← Product Code LCM · K181135

# Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay (K181135)

_Immunalysis Corporation · LCM · Jan 24, 2019 · TX · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181135

## Device Facts

- **Applicant:** Immunalysis Corporation
- **Product Code:** LCM
- **Decision Date:** Jan 24, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** TX

## Intended Use

The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10 ng/mL in neat oral fluid collected with the Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of PCP in human oral fluid with clinical analyzers. This assay is calibrated against PCP. This in vitro diagnostic device is for prescription use only. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures. The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

## Device Story

Device is an in vitro diagnostic homogeneous enzyme immunoassay for PCP detection in human oral fluid. Input: oral fluid collected using Quantisal II dual-pad system; sample split into two vials for screening and confirmation. Operation: assay performed on clinical chemistry analyzers (e.g., Beckman Coulter AU480); utilizes PCP-specific antibodies to detect drug presence via competitive binding. Output: qualitative (positive/negative) or semi-quantitative (concentration estimate) results. Used in clinical laboratories by technicians/scientists. Output informs clinical decision-making by identifying potential PCP presence, necessitating confirmatory testing via GC-MS or LC-MS/MS. Benefits: enables rapid screening of oral fluid samples for drug abuse monitoring.

## Clinical Evidence

Bench testing only. Precision study (N=60 per concentration) over 15 days confirmed repeatability at cutoff ±25-100%. Method comparison study (N=80 clinical samples) against LC-MS/MS showed 100% agreement in both qualitative and semi-quantitative modes. Interference studies confirmed no impact from endogenous compounds, exogenous substances, or pH (3.0-11.0). Stability studies confirmed sample integrity in Quantisal II devices for 10 days at 30°C and 2 months at 2-8°C.

## Technological Characteristics

Homogeneous enzyme immunoassay; utilizes PCP-specific antibodies and drug conjugate reagents. Form factor: liquid reagents for use on automated clinical chemistry analyzers. Sample collection: Quantisal II dual-pad collector with preservative buffer. Calibration: 14-day stability. Connectivity: compatible with standard clinical laboratory analyzers (e.g., Beckman Coulter AU480).

## Predicate Devices

- RapidFRET Oral Fluid Assay for PCP, PCP Calibrator Set, PCP Control Set and RapidEase Oral Fluid Collector ([K122703](/device/K122703.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 24, 2019

Immunalysis Corporation Michelle Bodien Associate Director, Regulatory Affairs 829 Towne Center Drive Pomona, CA 91767

Re: k181135

Trade/Device Name: Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay Regulatory Class: Unclassified, 510(k) required Product Code: LCM Dated: December 13, 2018 Received: December 14, 2018

Dear Michelle Bodien:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn

(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) k181135

#### Device Name

Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay

#### Indications for Use (Describe)

The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10 ng/mL in neat oral fluid collected with the Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of PCP in human oral fluid with clinical analyzers. This assay is calibrated against PCP. This in vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)

| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image. The font is sans-serif and the letters are bolded.

#### 510(k) SUMMARY k181135

#### A. GENERAL INFORMATION

| Applicant Name:  | Immunalysis Corporation<br>829 Towne Center Drive<br>Pomona, CA 91767<br>Establishment # 2020952                  |
|------------------|-------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Michelle Bodien<br>Assoc. Director, Regulatory Affairs<br>Phone: (210)4058453<br>Email: michelle.bodien@alere.com |
| Date Prepared:   | January 22, 2019                                                                                                  |

#### B. DEVICE IDENTIFICATION

| Trade or Proprietary Names: | Immunalysis SEFRIA™ PCP Oral Fluid Enzyme Immunoassay |
|-----------------------------|-------------------------------------------------------|
|-----------------------------|-------------------------------------------------------|

# C. REGULATORY INFORMATION

| Device Classification Name: | Enzyme Immunoassay, Phencyclidine                                                                                    |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------|
| Product Codes:              | LCM                                                                                                                  |
| Regulatory Class:           | Class II                                                                                                             |
| Classification Regulation:  | Unclassified                                                                                                         |
| Panel:                      | Toxicology (91)                                                                                                      |
| Predicate Device:           | RapidFRET Oral Fluid Assay for PCP, PCP Calibrator Set, PCP Control Set and RapidEase Oral Fluid Collector [K122703] |

#### D. DEVICE DESCRIPTION

The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a sensitive in vitro diagnostic test to detect the presence of PCP in human oral fluid samples collected with the Quantisal II Oral Fluid Collection Device.

