← Product Code [ORR](/productcode/ORR) · K181195

# ClearPoint System (K181195)

_Mri Interventions, Inc. · ORR · Nov 2, 2018 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181195

## Device Facts

- **Applicant:** Mri Interventions, Inc.
- **Product Code:** [ORR](/productcode/ORR.md)
- **Decision Date:** Nov 2, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The ClearPoint System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners.

## Device Story

ClearPoint System provides MRI-guided stereotactic navigation for neurosurgical procedures; includes workstation laptop with software, SMARTGrid planning grid, and SMARTFrame trajectory frame. Physician uses system in MRI suite to plan and monitor instrument placement; system integrates with 1.5T/3.0T MRI scanners. Software processes DICOM images to display 3D renderings of patient anatomy and disposable hardware; enables image fusion/registration, volume-of-interest definition, and trajectory planning. Hardware provides physical guidance and fixation for tools; MRI-visible fiducials allow real-time trajectory feedback via subsequent MR imaging. System assists in precise burr hole positioning and instrument insertion; benefits include enhanced visualization and accuracy during minimally invasive neurosurgery.

## Clinical Evidence

Bench testing only. Verification testing included integrated system tests with live MRI scanner (n=9 placements, accuracy ≤1.0 mm), regression testing against predicate output, functional verification, and automated UI testing. All tests met acceptance criteria.

## Technological Characteristics

System comprises workstation laptop, SMARTGrid, and SMARTFrame. Materials include Ultem and PEEK. Operates in 1.5T/3.0T MRI environments. Connectivity via DICOM. Hardware is single-use, sterile (EO 10^-6 SAL). Software v2.0 features 3D visualization, image fusion, and volumetric rendering. Targeting accuracy ±1.5mm @ ≤125mm.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- ClearPoint System ([K171257](/device/K171257.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 2, 2018

MRI Interventions, Inc. % John Smith Partner Hogan Lovells US LLP 555 Thirteenth St. NW Washington, District of Columbia 20004

Re: K181195

Trade/Device Name: ClearPoint System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: ORR, HAW Dated: October 5, 2018 Received: October 5, 2018

Dear John Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Matthew C. Krueger -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181195

Device Name ClearPoint System

#### Indications for Use (Describe)

The ClearPoint System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners.

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) SUMMARY MRI Intervention. Inc.'s ClearPoint Device

#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Hogan Lovells, US LLP 555 Thirteenth Street. NW Washington, DC 20004 Phone: 202-637-5600 Fax: 202-637-5910

Contact Person: John J. Smith. M.D., J.D.

Date Prepared: October 5, 2018

#### Name of Device and Name/Address of Sponsor

ClearPoint System™ MRI Interventions, Inc. 5 Musick Irvine, CA 92618

Common or Usual Name: Neurological Stereotaxic Instrument

Classification: 21 C.F.R. §882.4560

Product Code: ORR, HAW

Predicate Device: MRI Interventions, Inc. ClearPoint System (K171257)

#### Indications for Use

The ClearPoint System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners.

#### Device Description

The ClearPoint System is comprised of a workstation laptop with software, the SMARTGrid™ MRI-Guided Planning Grid, the SMARTFrame™ MRI-Guided Trajectory Frame, the SMARTFrame™ Accessory Kit and the SMARTFrame™ Thumbwheel Extension.

The SMARTGrid and associated Marking Tool are designed to assist the physician to precisely position the entry hole as called out in the traiectory planning software. The SMARTFrame is an Adjustable Trajectory Frame (ATF) that provides the quidance and fixation for neurosurqical tools. The MRI visible fluids of the Targeting Cannula along with the fiducial markers in the base of the frame allows for trajectory feedback when the physician views the MRI images, makes changes and confirms with subsequent MR images.

The ClearPoint System can be used with any MRI-compatible head fixation frame to immobilize the patient's head with respect to the scanner table, as well as with any imaging coil(s) (supplied by scanner manufacturers) that meet the physician's desired imaging quality. MRI Interventions also supplies an optional head fixation frame that can be used with the ClearPoint System.

