K181044 · AngioDynamics, Inc. · GEX · Jul 5, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181044
Device Name
400um Perforator and Accessory Vein Ablation Kit
Applicant
AngioDynamics, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 5, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 400 µm Perforator and Accessory Vein Ablation Kit is indicated for use in the treatment of superficial vein reflux of greater saphenous vein associated with varicosities. The 400 m Perforator and Accessory Vein Ablation Kit is indicated for treatment of incompetence and reflux of superficial veins in the lower extremity, and for the treatment of Incompetent (i.e. refluxing) Perforator Veins (IPVs).
Device Story
The 400um Perforator and Accessory Vein Ablation Kit is a sterile procedure kit used for endovascular laser ablation of varicose and incompetent perforator veins (IPVs). The kit includes a 400μm laser fiber, 21G entry needle, 4F introducer sheath/dilator, and 0.018" guidewire. It is intended for use in a clinical setting by a physician to treat venous insufficiency. The device functions by delivering laser energy to the vein, causing thermal ablation and closure of the vessel. This procedure helps alleviate symptoms associated with varicose veins and venous reflux. The kit provides the necessary access tools for the procedure, excluding standard minor treatment room supplies.
Clinical Evidence
The SECURE Study, a single-arm, prospective, multi-center, non-blinded clinical trial, evaluated the safety and effectiveness of the kit for treating IPVs. The study included 125 treated IPVs. The primary endpoint was acute primary ablation success at the 10-day visit. Results showed 96 of 125 IPVs (76.8%) met the primary endpoint. The GEE model success rate was 76.9%, which was statistically significantly above the 70% performance goal (p=0.033). Six adverse events were reported, including DVT, venous thrombosis, superficial thrombophlebitis, skin ulcer, wound, and procedural pain.
Technological Characteristics
The kit consists of a 400μm laser fiber and accessory components (21G needle, 4F sheath/dilator, 0.018" guidewire). It is a sterile, single-use procedure kit. The device operates via laser-induced thermal ablation. No software or electronic components are integrated into the kit itself; it is designed for use with compatible diode lasers.
Indications for Use
Indicated for patients with superficial vein reflux of the greater saphenous vein associated with varicosities, incompetence and reflux of lower extremity superficial veins, and incompetent (refluxing) perforator veins (IPVs).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
AngioDynamics, Inc. (Diomed, Ltd.) EVLT Kit and the D15 Plus and D30 Plus Diode Lasers (K041957)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
AngioDynamics, Inc. % Mr. Deepa Godhiya Specialist, Global Regulatory Affairs 26 Forest Street Marlborough, Massachusetts 01752
# Re: K181044
Trade/Device Name: 400um Perforator and Accessory Vein Ablation Kit Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and n Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 17, 2018 Received: April 19, 2018
# Dear Mr. Godhiya:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Peter L. Hudson -S 2018.07.03 14:06:42 -04'00'
For: Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Indications for Use | See PRA Statement below. |
| 510(k) Number (if known)<br>Unknown | |
| | |
| Device Name<br>400um Perforator and Accessory Vein Ablation Kit | |
| Indications for Use (Describe)<br>The 400um Perforator and Accessory Vein Ablation Kit is intended for use in the treatment of superficial vein reflux of<br>the greater saphenous vein associated with varicosities. The 400um Perforator and Accessory Vein Ablation Kit is<br>indicated for treatment of incompetence and reflux of superficial veins in the lower extremity, and for the treatment of<br>incompetent (i.e. refluxing) perforator veins (IPVs). | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| Type of Use (Select one or both, as applicable) | |
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer | |
| Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | |
{3}------------------------------------------------
## 510(K) SUMMARY – 400μ.M PERFORATOR AND ACCESSORY VEIN ABLATION KIT Date Prepared: April 17, 2018
### A. SPONSOR
AngioDynamics, Inc. 26 Forest Street Marlborough, MA 01752
## B. CONTACT
Deepa R. Godhiya Specialist, Global Regulatory Affairs T: 1-508-658-7923 F: 1-508-658-7976 E: dgodhiya@angiodynamics.com
#### OR
Brandon M. Brackett, RAC Manager, Global Regulatory Affairs T: 1-508-658-7984 F: 1-508-658-7976 E: bbrackett@angiodynamics.com
#### C. DEVICE NAME
| Trade Name: | 400μm Perforator and Accessory Vein Ablation Kit |
|-----------------------|--------------------------------------------------|
| Common/Usual Name: | Laser Vein Ablation Procedure Kit |
| Classification Name: | Power Laser Surgical Instrument (ProCode: GEX) |
| Classification Panel: | General and Plastic Surgery |
| D. PREDICATE DEVICE(S) | |
|------------------------|------------------------------------------------------------------------------------------|
| 510(k) Number | K041957 |
| Trade Name: | AngioDynamics, Inc. (Diomed, Ltd.) EVLT Kit and th<br>D15 Plus and D30 Plus Diode Lasers |
| Common/Usual Name: | Laser Vein Ablation Procedure Kit |
| Classification Name: | Power Laser Surgical Instrument (ProCode: GEX) |
| Classification Panel: | Dermatology and Plastic Surgery |
{4}------------------------------------------------
K181044, pg. 2 of 3
| 510(k) Number | K052003 |
|-----------------------|------------------------------------------------------------------------|
| Trade Name: | VNUS RFS and VNUS RFS Flex |
| Common/Usual Name: | Bipolar Electrosurgical Instrument |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories<br>(ProCode: GEI) |
| Classification Panel: | Dermatology and Plastic Surgery |
## E. DEVICE DESCRIPTION
The 400μm Perforator and Accessory Vein Ablation Kit is used to treat patients with varicose veins. Varicose veins are large, twisted, visibly bulging veins usually located in the legs. Varicose Veins occur when either the blood flow is too slow or valves are not functioning correctly resulting in "pile-up" of blood within the lower extremity veins. AngioDynamics proposes that this include the Incompetent Perforator Veins (IPVs). The 400um Perforator and Accessory Vein Ablation Kit is provided in a procedure kit which includes a 400μm Fiber along with the following accessory components:
- 21G Entry Needle
- . 4F Introducer Sheath/Dilator
- 0.018" Guidewire ●
The 400μm Perforator and Accessory Vein Ablation Kit is presented sterile and contains all of the accessories needed to perform the EVLT procedure, with the exception of generic disposable items routinely found in a well-equipped minor treatment room such as drapes, swabs, bandages etc.
