K181177 · Philips Medical Systems Nederland B.V. · LLZ · May 31, 2018 · Radiology
Device Facts
Record ID
K181177
Device Name
Interventional Workspot
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LLZ · Radiology
Decision Date
May 31, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The Interventional Workspot has the following medical purpose: · import, export and storage of digital clinical images, · manage the patient information associated with those images.
Device Story
Interventional Workspot is a software platform hosting Interventional Tools for use in interventional suites or operating rooms. It acts as a data handler for X-ray, CT, and MR images acquired on compatible Philips interventional systems. The device provides common functionalities including image import, export, storage, and patient information management. It is operated by clinicians with basic understanding of medical software. The platform supports physicians during interventional procedures by hosting tools that process and display clinical images. It does not perform clinical image analysis itself but facilitates the workflow for tools that do. The device is installed on specific manufacturer-provided PC hardware.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical performance testing, including software verification and validation against predetermined specifications, risk management, and compliance with FDA-recognized consensus standards (IEC 62304, IEC 62366-1, ISO 14971, NEMA PS 3.1-3.20).
Technological Characteristics
Software platform running on PC-based hardware (e.g., Dell T3600, Haswell, Radysis). Connectivity via DICOM. Features include data handling, whitelisting, and malware protection. Complies with IEC 62304 for software lifecycle, ISO 14971 for risk management, and NEMA PS 3.1-3.20 for DICOM interoperability.
Indications for Use
Indicated for import, export, and storage of digital clinical images and management of associated patient information. No specific patient population defined as it is a hosting platform.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
May 31, 2018
Philips Medical Systems Nederland BV % Chandrika Srinivasan Senior Regulatory Affairs Specialist Veenpluis 4-6 5684PC Best, THE NETHERLANDS
Re: K181177
Trade/Device Name: Interventional Workspot Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, OWB Dated: April 30, 2018 Received: May 2, 2018
Dear Chandrika Srinivasan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Chandrika Srinivasan
K181177
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hole 2. Molds
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181177
Device Name Interventional Workspot
Indications for Use (Describe)
The Interventional Workspot has the following medical purpose:
· import, export and storage of digital clinical images,
· manage the patient information associated with those images.
Patient Population:
Not applicable because Interventional Workspot is only a hosting platform.
Operating Profile:
The operator of the Interventional Workspot has basic understanding of the operating principle of medical computer software.
Clinical Environment:
The software can be used in the control room and in the exam room of an interventional suite and/or operating room.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | May 24, 2018 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277 |
| Primary Contact<br>Person: | Chandrika Srinivasan<br>Regulatory Affairs Officer<br>Phone: +9180 41894132<br>E-mail: chandrika.s@philips.com |
| Secondary Contact<br>Person: | Marta Walker<br>Head of Regulatory Affairs IGT Systems<br>Phone: +31 631978546<br>E-mail: marta.walker@philips.com |
| Device: | Trade Name: Interventional Workspot<br>Release Number: Version 1.4.5.1<br>Classification Name: Picture archiving and communications system<br>Classification Regulation: 21 CFR, Part 892.2050<br>Classification Panel: Radiology<br>Device Class: Class II<br>Primary Product Code: LLZ (System, Image Processing, Radiological)<br>Secondary Product Code: OWB (Interventional Fluoroscopic X-Ray system) |
| Predicate Device: | Trade Name: Interventional Workspot Release 1<br>Manufacturer: Philips Medical Systems Nederland B.V<br>510(k) Clearance: K121296 (Jan 02, 2013)<br>Classification Name: Picture archiving and communications system<br>Classification Regulation: 21 CFR, Part 892.2050<br>Classification Panel: Radiology<br>Device Class: Class II<br>Primary Product Code: LLZ (System, Image Processing, Radiological)<br>Secondary Product Code: OWB (Interventional Fluoroscopic X-Ray system) |
{4}------------------------------------------------
#### The Interventional Workspot is a software platform to host Interventional Tools. Device description: It provides common functionalities (e.g. import / export and data handling functions) that are required by the Interventional Tools to support the physician with performing the interventional procedure.
The images made with the Interventional Tools use X-ray imaging and CT/MR imaging as the source data, acquired on compatible equipment. The Interventional Workspot acts as a data handler for Interventional tools.
