Balego TENS Digital Edition with Accessories for Pain Relief (OTC)
K181207 · Everyway Medical Instruments Co., Ldt. · NUH · Jun 6, 2018 · Neurology
Device Facts
Record ID
K181207
Device Name
Balego TENS Digital Edition with Accessories for Pain Relief (OTC)
Applicant
Everyway Medical Instruments Co., Ldt.
Product Code
NUH · Neurology
Decision Date
Jun 6, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Balego™ TENS Digital Edition with Accessories for Pain Relief (OTC) is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
Device Story
Dual-channel transcutaneous electrical nerve stimulator (TENS); applies electrical current to skin via electrodes for pain relief. Device consists of stimulus generator and electrodes. User selects pre-programmed modes with fixed pulse width, rate, and frequency; intensity is manually adjustable. Used by patients in home or non-clinical settings for temporary muscle pain relief. Output is biphasic asymmetric rectangular waveform. Device features automatic shut-off and no-load trip. Healthcare provider/patient uses output to manage pain symptoms; device provides non-pharmacological pain relief.
Clinical Evidence
No clinical data; bench testing only. Substantial equivalence established via descriptive comparison of technological characteristics and intended use.
Technological Characteristics
Dual-channel TENS; 9V battery powered; ABS housing. Biphasic asymmetric rectangular waveform. Max output: 80mA @ 500Ω. Fixed pulse width/rate/frequency per mode; adjustable intensity. Type BF applied part. Connectivity: None. Sterilization: Not specified. Software: Microprocessor-controlled.
Indications for Use
Indicated for temporary relief of pain associated with sore and aching muscles in the low back, arms, and legs due to exercise or normal household/work activities. For OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Everyway Electrode, Wire Series/ model no. KF5050 (K083302)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo.
June 6, 2018 Everyway Medical Instruments CO., LDT. % Aaron Hage Associate Attorney DuVal & Associates 825 Nicollet Mall, Suite 1820 Minneapolis, Minnesota 55402
Re: K181207
Trade/Device Name: Balego TENS Digital Edition with Accessories for Pain Relief (OTC) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: May 4, 2018 Received: May 7, 2018
Dear Aaron Hage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
William J. Heetderks -S 2018.06.06 16:37:48 -04'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181207
#### Device Name
Balego™ TENS Digital Edition with Accessories for Pain Relief (OTC)
Indications for Use (Describe)
The Balego™ TENS Digital Edition with Accessories for Pain Relief (OTC) is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.'
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "EVERYWAY" in a teal color, sandwiched between horizontal teal lines. There are three lines above the word and three lines below the word. The lines and the word are all the same color.
EVERYWAY MEDICAL INSTRUMENTS CO., LTD. 3 FL., NO. 5, Lane 155, Section 3, Beishen Rd. Shenkeng Dist., New Taipei City, 22203, Taiwan
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being prepared in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) number is: K181207.
- 1. Submitter's Identifications:
| Establishment: | Everyway Medical Instrument CO., LTD. |
|--------------------------------------------|--------------------------------------------------------------------------------------------------|
| Address: | 3 FL., NO. 5, Lane 155, Section 3, Beishen Rd.<br>Shenkeng Dist., New Taipei City, 22203, Taiwan |
| Registration Number: | 9616877 |
| Operations Manufacturer<br>Owner/Operator: | Everyway Medical Instrument CO., LTD. |
| Contact Person: | Robert Tu |
| Phone: | 886-2-2662-0038 |
| Fax No: | 886-2-2664-5566 |
| e-mail: | robert@everyway-medical.com |
### 2. Date 510(k) Summary Prepared: April 25, 2018
### 3. Name of the Subject Device and Classification Information:
| Tradename/Device<br>Name | Balego™ TENS Digital Edition with Accessories for Pain<br>Relief (OTC) |
|--------------------------|------------------------------------------------------------------------|
| Regulation<br>Number | 21 CFR 882.5890 |
| Classification Name | Transcutaneous electrical nerve stimulator for pain relief. |
| Regulatory Class | Class II |
| Product Code | NUH |
- 4. Information of the 510(k) Cleared Device (Predicate Device):
| Tradename/Device | EV-804 OTC Pain Relief TENS |
|---------------------|-------------------------------------------------------------|
| Name | |
| 510(k) Number | K172919 |
| Regulation Number | 21 CFR 882.5890 |
| Classification Name | Transcutaneous electrical nerve stimulator for pain relief. |
| Regulatory Class | Class II |
| Product Code | NUH |
{5}------------------------------------------------
### 5. Device Description:
The Balego™ TENS Digital Edition with Accessories for Pain Relief (OTC) is a dual channel transcutaneous electrical nerve stimulator used for pain relief by applying an electrical current to electrodes, which are attached on the user's skin. The output and waveform characteristic are fixed for every operation mode, only the intensity is adjustable within specified limit. The Balego™ TENS Digital Edition consists mainly of two parts: the stimulus generator, electrode. The stimulus generator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the user's skin so as to transmit this stimulus current for pain relief.
The stimulation modes for Balego™ TENS Digital Edition are pre-program modes with fixed pulse width, pulse rate, frequency. The amplitude for each program are adjustable. The subject and predicate devices both use Everyway Electrode, Wire Series/ model no. KF5050, size 50x50mm, wire type as a standard accessory. The Electrode was cleared under K083302.
