← Product Code [CBK](/productcode/CBK) · K181216

# HAMILTON-T1, HAMILTON-C1 (K181216)

_Hamilton Medical AG · CBK · Aug 2, 2019 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181216

## Device Facts

- **Applicant:** Hamilton Medical AG
- **Product Code:** [CBK](/productcode/CBK.md)
- **Decision Date:** Aug 2, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5895
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The HAMILTON-C1 ventilator is intended to provide positive pressure ventilatory support or continuous flow of respiratory gases to adults and pediatrics, and optionally infants and neonates. Intended areas of use: · In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room · During transfer of ventilated patients within the hospital The HAMILTON-C1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications. The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support or continuous flow of respiratory gases to adults and pediatrics, and optionally infants and neonates. Intended areas of use: · In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room · For emergency medical care · During transport within and outside the hospital · During transfer by rescue vehicles, fixed wing aircraft, helicopter or ship The HAMILTON-T1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

## Device Story

HAMILTON-C1/T1 are continuous ventilators providing invasive/non-invasive support; inputs include air/gas mixture, SpO2 sensor data (Masimo or Nihon Kohden), and patient physiological parameters; device transforms inputs via automated ventilation (Adaptive Support Ventilation - ASV) and control settings to deliver respiratory gases; outputs include ventilatory support, monitoring data, and alarms; used in hospitals, emergency wards, and transport (T1 only) by trained personnel; healthcare providers use output to adjust ventilation parameters; benefits include optimized respiratory support for patients ranging from neonates to adults; new features include cFlow (continuous gas delivery), speaking valve compatibility for tracheostomized patients, and Nihon Kohden SpO2 sensor integration.

## Clinical Evidence

Bench testing only. Performance and non-clinical data provided, including software verification/validation, waveform comparison for speaking valve feature, and compatibility testing for cFlow and SpO2 sensors. Compliance with standards including IEC 60601-1, ISO 80601-2-12, and IEC 62304 demonstrated.

## Technological Characteristics

Critical care ventilators; invasive/non-invasive modes; ASV algorithm; SpO2 monitoring (Masimo/Nihon Kohden); cFlow gas delivery; speaking valve support. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 80601-2-12, IEC 60601-1-8, IEC 60601-1-6, IEC 62366, IEC 62304, ISO 80601-2-55, ISO 80601-2-61. Software level of concern: Major.

## Regulatory Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

## Predicate Devices

- HAMILTON-C1/T1 ([K140939](/device/K140939.md))

## Reference Devices

- HAMILTON-C3 ([K161450](/device/K161450.md))
- Nihon Kohden NKV-550 Series Ventilator System ([K181695](/device/K181695.md))
- Esprit Ventilator Speaking mode Option ([K071212](/device/K071212.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 2, 2019

Hamilton Medical AG Annemarie Weideli Quality Engineer Via Crusch 8 Bonaduz, 7402 Ch

Re: K181216

Trade/Device Name: Hamilton-C1, Hamilton-T1 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK, DQA Dated: November 7, 2018 Received: November 13, 2018

## Dear Annemarie Weideli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

James Lee, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K181216

Device Name HAMILTON-C1

### Indications for Use (Describe)

The HAMILTON-C1 ventilator is intended to provide positive pressure ventilatory support or continuous flow of respiratory gases to adults and pediatrics, and optionally infants and neonates.

Intended areas of use:

· In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room

· During transfer of ventilated patients within the hospital

The HAMILTON-C1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (701)-443-6740 EB

{3}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K181216

Device Name HAMILTON-T1

Indications for Use (Describe)

The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support or continuous flow of respiratory gases to adults and pediatrics, and optionally infants and neonates.

Intended areas of use:

· In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room

- · For emergency medical care
- · During transport within and outside the hospital
- · During transfer by rescue vehicles, fixed wing aircraft, helicopter or ship

The HAMILTON-T1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

[ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

PSC Publishing, Services (701) 443-6740 EF

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Hamilton Medical. The words "HAMILTON" and "MEDICAL" are stacked on top of each other in a blue sans-serif font. To the right of the word "HAMILTON" is a blue circle with a stylized "H" inside. Below the logo is the tagline "Intelligent Ventilation since 1983" in a smaller blue font.