Quantisal II Oral Fluid Collection Device is a collection system comprised of a dual pad collector and transport vials. The dual pad collector is separated after collection of oral fluid from a subject's mouth enabling each specimen-saturated collection pad to be placed into its own transport vial. The split specimen (referred to as "A" and "B") allows for one sample to be tested in a screening assay and confirmed by a quantitative laboratory method (such as liquid chromatography tandem mass spectrometry [LC-MS/MS] and the second sample to be stored for secondary confirmation if needed.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters against a red background. The letters are bold and sans-serif. The background is a solid red color, and the word is centered in the image.

PCP was first synthesized in 1926 and later tested after World War II as a surgical anesthetic. Because of its adverse side effects, such as hallucinations, mania, delirium, and disorientation, it was shelved until the 1950s. The drug is easily synthesized by anyone with a basic knowledge of chemistry and has become one of the drugs most frequently used by drug abusers. It has a variety of street names, including "angel dust," "animal tranquilizer," "PCP," "peace pill," "crystal joints," and "peace weed," with the name often reflecting the form in which it is taken. It can be smoked, "snorted" through the nose, ingested, or taken intravenously. Phencyclidine has also been shown to cause schizophrenia-like changes in Nacetylaspartate and Nacetylaspartylglutamate in the rat brain, which are detectable both in living rats and upon necropsy examination of brain tissue. It also induces symptoms in humans that mimic schizophrenia. Behavioral effects can vary by dosage. Low doses produce numbness in the extremities and intoxication, characterized by staggering, unsteady gait, slurred speech, bloodshot eyes, and loss of balance. Moderate doses (5-10 mg intranasal, or 0.01-0.02 mg/kg intramuscular or intravenous) will produce analgesia and anesthesia. High doses may lead to convulsions. Users frequently do not know how much of the drug they are taking due to the tendency of the drug to be made illegally in uncontrolled conditions.

# E. INTENDED USE

# Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay

The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a homogenous enzyme immunoassay with a cutoff of 10 ng/mL in neat oral fluid collected with the Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of PCP in human oral fluid with clinical analyzers. This assay is calibrated against PCP. This in vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

| Attribute      | Predicate Device<br>RapidFRET Oral Fluid Assay<br>for PCP [k122703] | Candidate Device<br>Immunalysis SEFRIA PCP<br>Oral Fluid Enzyme<br>Immunoassay |
|----------------|---------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Similarities   |                                                                     |                                                                                |
| Test Principle | Homogeneous enzyme<br>immunoassay                                   | Same                                                                           |
| Antibody       | Antibodies to PCP                                                   | Same                                                                           |

# F. COMPARISON WITH PREDICATE

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a red arrow pointing to the right with the word "IMMUNALYS" written in white letters inside the arrow. The word is written in all capital letters and is centered within the arrow. The arrow is a bright red color, and the white letters stand out against the red background.

| Attribute                | Predicate Device<br>RapidFRET Oral Fluid Assay<br>for PCP [k122703]                                                                                                                                   | Candidate Device<br>Immunalysis SEFRIA PCP<br>Oral Fluid Enzyme<br>Immunoassay                                                                                              |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Materials                | PCP specific antibody reagent,<br>PCP drug conjugate reagent                                                                                                                                          | Same                                                                                                                                                                        |
| Cutoff Level             | 10 ng/mL                                                                                                                                                                                              | Same                                                                                                                                                                        |
| User Environment         | For use in laboratories                                                                                                                                                                               | Same                                                                                                                                                                        |
| Sample Matrix            | Human oral fluid                                                                                                                                                                                      | Same                                                                                                                                                                        |
| Reagent Storage          | 2-8°C until expiration date                                                                                                                                                                           | Same                                                                                                                                                                        |
|                          | Differences                                                                                                                                                                                           |                                                                                                                                                                             |
| Intended Use             | Qualitative determination of<br>phencyclidine in human oral<br>fluid                                                                                                                                  | Qualitative and semi-<br>quantitative determination of<br>phencyclidine in human oral<br>fluid.                                                                             |
| Sample Collection Device | Neat oral fluid is collected with<br>the RapidEASE Oral Fluid<br>Collector via direct<br>expectoration. No diluent is used<br>and sample is stored in a glass<br>sample tube with inert screw<br>cap. | Oral fluid is collected with the<br>Quantisal II Oral Fluid<br>Collection Device. Sample is<br>stored in a plastic tube<br>containing preservative buffer<br>with snap cap. |

# G. PERFORMANCE CHARACTERISTICS

The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay and Quantisal II Oral Fluid Collection Devices to the predicate device. Assay performance was established using the Beckman Coulter AU480 chemistry analyzer.