With the combination of the main device parts, the device can be placed on the treatment locations as recommended in the user manual for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
### 6. Indications for Use:
The Balego™ TENS Digital Edition with Accessories for Pain Relief (OTC) is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
{6}------------------------------------------------
# 7. Comparison to the 510(k) Cleared Device (Predicate Device):
| Features | Predicate Model | Proposed Model |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | EV-804 | BAL-820 |
| 510(K) No. | K172919 | K18XXXX |
| Prescription or OTC | OTC | OTC |
| Indication for use | The OTC Pain Relief TENS, model EV-804, is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. | The Balego™ TENS Digital Edition, model BAL-820, is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. |
| FDA product code | NUH | NUH |
| Accessory | Self Adhesive Electrode<br>(5x5 cm)/K083302 | Self Adhesive Electrode<br>(5x5 cm)/K083302 |
# Table S-1. Overview of the predicate and proposed device model
### Table S-2. Comparison of significant output characteristics.
| Comparison feature | Predicate Model | New Model |
|-----------------------------------|--------------------|--------------------|
| | EV-804 | Balego TENS |
| Net charge | 0 | 0 |
| Max. phase charge | 20.8 uc | 20.8 uc |
| Max. current Density | 0.0998 mA/cm2 | 0.0998 mA/cm2 |
| Max. Average<br>current<br>(RMSA) | 500Ω | 80 mA |
| | 2KΩ | 40 mA |
| | 10KΩ | 10 mA |
| Max. Power Density | 0.00399 Watts/ cm2 | 0.00399 Watts/ cm2 |
| Burst Mode | Yes | Yes |
# Table S-3. Comparison of output specifications.
| | | Predicate Device | New Device |
|--------------------------------------------------------|----------------|------------------------------------------|------------------------------------------|
| Mode or Program Name | | EV-804 | Balego TENS |
| 510(K) Number | | K172919 | K18XXXX |
| Waveform (e.g., pulsed monophasic, biphasic) | | Biphasic asymmetric | Biphasic asymmetric |
| Shape (e.g., rectangular, spike, rectified sinusoidal) | | Rectangular | Rectangular |
| Maximum Output Voltage<br>(volts) (+/- 20 %) | | 40V @500Ω<br>80V @2KΩ<br>100V @10KΩ | 40V @500Ω<br>80V @2KΩ<br>100V @10KΩ |
| Maximum Output Current<br>(mA) (+/- 20 %) | | 80mA @500Ω<br>40mA @2KΩ<br>10mA<br>@10KΩ | 80mA @500Ω<br>40mA @2KΩ<br>10mA<br>@10KΩ |
| Duration of primary phase (usec) | | 260 max | 260 max |
| Pulse Duration (usec) | | 8333 max | 8333 max |
| Frequency (Hz) [or Rate (pps)] | | 120 max | 120 max |
| For | | | |
| multiphasic<br>waveforms<br>only: | Yes | Yes | Yes |
| | Not applicable | Not applicable | Not applicable |
{7}------------------------------------------------
| | | Predicate Device | New Device |
|----------------------------------------------------------------------------------------------------|-------------------------------------------|--------------------------------------|-----------------------------|
| Power Source(s) | | 9V x 1<br>(6F22<br>Size) | 9V x 1<br>(6F22<br>Size) |
| | | Type BF | Type BF |
| - Method of Line current Isolation | | --- | --- |
| - Patient Leakage Current | | --- | --- |
| - Normal condition (uA) | | Under 0.1 | Under 0.1 |
| - single Fault condition (uA) | | Under 0.5 | Under 0.5 |
| Average DC current through electrodes when<br>device is on but no pulses are being applied<br>(uA) | | Not applicable | Not applicable |
| Number of Output Modes | | 8 | 8 |
| Number of Output | Synchronous | Synchronous | Synchronous |
| Channels: | Output<br>Coil | Transformer | Transformer |
| Regulated Current or Regulated Voltage? | | Current | Current |
| | Software/Firmware/Microprocessor control? | Yes | Yes |
| Automatic Overload Trip? | | No | No |
| Automatic No-Load Trip? | | Yes | Yes |
| Automatic Shut Off? | | Yes | Yes |
| User Override control? | | No | No |
| | Yes | Yes | Yes |
| Indicator Display: | Yes | Yes | Yes |
| | Yes | Yes | Yes |
| Timer Range (Minutes) | | 15, 30,60 and<br>Continuous | 15, 30,60 and<br>Continuous |
| | | Compliance with Voluntary Standards? | |
| Compliance with 21 CFR 898? | | Yes | Yes |
| Weight (g) including battery | | 155 | 155 |
| Dimensions (mm.) [W x H x D] | | 101×68×21.5 | 101×68×21.5 |
| Housing Materials and construction | | ABS | ABS |
| Pulse per burst | | 6 | 6 |
| Burst per second | | 2 | 2 |
| Bust duration | | 260us | 260us |
| Duty Cycle | | Same for each<br>program | Same for each<br>program |
| | | | |
| Method of achieving zero net charge for net<br>charge/pulse | | wave for each pulse | wave for each pulse |
### 8. Non-Clinical Tests Performed to Determine the Safety and Performance of the Device:
Non-clinical performance data is not required to determine the safety and performance of the device. The subject device is identical to the predicate device with regards to intended use, technological characteristics, manufacturing, and labeling (apart from the different tradename). Therefore, descriptive information alone is sufficient to address the substantial equivalence determination.
{8}------------------------------------------------
- 9. Clinical Test Validation Activities Performed to Determine the Effectiveness of Device:
Clinical tests were not conducted for the Balego TENS device because there were no changes to the indications for use or technological characteristics as compared to the predicate device that would warrant clinical testing.
# 10. Summary for the technology comparison.
As evidence by The Balego™ TENS device has the same technological characteristics as the predicate device regarding product design, material, and energy source type, and all other technological features.
## 11. Conclusions
The Balego TENS device is the identical device to the predicate device, apart from its tradename. The Balego TENS device has the same intended use and the same technological characteristics as the legally marketed predicate device. Therefore, the Balego TENS device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.