# 510(k) SUMMARY

#### l. SUBMITTER:

Hamilton Medical AG Via Crusch 8 Bonaduz, 7402 SWITZERLAND

Phone: +41 58 610 15 68 Fax: +41 58 610 00 20

Contact Person: Frederike Brühschwein-Mandic, Senior Manager Regulatory Affairs Preparation Date: March 20, 2018

Establishment Registration Number: 3001421318

#### II. Device:

| Trade Name:            | HAMILTON-C1, HAMILTON-T1<br>Collectively referred to as the HAMILTON-C1/T1 ventilators |
|------------------------|----------------------------------------------------------------------------------------|
| Classification Name:   | Ventilator, Continuous                                                                 |
| Regulation:            | 21 CFR 868.5895                                                                        |
| Product Code:          | CBK (secondary: DQA)                                                                   |
| Device Classification: | 2                                                                                      |

#### lll. PREDICATE DEVICES:

HAMILTON-C1/T1 (K140939) (Predicate Device)

#### IV. REFERENCE DEVICE

HAMILTON-C3 (K161450) Nihon Kohden NKV-550 Series Ventilator System (K181695) Esprit Ventilator Speaking mode Option (K071212)

#### V. DEVICE DESCRIPTION:

The HAMILTON-C1 and HAMILTON-T1 ventilators are designed for adults, pediatrics, infants and neonatal patients requiring invasive or non-invasive ventilation support. All ventilators cover a range of clinical requirements, including invasive ventilation, automated ventilation with Adaptive Support Ventilation (ASV), and Non-Invasive Ventilation.

The previously cleared ventilators, HAMILTON-C1 and HAMILTON-T1 (both K140939), have been bundled together in this 510(k) submission in order to add the following new features:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Hamilton Medical. The words "HAMILTON" and "MEDICAL" are stacked on top of each other in blue font. Below the words is the phrase "Intelligent Ventilation since 1983" in a smaller blue font.

A modification to the software allows compatibility with the Nihon Kohden -SpO2 sensors to be used with HAMIILTON-C1 and HAMILTON-T1 ventilators. These ventilators are already compatible and cleared for use with Masimo SpO2 sensors (K140939).

- cFlow was added, which continuously delivers an air/gas mixture.
- -A modification to the software, which allows HAMILTON-C1/T1 ventilators to be compatible for use with speaking valves. A speaking valve allows tracheostomized adult and pediatric patients to communicate verbally.

#### VI. INDICATIONS FOR USE

The HAMILTON-C1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates.

Intended areas of use:

- -In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room
- -During transfer of ventilated patients within the hospital

The HAMILTON-C1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

The HAMILTON-T1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates.

Intended areas of use:

- -In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room
- । For emergency medical care
- -During transport within and outside the hospital
- During transfer by rescue vehicles, fixed wing aircraft, helicopter or ship -