# 1. Precision/ Cutoff Characterization

Precision/ Cutoff Characterization study was performed over 15 days, two collection device runs per day with two collections per run (N=60) on three lots of Quantisal II oral fluid collection devices. Drug free negative urine was spiked to concentrations of assay cutoff and ±25%, ±50%, ±75%, ±100% of the cutoff and transferred to the collection devices. The spiked concentrations were confirmed by mass spectrometry (LC-MS/MS). The study established the repeatability of the testing system, including assay and oral fluid collection device. Test results in qualitative and semi-quantitative modes for a representative lot are presented in Tables 1 to 6.

| Concentration<br>(ng/mL) | % of cutoff | # of<br>determinations | Result      |
|--------------------------|-------------|------------------------|-------------|
| 0                        | -100%       | 60                     | 60 Negative |

# Table 1. Precision - Quantisal II "A" - Qualitative

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

| Concentration<br>(ng/mL) | % of cutoff | # of<br>determinations | Result         |
|--------------------------|-------------|------------------------|----------------|
| 2.5                      | -75%        | 60                     | 60 Negative    |
| 5                        | -50%        | 60                     | 60 Negative    |
| 7.5                      | -25%        | 60                     | 60 Negative    |
| 10                       | Cutoff      | 60                     | 29 Neg /31 Pos |
| 12.5                     | 25%         | 60                     | 60 Positive    |
| 15                       | 50%         | 60                     | 60 Positive    |
| 17.5                     | 75%         | 60                     | 60 Positive    |
| 20                       | 100%        | 60                     | 60 Positive    |

# Table 2. Precision – Quantisal II "A" - Semi-Quantitative

| Concentration<br>(ng/mL) | % of cutoff | # of<br>determinations | Mean Conc.<br>(ng/mL) | Result          |
|--------------------------|-------------|------------------------|-----------------------|-----------------|
| 0                        | -100%       | 60                     | 0.5                   | 60 Negative     |
| 2.5                      | -75%        | 60                     | 2.7                   | 60 Negative     |
| 5                        | -50%        | 60                     | 5.5                   | 60 Negative     |
| 7.5                      | -25%        | 60                     | 7.8                   | 60 Negative     |
| 10                       | Cutoff      | 60                     | 10.1                  | 28 Neg / 32 Pos |
| 12.5                     | 25%         | 60                     | 12.5                  | 60 Positive     |
| 15                       | 50%         | 60                     | 15.5                  | 60 Positive     |
| 17.5                     | 75%         | 60                     | 17.4                  | 60 Positive     |
| 20                       | 100%        | 60                     | 20.2                  | 60 Positive     |

# Table 3. Precision – Quantisal II "B" - Qualitative

| Concentration<br>(ng/mL) | % of cutoff | # of<br>determinations | Result         |
|--------------------------|-------------|------------------------|----------------|
| 0                        | -100%       | 60                     | 60 Negative    |
| 2.5                      | -75%        | 60                     | 60 Negative    |
| 5                        | -50%        | 60                     | 60 Negative    |
| 7.5                      | -25%        | 60                     | 60 Negative    |
| 10                       | Cutoff      | 60                     | 32 Neg /28 Pos |
| 12.5                     | 25%         | 60                     | 60 Positive    |
| 15                       | 50%         | 60                     | 60 Positive    |
| 17.5                     | 75%         | 60                     | 60 Positive    |
| 20                       | 100%        | 60                     | 60 Positive    |

# Table 4. Precision – Quantisal II "B" - Semi-Quantitative

| Concentration<br>(ng/mL) | % of cutoff | # of<br>determinations | Mean Conc.<br>(ng/mL) | Result      |
|--------------------------|-------------|------------------------|-----------------------|-------------|
| 0                        | -100%       | 60                     | 0.5                   | 60 Negative |
| 2.5                      | -75%        | 60                     | 2.7                   | 60 Negative |
| 5                        | -50%        | 60                     | 5.5                   | 60 Negative |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

| Concentration<br>(ng/mL) | % of cutoff | # of<br>determinations | Mean Conc.<br>(ng/mL) | Result          |
|--------------------------|-------------|------------------------|-----------------------|-----------------|
| 7.5                      | -25%        | 60                     | 8.0                   | 60 Negative     |
| 10                       | Cutoff      | 60                     | 10.1                  | 29 Neg / 31 Pos |
| 12.5                     | 25%         | 60                     | 12.7                  | 60 Positive     |
| 15                       | 50%         | 60                     | 15.5                  | 60 Positive     |
| 17.5                     | 75%         | 60                     | 17.5                  | 60 Positive     |
| 20                       | 100%        | 60                     | 20.2                  | 60 Positive     |

