← Product Code [MQB](/productcode/MQB) · K180976

# PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-ray Receptor Panels (K180976)

_Pixxgen Corporation · MQB · May 24, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180976

## Device Facts

- **Applicant:** Pixxgen Corporation
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** May 24, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic, and mammographic applications

## Device Story

PIXX series (1717, 1417, 1212) are digital radiography systems featuring integrated flat panel digital detectors (FPD). Devices capture X-ray signals to produce digital radiographic images, replacing conventional film. Operated by clinicians in diagnostic settings; panels interface with PACS for image management. Systems utilize CsI or GOS scintillators to convert X-rays to light, then to digital signals. Output is a DICOM image for radiologist review to support clinical diagnosis. Benefits include filmless workflow, improved cost-effectiveness, and digital image quality. Wireless (IEEE 802.11ac) and wired (Gigabit Ethernet) connectivity provided.

## Clinical Evidence

Clinical image evaluation performed by a board-certified radiologist. Images acquired with the new panels were compared to those from the predicate device. Results concluded that image quality is equal to or better than the predicate. No formal clinical trial with primary endpoints or statistical power analysis reported; evidence based on qualitative clinical image assessment.

## Technological Characteristics

Flat panel digital detectors (FPD) using CsI or GOS scintillators. 16-bit A/D conversion. Connectivity via Gigabit Ethernet or IEEE 802.11ac. Electrical safety per IEC 60601-1:2012; EMC per IEC 60601-1-2:2007+AC:2010. Dimensions vary by model (17x17, 14x17, 12x12 inch active areas). Software is identical to predicate K152151.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- ATAL 9 ([K152151](/device/K152151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

May 24, 2018

PIXXGEN Corporation % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

Re: K180976

Trade/Device Name: PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: April 25, 2018 Received: April 27, 2018

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

### Page 2 - Daniel Kamm

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hole 2. Nils

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180976

Device Name

PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels

Indications for Use (Describe)

Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic, and mammographic applications

Type of Use (Select one or both, as applicable)

| <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> |
|------------------------------------------------------------------------------------------------------------------|
| <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span>                                                |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary 510(k) Number K180976 PIXXGEN Corporation 5F, SMART BAY, 123, Beolmal-ro, Dongan-gu, Anyang-si, Gyeonggi-do, 14056, Korea Tel +82.70.4846.8888 Date Prepared: May 18, 2018 Contact: Young Kim, President

- 1. Identification of the Device: Proprietary-Trade Names: PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels Device: Solid State X-Ray Imager (Flat Panel/Digital Imager) Regulation Name: Stationary x-ray system. Product Code: MQB Regulation Number: 892.1680 Device Class: 2
- 2. Equivalent legally marketed device: ATAL 9, K152151, Atlaim Corporation. (Company name changed now to Pixxgen Corporation)

| Device:            | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
|--------------------|------------------------------------------------------|
| Regulation Name:   | Stationary x-ray system.                             |
| Product Code:      | MQB                                                  |
| Regulation Number: | 892.1680                                             |
| Device Class:      | 2                                                    |

- 3. Indications for Use (intended use) Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, and mammographic applications.
- 4. Description of the Device:

The PIXX series are digital radiography systems, featuring an integrated flat panel digital detector (FPD). They are designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the updated panels are the same as our previous panel retaining the Wi-Fi wireless features.

- 5. Safety and Effectiveness, comparison to predicate device. The results of clinical image inspection, bench, and test laboratory results indicates that the new device is as safe and effective as the predicate device. Clinical images collected demonstrate equal or better image quality as compared to our predicate.

