Ultra ICE Plus – PI 9 MHz Peripheral Imaging Catheter
K181042 · Boston Scientific Corporation · DXK · May 17, 2018 · Cardiovascular
Device Facts
Record ID
K181042
Device Name
Ultra ICE Plus – PI 9 MHz Peripheral Imaging Catheter
Applicant
Boston Scientific Corporation
Product Code
DXK · Cardiovascular
Decision Date
May 17, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2330
Device Class
Class 2
Indications for Use
The Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter is intended for intravascular ultrasonic visualization.
Device Story
Intravascular ultrasound (IVUS) catheter for peripheral vascular imaging. Input: reflected acoustic energy from internal vascular structures. Operation: 9 MHz piezoelectric (PZT) transducer at distal tip rotates via drive cable to provide 360° imaging; transducer converts electrical impulses to acoustic energy and back to electrical signals. Output: live visualization of intravascular structures processed by Boston Scientific iLab equipment. Used in clinical settings by physicians; catheter connects to MDU5 PLUS motor drive unit. Provides real-time imaging to guide vascular interventions like angioplasty or stenting; assists clinicians in assessing vessel morphology and procedural planning.
Clinical Evidence
No clinical data. Substantial equivalence based on non-clinical performance evaluations including bench-top testing (deliverability, image quality, measurement accuracy, fatigue), biocompatibility (ISO 10993-1), acoustic output (FDA Track 1 limits), electromagnetic compatibility (IEC 60601-1-2), and packaging validation (ISO 11607).
Technological Characteristics
9 MHz PZT transducer; 110 cm usable length; braided proximal shaft; sonolucent distal tip; self-sealing septum for acoustic coupling media (water flush). Connectivity: interfaces with MDU5 PLUS motor drive unit and iLab equipment. Sterilization: E-Beam irradiation. Compliance: IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), ISO 11607 (packaging).
Indications for Use
Indicated for patients with vascular occlusive disease for which angioplasty, atherectomy, the placement of stents, or other intervention is contemplated.
Regulatory Classification
Identification
An echocardiograph is a device that uses ultrasonic energy to create images of cardiovascular structures. It includes phased arrays and two-dimensional scanners.
Ultra ICE Plus 9 MHz IntraCardiac Echo Catheter (K160173)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 17, 2018
Boston Scientific Corporation Jennifer Foley Regulatory Affairs Specialist 3 Scimed Place Maple Grove, Minnesota 55311
Re: K181042
Trade/Device Name: Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter Regulation Number: 21 CFR 870.2330 Regulation Name: Echocardiograph Regulatory Class: Class II Product Code: DXK, ITX Dated: April 18, 2018 Received: April 19, 2018
Dear Jennifer Foley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Hillebrand
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181042
### Device Name
Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter
Indications for Use (Describe)
The Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter is indicated for patients with vascular occlusive disease for which angioplasty, atherectomy, the placement of stents, or other intervention is contemplated.
Type of Use (Select one or both, as applicable)
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpa</label> |
## |__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 3. 510(k) Summary Per CFR 807.92
| Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>Three Scimed Place<br>Maple Grove, MN 55311<br>USA |
|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name<br>and Information | Jennifer L. Foley, DVM<br>Regulatory Affairs Specialist<br>Tel: 763-955-8049<br>Fax: 763-494-2222<br>E-mail: jennifer.foley@bsci.com |
| Date Prepared | May 17, 2018 |
| Trade Name | Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter |
| Common Name | Intravascular Ultrasound Transducer |
| Classification<br>Name | Echocardiograph (DXK) is Class II per 21 CFR 870.2330<br>Diagnostic ultrasonic transducer (ITX) is Class II per 21 CFR 892.1570 |
| Predicate Device | Sonicath 9F, 12 MHZ (K902245) cleared October 9, 1990, amended<br>February 28, 1996 and marketed as Ultra ICE 9 MHz IntraCardiac Echo<br>Catheter. |
| Reference<br>Device | Ultra ICE Plus 9 MHz IntraCardiac Echo Catheter (K160173), cleared<br>February 26, 2016 |
| Description of<br>Device | Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter is intended for use<br>with the Boston Scientific iLab™ equipment and motor drive unit, MDU5<br>PLUS™. When used together, the catheter, motor drive unit (MDU), and<br>iLab equipment form a complete imaging system that allows for ultrasonic<br>visualization of peripheral intravascular structures. |
| | The catheter consists of two main components: the catheter body and the<br>imaging core. |
| | The catheter body consists of three sections: the braided proximal shaft,<br>single lumen mid-shaft, and the sonolucent distal tip. The catheter body<br>comprises the usable length of the catheter (110 cm). |