# 2. Specificity and Cross-Reactivity

Structurally and functionally similar compounds were spiked into drug free oral fluid at levels that will yield a result that is equivalent to the cutoff. The study verified assay performance relative to the ability of the device to exclusively determine certain drugs, in both the qualitative and semi-quantitative modes. Cross-reactivity test results for qualitative and semi-quantitative modes are presented in Table 7.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The red background is a solid color, and the white letters are easy to read. The image is simple and straightforward.

| Compound                         | Compound<br>Conc.<br>(ng/mL) | PCP<br>Equivalent<br>Conc.<br>(ng/mL) | Result | Cross-<br>Reactivity<br>(%) |
|----------------------------------|------------------------------|---------------------------------------|--------|-----------------------------|
| Amitriptyline                    | 22,000                       | 10                                    | POS    | 0.05                        |
| Chlorpromazine                   | 5,200                        | 10                                    | POS    | 0.19                        |
| Clomipramine                     | 22,000                       | 10                                    | POS    | 0.05                        |
| Cyclobenzaprine                  | 1,900                        | 10                                    | POS    | 0.53                        |
| Desipramine                      | 40,000                       | <10                                   | NEG    | <0.03                       |
| Dextromethorphan                 | 40,000                       | <10                                   | NEG    | <0.03                       |
| Diphenhydramine                  | 37,000                       | 10                                    | POS    | 0.03                        |
| Doxepin                          | 5,600                        | 10                                    | POS    | 0.18                        |
| Doxylamine                       | 40,000                       | <10                                   | NEG    | <0.03                       |
| EDDP                             | 40,000                       | <10                                   | NEG    | <0.03                       |
| 4-Hydroxyphencyclidine<br>(PCHP) | 85                           | 10                                    | POS    | 11.76                       |
| Imipramine                       | 13,400                       | 10                                    | POS    | 0.07                        |
| Methoxetamine                    | 34,000                       | 10                                    | POS    | 0.03                        |
| Nortriptyline                    | 40,000                       | <10                                   | NEG    | <0.03                       |
| Protriptyline                    | 40,000                       | <10                                   | NEG    | <0.03                       |
| Thioridazine                     | 8,600                        | 10                                    | POS    | 0.12                        |
| Trimipramine                     | 40,000                       | <10                                   | NEG    | <0.03                       |
| Venlafaxine                      | 40,000                       | <10                                   | NEG    | <0.03                       |

# Table 5. Cross-Reactivity - Qualitative

# 3. Interference – Structurally Unrelated Compounds

Structurally unrelated compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free oral fluid containing PCP at ±25% of the cutoff. All potential interferents analyzed verified that assay performance is unaffected by externally ingested compounds. The levels of structurally unrelated compounds that did not interfere in the assay are presented in Table 8.

| Compound                                | Conc. Tested (ng/mL) |
|-----------------------------------------|----------------------|
| 4-Bromo-<br>2,5,Dimethoxyphenethylamine | 5,000                |
| 6-Acetylcodeine                         | 40,000               |
| 6-Acetylmorphine                        | 40,000               |
| Alprazolam                              | 40,000               |
| 7-Aminoclonazepam                       | 40,000               |
| 7-Aminoflunitrazepam                    | 40,000               |
| 7-Aminonitrazepam                       | 40,000               |
| S-(+) Amphetamine                       | 40,000               |

| Table 6. Non-Interfering Structurally Unrelated Compounds- Qualitative and Semi-quantitative |  |  |  |
|----------------------------------------------------------------------------------------------|--|--|--|
|                                                                                              |  |  |  |