Interventional Workspot can be used with compatible Philips interventional Xray systems that is equipped with appropriate options. Interventional Workspot is only intended to be used with the PC hardware configuration on which it is initially installed by the manufacturer.
Indications for Use: Interventional Workspot is provided as accessory to the Philips Interventional X-ray system, has the following indications for use:
# Indications for use/ Medical Purpose:
The Interventional Workspot has the following medical purpose:
- · import, export and storage of digital clinical images,
- · manage the patient information associated with those images.
# Patient Population:
Not applicable because Interventional Workspot is only a hosting platform.
# Operating Profile:
The operator of the Interventional Workspot has basic understanding of the operating principle of medical computer software.
# Clinical Environment:
The software can be used in the control room and in the exam room of an interventional suite and/or operating room.
| Interventional Workspot<br>R1.0 (predicate device)<br>(K121296) | Interventional Workspot<br>R1.4.5.1<br>(proposed device) | Comparison |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| The Interventional<br>Workspot has the<br>following medical purpose:<br>• import, export and storage<br>of<br>digital clinical images,<br>• manage the patient<br>Information associated with<br>those images | The Interventional Workspot<br>has the<br>following medical purpose:<br>• import, export and storage<br>of<br>digital clinical images,<br>• manage the patient<br>information<br>associated with those images | Same |
Table 1: Indications for Use comparison with predicate device
{5}------------------------------------------------
The Indications for use is same for both the predicate and proposed device.
Technological Interventional Workspot software is executed on a PC based hardware platform. characteristics: There is no difference is Technological characteristics between the Interventional Workspot Rel. 1.0 and the proposed Interventional Workspot IW1.4.5.1.
| Parameter | Interventional Workspot<br>R1<br>(K121296) 21 CFR<br>sec.892.2050 | Interventional<br>Workspot R1.4.5.1<br>(proposed device) 21<br>CFR sec.892.2050 | Comparison |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics | The interventional<br>workspot is a software<br>platform to host<br>Interventional Tools. It<br>provides common<br>functionalities (e.g. import<br>/ export and data handling<br>functions) that are required<br>by the Interventional Tools<br>to<br>support the physician with<br>performing the<br>interventional procedure. | The interventional<br>workspot is a software<br>platform to host<br>Interventional Tools.<br>It provides common<br>functionalities (e.g.<br>import / export and<br>data handling<br>functions) that are<br>required by the<br>Interventional Tools to<br>support the physician<br>with performing the<br>interventional<br>procedure. | Same |
| Operating<br>System | Windows XP professional | Windows 7 | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design.<br>function and<br>application. |
| DICOM | DICOM compatible | DICOM compatible | Same |
| Compatibility<br>with currently<br>marketed<br>Philips<br>Interventional<br>X-ray systems. | Compatible with currently<br>marketed Philips<br>Interventional X-ray<br>systems- Allura Series. | Compatible with<br>currently marketed<br>Philips Interventional<br>X-ray systems- Allura<br>Series & Azurion<br>series. | Substantially<br>equivalent.<br>Note: for ease of<br>review, the<br>differences have<br>been underlined.<br>SE analysis:<br>Both the<br>predicate and<br>proposed device<br>are accessories<br>to the Philips |
| | | Interventional<br>X-ray systems. | |
| Hardware<br>obsolescences | - | Changes to support<br>different hardware<br>such as: PC Dell<br>T3600, Haswell PC,<br>Radysis 7 &8 | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design,<br>function and<br>application. |
| Bug fixes | - | Minor bug fixes | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design,<br>function and<br>application. |
| OS & network<br>security<br>features | - | Changes to support<br>network security<br>features like<br>whitelisting, and<br>technology to protect<br>malware. | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design,<br>function and<br>application. |
| Feature<br>enhancement | - | Changes to support<br>resizing larger data<br>sets;<br>Support for direct<br>upgrade from older<br>versions, software<br>installer | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design,<br>function and<br>application. |
| Code<br>refactoring | - | Code refactoring to<br>improve software<br>maintenance. | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design,<br>function and<br>application. |
| Labeling<br>changes | - | Adapting the labeling<br>to include the minor<br>changes implemented. | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design, |
| | | function and<br>application. | |