The HAMILTON-T1 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

#### VII. SUMMARY OF THE TECHNOLOGY AND PERFORMANCE SPECIFICATIONS COMPARISON WITH THE PREDICATED DEVICES

## Table 1: Comparison of the HAMILTON-C1/T1 with the predicate devices

| Parameter                                           | HAMILTON-C1/T1<br>(Proposed device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HAMILTON-C1/T1<br>(Predicate device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                            |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Indications of Use                                  | <b>HAMILTON-C1:</b><br>The HAMILTON-C1 ventilator is<br>intended to provide positive<br>pressure ventilatory support to<br>adults and pediatrics, and<br>optionally infants and neonates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | <b>HAMILTON-C1:</b><br>The HAMILTON-C1 ventilator is<br>intended to provide positive pressure<br>ventilatory support to adults and<br>pediatrics, and optionally infants and<br>neonates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | <b>HAMILTON-C1</b><br>and<br><b>HAMILTON-<br/>T1:</b><br>Equivalent   |
| Parameter                                           | HAMILTON-C1/T1<br>(Proposed device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HAMILTON-C1/T1<br>(Predicate device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                            |
|                                                     | Intended areas of use:<br>• In the intensive care ward,<br>intermediate care ward,<br>emergency ward, long term acute<br>care hospital, or in the recovery<br>room<br>• During transfer of ventilated<br>patients within the hospital<br>The HAMILTON-C1 ventilator is a<br>medical device intended for use<br>by qualified, trained personnel<br>under the direction of a physician<br>and within the limits of its stated<br>technical specifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Intended areas of use:<br>• In the intensive care ward,<br>intermediate care ward, emergency<br>ward, long term acute care hospital, or<br>in the recovery room<br>• During transfer of ventilated patients<br>within the hospital<br>The HAMILTON-C1 ventilator is a<br>medical device intended for use by<br>qualified, trained personnel under the<br>direction of a physician and within the<br>limits of its stated technical<br>specifications.                                                                                                                                                                                                                                                                                                                             |                                                                       |
|                                                     | HAMILTON-T1:<br>The HAMILTON-T1 ventilator is<br>intended to provide positive<br>pressure ventilatory support to<br>adults, pediatrics, and optionally<br>infants and neonates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | HAMILTON-T1:<br>The HAMILTON-T1 ventilator is<br>intended to provide positive pressure<br>ventilatory support to adults,<br>pediatrics, and optionally infants and<br>neonates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                       |
|                                                     | Intended areas of use:<br>• In the intensive care ward,<br>intermediate care ward,<br>emergency ward, long term<br>acute care hospital or in the<br>recovery room<br>• For emergency medical care<br>• During transport within and<br>outside the hospital<br>• During transfer by rescue<br>vehicles, fixed wing aircraft,<br>helicopter or ship                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Intended areas of use:<br>• In the intensive care ward,<br>intermediate care ward,<br>emergency ward, long term acute<br>care hospital or in the recovery<br>room<br>• For emergency medical care<br>• During transport within and<br>outside the hospital<br>• During transfer by rescue vehicles,<br>fixed wing aircraft, helicopter or<br>ship                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                       |
|                                                     | The HAMILTON-T1 ventilator is a<br>medical device intended for use<br>by qualified, trained personnel<br>under the direction of a physician<br>and within the limits of its stated<br>technical specifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | The HAMILTON-T1 ventilator is a<br>medical device intended for use by<br>qualified, trained personnel under the<br>direction of a physician and within the<br>limits of its stated technical<br>specifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                       |
