Infrared Ear Thermometer

K181081 · Joytech Healthcare Co. , Ltd. · FLL · Oct 2, 2018 · General Hospital

Device Facts

Record IDK181081
Device NameInfrared Ear Thermometer
ApplicantJoytech Healthcare Co. , Ltd.
Product CodeFLL · General Hospital
Decision DateOct 2, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Intended Use

Infrared Ear Thermometers are intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only.

Device Story

Infrared ear thermometers (models DET-101, DET-102, DET-103, DET-105) measure human body temperature via infrared radiation emitted from the tympanic membrane. The device probe is inserted into the ear canal; a thermopile detector captures thermal radiation; internal circuitry converts this signal into a temperature value. Used in home settings by patients of all ages. Output is displayed on an integrated screen; DET-105 includes a voice function. Probe covers are used as sanitary barriers. Data aids in monitoring body temperature for clinical decision-making.

Clinical Evidence

Clinical performance testing conducted per ASTM E1965-98 (2016) requirements. Results demonstrate compliance with standard specifications for clinical accuracy in intermittent temperature measurement.

Technological Characteristics

Materials: ABS enclosure, ABS/glass probe, PE probe cover. Sensing: Thermopile infrared detector. Energy: CR2032 or AAA batteries. Connectivity: Standalone. Standards: ASTM E1965-98, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5/10, ISO 80601-2-56.

