Aveo (omafilcon A) Soft (Hydrophilic) Contact Lenses
K180985 · Supervision Optimax Sdn Bhd · LPL · Jun 8, 2018 · Ophthalmic
Device Facts
Record ID
K180985
Device Name
Aveo (omafilcon A) Soft (Hydrophilic) Contact Lenses
Applicant
Supervision Optimax Sdn Bhd
Product Code
LPL · Ophthalmic
Decision Date
Jun 8, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Aveo (omafilcon A) Aspheric Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic or hyperopic and exhibit astigmatism of 1.00D or less that does not interfere with visual acuity. Aveo ( omafilcon A) Toric Soft ( Hydrophilic) Contact Lenses are indicated the daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic or hyperopic and exhibit astigmatism of between 0.50D and 2.50D or less that does not interfere with visual acuity. Daily wear replacement schedules may vary from patient to patient and should be decided by eyecare practitioners in consultation with their patients. Frequent Planned Replacement Wear When prescribed for frequent planned replacement wear the lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eve and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system. Disposable Wear When prescribed for Daily Disposable Wear the lens is to be discarded after each removal.
Device Story
Aveo (omafilcon A) soft contact lenses; daily wear; indicated for visual acuity correction in non-aphakic, non-diseased eyes with myopia, hyperopia, and astigmatism. Material: HEMA-MPC copolymer; 58% water content; includes RB 246 pigment for handling tint and RUVA-93 UV absorber. Aspheric front curve (tri-curve) and spherical base curve (aspheric model); multi-radius base curve (toric model). Supplied sterile in blister packs with phosphate-buffered saline. Prescribed by eyecare practitioners; worn by patients. Frequent planned replacement (cleaned/disinfected) or daily disposable (discarded after removal). Benefits: biocompatibility via MPC (phospholipid-like) material; vision correction.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via bench testing, including physiochemical studies (ISO 18369-3/4) and toxicology studies (ISO 10993-5/10/11: cytotoxicity, ocular irritation, skin sensitization, acute systemic toxicity).
Indicated for daily wear correction of visual acuity in non-aphakic, non-diseased eyes. Aspheric lenses for myopia/hyperopia with astigmatism ≤1.00D. Toric lenses for myopia/hyperopia with astigmatism 0.50D to 2.50D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Proclear (omafilcon A) Soft Contact Lenses (K112302)
Aveo (omafilcon A) 1-day Aspheric Soft (Hydrophilic) Contact Lens (K162223)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in blue and white. The text is arranged in three lines, with "FDA" on the top line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
June 8, 2018
Supervision Optimax SDN BHD Yap Peak Geeh Regulatory Affairs Manager Lot 38. Putra Industrial Park. Bukit Rahman Putra 40160 Sungai Buloh, Selangor Dural Ehsan, Malaysia
Re: K180985
Trade/Device Name: Aveo (omafilcon A) Soft (Hydrophilic) Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: March 30, 2018 Received: April 13, 2018
Dear Yap Peak Geeh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J Angelo Green -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180985
#### Device Name
Aveo (omafilcon A) Soft (Hydrophilic) Contact Lenses
#### Indications for Use (Describe)
Aveo (omafilcon A) Aspheric Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic and exhibit astigmatism of 1.00D or less that does not interfere with visual acuity.
Aveo ( omafilcon A) Toric Soft ( Hydrophilic) Contact Lenses are indicated the daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic and exhibit astigmatism of between 0.50D and 2.50D or less that does not interfere with visual acuity.
Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.
#### Frequent Planned Replacement Wear
When prescribed for frequent planned replacement wear the lenses are to be cleaned, rinsed and time they are removed from the patients' eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.