| Control settings<br>(Ranges) Adult/Ped              | HAMILTON-C1 and HAMILTON-<br>T1:<br>• Apnea backup (On, automatic,<br>off)<br>• ETS (5 to 80%)<br>• Flow (2-60l/min)<br>• Flow trigger (0.1 to 20 l/min,<br>off)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea backup (On, automatic, off)<br>• ETS (5 to 80%)<br>• Flow trigger (0.1 to 20 l/min, off)<br>• Gender (Male, female)<br>• Loudness, alarm signals (1 to 10)<br>• %MinVol (25 to 350%)<br>• O2 monitoring (On, off)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | HAMILTON-C1<br>and<br>HAMILTON-<br>T1:<br>Substantially<br>Equivalent |
| Parameter                                           | HAMILTON-C1/T1<br>(Proposed device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HAMILTON-C1/T1<br>(Predicate device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                            |
|                                                     | • Gender (Male, female)<br>• Loudness, alarm signals (1 to 10)<br>• %MinVol (25 to 350%)<br>• O2 monitoring (On, off)<br>• CO2 monitoring (On, off)<br>• SpO2 monitoring (On, off)<br>• Oxygen (21 to 100%)<br>• Pasvlimit (5 to 60 cm)<br>• Patient height (30 to 250 cm)<br>• Pcontrol (5 to 60 cmH2O)<br>• PEEP/CPAP (0 to 35 cmH2O)<br>• Pinsp (5 to 60 cmH2O)<br>• P-ramp (0 to 2000 ms)<br>• Psupport (0 to 60 cmH2O)<br>• Phigh (0 to 60 cmH2O)<br>• Plow (0 to 35 cmH2O)<br>• Rate (1 to 80 b/min)<br>• Sigh (On, off)<br>• Thigh (0.1 to 40 s)<br>• TI (0.1 to 12 s)<br>• Tlmax (1 to 3 s)<br>• Tlow (0.2 to 40 s)<br>• Vt (20 to 2000 ml)                                                                                                                                                                                                                                                                                           | • CO2 monitoring (On, off)<br>• SpO2 monitoring (On, off)<br>• Oxygen (21 to 100%)<br>• Pasvlimit (5 to 60 cm)<br>• Patient height (30 to 250 cm)<br>• Pcontrol (5 to 60 cmH2O)<br>• PEEP/CPAP (0 to 35 cmH2O)<br>• Pinsp (5 to 60 cmH2O)<br>• P-ramp (0 to 2000 ms)<br>• Psupport (0 to 60 cmH2O)<br>• Phigh (0 to 60 cmH2O)<br>• Plow (0 to 35 cmH2O)<br>• Rate (1 to 80 b/min)<br>• Sigh (On, off)<br>• Thigh (0.1 to 40 s)<br>• TI (0.1 to 12 s)<br>• Tlmax (1 to 3 s)<br>• Tlow (0.2 to 40 s)<br>• Vt (20 to 2000 ml)                                                                                                                                                                                                                                                        |                                                                       |
| Control settings<br>(Ranges) NEO                    | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea backup (On, automatic, off)<br>• ETS (5 to 80%)<br>• Flow (2-12 l/min)<br>• Flow trigger (0.1 to 20 l/min, off)<br>• Loudness, alarm signals (1 to 10)<br>• O2 monitoring (On, off)<br>• CO2 monitoring (On, off)<br>• SpO2 monitoring (On, off)<br>• Oxygen (21 to 100%)<br>• Pcontrol (0 to 45 cmH2O)<br>• PEEP/CPAP (3 to 25 cmH2O)<br>• Pinsp (3 to 45 cmH2O)<br>• P-ramp (0 to 600 ms)<br>• Psupport (0 to 45 cmH2O)<br>• Phigh (0 to 45 cmH2O)<br>• Plow (0 to 25 cmH2O)<br>• Rate (1 to 80 b/min)<br>• Thigh (0.1 to 40 s)<br>• TI (0.1 to 12 s)<br>• Tlmax (1 to 3 s)<br>• Tlow (0.25 to 40 s)                                                                                                                                                                                                                                                                                                | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea backup (On, automatic, off)<br>• ETS (5 to 80%)<br>• Flow trigger (0.1 to 20 l/min, off)<br>• Loudness, alarm signals (1 to 10)<br>• O2 monitoring (On, off)<br>• CO2 monitoring (On, off)<br>• SpO2 monitoring (On, off)<br>• Oxygen (21 to 100%)<br>• Pcontrol (0 to 45 cmH2O)<br>• PEEP/CPAP (3 to 25 cmH2O)<br>• Pinsp (3 to 45 cmH2O)<br>• P-ramp (0 to 600 ms)<br>• Psupport (0 to 45 cmH2O)<br>• Phigh (0 to 45 cmH2O)<br>• Plow (0 to 25 cmH2O)<br>• Rate (1 to 80 b/min)<br>• Thigh (0.1 to 40 s)<br>• TI (0.1 to 12 s)<br>• Tlmax (1 to 3 s)<br>• Tlow (0.25 to 40 s)<br>• Vt (2 to 300 ml)<br>• Weight (0.2 kg to 30 kg)                                                                                                       | HAMILTON-C1<br>and<br>HAMILTON-<br>T1:<br>Substantially<br>Equivalent |