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

| Compound                         | Conc. Tested (ng/mL) |
|----------------------------------|----------------------|
| Benzylpiperazine                 | 40,000               |
| Bromazepam                       | 40,000               |
| Buprenorphine                    | 40,000               |
| Bupropion                        | 40,000               |
| Butabarbital                     | 40,000               |
| Butalbital                       | 40,000               |
| Caffeine                         | 40,000               |
| Cannabidiol                      | 40,000               |
| Cannabinol                       | 40,000               |
| Carbamazepine                    | 40,000               |
| Carisoprodol                     | 40,000               |
| Chlordiazepoxide                 | 40,000               |
| cis-Tramadol                     | 40,000               |
| Clobazam                         | 40,000               |
| Clonazepam                       | 40,000               |
| Clozapine                        | 40,000               |
| Cocaine                          | 40,000               |
| Codeine                          | 40,000               |
| Cotinine                         | 40,000               |
| Demoxepam                        | 40,000               |
| Desalkylflurazepam               | 40,000               |
| Dihydrocodeine                   | 40,000               |
| Diazepam                         | 40,000               |
| Digoxin                          | 40,000               |
| Dehydronorketamine               | 40,000               |
| Delta-9-THC                      | 40,000               |
| Ecgonine                         | 40,000               |
| Ecgonine Methyl Ester            | 40,000               |
| EMDP                             | 40,000               |
| 1R,2S(-)-Ephedrine               | 40,000               |
| 1S,2R(+)-Ephedrine               | 40,000               |
| Ethyl-β-D-Glucuronide            | 40,000               |
| Ethylmorphine                    | 40,000               |
| Fenfluramine                     | 40,000               |
| Fentanyl                         | 20,000               |
| Flunitrazepam                    | 40,000               |
| Fluoxetine                       | 40,000               |
| Flurazepam                       | 40,000               |
| Haloperidol                      | 40,000               |
| Heroin                           | 40,000               |
| Hexobarbital                     | 40,000               |
| Compound                         | Conc. Tested (ng/mL) |
| Hydrocodone                      | 40,000               |
| Hydromorphone                    | 40,000               |
| 11-hydroxy-delta-9-THC           | 40,000               |
| Ibuprofen                        | 40,000               |
| Ketamine                         | 40,000               |
| Lamotrigine                      | 40,000               |
| Levorphanol                      | 40,000               |
| Lidocaine                        | 40,000               |
| Lorazepam                        | 40,000               |
| Lorazepam Glucuronide            | 40,000               |
| Lormetazepam                     | 40,000               |
| LSD                              | 40,000               |
| Maprotiline                      | 40,000               |
| MDA                              | 40,000               |
| MDEA                             | 40,000               |
| MDMA                             | 40,000               |
| Meperidine                       | 40,000               |
| Meprobamate                      | 40,000               |
| S(+)-Methamphetamine             | 40,000               |
| Methadone                        | 40,000               |
| Methaqualone                     | 40,000               |
| Methylone                        | 40,000               |
| Methylphenidate                  | 40,000               |
| Midazolam                        | 40,000               |
| Morphine                         | 40,000               |
| Morphine-3-Glucuronide           | 40,000               |
| Morphine-6-Glucuronide           | 40,000               |
| N-desmethyltapentadol            | 40,000               |
| N-desmethyl tramadol             | 40,000               |
| N-desmethyl venlafaxine          | 40,000               |
| Nalorphine                       | 40,000               |
| Naloxone                         | 40,000               |
| Naltrexone                       | 40,000               |
| Naproxen                         | 40,000               |
| Nitrazepam                       | 40,000               |
| 11-nor-9 carboxy THC             | 40,000               |
| Norbuprenorphine                 | 40,000               |
| Norcodeine                       | 40,000               |
| Nordiazepam                      | 40,000               |
| Norketamine                      | 40,000               |
| Normorphine                      | 40,000               |
| Compound                         | Conc. Tested (ng/mL) |
| Noroxycodone                     | 40,000               |
| Noroxymorphone                   | 40,000               |
| Norpropoxyphene                  | 20,000               |
| Norpseudoephedrine               | 2,000                |
| O-desmethyl tramadol             | 40,000               |
| O-desmethyl venlafaxine          | 40,000               |
| Oxycodone                        | 40,000               |
| Oxymorphone                      | 40,000               |
| Olanzapine                       | 40,000               |
| Oxazepam                         | 40,000               |
| Pentazocine                      | 40,000               |
| Pentobarbital                    | 40,000               |
| Phenobarbital                    | 40,000               |
| Phentermine                      | 40,000               |
| Phenylephrine                    | 40,000               |
| Phenytoin                        | 40,000               |
| Phenylpropanolamine              | 40,000               |
| PMA                              | 40,000               |
| Prazepam                         | 40,000               |
| Propranolol                      | 40,000               |
| Propoxyphene                     | 40,000               |
| R,R(-)-Pseudoephedrine           | 40,000               |
| S,S(+)-Pseudoephedrine           | 40,000               |
| Ritalinic Acid                   | 40,000               |
| Salicylic Acid                   | 40,000               |
| Secobarbital                     | 40,000               |
| Sertraline                       | 40,000               |
| Sufentanil                       | 40,000               |
| Tapentadol                       | 40,000               |
| Temazepam                        | 40,000               |
| Theophylline                     | 40,000               |
| Trazadone                        | 40,000               |
| Triazolam                        | 40,000               |
| Trifluoromethylphenyl-piperazine | 40,000               |
| Verapamil                        | 40,000               |
| Zolpidem Tartrate                | 40,000               |