{4}------------------------------------------------

The ClearPoint Workstation includes the following:

- 1. ClearPoint Workstation Software (for trajectory planning and monitoring)
- 2. Laptop Computer

The hardware components of the current ClearPoint System are the SMARTFrame and Accessories. They are all single use devices that are provided sterile. They include the following:

- 1. SMARTGrid MRI Planning Grid (interacts with the software to determine the desired location of the burr hole)
	- a. Marking Grid
	- b. Marking Tool
- 2. SMARTFrame Pack (SMARTFrame or SMARTFrame XG)
	- a. SMARTFrame ("ATF") with Base
	- Centering Device and Wharen Centering Guide ﻓ
	- c. Dock
	- d. Device Lock (2 different diameters)
	- e. Screwdriver
	- f. Roll Lock Screw and Washer
- 3. Rescue Screws (Extra Titanium Screws)
- 4. Thumbwheel Extension
- 5. Accessory Kit
	- a. Peel-away Sheath
	- b. Stylet
	- c. Lancet
	- d. Depth Stop
	- e. Ruler
- 6. Scalp Mount Base
- 7. Guide Tubes and Device Guide Packs (Guide Cannulas)

In addition, the ClearPoint System is used with the following separately cleared or Class I, 510(k)exempt products:

- SmartTip MRI Hand Drill and Drill Bit Kit
- MRI Neuro Procedure Drape, with Marker Pen and Cover ●
- SmartFrame MR Fiducial .
- Camera Fiberscope Accessory Kit ●

#### Purpose of 510(k)

The primary purpose of this 510(k) is to modify the ClearPoint System software to version 2.0 by (1) changing the underlying software visualization toolkit used to display reformatted and threedimensional image renderings; (2) introducing the ability to provide three-dimensional model renderings of the MRI Interventions SMARTGrid and SMARTFrame disposable hardware in combination with 3D volumetric renderings of images that are loaded into the software; (3) changing the DICOM toolkit used to receive images from the MR scanner and loading images from media; (4) adding the ability to fuse/register images together that were not acquired in the same frame of reference; (5) introducing the ability to define three-dimensional volumes of interest within image sets that are loaded into the software; and (6) modifying the software workflow to include new software features and make the existing workflow outlined in the predicate device more flexible.

The table below presents a detailed comparison of the technological characteristics of the updated ClearPoint System and those of the predicate device.

{5}------------------------------------------------

# Summary of the Technological Characteristics of the Device Compared to the Predicate Device

|                                                          | ClearPoint System (Predicate): K171257                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | ClearPoint System (Subject)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification                                           | 21 C.F.R. § 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 21 C.F.R. § 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code                                             | ORR, HAW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | ORR, HAW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Intended Use                                             | The ClearPoint System is intended to provide stereotactic guidance and operation of instruments or devices during the planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter, and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners. | The ClearPoint System is intended to provide stereotactic guidance and operation of instruments or devices during the planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter, and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners. |
| Environment                                              | MRI Suite<br>EO 10 -6 SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | MRI Suite<br>EO 10 -6 SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterilization                                            | MRI Planning Grid & Marking Tool                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | MRI Planning Grid & Marking Tool                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| SmartGrid Pack                                           | SmartFrame XG, Scalp Mount Base, Bone<br>Screws, Stand-Off Pins, Screwdriver,<br>Centering Tool, Wharen Centering Guide<br>(packaged and sold separately), Dock and<br>Lock, Roll Lock Screw with Washer, Rescue<br>Screws (packaged separately)                                                                                                                                                                                                                                                                                                                        | SmartFrame XG, Scalp Mount Base, Bone<br>Screws, Stand-Off Pins, Screwdriver,<br>Centering Tool, Wharen Centering Guide<br>(packaged and sold separately), Dock and<br>Lock, Roll Lock Screw with Washer,<br>Rescue Screws (packaged separately)                                                                                                                                                                                                                                                                                                                        |
| SmartFrame<br>Pack                                       | Thumbwheel Extension (Light Hand<br>Controller)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Thumbwheel Extension (Light Hand<br>Controller)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Hand<br>Controller                                       | Peel-away Sheath, Lancet, Stylet, Depth<br>Stop, Ruler                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Peel-away Sheath, Lancet, Stylet, Depth<br>Stop, Ruler                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Accessory Kit                                            | 4.5mm & 6.0mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 4.5mm & 6.0mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Drill Guides                                             | 0.0825"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 0.0825"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Targeting<br>Cannula ID                                  | Ultem and PEEK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Ultem and PEEK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Targeting<br>Cannula<br>Material                         | 0.0938, 0.141, 0.191, 0.250"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 0.0938, 0.141, 0.191, 0.250"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Guide Tube /<br>Device Guide /<br>Drill Guide ID         | Ultem                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Ultem                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Guide Tube/<br>Device Guide /<br>Drill Guide<br>Material | Sterile, Sealed Tray, Inside<br>Sterile Tyvek Pouch<br>(Wharen Centering Guide and Drill and<br>Device Guides are Sterile in a double Tyvek<br>Pouch Without a Tray; Wharen Centering<br>Guide Packaging includes PVC)                                                                                                                                                                                                                                                                                                                                                  | Sterile, Sealed Tray, Inside<br>Sterile Tyvek Pouch<br>(Wharen Centering Guide and Drill and<br>Device Guides are Sterile in a double<br>Tyvek Pouch Without a Tray; Wharen<br>Centering Guide Packaging includes PVC)                                                                                                                                                                                                                                                                                                                                                  |
| Packaging                                                | ±1.5mm @ ≤125mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | ±1.5mm @ ≤125mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Targeting<br>Accuracy                                    | 1.6.2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Targeting<br>Software                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{6}------------------------------------------------