|                                                     | • Vt (2 to 300 ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                       |
| Parameter                                           | HAMILTON-C1/T1<br>(Proposed device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | HAMILTON-C1/T1<br>(Predicate device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparison                                                            |
|                                                     | • Weight (0.2 kg to 30 kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                       |
| Control settings<br>(Default settings)<br>Adult/Ped | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea backup (On)<br>• ETS (25%)<br>• Flow (15 l/min)<br>• Flow trigger (5 l/min)<br>• Gender (Male)<br>• Loudness, alarm signals (5)<br>• %MinVol (100%)<br>• O2 monitoring (On)<br>• CO2 monitoring (Off)<br>• SpO2 monitoring (Off)<br>• Oxygen (50%)<br>• Pasvlimit (30 cm)<br>• Patient height (174 cm)<br>• Pcontrol (15 cmH2O)<br>• PEEP/CPAP (5 cmH2O)<br>• Pinsp (15 cmH2O)<br>• P-ramp (50 ms)<br>• Psupport (15 cmH2O)<br>• Phigh (20 cmH2O)<br>• Plow (5 cmH2O)<br>• Rate (10 b/min)<br>• Sigh (Off)<br>• Thigh (2.0 s)<br>• TI (1 s)<br>• Tlmax (1.5 s)<br>• Tlow (4.0 s)<br>• Vt (700 ml)                                                                                                                                                                                                                                                                                                     | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea backup (On)<br>• ETS (25%)<br>• Flow trigger (5 l/min)<br>• Gender (Male)<br>• Loudness, alarm signals (5)<br>• %MinVol (100%)<br>• O2 monitoring (On)<br>• CO2 monitoring (Off)<br>• SpO2 monitoring (Off)<br>• Oxygen (50%)<br>• Pasvlimit (30 cm)<br>• Patient height (174 cm)<br>• Pcontrol (15 cmH2O)<br>• PEEP/CPAP (5 cmH2O)<br>• Pinsp (15 cmH2O)<br>• P-ramp (50 ms)<br>• Psupport (15 cmH2O)<br>• Phigh (20 cmH2O)<br>• Plow (5 cmH2O)<br>• Rate (10 b/min)<br>• Sigh (Off)<br>• Thigh (2.0 s)<br>• TI (1 s)<br>• Tlmax (1.5 s)<br>• Tlow (4.0 s)<br>• Vt (700 ml)                                                                                                                                                              | HAMILTON-C1<br>and<br>HAMILTON-<br>T1:<br>Substantially<br>Equivalent |
| Control settings<br>(Default settings)<br>Neo       | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea backup (On)<br>• ETS (25%)<br>• Flow (2 l/min)<br>• Flow trigger (0.5 l/min)<br>• Loudness, alarm signals (3)<br>• O2 monitoring (On)<br>• CO2 monitoring (Off)<br>• SpO2 monitoring (Off)<br>• Oxygen (40%)<br>• Pcontrol (15 cmH2O)<br>• PEEP/CPAP (5 cmH2O)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea backup (On)<br>• ETS (25%)<br>• Flow trigger (0.5 l/min)<br>• Loudness, alarm signals (3)<br>• O2 monitoring (On)<br>• CO2 monitoring (Off)<br>• SpO2 monitoring (Off)<br>• Oxygen (40%)<br>• Pcontrol (15 cmH2O)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | HAMILTON-C1<br>and<br>HAMILTON-<br>T1:<br>Substantially<br>Equivalent |
| Parameter                                           | HAMILTON-C1/T1<br>(Proposed device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HAMILTON-C1/T1<br>(Predicate device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                            |
|                                                     | • Pinsp (15 cmH2O)<br>• P-ramp (50 ms)<br>• Psupport (15 cmH2O)<br>• Phigh (20 cmH2O)<br>• Plow (5 cmH2O)<br>• Rate (derived from the body<br>weight setting b/min)<br>• Thigh (based on rate s)<br>• TI (1 s)<br>• Tlmax (based on weight s)<br>• Tlow (based on weight s)<br>• Vt (based on weight ml)<br>• Weight (2.0 kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | • Psupport (15 cmH2O)<br>• Phigh (20 cmH2O)<br>• Plow (5 cmH2O)<br>• Rate (derived from the body weight<br>setting b/min)<br>• Thigh (based on rate s)<br>• TI (1 s)<br>• Tlmax (based on weight s)<br>• Tlow (based on weight s)<br>• Vt (based on weight ml)<br>• Weight (2.0 kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                       |
| Modes of ventilation                                | HAMILTON-C1 and HAMILTON-T1:<br>• ASV<br>• (S)CMV+<br>• SIMV+<br>• PCV+<br>• PSIMV+<br>• SPONT<br>• DuoPAP<br>• APRV<br>• NIV<br>• NIV-ST<br>• nCPAP<br>• nCPAP-PC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | HAMILTON-C1<br>and<br>HAMILTON-<br>T1:<br>Equivalent                  |