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.

# 4. Interference - Endogenous Compounds and Exogenous Compounds

Endogenous compounds and exogenous compounds were evaluated in qualitative and semiquantitative modes by spiking the potential interferent into drug free oral fluid containing PCP at ±25% of the cutoff. Additional orally used products were tested by collecting oral fluid using

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color. The image is simple and easy to read.

Quantisal II Oral Fluid Collection Devices from volunteers after use of the substances. Assay performance is unaffected by all of the substances tested. Endogenous compounds and exogenous compounds tested are presented in Tables 9 and 10. Orally used compounds tested are presented in Table 11.

| Compound            | Concentration Tested |
|---------------------|----------------------|
| Ascorbic Acid       | 3 mg/mL              |
| Bilirubin           | 0.15 mg/mL           |
| Cholesterol         | 0.45 mg/mL           |
| γ-Globulin          | 0.8 mg/mL            |
| Hemoglobin          | 3 mg/mL              |
| Human Serum Albumin | 15 mg/mL             |
| IgA                 | 1 mg/mL              |
| IgG                 | 1 mg/mL              |
| IgM                 | 0.5 mg/mL            |
| Salivary-α-amylase  | 1000 U/mL            |

Table 7. Non-interfering Endogenous Compounds

#### Table 8. Non-interfering Exogenous Compounds

| Compound                   | Concentration Tested |
|----------------------------|----------------------|
| Acetaminophen              | 0.1 mg/mL            |
| Acetylsalicylic Acid       | 0.1 mg/mL            |
| Baking Soda                | 0.6% v/v             |
| Cotinine                   | 0.03 mg/mL           |
| Denture Adhesive           | 0.6% w/v             |
| Ibuprofen                  | 0.1 mg/mL            |
| Alcohol (Ethanol)          | 6% v/v               |
| Caffeine                   | 0.1 mg/mL            |
| Coffee                     | 6% v/v               |
| Cranberry Juice            | 6% v/v               |
| Hydrogen Peroxide (3% OTC) | 0.5% v/v             |
| Milk                       | 1% v/v               |
| Mouthwash                  | 6% v/v               |
| Naproxen                   | 0.1 mg/mL            |
| Orange Juice               | 6% v/v               |
| Soft Drink (Pepsi)         | 6% v/v               |
| Sodium Chloride            | 18 mg/mL             |
| Sugar                      | 20 mg/mL             |
| Tea                        | 6% v/v               |
| Toothpaste                 | 6% v/v               |

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The font is bold and sans-serif. The red background is a solid color and there are no other objects in the image.

| Compound                   | Concentration Tested     |
|----------------------------|--------------------------|
| Teeth Whitener             | 2 strips                 |
| Hydrogen Peroxide (3% OTC) | Neat (2 min mouth rinse) |
| Cigarette                  | 1 cigarette              |
| Hard Candy                 | 1 piece                  |
| Chewing Gum                | 1 piece                  |
| Cough Syrup                | 2 Teaspoons              |

Table 9. Non-interfering Orally Used Exogenous Products

# 5. Interference - pH

To evaluate potential interference from the effect of oral fluid pH, device performance in the qualitative and semi-quantitative modes was tested using a range of oral fluid pH values (3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0, 10.0 and 11.0). All test samples were prepared in drug free oral fluid containing PCP at ±25% of the cutoff. No positive or negative interference was observed at oral fluid pH values ranging from 3.0 to 11.0 for each test mode.

