in unsupervised environment, and
- modifications to the optional demographics portion of the test to shorten and streamline the test administration.
There are no changes to the intended use, users, patient population or the conditions assessed. Further, there are no changes to the design of neurocognitive test battery; all tasks, stimuli, and captured information remain identical to the original version.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for ImPACT Applications, Inc. The logo features a stylized drawing of a human head with a brain inside, followed by the text "ImPACT" in blue and light blue. Below the text is the phrase "APPLICATIONS, INC." in smaller, uppercase letters. The logo is clean and professional, suggesting a focus on brain-related applications.
510(k) Summary
| Table 1. Predicate Comparison. | | |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate Device: ImPACT (K170209) | Modified Device: ImPACT |
| Intended Use | ImPACT is intended for use as a computer-based neurocognitive test<br>battery to aid in the assessment and management of concussion.<br>ImPACT is a neurocognitive test battery that provides healthcare<br>professionals with objective measure of neurocognitive functioning as<br>an assessment aid and in the management of concussion in individuals<br>ages 12-59. | Same as predicate |
| Patient Population | 12-59 | Same as predicate |
| Use Environment | Recommended supervised environment for baseline and post-injury<br>testing. | Similar to predicate<br>Unsupervised and supervised<br>environment for baseline<br>testing.<br>Supervised environment only<br>for post-injury testing. |
| Neurocognitive test<br>battery | 1. Demographic data, (age, gender, concussion history, relevant<br>medical information)<br>2. Symptoms list and questionnaires<br>3. Neurocognitive test battery consisting of 6 modules:<br>Module 1: Word Memory and Delayed Memory Recognition Module 2: Design Memory and Delayed Design Recognition Module 3: X's and O's Module 4: Symbol Matching Module 5: Color Match Module 6: Three Letter Memory | Same as predicate |
| Results | 1. Recording and scoring of symptoms<br>2. Raw Scores and Composite Scores<br>3. Normative data | Same as predicate |
| Suggest options or<br>treatment | No | Same as predicate |
| User Interface | Desktop or laptop computer screen to present stimuli. | Same as predicate |
| Platform | Stand-alone software running on general purpose commercial off-the-<br>shelf personal computers (desktops, laptops), with a modern web<br>browser connected to the internet. | Same as predicate |
| Software Technology | Software application, written in HTML5, accessed via standard web<br>browser | Same as predicate |
| Stimulus<br>presentation | Information and stimulus displayed on a desktop or laptop computer<br>screen | Same as predicate |
| Stimulus capture<br>(test taker response) | ImPACT uses computer peripherals to capture test taker's response | Same as predicate |
| Data Storage | Remote central database | Same as predicate |
| Standards Used | ISO 14971 and IEC 62304 | Same as predicate |
# Summary of Performance Testing:
Software verification and validation activities including code reviews, evaluations, analyses, traceability assessment, and manual testing were performed in accordance with IEC 62304 and other software standards to demonstrate device performance and functionality. All tests met the required acceptance criteria. Risk Management activities conducted in accordance with ISO 14971 assure all risk related to use of a computerized neurocognitive test, including use related risks and security risks, are appropriately mitigated.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ImPACT Applications, Inc. The logo features a stylized outline of a human head with a brain inside, followed by the company name in a bold, two-tone blue font. The "Im" is a darker blue, while the "PACT" is a lighter blue. Below the company name, the words "APPLICATIONS, INC." are written in a smaller, sans-serif font.
Additionally, to validate test takers' ability to perform the baseline test in an unsupervised environment, a usability assessment and a usability study was performed on 162 subjects. The sample included 74 college students, 44 Middle and High School students, and 44 adults. 5.8% of subjects reported invalid results. The results of the studies indicate that the number of invalid self-administered tests are not different when compared to the supervised environment reported in the literature.
# Summary of Clinical Data:
The goal of this is to demonstrate that performing a baseline test does without direct oversight of a baseline test administrator does not affect the test-retest reliability of ImPACT.
A total of 50 participants completed the ImPACT test and brief survey on two occasions, approximately 80 days between assessments (Mean=80 days, S.D.=17 days, Range=52 to 108 days). None of the participants were unable to complete the test independently, or in one test session, with no "Invalid Baselines" obtained.
Pearson correlations between baseline assessments ranged from .43 to .78. ICCs reflected higher reliability than Pearson's r, across all measures. Visual Motor Speed scores showed the most stability (mean ICC=.91, .84 to .95 (lower and upper 95% confidence intervals); UER=.91), followed by Reaction Time (.78, .61-.87; UER=.79), Visual Memory (.62, .34-.77; UER=.64), and Verbal Memory (.55, .20-.74; UER=.56). Mean ImPACT composite and symptom scores showed no significant improvement between the two assessments on any of the Composite Scores. There were no significant practice effects across the two assessments, at a mean of 80 days, and scores reflected considerable stability as reflected in ICCs and UERs.
# Substantial Equivalence:
The differences in the use environment and software described above do not affect the safety or effectiveness of lmPACT for its intended use, which was demonstrated through risk management, usability assessment and performance testing. Therefore, the modified device is substantially equivalent to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.