← Product Code [NKB](/productcode/NKB) · K181119

# Range (Denali and Mesa) Spinal Systems (K181119)

_K2m, Inc. · NKB · Jun 26, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K181119

## Device Facts

- **Applicant:** K2m, Inc.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jun 26, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The MESA and DENALI Systems (including ARI Staples) and the EVEREST Spinal System are cleared for the following indications: Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudoarthosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion. Except for the ARI staples, the MESA, DENALI and EVEREST Spinal Systems are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis when used for posterior non-cervical fixation in pediatric patients. The MESA, DENALI and EVEREST Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

Range (Denali and Mesa) Spinal Systems are top-loading, multi-component posterior spinal fixation systems; components include pedicle screws, rods, hooks, and rod connectors. Device provides support and stabilization of posterior thoracic and lumbar spine as adjunct to fusion. Used by surgeons in clinical/OR settings. This submission adds revision connectors and rods to existing systems. Implants are surgically placed to stabilize spinal segments; intended to facilitate fusion. Benefits include structural support for spinal pathologies.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1717 (static torsion, static compression, dynamic compression bending) on worst-case constructs. Bacterial endotoxin testing also performed. No clinical data provided.

## Technological Characteristics

Top-loading spinal fixation system; components include pedicle screws, rods, hooks, and rod connectors. Materials and design features are substantially the same as predicate systems. Mechanical testing conducted per ASTM F1717. Sterilization method not specified.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- K2M Range (Denali and Mesa) Spinal Systems ([K171832](/device/K171832.md))
- K2M EVEREST Spinal System ([K173508](/device/K173508.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 26, 2018

K2M, Inc. Ms. Nancy Giezen Manager, Regulatory Affairs 600 Hope Parkway Southeast Leesburg, Virginia 20175

Re: K18119

Trade/Device Name: Range (Denali and Mesa) Spinal Systems Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWO, KWP Dated: April 26, 2018 Received: April 27, 2018

Dear Ms. Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Ronald P. Jean -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

#### K181119

Device Name Range (Denali and Mesa) Spinal Systems

#### Indications for Use (Describe)

The MESA and DENALI Systems (including ARI Staples) and the EVEREST Spinal System are cleared for the following indications:

Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudoarthosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.

Except for the ARI staples, the MESA. DENALI and EVEREST Spinal Systems are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis when used for posterior non-cervical fixation in pediatric patients. The MESA, DENALI and EVEREST Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (*Select one or both, as applicable*)

| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> | <span>      ☐ Over-The-Counter Use (21 CFR 801 Subpart C)     </span> |
|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY Range (Denali and Mesa) Spinal Systems K2M, Inc.

# Submitter

K2M, Inc. 600 Hope Parkway SE Leesburg, VA 20175

Contact Person: Nancy Giezen Telephone: 571 919-2168 Date Prepared: 06/26/2018

# Classification

Trade Name: Range (Denali and Mesa) Spinal Systems Common Name: Spinal Fixation System Regulatory Class: Class II

Classification Name(s):

Thoracolumbosacral Pedicle Screw System (21 CFR 888.3070. Product Codes: NKB) Spinal Intervertebral body fixation Orthosis (21 CFR 888.3060, Product Code: KWQ) Spinal Interlaminal fixation Orthosis (21 CFR 888.3050, Product Code: KWP)

# Predicate Device(s)

Primary Predicates: K2M Range (Denali and Mesa) Spinal Systems (K171832)

Additional Predicates: K2M EVEREST Spinal System (K173508)

# Device Description

The Range (Denali and Mesa) Spinal Systems are top-loading, multiple component, posterior (thoracic-lumbar) spinal fixation systems consisting of pedicle screws, rods, hooks and rod connectors. The purpose of this 510(k) is to add revision connectors and rods to the system.

Function: The systems function as spinal fixation devices to provide support and stabilization of the posterior thoracic and lumbar spine.

# Indications For Use

The MESA and DENALI Spinal Systems (including ARI Staples) and the EVEREST Spinal System are cleared for the following indications:

Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation);

{4}------------------------------------------------

spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudoarthosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.

Except for the ARI staples, the MESA, DENALI and EVEREST Spinal Systems are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior non-cervical fixation in pediatric patients. The MESA, DENALI and EVEREST Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

# Technological Comparison to Predicate(s)

The Range (Denali and Mesa) Spinal Systems were compared to predicate systems and the design features, materials and sizes were found to be substantially the same as these systems.

# Non-clinical Performance Evaluation

Mechanical testing in accordance with ASTM F1717 (including static torsion, static compression and dynamic compression bending) was conducted on constructs representing the worst case components, The test results revealed that the proposed implants were substantially the same as the predicate devices. The results of bacterial endotoxin testing were also provided to support substantial equivalence.

# Conclusion

There are no significant differences between the Range (Denali and Mesa) Systems and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.

---

**Source:** [https://fda.innolitics.com/device/K181119](https://fda.innolitics.com/device/K181119)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).