This device is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions. Do not use for any purpose other than its intended use.
Device Story
Single-use aspiration needle for tissue biopsy; used in surgical, endoscopy, or bronchoscopy suites. Device consists of handle, sheath, needle, stylet, and removable stopper. Physician inserts device through bronchoscope working channel to target tissue; needle deployed manually by advancing slider. Stopper limits needle deployment depth by 10 mm. Provides tissue samples for diagnostic evaluation. Benefits patient by enabling minimally invasive intrapulmonary tissue collection.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence supported by bench testing, including stopper width, removal force, bubble leak (ASTM F2096-11), and accelerated aging (ASTM F1980-16).
Technological Characteristics
Manual aspiration needle; materials include stainless steel, PTFE, PeBax, and Nitinol. Dimensions: 1.5 mm max OD, 115 cm length, 21G needle. Deployment: 20 mm max extension (10 mm with stopper). Sterilization: Ethylene oxide. Single-use.
Indications for Use
Indicated for the collection of tissue from intrapulmonary regions via a compatible bronchoscope. Intended for use by or under the supervision of a physician.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2018
Olympus Surgical Technologies America Mary Anne Patella Senior Specialist, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772-2104
Re: K181193
Trade/Device Name: PeriView FLEX Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: KTI Dated: June 1, 2018 Received: June 4, 2018
Dear Ms. Patella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181193
Device Name PeriView FLEX
#### Indications for Use (Describe)
This device is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions. Do not use for any purpose other than its intended use.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### K181193 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. PeriView FLEX
#### General Information
| Manufacturer: | Olympus Surgical Technologies<br>America<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 508-804-2600<br>Fax: 508-804-2624 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Mary Anne Patella<br>Senior Specialist, Regulatory Affairs<br>508-804-2771<br>Maryanne.patella@olympus-osta.com |
| Date Prepared: | May 3, 2018 |
| <b>Device Description</b> | |
| Classification Name: | Bronchoscope (flexible or rigid) and<br>accessories |
| CFR Citation Number: | 21 CFR 874.4680 |
| Product Code: | KTI |
| Classification: | Class II |
| Review Panel: | Ear Nose & Throat |
| Trade Name: | PeriView FLEX |
| Generic/Common Name: | Aspiration Needle |
| <b>Predicate Device</b> | |
| K171232 | Gyrus ACMI, Inc. PeriView FLEX |
#### Comparison to Predicate Device:
The PeriView FLEX has been compared to the predicate PeriView FLEX with respect to intended use, design and fundamental scientific technology. The comparisons and summary of testing results presented in this Special 510(k)
{5}------------------------------------------------
Notification show this device to be substantially equivalent to the predicate PeriView FLEX and raises no new concerns of safety or effectiveness.
Like the predicate PeriView FLEX. the modified PeriView FLEX is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions.
### Product Description
The PeriView FLEX needle is a single use tissue biopsy device that consists of five parts: Handle, Stopper, Sheath, Needle, and, Stylet. The sheath and needle (together termed the insertion section) are attached to the handle. The Stopper is removable and is attached between the Handle and the Needle Slider. The removable Stylet runs the full length of the device and is located within the inner lumen of the needle.
#### Technological Characteristics
PeriView FLEX, with the Stopper, is identical in design to its predicate, PeriView FLEX, except for the added Stopper. They both operate similarly to obtain a tissue sample by inserting the insertion section (sheath and needle) of the device into a bronchoscope's working channel or guide sheath. It is then pushed forward to position the device to the tissue target. The Needle is thenadvanced out of the sheath and into the tissue by pushing the Needle Slider forward while holding the handle. When the Needle Slider is pushed forward (distally), the Needle can be deployed up to 20 mm from the sheath's distal tip. With the Stopper attached between the Handle and the Needle Slider, it blocks the forward movement of the Needle Slider, reducing needle deployment by 10mm. The Stopper is removable/replaceable by the user as desired.
## Material
No changes were made to patient contacting material or packaging to the modified PeriView FLEX since the predicate system was cleared in K171232.
#### Intended Uses
The intended use of the modified device, as described in its labeling, has not changed as a result of the addition of the stopper.
This device is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions. Do not use for any purpose other than its intended use.
#### Summary of Sterilization and Shelf Life Discussion
{6}------------------------------------------------
Like the predicate PeriView FLEX, (K171232), the modified PeriView FLEX will be distributed in a sterile state and is intended for single patient use only. The sterilization method used continues to be ethylene oxide.
#### Summary of All Performance Testing (no clinical testing was conducted)
| Description | Specification/objective |
|-------------------|--------------------------|
| Stopper Width | Meet width specification |
| Removal Force | Meet specification |
| Bubble Leak | ASTM F2096-11 |
| Accelerated Aging | ASTM F1980-16 |
All performance testing passed or met prescribed acceptance criteria
#### Substantial Equivalence
The modified PeriView FLEX has the same intended use, scientific technology and similar design as its predicate PeriView FLEX device. The predicate and modified PeriView FLEX devices are both bronchial aspiration needles consisting of a handle, sheath, needle, and stylet. The predicate and modified devices were shown to perform substantially equivalent in bench testing. There were no new issues of safety or effectiveness with the proposed device. Please see the following substantial equivalence comparison table.
| | Predicate Device<br>(K171232) | Proposed Device |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name/<br>Characteristics | PeriView FLEX | PeriView FLEX |
| Indications for Use | This device is intended to be used<br>through a compatible bronchoscope for<br>the collection of tissue from the<br>intrapulmonary regions. Do not use for<br>any purpose other than its intended use. | This device is intended to be used through<br>a compatible bronchoscope for the<br>collection of tissue from the<br>intrapulmonary regions. Do not use for<br>any purpose other than its intended use. |
| Anatomical Site | Lung | Identical |
| Use Conditions | Surgical suite, endoscopy or<br>bronchoscopy suite, used with a<br>bronchoscope | Identical |
| Intended User | By or under the supervision of a<br>physician | Identical |
| Mechanics of Action | Manual | Identical |
| Mode of Action | Single/multiple puncture and aspirate | Identical |
| General design | Handle, Sheath, Needle, Stylet | Similar<br>A Stopper will be added to the shaft on the<br>piston of the needle assembly. |
| Patient Contacting | Stainless Steel, PTFE, PeBax ™, Nitinol | Identical |
{7}------------------------------------------------
| | Predicate Device<br>(K171232) | Proposed Device |
|-------------------------------------------|-------------------------------|-----------------|
| Device Name/<br>Characteristics | PeriView FLEX | PeriView FLEX |
| Materials | | |
| Biocompatible | Yes | Yes |
| Sterilization | Ethylene Oxide | Identical |
| Single Use Only | Yes | Identical |
| Maximum Working<br>OD (mm) | 1.5 | Identical |
| Catheter Length<br>(cm) | 115 | Identical |
| Needle Gauge | 21G | Identical |
| Maximum Needle<br>Extended Length<br>(mm) | 20 | Identical |
| Stylet OD (mm) | 0.47 | Identical |
#### Conclusion:
In summary, the PeriView FLEX is substantially equivalent to the predicate device and presents no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.