# 6. Linearity/Recovery

A linearity study in the semi-quantitative mode was conducted by spiking a drug free oral fluid pool with a high concentration of PCP above the highest calibrator. Additional pools were made by serially diluting the high concentration specimen with drug free oral fluid to achieve concentrations ranging from 4 ng/mL to 44 ng/mL specimen was made from drug free oral fluid. Each pool was collected using Quantisal II oral fluid collection devices and tested in triplicate to calculate the mean concentration values that were used to calculate drug recovery. Linearity test results in semi-quantitative mode are presented in Tables 12 to 14.

| Expected Concentration<br>(ng/mL) | Mean Concentration<br>(ng/mL) | Recovery (%) |
|-----------------------------------|-------------------------------|--------------|
| 0                                 | 1.5                           | N/A          |
| 4                                 | 4.2                           | 105.0        |
| 8                                 | 8.5                           | 105.8        |
| 10                                | 10.0                          | 100.0        |
| 12                                | 11.7                          | 97.2         |
| 16                                | 15.8                          | 98.5         |
| 20                                | 21.2                          | 105.8        |
| 24                                | 24.6                          | 102.4        |
| 28                                | 28.5                          | 101.9        |
| 32                                | 31.5                          | 98.3         |
| 36                                | 35.5                          | 98.6         |
| 40                                | 42.1                          | 105.3        |
| 44                                | 46.0                          | 104.5        |

Table 10. Linearity/Recovery - Quantisal II "A"

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

| Expected Concentration<br>(ng/mL) | Mean Concentration<br>(ng/mL) | Recovery (%) |
|-----------------------------------|-------------------------------|--------------|
| 0                                 | 1.3                           | N/A          |
| 4                                 | 3.7                           | 92.5         |
| 8                                 | 8.2                           | 102.1        |
| 10                                | 10.3                          | 102.7        |
| 12                                | 12.3                          | 102.5        |
| 16                                | 16.0                          | 99.8         |
| 20                                | 19.0                          | 94.8         |
| 24                                | 24.1                          | 100.6        |
| 28                                | 27.7                          | 98.9         |
| 32                                | 33.9                          | 105.8        |
| 36                                | 36.6                          | 101.8        |
| 40                                | 42.0                          | 105.1        |
| 44                                | 46.4                          | 105.5        |

# Table 11. Linearity/Recovery - Quantisal II "B"

# 7. Calibration Duration

Drug free negative oral fluid spiked with PCP at ±25% of the cutoff were tested in qualitative mode at time points up to 14 days. At the initial time point calibration curve was established. This calibration was used through the duration of this study. The test results met acceptance criteria at each time point. The recommended frequency of calibration is 14 days.

# 8. PCP Stability in Oral Fluid

Drug free negative oral fluid spiked with PCP at +50% of the cutoff were collected and stored in Quantisal II Oral Fluid Collection Device and tested by LC-MS/MS at each time point at 30°C and at 2℃ - 8℃. Test results indicated that oral fluid samples containing PCP are stable for up to 10 days stored in Quantisal II transport tubes at ambient temperature up to 30°C and up to 2 months stored in Quantisal II transport tubes at 2℃ - 8℃.

#### 9. Sample Transportation Stability

Drug free negative oral fluid spiked with PCP at ±50% of the cutoff were collected and stored in Quantisal II Oral Fluid Collection Device and packed in standard boxes used by common carrier (FedEx). During the 4-day (96 hours) simulated transportation study, the samples were stored in oven/freezer at temperatures ranged from -20°C to 40°C to encompass the temperatures likely to occur during shipment of products. The device used as the reference (unstressed) condition was stored continuously at the recommended storage condition at 2℃ - 8℃. LC-MS/MS testing was performed in replicates of two and compared to the reference sample. The studies demonstrated the PCP concentration of the sample collected by Quantisal II Oral Fluid Collection Device is within 20% of initial value during transportation.

#### 10. Sample Recovery

Drug free negative oral fluid spiked with PCP at ±25%, +50% of the cutoff were collected and stored in Quantisal II Oral Fluid Collection Devices overnight at room temperature. LC-MS/MS testing was performed the next day to determine percentage recovery. The studies demonstrated

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

that the Quantisal II Collection Device recovers PCP at greater than 80% of the original concentration.

# 11. Quantisal II Sample Volume

Fifty oral fluid samples were collected using Quantisal II collectors (collection pad with plastic stem) from fifty volunteers. Prior to collection, each collector (A and B) was weighed. After the volume adequacy indicator turned blue on both A and B collector stems, each collector was weighed again. The difference in weight was noted. Specific gravity of saliva was rounded to 1.000 to compute the volume collection. The results confirmed consistency of sample volume of 1 mL collected by each of Quantisal II collectors in this population.