| | The braided proximal shaft provides the pushability to the catheter and<br>serves as a lumen to the imaging core. The mid-shaft provides a flexible<br>transition between the stiffer proximal shaft and the acoustically transparent<br>distal tip. The distal tip serves as the imaging window and houses a septum<br>situated between the inner lumen and the atraumatic rounded tip of the<br>catheter. The self-sealing septum serves as the distal-flush entry point, as<br>the catheter is flushed with water prior to use. This provides the acoustic<br>coupling media required for ultrasonic imaging.<br>The imaging core consists of a proximal hub assembly, a rotating drive |
| | cable, and a radiopaque tip. The hub assembly provides an electro- |
| | mechanical interface between the catheter and the motor drive unit. The<br>drive cable houses the low frequency piezoelectric (PZT) transducer at the<br>distal imaging window.<br>The PZT transducer and the drive cable rotate independently from the<br>sheath to provide 360° image resolution. The transducer converts electrical<br>impulses sent by the motor drive into transmittable acoustic energy.<br>Reflected ultrasound signals are converted back to electrical impulses,<br>returned to the motor drive unit, and are ultimately processed by the iLab<br>equipment for live visualization of intravascular structures. |
| Intended Use | The Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter is intended for<br>intravascular ultrasonic visualization. |
| Indications for<br>Use | The Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter is indicated for<br>patients with vascular occlusive disease for which angioplasty,<br>atherectomy, the placement of stents, or other intervention is contemplated. |
| Device<br>Technology<br>Characteristics<br>and Comparison<br>to Predicate<br>Device | The Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter maintains the<br>same fundamental scientific technology and operating principles as the<br>predicate Sonicath/Ultra ICE (K902245). The transducer, sheath design,<br>and indications for use remain unchanged. |
| | Modifications with respect to the predicate device include a new hub<br>interface for compatibility with the MDU5 PLUS™ motor drive unit as<br>similarly adopted by the Ultra ICE Plus 9 MHz IntraCardiac Echo Catheter<br>(K160173). |
| | Ultra ICE Plus - PI will be sterilized using electron beam (E-Beam)<br>irradiation. |
| | Non-clinical performance evaluations demonstrate the Ultra ICE Plus - PI<br>9 MHz Peripheral Imaging Catheter is substantially equivalent to, and at<br>least as safe and effective as the Sonicath/Ultra ICE (K902245). |
| Non-Clinical<br>Performance<br>Data | Determination of substantial equivalence is based on an assessment of<br>non-clinical performance data. |
| | Non-clinical data includes bench-top performance evaluations, packaging<br>validation, biological safety, electromagnetic compatibility, and acoustic<br>output testing. |
| | Bench Testing: |
| | Bench testing was performed to evaluate physical integrity, functionality,<br>and performance of the catheter. Performance criteria includes:<br>deliverability, guide catheter compatibility, image quality, non-uniform<br>rotational distortion, measurement accuracy, general imaging capabilities,<br>dimensional requirements, visibility under fluoroscopy, interface with<br>ancillary devices, environmental requirements, user interface requirements,<br>catheter fatigue, and bending stiffness. |
| | Biological Safety Testing: |
| | Biocompatibility testing in accordance with ISO 10993-1, microbial<br>assessments including bioburden and endotoxin, and pyrogenicity and<br>sterility assurance testing show the device is biocompatible for its intended<br>use. |
| | Electrical and Mechanical Safety: |
| | Acoustic Output was evaluated in accordance with FDA Guidance,<br>Information for Manufacturers Seeking Marketing Clearance of Diagnostic<br>Ultrasound Systems and Transducers (September 9, 2008). Acoustic<br>Output test results for the Ultra ICE Plus - PI 9 MHz Peripheral Imaging<br>Catheter are below the FDA Track 1 limits. Electromagnetic compatibility<br>testing was also conducted demonstrating compliance to IEC 60601-1-2<br>(3rd Edition). |
| | Packaging Validation: |
| | The integrity of the packaging configuration was evaluated in accordance<br>with ISO 11607-1 and ISO 11607-2. Testing was conducted on fully<br>packaged units after electron beam sterilization, climatic conditioning, and<br>distribution challenge conditioning. |
| Clinical<br>Performance<br>Data | Not applicable; determination of substantial equivalence is based on an<br>assessment of non-clinical performance data. |
| Conclusion | With respect to the predicate, design modifications incorporated by Ultra<br>ICE Plus - PI 9 MHz Peripheral Imaging Catheter do not alter the<br>fundamental scientific technology or the indications for use. Based on the<br>Failure Mode Effects Analysis, comprehensive verification and validation<br>activities were successfully completed, generating no new issues of safety<br>or effectiveness. |
| | Non-clinical performance data supports a determination that the subject<br>device, Ultra ICE Plus - PI 9 MHz Peripheral Imaging Catheter, is<br>substantially equivalent to the predicate device, Sonicath/Ultra ICE<br>(K902245), and it is at least as safe and effective for its intended use. |
{4}------------------------------------------------
{5}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.