## Performance Data

The company performed verification testing for all software changes in the ClearPoint 2.0 software release. Specifically, the following tests were performed to support this 510(k) notice:

| Test                      | Test Method Summary                                                                                                                                                                            | Results                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Integrated<br>System Test | This test utilized a live scanner<br>connection to verify correct<br>interoperation with scanner and<br>disposable hardware kit.                                                               | Pass. The design has been verified to satisfy all<br>acceptance criteria. All functional requirements<br>necessitating interfacing with a live MR scanner were<br>satisfied and validated using the software. Nine device<br>placements were evaluated and had accuracy of 1.0 mm or<br>less.                                                                                                                                                 |
| Regression<br>Test        | This test used data from the<br>predicate device and included<br>head-to-head comparison with<br>output from the predicate device<br>to support the claim of<br>substantive equivalence.       | Pass. The design has been verified to satisfy all<br>acceptance criteria. The following underlying system<br>components still produce output that is consistent with the<br>previously released software version: image<br>display/visualization, image transfer, network<br>communication verification, segmentation library output,<br>scan plane parameter computations, hardware adjustment<br>computations and low-level math utilities. |
| Verification<br>test      | This test used data from the<br>predicate device to execute all<br>areas of functionality of the<br>workstation.                                                                               | Pass. The design has been verified to satisfy all<br>acceptance criteria. All functional areas within the software<br>behaved as described in the requirements.                                                                                                                                                                                                                                                                               |
| Automated<br>UI test      | This is an automated test that<br>simulates user input to the<br>ClearPoint software to test<br>functionality that is otherwise<br>difficult or time-consuming to<br>execute as a manual test. | Pass. The design has been verified to satisfy all<br>acceptance criteria. The ClearPoint 2.0 Software has<br>passed the Automated UI test protocol.                                                                                                                                                                                                                                                                                           |

The ClearPoint System hardware is identical to that of the predicate; no additional tests for biocompatibility or sterility were provided in support of this 510(k) notice. Biocompatibility data in accordance with ISO 10993 and Sterilization Validation using the overkill method of ISO 11135-1:2007 utilizing the half cycle method was provided in support of the predicate device.

# Conclusions

The ClearPoint System is as safe and effective as the predicate ClearPoint System. The ClearPoint System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the ClearPoint System and its predicate devices raise no new issues of safety or effectiveness. Thus, the ClearPoint System is substantially equivalent.

---

**Source:** [https://fda.innolitics.com/device/K181195](https://fda.innolitics.com/device/K181195)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