| Alarms, non-<br>adjustable (Priorities)             | HAMILTON-C1 and HAMILTON-T1:<br>• Air supply (high)<br>• Apnea (high)<br>• Apnea ventilation (low)<br>• Apnea ventilation ended (low)<br>• ASV: Cannot meet the target<br>(low)<br>• Battery power loss (high)<br>• Battery low (high)<br>• Battery 1/2 wrong (low)<br>• Battery 1/2 calibration required<br>(low)<br>• Battery 1/2 temperature high<br>(high)<br>• Blower service required (low)<br>• Check Flow Sensor (high)<br>• Check Flow Sensor tubing (high)<br>• Check for blockage (high)<br>• Device temperature high (high)<br>• Disconnection (high)<br>• Disconnection on patient side<br>(high)                                                                                                                                                                                                                                                                                                                                | • Air supply (high)<br>• Apnea (high)<br>• Apnea ventilation (low)<br>• Apnea ventilation ended (low)<br>• ASV: Cannot meet the target (low)<br>• Battery power loss (high)<br>• Battery low (high)<br>• Battery 1/2 wrong (low)<br>• Battery 1/2 calibration required<br>(low)<br>• Battery 1/2 temperature high (high)<br>• Blower service required (low)<br>• Check Flow Sensor (high)<br>• Check Flow Sensor tubing (high)<br>• Device temperature high (high)<br>• Disconnection (high)<br>• Disconnection on patient side (high)<br>• Disconnection on ventilator side<br>(high)<br>• Exhalation obstructed (high)<br>• Exhalation port occluded (high)                                                                                                                     | HAMILTON-C1<br>and<br>HAMILTON-<br>T1:<br>Substantially<br>Equivalent |
| Parameter                                           | HAMILTON-C1/T1<br>(Proposed device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HAMILTON-C1/T1<br>(Predicate device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                            |
|                                                     | • Disconnection on ventilator side (high)<br>• Exhalation obstructed (high)<br>• Exhalation port occluded (high)<br>• External flow sensor failed (high)<br>• Fan failure (medium)<br>• Flow (high, low after silencing)<br>• Flow sensor calibration needed (High)<br>• High pressure during sigh (high)<br>• High tidal volume (medium)<br>• IRV (low)<br>• Loss of external power (low)<br>• Loss of PEEP (medium)<br>• Obstruction (high) (only for nCPAP and nCPAP-PC modes)<br>• O2 sensor calibration needed (low)<br>• O2 sensor defective (low)<br>• O2 sensor missing (low)<br>• O2 sensor not system-compatible (low)<br>• Oxygen supply failed (high)<br>• SpO2 adapter missing (medium)<br>• SpO2 sensor error (medium)<br>• SpO2 light interference (medium)<br>• SpO2 low perfusion index (medium)<br>• SpO2 Patient disconnected (medium)<br>• SpO2 Probe missing (medium)<br>• Performance limited by high altitude (medium) | • Fan failure (medium)<br>• Flow (high, low after silencing)<br>• Flow sensor calibration needed (High)<br>• High pressure during sigh (high)<br>• High tidal volume (medium)<br>• IRV (low)<br>• Loss of external power (low)<br>• Loss of PEEP (medium)<br>• Obstruction (high) (only for nCPAP and nCPAP-PC modes)<br>• O2 sensor calibration needed (low)<br>• O2 sensor defective (low)<br>• O2 sensor missing (low)<br>• O2 sensor not system-compatible (low)<br>• Oxygen supply failed (high)<br>• SpO2 adapter missing (medium)<br>• SpO2 sensor error (medium)<br>• SpO2 light interference (medium)<br>• SpO2 low perfusion index (medium)<br>• SpO2 Patient disconnected (medium)<br>• SpO2 Probe missing (medium)<br>• Performance limited by high altitude (medium) |                                                                       |