| Support for<br>new<br>Interventional<br>tools | - | Hosting platform<br>when new<br>Interventional tools<br>are released. | Substantially<br>equivalent.<br>SE analysis:<br>The predicate<br>and proposed<br>device are same<br>in design,<br>function and<br>application. |
Table 2: Technological characteristics comparison-Classification & Design attributes and functionality
{6}------------------------------------------------
{7}------------------------------------------------
Summary of Non-Non-clinical performance testing has been performed on the modified Clinical Performance Interventional Workspot R1.4.5.1 and demonstrates compliance with the Data: following FDA recognized consensus standards and FDA guidance documents:
- IEC 62304:2006 Medical device software Software life cycle processes. ● FDA/CDRH recognition number 13-32;
- IEC 62366-1 Medical devices - Part 1: Application of usability engineering to medical devices (Edition 1.0 including corrigendum, 2015). FDA/CDRH recognition number 5-114;
- ISO 14971 Medical devices Application of risk management to medical . devices (Edition 2.0, corrected version, 2007). FDA/CDRH recognition number 5-40;
- ISO15223-1 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements (Second edition, 2012). FDA/CDRH recognition number 5-90:
- NEMA PS 3.1 - 3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2016). FDA/CDRH recognition number 12-300
- . "Guidance for Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005 (document number 337);
- . "Guidance for Industry and FDA Staff - Applying Human Factors and Usability Engineering to Medical Devices", February 3, 2016 (document number 1757);
- . "Guidance for Industry and FDA Staff - Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", October 2, 2014 (document number 1825);
- "Guidance for Industry and FDA Staff The 510(k) Program: ● Evaluating Substantial Equivalence in Premarket Notifications [510(k)] ", July 28, 2014 (document number 1766).
- . Guidance for Industry, FDA Reviewers and Compliance on "Guidance for Off-the-Shelf Software Use in Medical Devices, September 1999"
Software verification testing has been performed to verify the modifications as per predetermined System Requirements Specification and acceptance critera. The verification tests and acceptance criteria were identified based on Risk Assessent. Verification tests were performed to verify safety risk control measures from the Detailed Risk Management Matrix and to verify the Privacy and Security
{8}------------------------------------------------
| | requirements for <b>Interventional Workspot 1.4.5.1</b> . have been implemented.<br>Verification results demonstrated that all executed tests were passed. |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Non clinical software validation testing has been performed to validate the<br>intended use, claims, user needs, service user needs, effectiveness of safety<br>measures and Instructions for use. |
| | Software validation testing has been performed to validate that <b>Interventional<br/>Workspot</b> conforms to its intended use, claims and user needs. |
| | All these tests were used to support substantial equivalence of the subject device<br>and demonstrate that <b>Interventional Workspot</b> :<br>complies with the aforementioned international and FDA-recognized<br>consensus standards and FDA guidance documents, andmeets the acceptance criteria and is adequate for its intended use. |
| | Based on the information provided above, <b>Interventional Workspot Release<br/>1.4.5.1</b> is substantially equivalent to the predicate device <i>Interventional Workspot<br/>Release 1.0 (K121296)</i> in terms of safety and effectiveness. |
| Summary of Clinical<br>Performance Data: | Clinical images are not necessary to establish substantial equivalence based on<br>the modifications to the predicate device (note that the <i>Interventional Workspot<br/>software</i> is strongly based on the predicate). Non-clinical performance data<br>provides sufficient evidence that the subject device works as intended. |
| Substantial<br>Equivalence<br>Conclusion: | <b>Interventional Workspot Release 1.4.5.1</b> is substantially equivalent to the<br>predicate device <i>Interventional Workspot Release 1.0 (K121296)</i> in terms of<br>indications for use, technological characteristics and safety and effectiveness.<br><br>The modifications made in the proposed device <b>Interventional Workspot<br/>Release 1.4.5.1</b> are within the controls and predetermined specifications.<br>Additionally, non-clinical performance tests provided in this 510(k) premarket<br>notification demonstrated substantial equivalence to the predicate device and<br>ensured that the modifications are implemented successfully. Verification and<br>Validation tests were conducted to verify the modifications listed and<br>conformance to international and FDA-recognized consensus standards and<br>guidance documents were provided. These tests demonstrate that <b>Interventional<br/>Workspot Release 1.4.5.1</b> is substantially equivalent to the predicate device and<br>is as safe and effective as its predicate device and does not raise any new safety<br>and/or effectiveness concerns. |