← Product Code [GEH](/productcode/GEH) · K181153

# ICEfx Cryoablation System (K181153)

_Galil Medical, Inc. · GEH · May 15, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181153

## Device Facts

- **Applicant:** Galil Medical, Inc.
- **Product Code:** [GEH](/productcode/GEH.md)
- **Decision Date:** May 15, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ICEfx Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, oncology, proctology and urology. This System is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions) by the application of extremely cold temperatures. The ICEfx Cryoablation System has the following specific indications: Urology Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH) Oncology Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention Dermatology Ablation or freezing of skin cancers and other cutaneous disorders: Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin Gynecology Ablation of malignant neoplasia or benign dysplasia of the female genitalia General surgery Palliation of tumors of the rectum, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas ENT Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth Thoracic surgery (with the exception of cardiac tissue) Proctology Ablation of benign or malignant growths of the anus or rectum

## Device Story

Mobile, computer-controlled cryoablation system; utilizes Joule-Thomson effect via high-pressure argon gas circulating through closed-tip needles to induce tissue freezing. Features Galil Medical CX technology; heating element inside needle enables active electrical tissue thawing. Operated by clinicians via touchscreen interface to monitor and control procedure. System provides localized tissue destruction for various surgical specialties. Benefits include minimally invasive tissue ablation. ICEfx is a smaller, lighter version of the predicate with fewer needle ports; does not support helium thaw.

## Clinical Evidence

Bench testing only. Verification and validation included system, electrical, mechanical, labeling, and software testing. Testing confirmed design, functionality, and performance met specifications. No clinical data presented.

## Technological Characteristics

Mobile console; Joule-Thomson cryoablation; high-pressure argon gas; closed-tip needles; electrical heating element (CX technology) for thawing; touchscreen interface; computer-controlled; electrical safety and EMC/EMI compliance.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- Visual-ICE Cryoablation System ([K113860](/device/K113860.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 15, 2018

Galil Medical Inc. Ms. Rachel Owens Sr. Regulatory Affairs Specialist 4364 Round Lake Road Arden Hills, Minnesota 55112

Re: K181153

Trade/Device Name: ICEfx Cryoablation System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: April 30, 2018 Received: May 1, 2018

Dear Ms. Owens:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application(PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

# Page 2 - Ms. Rachel Owens

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Jennifer R. Stevenson For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181153

Device Name ICEfx™ Cryoablation System

#### Indications for Use (Describe)

The ICEfx Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, proctology and urology. This System is destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions) by the application of extremely cold temperatures.

The ICEfx Cryoablation System has the following specific indications:

· Urology Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasja (BPH)

· Oncology Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention

· Dermatology Ablation or freezing of skin cancers and other cutaneous disorders; Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin

- · Gynecology Ablation of malignant neoplasia or benign dysplasia of the female genitalia
· General surgery Palliation of tumors of the rectum, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas

· ENT Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth

· Thoracic surgery (with the exception of cardiac tissue)

· Proctology Ablation of benign or malignant growths of the anus or rectum

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### K181153 510(k) SUMMARY ICEfx™ Cryoablation System

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

| Submitter:                  | Galil Medical Inc.<br>4364 Round Lake Road<br>Arden Hills, MN 55112     |
|-----------------------------|-------------------------------------------------------------------------|
| Company Contact Person:     | Rachel Owens<br>Sr. Regulatory Affairs Specialist<br>Galil Medical Inc. |
| Phone:                      | 651-287-5024                                                            |
| Fax:                        | 651-203-7392                                                            |
| Email:                      | rachel.owens@btgplc.com                                                 |
| Alternate Contact Person:   | Rebeka Stoltman<br>Director, Regulatory Affairs<br>Galil Medical Inc.   |
| Phone:                      | 651-287-5020                                                            |
| Fax:                        | 877-510-7757                                                            |
| Email:                      | Rebeka.stoltman@btgplc.com                                              |
| Device Name:                | ICEfx™ Cryoablation System                                              |
| Device Classification Name: | Cryosurgical unit and accessories                                       |
| Regulation Number:          | 21 CFR 878.4350                                                         |
| Product Code:               | GEH                                                                     |
| Predicate Device 510(k):    | Visual-ICE Cryoablation System (K113860)                                |
| Date of Preparation:        | April 30, 2018                                                          |

### Device Description:

The ICEfx™ Cryoablation System is a mobile system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The ICEfx System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by the use of Galil Medical CX technology in which a heating element inside the cryoablation needle can be energized to cause thawing.