Disposable Wear When prescribed for Daily Disposable Wear the lens is to be discarded after each removal.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary K180985
### Preparation Date: 05th June 2018
#### 1.0 Submitter:
| Name | : Supervision Optimax Sdn Bhd |
|----------------|---------------------------------------------------------------------------------------------------|
| Address | : Lot 38, Putra Industrial Park,<br>Bukit Rahman Putra,<br>40160 Sungai Buloh, Selangor, Malaysia |
| Phone No. | : 603-6145 2328 |
| Fax No. | : 603-6156 2191 |
| Contact Person | : Yap Peak Geeh |
| Email Address | : pg.yap@supervision.com.my |
# 2.0 Device Identification:
Proprietary Name : Aveo (omafilcon A) Soft (Hydrophilic) Contact Lenses Common Name : Soft (Hydrophilic) Contact Lenses Classification Name: Lenses, Soft Contact, Daily Wear (21 CFR 886.5925, Product Code LPL) Lenses, Soft Contact, (Disposable) (21 CFR 886.5925, Product Code MVN) Classification : Class II
### 3.0 Identification Of The Legally Marketed Devices that equivalency is claimed:
| | Predicate Device | |
|-------------------|--------------------------------------------|-------------------------------------------------------------------|
| Manufacturer | CooperVision, Inc | Supervision Optimax Sdn Bhd |
| Device Name | Proclear (omafilcon A) Soft Contact Lenses | Aveo (omafilcon A) 1-day Aspheric Soft (Hydrophilic) Contact Lens |
| 510(k) Number | K112302 | K162223 |
| Regulation Number | 21 CFR 886.5925 | 21CFR 886.5925 |
| Regulatory Name | Soft (hydrophilic) contact lens | Soft (hydrophilic) contact lens |
| Regulatory Class | II | II |
| Product Code | LPL and MVN | LPL and MVN |
{4}------------------------------------------------
### 4.0 Description of the Device:
The Aveo (Omafilcon A) Soft (Hydrophilic) Contact Lenses are daily wear soft contact lenses produced from the HEMA-MPC copolymer material. MPC is similar to phospholipids (e.g., phosphatidylcholine) where molecules are found naturally in human cell membranes that improves the lens biocompatibility.
The contact lenses contain 58% water by weight and are sold in the blister package immersed in phosphate buffered packaging saline. RB 246 pigment conforms to 21 CFR Part 73.3106 is used to provide the handling blue tint for the lens. Reactive ultraviolet absorber RUVA-93 is used to block UV radiation.
The device is a corneal contact lens having a total diameter less than the visible iris diameter and is designed to be worn in its entirety on the cornea.
Aveo (Omafilcon A) Aspheric Soft (Hydrophilic) Contact Lenses have an aspheric front curve (external curve) which is tri-curve and spherical base curve (internal curve).
Aveo (Omafilcon A) Toric Soft (Hydrophilic) Contact Lenses have an aspheric front curve (external curve) which is tri-curve for the sphere power and multi-radius base curve for the cylinder power (internal curve).
The contact lenses are hydrophilic, soft and it is supplied in sterile state.
#### 5.0 Intended Use of the Device:
Aveo (omafilcon A) Aspheric Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic or hyperopic and exhibit astigmatism of 1.00D or less that does not interfere with visual acuity.
Aveo ( omafilcon A) Toric Soft ( Hydrophilic) Contact Lenses are indicated the daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic or hyperopic and exhibit astigmatism of between 0.50D and 2.50D or less that does not interfere with visual acuity.
Daily wear replacement schedules may vary from patient to patient and should be decided by eyecare practitioners in consultation with their patients.
#### Frequent Planned Replacement Wear
When prescribed for frequent planned replacement wear the lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eve and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.
{5}------------------------------------------------
## Disposable Wear
When prescribed for Daily Disposable Wear the lens is to be discarded after each removal.
## 6.0 Summary of the Technological Characteristics of the Device:
Below is the summary of the technological characteristics of the Aveo (omafilcon A) Soft (Hydrophilic) Contact Lenses as compared to the predicate device.
| Technological Characteristics | | | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Subject Device | Predicate Device | Predicate Device |
| Product Name | Aveo | Proclear Asphere | Aveo |
| Manufacturer | Supervision Optimax Sdn<br>Bhd | CooperVision, Inc. | Supervision Optimax<br>Sdn Bhd |
| 510(K) Number | K180985 | K112302 | K162223 |
