ADVIA Chemistry Cystatin C_2 Assay (CYSC_2)
Device Facts
| Record ID | K181082 |
|---|---|
| Device Name | ADVIA Chemistry Cystatin C_2 Assay (CYSC_2) |
| Applicant | Siemens Healthcare Diagnostics, Inc. |
| Product Code | NDY · Clinical Chemistry |
| Decision Date | May 25, 2018 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1225 |
| Device Class | Class 2 |
Indications for Use
The ADVIA® Chemistry Cystatin C 2 (CYSC 2) assay is for the in vitro diagnostic use in the quantitative determination of cystatin C in human serum and plasma (lithium heparin, potassium EDTA) on ADVIA Chemistry systems. Measurement of cystatin C aids in the diagnosis and treatment of renal disease.
Device Story
The ADVIA Chemistry Cystatin C 2 (CYSC 2) assay is an in vitro diagnostic kit for quantitative cystatin C measurement in human serum and plasma. The device uses particle-enhanced turbidimetry; latex particles coated with anti-cystatin-C antibodies agglutinate in the presence of cystatin C, increasing sample turbidity. Turbidity is measured at 571 and 805 nm on ADVIA Chemistry systems. Concentration is determined via a 5-level calibration curve. Used in clinical laboratories by technicians to assist physicians in diagnosing and treating renal disease; results are interpreted alongside other clinical and laboratory findings.
Clinical Evidence
Bench testing only. Method comparison study (n=158) against the predicate showed high correlation (r=1.000) with Passing-Bablok slope 1.008 and intercept 0.101 mg/L. Precision testing (n=80 replicates) demonstrated repeatability CVs of 0.9-2.0% and within-lab CVs of 1.2-3.7%. Linearity confirmed across 0.25-8.93 mg/L range. LoB/LoD/LoQ established at 0.10 mg/L, 0.25 mg/L, and 0.25 mg/L respectively. Interference testing showed no significant interference from bilirubin, lipemia, hemolysis, or rheumatoid factor.
Technological Characteristics
Particle-enhanced turbidimetry assay. Reagents consist of latex particles coated with rabbit anti-cystatin-C antibodies. Measures turbidity at 571 and 805 nm. Calibrators are traceable to IFCC ERM-DA471/IFCC. Compatible with ADVIA Chemistry systems. Measuring range 0.25-8.93 mg/L (up to 26.79 mg/L with dilution).
Indications for Use
Indicated for the quantitative determination of cystatin C in human serum and plasma (lithium heparin, potassium EDTA) to aid in the diagnosis and treatment of renal disease.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.