{4}------------------------------------------------

# 6. Substantial Equivalence Chart

|                                    | ATAL 9, K152151                                                                                                                                                                                                                                                                             | PIXX 1717                                                                                                                                        | PIXX 1417                                                                                                                                           | PIXX 1212                                                                                                                                        |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                       | Indicated for use in general<br>radiographic images of human<br>anatomy. It is intended to<br>replace radiographic film/screen<br>systems in all general-purpose<br>diagnostic procedures, excluding<br>fluoroscopic, angiographic, and<br>mammographic applications                        | UNCHANGED                                                                                                                                        | UNCHANGED                                                                                                                                           | UNCHANGED                                                                                                                                        |
| Configuration                      | This submission is for the<br>Digital Panel and Software<br>only, no generator or stand<br>provided.                                                                                                                                                                                        | UNCHANGED                                                                                                                                        | UNCHANGED                                                                                                                                           | UNCHANGED                                                                                                                                        |
| Pixel Pitch                        | 139um                                                                                                                                                                                                                                                                                       | 140um/168um                                                                                                                                      | 140 um                                                                                                                                              | 140 um                                                                                                                                           |
| Limiting<br>Resolution             | Over 3 lp/mm                                                                                                                                                                                                                                                                                | SAME                                                                                                                                             | SAME                                                                                                                                                | SAME                                                                                                                                             |
| DQE(CSI)                           | At 2 lp/mm 29.5%                                                                                                                                                                                                                                                                            | At 2 lp/mm 26.5%                                                                                                                                 | At 2 lp/mm 26.5%                                                                                                                                    | At 2 lp/mm 26.5%                                                                                                                                 |
| MTF(CSI)                           | At 2 lp/mm 42%                                                                                                                                                                                                                                                                              | At 2 lp/mm 44%                                                                                                                                   | At 2 lp/mm 44%                                                                                                                                      | At 2 lp/mm 44%                                                                                                                                   |
| DQE(GOS)                           | At 2 lp/mm 18                                                                                                                                                                                                                                                                               | At 2 lp/mm 21                                                                                                                                    | At 2 lp/mm 21                                                                                                                                       | At 2 lp/mm 21                                                                                                                                    |
| MTF(GOS)                           | At 2 lp/mm 33%                                                                                                                                                                                                                                                                              | At 2 lp/mm 35%                                                                                                                                   | At 2 lp/mm 35%                                                                                                                                      | At 2 lp/mm 35%                                                                                                                                   |
| DQE and MTF<br>Comments            | The MTF and DQE measurements are very similar between our predicate and the new PIXX series of<br>digital panels. Some measurements are slightly higher, and some are very slightly lower. The 7% MTF<br>difference for CSI performance is comparable to the<br>possible measurement error. |                                                                                                                                                  |                                                                                                                                                     |                                                                                                                                                  |
| A/D Conversion                     | 16 bits                                                                                                                                                                                                                                                                                     | 16 bits                                                                                                                                          | 16 bits                                                                                                                                             | 16 bits                                                                                                                                          |
| Active Area                        | 17 x 17 inch                                                                                                                                                                                                                                                                                | 17 x 17 inch                                                                                                                                     | 14 x 17 inch                                                                                                                                        | 12 x 12 inch                                                                                                                                     |