| Alarm settings<br>(Ranges), Adult                   | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea time (15 to 60 s)<br>• ExpMinVol low (Off, 0.1 to 50 l/min)<br>• ExpMinVol high (0.1 to 50, Off l/min)<br>• fTotal low (0 to 99 b/min)<br>• fTotal high (0 to 99 b/min)<br>• Oxygen low (18 to 97%)<br>• Oxygen high (18 to 105%)<br>• PetCO2 low (Off, 0 to 100 mmHg)<br>• PetCO2 high (Off, 1 to 100 mmHg)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HAMILTON-C1 and HAMILTON-T1:<br>• Apnea time (15 to 60 s)<br>• ExpMinVol low (Off, 0.1 to 50 l/min)<br>• ExpMinVol high (0.1 to 50, Off l/min)<br>• fTotal low (0 to 99 b/min)<br>• fTotal high (0 to 99 b/min)<br>• Oxygen low (18 to 97%)<br>• Oxygen high (18 to 105%)<br>• PetCO2 low (Off, 0 to 100 mmHg)<br>• PetCO2 high (1 to 100 mmHg)<br>• Pressure high (15 to 70 cmH2O)<br>• Pressure low (4 to 60 cmH2O)<br>• Pressure limitation (5 to 60 cmH2O)                                                                                                                                                                                                                                                                                                                    | HAMILTON-C1<br>and<br>HAMILTON-<br><b>T1:</b><br>Equivalent           |
| Parameter                                           | HAMILTON-C1/T1<br>(Proposed device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HAMILTON-C1/T1<br>(Predicate device )                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Comparison                                                            |
|                                                     | Pressure high (15 to 70 cmH2O) Pressure low (4 to 60 cmH2O) Pressure limitation (5 to 60 cmH2O) Vt low (10 to 3000, Off ml) Vt high (Off, 10 to 3000 ml) SpO2 high (71 to 100, Off %) SpO2 low (70 to 99, Off %) Pulse high (35 to 235 1/min) Pulse low (30 to 230 1/min) PI high (0.04 to 19, Off %) PI low (Off/0.03 to 18 %) PVI high (2 to 100, Off) PVI low (Off, 1 to 99)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Vt low (10 to 3000, Off ml) Vt high (Off, 10 to 3000 ml) SpO2 high (71 to 100, Off %) SpO2 low (70 to 99, Off %) Pulse high (35 to 235 1/min) Pulse low (30 to 230 1/min) PI high (0.04 to 19, Off %) PI low (Off/0.03 to 18 %) PVI high (2 to 100, Off) PVI low (Off, 1 to 99)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                       |
| Alarm settings<br>(Ranges), Neonatal                | Apnea time (5 to 60 s) ExpMinVol low (Off, 0.01 to 10, Off l/min) ExpMinVol high (0.03 to 10 l/min, Off) Flow (8 to 30 l/min) fTotal low (0 to 200 b/min) fTotal high (2 to 210 b/min) Oxygen low (18 to 97%) Oxygen high (18 to 105%) PetCO2 low (Off, 0 to 100 mmHg) PetCO2 high (1 to 100 mmHg) Pressure high (10 to 55 cmH2O) Pressure low (2 to 55 cmH2O) Pressure limitation (5 to 60 cmH2O) Vt low (Off, 0.1 to 300 ml) Vt high (0.1 to 300, Off ml) SpO2 high (71 to 100, Off %) SpO2 low (70 to 99, Off %) Pulse high (35 to 235 1/min) Pulse low (30 to 230 1/min) PI high (0.04 to 19, Off %) PI low (Off/0.03 to 18 %) PVI high (2 to 100, Off) PVI low (Off, 1 to 99)                                                                                                                                                                                                                                                            | Apnea time (5 to 60, Off s) ExpMinVol low (Off, 0.01 to 10 l/min) ExpMinVol high (0.03 to 10 l/min, Off) Flow (8 to 30 l/min) fTotal low (0 to 200 b/min) fTotal high (2 to 210 b/min) Oxygen low (18 to 97%) Oxygen high (18 to 105%) PetCO2 low (Off, 0 to 100 mmHg) PetCO2 high (1 to 100 mmHg) Pressure high (10 to 55 cmH2O) Pressure low (2 to 55 cmH2O) Pressure limitation (5 to 60 cmH2O) Vt low (Off, 0.1 to 300 ml) Vt high (0.1 to 300, Off ml) SpO2 high (71 to 100, Off %) SpO2 low (70 to 99, Off %) Pulse high (35 to 235 1/min) Pulse low (30 to 230 1/min) PI high (0.04 to 19, Off %) PI low (Off/0.03 to 18 %) PVI high (2 to 100, Off) PVI low (Off, 1 to 99)                                                                                                | HAMILTON-C1<br>and<br>HAMILTON-<br>T1:<br>Equivalent                  |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and consists of the words "HAMILTON" and "MEDICAL" stacked on top of each other. To the right of the word "HAMILTON" is a stylized "H" symbol. The font is bold and sans-serif.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is blue and features the word "HAMILTON" in large, bold letters on the top line. To the right of the word "HAMILTON" is a stylized "H" symbol. The word "MEDICAL" is on the second line in large, bold letters.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and consists of the word "HAMILTON" in large, bold letters, with a stylized "H" symbol to the right. Below "HAMILTON" is the word "MEDICAL" in a similar font and color, but slightly smaller in size. The overall design is clean and professional.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is in blue and consists of two lines of text. The first line says "HAMILTON" with a stylized "H" in a circle. The second line says "MEDICAL".