Additional seventy-five oral fluid samples from known drug users were collected using Quantisal II collector. After the volume adequacy indicator turned blue on both A and B collector stems, each collector was weighed and compared to the average weight of collector before collection. The difference in weight was noted. Specific gravity of saliva was rounded to 1.000 to compute the volume collection. The results confirmed consistency of sample volume of 1 mL collected by each of Quantisal II collectors in this population.

# 12. Quantisal II Sample Collection Time

Fifty oral fluid samples from volunteers and seventy-five oral fluid samples from known drug users were collected using Quantisal II collection pad with plastic stem). The collection time was documented. The results verified the sample collection time for Quantisal II Oral Fluid Collection Device is within the claimed time of 10 minutes in over 90% of subjects.

# 13. Method Comparison

Eighty deidentified, unaltered clinical oral fluid samples collected by Quantisal II Oral Fluid Collection Devices were obtained from drug treatment facilities, analyzed for PCP at assay cutoff with the Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay in both qualitative and semi-quantitative modes and compared to Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) results with neat oral fluid collected by expectoration (spitting). The instruments used were the Beckman Coulter AU480 chemistry analyzer and an Agilent 6430 Liquid Chromatography-Tandem Mass Spectrometry. Method comparison test results in qualitative and semi-quantitative modes are presented from Tables 15 to 17.

| Immunalysis SEFRIA<br>PCP Oral Fluid<br>Enzyme Immunoassay<br>Result |          | LC-MS/MS PCP Neat Oral Fluid Concentration    |                                                        |                                                            |                                                | Agreement<br>(%) |
|----------------------------------------------------------------------|----------|-----------------------------------------------|--------------------------------------------------------|------------------------------------------------------------|------------------------------------------------|------------------|
|                                                                      |          | < 5<br>ng/mL<br>(less than<br>-50%<br>cutoff) | 5 - 9 ng/mL<br>(between -<br>50% cutoff<br>and cutoff) | 10 - 15<br>ng/mL<br>(between<br>cutoff and<br>+50% cutoff) | > 15 ng/mL<br>(greater<br>than +50%<br>cutoff) |                  |
| Qual.                                                                | Positive | 0                                             | 0                                                      | 5                                                          | 35                                             | 100% (40/40)     |
|                                                                      | Negative | 36                                            | 4                                                      | 0                                                          | 0                                              | 100% (40/40)     |
| Semi-<br>Quant.                                                      | Positive | 0                                             | 0                                                      | 5                                                          | 35                                             | 100% (40/40)     |
|                                                                      | Negative | 36                                            | 4                                                      | 0                                                          | 0                                              | 100% (40/40)     |

Table 12. Method Comparison - Quantisal II "A"

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered and in all caps. The background is a solid red color.

| Immunalysis SEFRIA<br>PCP Oral Fluid<br>Enzyme Immunoassay<br>Result |          | LC-MS/MS PCP Neat Oral Fluid Concentration |                                                        |                                                         |                                                |                  |
|----------------------------------------------------------------------|----------|--------------------------------------------|--------------------------------------------------------|---------------------------------------------------------|------------------------------------------------|------------------|
|                                                                      |          | < 5 ng/mL<br>(less than<br>-50%<br>cutoff) | 5 – 9 ng/mL<br>(between -<br>50% cutoff<br>and cutoff) | 10 - 15 ng/mL<br>(between<br>cutoff and<br>+50% cutoff) | > 15 ng/mL<br>(greater<br>than +50%<br>cutoff) | Agreement<br>(%) |
| Qual.                                                                | Positive | 0                                          | 0                                                      | 5                                                       | 35                                             | 100% (40/40)     |
|                                                                      | Negative | 36                                         | 4                                                      | 0                                                       | 0                                              | 100% (40/40)     |
| Semi-<br>Quant.                                                      | Positive | 0                                          | 0                                                      | 5                                                       | 35                                             | 100% (40/40)     |
|                                                                      | Negative | 36                                         | 4                                                      | 0                                                       | 0                                              | 100% (40/40)     |

# H. CONCLUSION

The information provided in this pre-market notification demonstrates that the Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay and Immunalysis Quantisal II Oral Fluid Collection Device are substantially equivalent to the legally marketed predicate device for their intended uses.

---

**Source:** [https://fda.innolitics.com/device/K181135](https://fda.innolitics.com/device/K181135)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