#### F. INTENDED USE/INDICATIONS FOR USE
"The 400 µm Perforator and Accessory Vein Ablation Kit is indicated for use in the treatment of superficial vein reflux of greater saphenous vein associated with varicosities. The 400 m Perforator and Accessory Vein Ablation Kit is indicated for treatment of incompetence and reflux of superficial veins in the lower extremity, and for the treatment of Incompetent (i.e. refluxing) Perforator Veins (IPVs)."
## G. SUMMARY OF SIMILARITIES AND DIFFERENCES IN TECHNOLOGICAL CHARACTERISTICS AND PERFORMANCE
The proposed 400μm Perforator and Accessory Vein Ablation Kit is substantially equivalent to the predicate devices, both previously reviewed and cleared by the Agency, based upon the following conclusions:
- . to K041957 (EVLT Kit and the D15 Plus and D30 Plus Diode Lasers) in terms of overall design, materials, kitted accessories, labeling, and other design-related attributes - please refer to the table below for a side-by-side comparison of the Kit Accessories of the proposed device to that cleared via this predicate 510(k),
| Proposed Kit | Predicate Kit (K041957) |
|------------------------------|-----------------------------------|
| 21G Entry Needle | 19G Entry Needle |
| 0.018" Guidewire | 0.035" Guidewire |
| 4F Introducer Sheath/Dilator | 4F & 5F Introducer Sheath/Dilator |
- . to K052003 (VNUS RFS and RFS Flex Devices) in regards to the Indications for Use, which includes for use in perforator veins – AngioDynamics' intended expansion in regards to its Indications for Use.
{5}------------------------------------------------
#### H. PERFORMANCE DATA
standard, including:
The performance evaluation of the proposed 400μm Perforator and Accessory Vein Ablation Kit is centered around the clinical study 'SECURE Study', which was performed in compliance with Good Clinical Practice (GCP) and in accordance to the international standard AAMI/ANSI/ISO 14155:2011 – "Clinical Investigation of Medical Devices for Human Subjects: Good Clinical Practice". Furthermore, the proposed 400µm Perforator and Accessory Vein Ablation Kit has demonstrated successful results based upon the conclusions of the clinical testing per the above guidance and
- . Primary objective of acute primary ablation success; and,
- Secondary objectives to evaluate the post-procedural clinical outcomes. ●
## I. CLINICAL STUDY (For Perforator Study)
The SECURE STUDY was conducted to evaluate the safety and effectiveness of the 400um Perforator and Accessory Vein Ablation Kit when used to treat Incompetent Perforator Veins (IPVs). This was a singlearm, prospective, multi-center, non-blinded clinical trial. The study population included patients diagnosed with perforating vein insufficiency who were deemed by their treating clinician to be eligible for treatment by endovascular ablation. The primary objective of the study was to evaluate the VenaCure EVLT 400 µm Fiber Procedure Kit when used to treat Incompetent Perforator Veins (IPVs). For the primary endpoint of "acute primary ablation success," 96 of 125 treated IPVs (ITT Population) at the 10-day visit met the primary endpoint of acute primary ablation success, for a success rate of 76.8%. In the generalized estimating equation (GEE) model specified in the protocol and used to evaluate the primary endpoint, this resulted in a 76.9% model success rate, statistically significantly above the performance goal of 70% (p=0.033). The table below provides details of the SECURE STUDY.
| Intent to Treat (ITT) Set | | | |
|---------------------------|--------------------|------------------------|--------------------------------------------|
| Site | Number of Patients | Number of IPVs Treated | Number of IPV Successes Range of % Success |
| 100 | 11 | 11 | 2 – 43<br>28.6 – 100.0% |
| 101 | 29 | 56 | |
| 102 | 9 | 11 | |
| 103 | 13 | 18 | |
| 104 | 11 | 15 | |
| 106 | 4 | 7 | |
| 107 | 6 | 7 | |
#### Adverse Events (Safety Population)
A total of 6 adverse events were reported during this clinical study, that were either procedure and/or device related. They include the following: deep vein thrombosis; venous thrombosis limb; thrombophlebitis superficial; skin ulcer; wound; procedural pain.
## J. CONCLUSION
Based upon successful results of clinical testing, in addition to the responses to questions posed within FDA's 510(k) Decision-Making Tree, the proposed device is determined to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.