| Dimensions(m<br>m)/<br>Weights(Kg) | 460(W)x461(L)x15(H)/(2.9Kg<br>w/o Battery)(up to 10 hours<br>of battery life)                                                                                                                                                                                                               | 460(W)x461(L)x15(H)/<br>3.0Kg                                                                                                                    | 385(W) x 460(L) x<br>15(H)/ 2.8Kg                                                                                                                   | 308.5(W) x 319.5(L) x<br>15(H)/ 1.9Kg                                                                                                            |
| Pixels                             | 3,072x3,072                                                                                                                                                                                                                                                                                 | 3,072 x 3,072                                                                                                                                    | 2,560 x 3,072                                                                                                                                       | 2,048 x 2,048                                                                                                                                    |
| Software                           | Outputs a DICOM image                                                                                                                                                                                                                                                                       | SAME as K152151<br>The software is<br>unchanged.                                                                                                 | SAME as K152151<br>The software is<br>unchanged.                                                                                                    | SAME as K152151<br>The software is<br>unchanged.                                                                                                 |
| DICOM                              | Yes                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                              | Yes                                                                                                                                                 | Yes                                                                                                                                              |
| Scintillator                       | Csl orGOS                                                                                                                                                                                                                                                                                   | UNCHANGED                                                                                                                                        | UNCHANGED                                                                                                                                           | UNCHANGED                                                                                                                                        |
| Interface                          | Wired: Gigabit Ethernet<br>(1000BASE-T)<br>Wireless:IEEE802.11ac,<br>backward compatible                                                                                                                                                                                                    | Wired: Gigabit<br>Ethernet<br>(1000BASE-T)<br>Wireless:<br>IEEE802.11ac,<br>backward compatible                                                  | Wired: Gigabit<br>Ethernet<br>(1000BASE-T)<br>Wireless:<br>IEEE802.11ac,<br>backward<br>compatible                                                  | Wired: Gigabit<br>Ethernet<br>(1000BASE-T)<br>Wireless:<br>IEEE802.11ac,<br>backward compatible                                                  |
| Power source                       | AC Line and/or Rechargeable<br>Lithium Battery                                                                                                                                                                                                                                              | AC Line                                                                                                                                          | AC Line                                                                                                                                             | AC Line                                                                                                                                          |
|                                    | ATAL 9, K152151                                                                                                                                                                                                                                                                             | PIXX 1717                                                                                                                                        | PIXX 1417                                                                                                                                           | PIXX 1212                                                                                                                                        |
| Standards                          | Electrical Safety per IEC<br>60601-1 and EMC per IEC<br>60601-1-2 as well as<br>IEEE802.11ac. Meets FCC<br>requirements.                                                                                                                                                                    | Electrical Safety per<br>IEC 60601-1:2012 and<br>EMC per IEC 60601-1-<br>22007+AC:2010<br>as well as<br>IEEE802.11ac. Meets<br>FCC requirements. | Electrical Safety per<br>IEC 60601-1:2012<br>and EMC per IEC<br>60601-1-<br>22007+AC:2010<br>as well as<br>IEEE802.11ac. Meets<br>FCC requirements. | Electrical Safety per<br>IEC 60601-1:2012 and<br>EMC per IEC 60601-1-<br>22007+AC:2010<br>as well as<br>IEEE802.11ac. Meets<br>FCC requirements. |
| Photos                             | Image: ATAL 9, K152151                                                                                                                                                                                                                                                                      | Image: PIXX 1717                                                                                                                                 | Image: PIXX 1417                                                                                                                                    | Image: PIXX 1212                                                                                                                                 |

{5}------------------------------------------------

- 7. Summary of Bench Testing Conducted: IEC Standards were employed for: Electrical Safety, IEC60601-1 and Electromagnetic Compatibility, IEC60601-1-2 as shown in the comparison table above. MTF and DQE measurements, Risk Analysis and Software verification were conducted in accordance with the FDA guidance documents: Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff and Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. As expected the DQE and MTF measurements show minimal differences, and the limiting resolution is the same. The software remains the same as in K152151. Wireless communication testing was performed to verify wireless connectivity. The device was also found to comply with FCC requirements for wireless operation. In addition, the following FDA Guidance Documents were consulted during the development of this device: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff and Pediatric Information for X-ray Imaging Device Premarket Notifications.
- 8. Summary of Clinical Testing: Clinical images were acquired and evaluated by a board certified radiologist who concluded the images from the new panels are as good as or better than the images acquired with the predicate panel, as required by the applicable FDA guidance on solid state imaging devices. The core technology of the panel remains the same.
- 9. Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable standards, it is the conclusion of Pixxgen that the PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels are as safe and effective as the predicate device, have few technological differences, and has the identical indications for use, thus rendering them substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K180976](https://fda.innolitics.com/device/K180976)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