Indications for Use

Indicated for intermittent measurement of human body temperature in people of all ages in a home use environment.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 2, 2018 JOYTECH Healthcare Co., Ltd. Yunhua Ren General manager No.365, Wuzhou Road, Yuhang Economic Development Zone Hangzhou City Zhejiang, 311100 China Re: K181081 Trade/Device Name: Infrared Ear Thermometer, Model: DET-102, DET-102, DET-103, DET-105 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 25, 2018 Received: August 29, 2018 Dear Yunhua Ren: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Geeta K. Pamidimukkala -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181081 Device Name Infrared Ear Thermometers DET-101,DET-102,DET-103,DET-105 Indications for Use (Describe) Infrared Ear Thermometers are intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The assigned 510(k) number is: k181081 #### Date Prepared: 2018.08.24 1. #### Submitter's Identification: 2. Name: JOYTECH Healthcare Co.,Ltd. Add.: No,365,Wuzhou Road,Yuhang Economic Development Zone, Hangzhou City,311100 Zhejiang,China Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: renyh@sejoy.com ### 3. Name of the Device: | | Device | Accessory | |-------------|--------------------------------------------------------------------|------------------| | Device Name | Infrared Ear Thermometer<br>Models:DET-101,DET-102,DET-103,DET-105 | Probe Cover | | Trade Name | Infrared Ear Thermometer<br>Models:DET-101,DET-102,DET-103,DET-105 | PC11 Probe Cover | ## 4. Classification Information: Product Code: FLL Device Class: II Panel: 80 Regulation number:21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer {4}------------------------------------------------ #### Predicate Device Information: 5. Manufacturer: SEJOY ELECTRONICS & INSTRUMENTS CO., LTD. Device: Infrared Ear Thermometer,Model ET-101H 510k number: K153146 ### 6. Intended use / Indication for Use: _ Infrared Ear Thermometers are intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only. ### 7. Device Description: #### Infrared Ear Thermometer: The Infrared Ear Thermometers are electronic thermometers using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. It's operation is based on measuring the natural thermal radiation emanating from the tympanic membrane.DET-103 and DET-105 include probe cover that is optional. The Infrared Ear Thermometers include probe which used to measure ear canal temperature,plastic enclosed the display window,buttons and the battery cover. #### Probe Cover: The probe cover is used as a sanitary barrier between the infrared ear thermometer and the ear canal to prevent any secretions or particulates from being transferred between different people. ## 8. Comparing to Predicate Device: {5}------------------------------------------------ | Comparisons | SUBJECT DEVICE:<br>Infrared Ear Thermometer<br>Model:DET-101,DET-102,DET-103,<br>DET-105<br>510k number:K181081 | PREDICATE DEVICE:<br>Infrared Ear Thermometer<br>Model:ET-101H<br>510k number:K153146 | Remark | | |-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|----------------------| | Indications for<br>Use | Infrared Ear Thermometers are intended<br>for the intermittent measurement of<br>human body temperature by people of<br>all ages. The devices are reusable for<br>home use only. | Infrared Ear Thermometers are indicated<br>for the intermittent measurement and<br>monitoring of human body temperature<br>by consumers in a home use<br>environment. It's intended for use on<br>people of all ages. | Same | | | Measurement<br>method | Infrared radiation detection | Infrared radiation detection | Same | | | Measuring Site | Ear canal | Ear canal | Same | | | Principle of<br>operation | The thermometer measures<br>temperature by reading infrared<br>radiation emitting from eardrum<br>tissue. The probe of the thermometer is<br>inserted into the ear canal, where<br>surrounding tissues give off heat. The<br>thermometer converts it into a<br>temperature value. | The thermometer measures temperature<br>by reading infrared radiation emitting<br>from eardrum tissue. The probe of the<br>thermometer is inserted into the ear<br>canal, where surrounding tissues give off<br>heat. The thermometer converts it into a<br>temperature value. | Same | | | Display | $0.1°C (0.1°F)$ | $0.1°C (0.1°F)$ | Same | | | Measurement<br>range | DET-101,<br>DET-102,<br>$32.0°C~43.0°C$<br>$(89.6°F~109.4°F)$ | Ear mode: $28.0°C~43.0°C$<br>$(82.4°F~109.4°F)$<br>Object mode: $-20°C~100°C(-4°F~212°F)$ | Similar<br>(Note1) | | | | DET-103,<br>DET-105<br>Ear mode: $32.0°C~43.0°C$<br>$(89.6°F~109.4°F)$<br>Object mode:<br>$0°C~100°C(32°F~212°F)$ | Ear mode: $28.0°C~43.0°C$<br>$(82.4°F~109.4°F)$<br>Object mode: $-20°C~100°C(-4°F~212°F)$ | Similar<br>(Note1) | | | Accuracy | DET-101,<br>DET-102<br>$±0.2°C (0.4°F)$ during<br>$35.0°C~42.0°C (95.0°F$<br>$~107.6°F)$ at $15°C~35°C$<br>$(59.0°F~95.0°F)$ operating | Ear mode:<br>$±0.2°C (0.4°F)$ during $35.5°C~42.0°C$<br>$(95.9°F~107.6°F)$ at $15°C~35°C (59.0°F$ | Similar<br>(Note1) | | | | | temperature range | ~95.0°F) operating temperature range | | | | | ±0.3℃ (0.5°F) for other<br>measuring and operating<br>temperature range | ±0.3℃ (0.5°F) for other measuring and<br>operating<br>temperature range | | | | | Ear mode:<br>±0.2℃ (0.4°F) during<br>35.5℃~42.0℃ (95.9°F<br>~107.6°F) at 15℃~35℃<br>(59.0°F~95.0°F) operating<br>temperature range | Object mode: ±4% or ±2℃(4°F)<br>whichever is greater | | | | DET-103,<br>DET-105 | ±0.3℃ (0.5°F) for other<br>measuring and operating<br>temperature range<br>Object mode: ±4% or ±<br>2°C(4°F) whichever is<br>greater | | | | Operating<br>temperature<br>range | | 10℃~40℃ (50°F~104°F),<br>15%~85%RH, non-condensing<br>Atmospheric Pressure:<br>700hPa ~ 1060hPa | 10℃~40℃ (50°F~104°F), 15%~85%RH,<br>non-condensing<br>Atmospheric Pressure:<br>700hPa ~ 1060hPa | Same | | Storage and<br>transport<br>environment | | -25℃~55℃(-13°F~131°F),<br>15%~95%RH, non-condensing<br>Atmospheric Pressure:<br>700hPa~1060hPa | -25℃~55℃ (-13°F~131°F),<br>15%~95%RH, non-condensing<br>Atmospheric Pressure:<br>700hPa~1060hPa | Same | | Power Source | DET-101,<br>DET-102 | One CR2032 battery | Two AAA battery | Different<br>(Note2) | | | DET-103,<br>DET-105 | Two AAA battery | Two AAA battery | Same | | Accessory | DET-101,<br>DET-102 | No accessory | Probe cover | Different<br>(Note3) | | | DET-103,<br>DET-105 | Probe cover | Probe cover | Same | | Response Time | | Approx. One second | Approx. One second | Same | | Material | | Enclosure;ABS<br>Probe:ABS & Glass<br>Probe Cover:PE | Enclosure;ABS<br>Probe:ABS & Glass<br>Probe Cover:PE | Same | | Memory | | 10 memories | 10 memories | Same | | Voice | DET-101,<br>DET-102,<br>DET-103 | No | No | Same | | | DET-105 | Yes | NO | Different<br>(Note4) | | Biocompatibility | Comply with ISO 10993-5 and ISO<br>10993-10 | | Comply with ISO 10993-5 and ISO<br>10993-10 | Same | | Electrical<br>Safety | Complied with IEC 60601-1 | | Complied with IEC 60601-1 | Same | | EMC | Complied with IEC 60601-1-2 | | Complied with IEC 60601-1-2 | Same | {6}------------------------------------------------ {7}------------------------------------------------ #### Analysis From the comparison table, the subject devices and predicate devices have the same Intended use & Indications for Use, Measurement place, Scale selection, Display screen, Auto power-off while no operation & Conformance standard. There are slightly differences between the devices and predicate devices as follows and these differences do not raise new performance questions. | Note1 | Measurement Range changes from 28.0 °C<br>~43.0 °C to 32.0 °C~43.0 °C & Accuracy changes<br>from ±0.2 °C (0.4 °F) during 35.5 °C~42.0 °C<br>(95.9 °F~107.6 °F) to ±0.2 °C (0.4 °F) during<br>35.0 °C~42.0 °C (95.0 °F~107.6 °F) | The changed measurement range and<br>accuracy of subject devices meet the<br>requirements of ISO80601-2-56 and<br>ASTM E1965-98. | |-------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------| | Note2 | DET-101, DET-102 Power Source changes from<br>Two AAA battery to CR2032 battery | Meet the requirement of IEC 60601-1. | | Note3 | DET-101, DET-102 accessory changes from<br>have probe cover to no probe cover. | Meet the requirement of ASTM E1965-98. | | Note4 | DET-105 add voice function | Software Validation passed. | {8}------------------------------------------------ # 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Infrared Ear Thermometer conforms to applicable standards that include: - ◆ ASTM E 1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature - ◆ IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance - � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - ◆ ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, - � ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization - � IEC 60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment - ◆ ISO 80601-2-56: 2017 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement - � Software Validation:FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued on May 11, 2005 {9}------------------------------------------------ # 10. Discussion of Clinical Tests Performed: The clinical performance test protocol and data analysis is conducted as the requirement of ASTM E1965-98 (2016). The test report showed the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016). # 11. Conclusions: Based on the information provided in this submission, we conclude that the submit Infrared Ear Thermometers, models: DET-101, DET-102,DET-103 and DET-105 are substantially equivalent to the predicate thermometer.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...