Image /page/9/Picture/1 description: The image contains two lines of text in a blue font. The first line reads "HAMILTON-C1", and the second line reads "HAMILTON-T1". The text appears to be a label or identifier, possibly related to a product or system. The font is sans-serif and the text is aligned to the left.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and features the words "HAMILTON" and "MEDICAL" stacked on top of each other. To the right of the word "HAMILTON" is a stylized "H" inside of a circle. The font is bold and sans-serif.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and features the word "HAMILTON" in large, bold letters, with a stylized "H" symbol to the right. Below "HAMILTON" is the word "MEDICAL", also in large, bold letters.

## Intelligent Ventilation since 1983

Hamilton Medical has demonstrated the modified HAMILTON-C1 and HAMILTON-T1 ventilators to be substantially equivalent to currently marketed predicate devices, which have been previously cleared by FDA.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and consists of the words "HAMILTON" and "MEDICAL" stacked on top of each other. To the right of the word "HAMILTON" is a stylized "H" symbol. The logo is simple and professional, and it is likely used to represent the company in its marketing and branding materials.

elligent Ventilation since 1983

VIII. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE REFERENCE DEVICES

The Nihon Kohden NKV-550 Series Ventilator System is used as a reference device for the proposed HAMILTON-C1/T1 as both the reference device and the proposed device are intensive care ventilators which have the O2 therapy/cFlow feature.

The HAMILTON-C3 is used as a reference device for the proposed HAMILTON-C1/T1 as both the reference device and the proposed device are intensive care ventilators which can be used with Nihon Kohden SpO2 sensors and accessories.

The Esprit Ventilator V200 with Speaking mode Option is used as a reference device for the proposed Speaking valve compatibility on the modified HAMILTON-C1/T1, which is only available in invasive modes for tracheostomized patients with the use of a special speaking valve, for example the Passy-Muir valve.

#### IX. NON-CLINICAL PERFORMANCE TESTS

The following performance and nonclinical data are provided in support of the substantial equivalence determination.

The Software Design and Validation process, together with the bench testing of the device, demonstrated that the HAMILTON-C1/T1 operates as intended.

In particular, testing demonstrated that the HAMILTON-C1/T1 is compliant with the following guidelines and standards:

- -ANSI/AAMI ES60601-1 (2005/ (R) 2012): Medical electrical equipment – General Requirements for Safety
- -IEC 60601-1-2 (2014): Medical electrical equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- -ISO 80601-2-12 (2011): Medical electrical equipment - Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators
- -IEC 60601-1-8 (2006 + Am.1: 2012): Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- -IEC 60601-1-6 (2010 + A1 :2013): Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- -IEC 62366 (2014): Medical devices - Application of usability engineering to medical de- vices
- -ANSI/AAMI HE75(2009(R) 2013): Human factors engineering - Design of medicalde- vices
- -IEC 62304 (2006): Medical device software - Software life-cycle processes
- -ISO 80601-2-55 (2011): Medical electrical equipment -- Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors
- -ISO 80601-2-61 (2011): Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and features the word "HAMILTON" in large, bold letters, with a stylized "H" symbol next to it. Below "HAMILTON" is the word "MEDICAL" in a similar font and color, but slightly smaller in size.

Image /page/13/Picture/1 description: The image shows two lines of text in blue font. The first line reads "HAMILTON-C1", and the second line reads "HAMILTON-T1". The text appears to be a label or identifier, possibly for equipment or products.

elligent Ventilation since 1983

AIM 7351731: Medical Electrical Equipment and System Electromagnetic Immunity Test for Exosure to Radio Frequency Identification Readers

Additional software verification and validation testing were completed as recommended by the FDA's "Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

Testing of the modified HAMILTON-C1/T1, with the new features, was conducted. The new features cFlow, Speaking Valve, and the use of Nihon Kohden SpO2 sensors were subjected to comparison testing with legally marketed devices. Furthermore, waveform comparison testing was completed for the Speaking Valve Feature. The data provided from these tests was shown to be equivalent to the legally marketed devices.

Since only materials already used in in the predicate (cleared under document number K140939) are described with this 510(k), Hamilton Medical did not conduct any additional biocompatibility testing.

#### X. CONCLUSION

The results of verification, validation, and testing activities demonstrate that the modified HAMILTON-C1 and HAMILTON-T1 ventilators are substantially equivalent to the legally marketed devices identified herein.

---

**Source:** [https://fda.innolitics.com/device/K181216](https://fda.innolitics.com/device/K181216)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