The table below provides a summary comparison of the submitted device compared to the predicate device.

| Description of Submitted<br>Device:<br>ICEfx Cryoablation System | Comments related to Predicate:<br>Visual-ICE Cryoablation System (K113860) |
|------------------------------------------------------------------|----------------------------------------------------------------------------|
| Design and Construction                                          |                                                                            |
| Console                                                          | Similar to predicate; ICEfx is smaller and lighter weight                  |
| Needle ports                                                     | ICEfx contains fewer needle ports                                          |
| Channel Lock                                                     | Same as predicate                                                          |

{4}------------------------------------------------

| Description of Submitted<br>Device:<br>ICEfx Cryoablation System | Comments related to Predicate:<br>Visual-ICE Cryoablation System (K113860) |
|------------------------------------------------------------------|----------------------------------------------------------------------------|
| User interface                                                   | Similar to predicate; touchscreen with equivalent<br>function and features |
| Needle Interface                                                 | Same as predicate                                                          |
| Performance and Function                                         |                                                                            |
| Principles of Operation:<br>Freezing / Thawing                   | Same as predicate; ICEfx does not support the option<br>for helium thaw    |
| CX Technology for Electrical<br>Thawing                          | Same as predicate                                                          |
| Needle Compatibility                                             | Same as predicate; ICEfx supports fewer needles                            |

In summary, the submitted ICEfx Cryoablation System has the same technology and principle of operation as the predicate device.

# Indications for Use / Intended Use:

The ICEfx Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, oncology, proctology and urology. This System is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, skin lesions) by the application of extremely cold temperatures.

The ICEfx Cryoablation System has the following specific indications:

- Urology Ablation of prostate tissue in cases of prostate cancer and Benign Prostate ● Hyperplasia (BPH)
- . Oncology Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention
- . Dermatology Ablation or freezing of skin cancers and other cutaneous disorders: Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin
- Gynecology Ablation of malignant neoplasia or benign dysplasia of the female genitalia ●
- General surgery Palliation of tumors of the rectum, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas
- ENT Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth
- Thoracic surgery (with the exception of cardiac tissue)
- . Proctology Ablation of benign or malignant growths of the anus or rectum

The intended use and indications for use are the identical to the Visual-ICE predicate.

## Summary of Performance Data and Substantial Equivalence:

A full battery of verification and validation testing was conducted on the ICEfx Cryoablation System to ensure that the design, functionality, and performance met all the specified requirements and that the features of the system satisfy its intended use. Testing was conducted according to protocols based on international standards and in-house requirements. Verification testing included system testing, electrical testing, mechanical

{5}------------------------------------------------

testing, labeling testing, and software testing. System testing assessed whether the functional requirements of the system as a whole were satisfied by the design. Mechanical testing evaluated the mechanical robustness of the system and sub-systems, functional testing of the gas system, and tests of mechanical safety requirements. Electrical testing assessed the functional aspects of the circuit assemblies, tests of electromagnetic compatibility and immunity (EMC/EMI), and tests of electrical safety requirements. Labeling verification evaluated user manual and labeling accuracy with respect to design requirements and risk mitigations. Software testing exercised individual units of software as well as tests of the functionality of the entire software package. Validation testing included design and usability testing. Test results demonstrated that the ICEfx Cryoablation System meets defined specifications, is substantially equivalent to the predicate device, and does not raise any new issues of safety and effectiveness for its intended use.

# Conclusion (Statement of Equivalence):

As the indications for use and fundamental scientific technology have not changed, the nonclinical performance data provided in this Special 510(k) Premarket Notification supports a determination that the ICEfx Cryoablation System is substantially equivalent to the legally marketed predicate device, with regard to performance, safety, and effectiveness for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K181153](https://fda.innolitics.com/device/K181153)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