| Intended Use | Aveo (omafilcon A)<br>Aspheric Soft (Hydrophilic)<br>Contact Lenses are<br>indicated for daily wear for<br>the correction of visual<br>acuity in not aphakic<br>persons with non-diseased<br>eyes that are myopic or<br>hyperopic and exhibit<br>astigmatism of 1.00D or less<br>that does not interfere with<br>visual acuity.<br><br>Aveo (omafilcon A) Toric<br>Soft (Hydrophilic) Contact<br>Lenses are indicated the<br>daily wear for the<br>correction of visual acuity in<br>not aphakic persons with<br>non-diseased eyes that are<br>myopic or hyperopic and<br>exhibit astigmatism of<br>between 0.50D and 2.50D<br>or less that does not<br>interfere with visual acuity. | Proclear Asphere<br>(omafilcon A) Soft<br>Contact Lenses are<br>indicated for daily wear<br>for the correction of<br>visual acuity in non-<br>aphakic persons with<br>non-diseased eyes that<br>are myopic or hyperopic<br>and exhibit astigmatism<br>of 2.00D or less that<br>does not interfere with<br>visual acuity.<br><br>Proclear Toric<br>(omafilcon A) Soft<br>Contact Lenses are<br>indicated for daily wear<br>for the correction of<br>visual acuity in non-<br>aphakic persons with<br>non-diseased eyes that<br>are myopia or hyperopic.<br>The lens may be worn by<br>persons who have<br>astigmatism of 5.00D or<br>less . | Aveo (omafilcon A)1-<br>Day Aspheric Soft<br>(Hydrophilic) Contact<br>Lenses are indicated<br>for daily wear for the<br>correction of visual<br>acuity in not aphakic<br>persons with non-<br>diseased eyes that<br>are myopic or<br>hyperopic and exhibit<br>astigmatism of 1.00D<br>or less that does not<br>interfere with visual<br>acuity. The contact<br>lenses are intended<br>for daily wear, single<br>use and are to be<br>discarded at the end<br>of the day. |
| Modality | Daily Wear | Daily Wear | Daily Wear |
| Lens Design | Aspherical | Aspherical | Aspherical |
| Material | Omafilcon A | Omafilcon A | Omafilcon A |
| Name | | | |
| FDA<br>(Group) | Group II<br>Non-ionic, High water | Group II<br>Non-ionic, High water | Group II<br>Non-ionic, High water |
| Manufacturing<br>Method | Cast Molded | Finished Inside<br>Polymerization System II | Cast Molded |
| Curing | Thermal Cure | Thermal Cure | Thermal Cure |
| Sterilization | Moist Heat (Steam) in<br>Validated Autoclave | Moist Heat (Steam) in<br>Validated Autoclave | Moist Heat (Steam) in<br>Validated Autoclave |
| Packaging | Blister Pack | Blister Pack | Blister Pack |
| Visibility Tint | Reactive Blue Dye 246 | VAT Blue 6 | Reactive Blue Dye 246 |
| Water Content | 59% ± 2% | 59% ± 2% | 59% ± 2% |
| Package Saline | Phosphate Buffered Saline | Phosphate Buffers<br>PEG200 and Tween 80 | Phosphate Buffered Saline |
| Refractive Index | 1.4002 | $1.395 \pm 0.005$ | 1.4002 |
| Oxygen<br>Permeability (Dk) x<br>10-11 | 25.68 | 21.05 | 25.68 |
| Light Transmission | 98% | >90% | 98% |
| Base Curve | 8.4mm to 8.8mm | 8.0mm to 9.3mm | 8.4mm to 8.8mm |
| Diameter Øт | 14.0mm to 14.7mm | 13.6mm to 15.2mm | 14.0mm to 14.4mm |
| Power | -10.00 to +6.00 | -20.00 to +20.00 | -10.00 to +6.00 |
{6}------------------------------------------------
## 7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data
## Physiochemical Studies
The physiochemical studies were conducted according to ISO 18369-4:2006 Ophthalmic Optics-Contact Lenses-Part 4: Physiochemical properties of contact lens materials and ISO 18369-3:2006 Ophthalmic Optic-Contact Lenses-Part 3: Measurement methods. The physical, optical and chemical properties of the lens are within established specifications for the lenses.
#### Toxicology Studies
Toxicology (in-vivo and in-vitro) studies reports show that the lenses are non-toxic and biocompatible with the ocular environment.
{7}------------------------------------------------
| Test | Performance of Subject<br>Device | Result |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|--------|
| Cytotoxicity Test<br>ISO 10993-5: 2009(E): Biological<br>Evaluation of Medical Devices-Part 5:<br>Tests for in vitro Cytotoxicity. | Non-cytotoxic. | Pass |
| Ocular Irritation Study in New Zealand<br>White Rabbit<br>ISO 10993-10: 2010(E): Biological<br>Evaluation of Medical Devices-Part 10:<br>Tests for Irritation and Skin Sensitization. | Non-irritant to eyes of<br>rabbits. | Pass |
| Skin Sensitization Study in Guinea Pigs<br>ISO 10993-10: 2010(E): Biological<br>Evaluation of Medical Devices-Part 10:<br>Tests for Irritation and Skin Sensitization. | Non-sensitizer | Pass |
| Acute Systemic Toxicity Study in Swiss<br>Albino Mice<br>ISO 10993-11: 2006(E): Biological<br>Evaluation of Medical Devices-Part 11:<br>Tests for in Systemic Toxicity. | Animals treated with the<br>extract of the subject<br>device did not show any<br>systemic toxicity. | Pass |
#### 8.0 Clinical Test
The technological characteristics, formulation, manufacturing and sterilization processes are the same as the predicate device, therefore, no clinical studies were required to demonstrate the safety or effectiveness of the subject device.
#### 9.0 Conclusion
The Aveo (Omafilcon A) Soft (Hydrophilic) Contact Lenses are substantially equivalent to the predicate device as they are produced from the same material (Omafilcon A), have the same functional and scientific technology, lens characteristics and the intended use is identical.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.